Retinitis PigmentosaNCT00407602

Argus® II Retinal Stimulation System Feasibility Protocol

A study testing Implant of Argus II Retinal Prosthesis in people with Retinitis Pigmentosa, run by Second Sight Medical Products at 12 sites in 5 countries.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Investigational Phase of the Study: The objective of this feasibility study is to evaluate the safety and utility of the Argus II Retinal Stimulation System in providing visual function to blind subjects with retinitis pigmentosa. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Retinitis Pigmentosa
What is being tested?
Implant of Argus II Retinal Prosthesis
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
30 participants
Where
12 sitesin 5 countries
Study dates
Sep 2006to 31 Dec 2019
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • A confirmed history of retinitis pigmentosa (all centers) or outer retinal degeneration (France, U.K., Switzerland, Mexico only) with remaining visual acuity of bare light perception (all centers) or 2.3 logMAR (France, U.K., Switzerland, Mexico only) or worse in both eyes.
  • Functional ganglion cells and optic nerve as determined by a measurable electrically evoked response or documented light perception.
  • A history of former useful form vision in the worse-seeing eye.
  • Must be at least the following age at the time of enrollment: 25 (USA, Switzerland) or 18 (France, U.K., and Mexico) years old
  • Must reside within 2 hours (USA, UK and Mexico) or 3 hours (France and Switzerland) distance (by ground transportation) of the investigational site.
  • Must be willing and able to comply with the protocol testing and follow-up requirements.

Who cannot join

  • Optic Nerve disease
  • History of glaucoma
  • Optic neuropathy or other confirmed damage to optic nerve or visual cortex damage
  • Diseases or conditions that effect retinal function including but not limited to:
  • Central retinal artery/vein occlusion (CRAO or CRVO)
  • End-stage diabetic retinopathy
Show all 28Show fewer
  • Retinal detachment or history of retinal detachment
  • Trauma
  • Infectious or inflammatory retinal diseases
  • Diseases or conditions that prevent adequate visualization of the retina including, but not limited to corneal degeneration that cannot be resolved before implant.
  • Diseases or conditions of the anterior segment that prevent the ability to adequately perform the physical examination including but not limited to trauma or lid malpositions.
  • Diseases of the ocular surface including but not limited to keratitis sicca.
  • An ocular condition that predisposes the subject to eye rubbing.
  • Any disease or condition that prevents understanding or communication of informed consent, study demands, and testing protocols, including:
  • Cognitive decline including diagnosed forms of dementia and/or progressive neurological disease
  • Psychiatric Disease including diagnosed forms of depression
  • Does not speak a principal language associated with the region
  • Deafness or selective frequency hearing loss that prevents hearing device alarms and alerts.
  • Pregnancy
  • Subject has another active implantable device (e.g. cochlear implant), or any form of metallic implant in the head
  • Conjunctival thinning which may predispose the subject to conjunctival erosion in the area where the implant will be installed extra-ocularly.
  • Subject is participating in another investigational drug or device study that may conflict with the objectives, follow-up or testing of this study
  • Any health concern that makes general anesthesia inadvisable.
  • Subject has unrealistic expectations of the implant.
  • Known allergy or contraindication to anticipated pre-operative, intra-operative or post-operative medications.
  • Conditions likely to limit life to less than 1 year from the time of screening.
  • Diseases or conditions that, in the judgement of the surgeon, impede the ability to implant the device or would prevent the system from functioning for the duration of the study (e.g. strabismus)
  • Axial eye length <21.5 mm or >26.0 mm in the implanted eye as measured by ultrasound (US only)

Trial locations and contacts

12 sites in 5 countries.

12 sites in 5 countries

United States7 sites
  • Johns Hopkins, Wilmer Eye Institute

    Baltimore, Maryland, 21205

  • Retina Foundation of the Southwest

    Dallas, Texas, 75231

  • Doheny Eye Institute

    Los Angeles, California, 90033

  • Columbia University, Department of Ophthalmology, Edward S. Harkness Eye Institute

    New York, New York, 10032

  • University of Pennsylvania, Scheie Eye Institute

    Philadelphia, Pennsylvania, 19104

  • Wills Eye Hospital

    Philadelphia, Pennsylvania, 19107

  • University of California San Francisco

    San Francisco, California, 94143

Other trials in United States

United Kingdom2 sites
  • Moorfields Eye Hospital, Vitreoretinal Research Unit

    London, EC1V 2PD

  • Manchester Royal Eye Hospital

    Manchester

Other trials in United Kingdom

France1 site
  • Centre d'Investigation Clinique, Service d'Ophtalmologie, Quinze-Vingts National Eye Hospital, 28 rue de Charenton, 75557 Cedex 12

    Paris

Other trials in France

Mexico1 site
  • Puerta de Hierro, Centro Medico, Centro de Retina

    Zapopan, Jalisco, 51116

Other trials in Mexico

Switzerland1 site
  • Clinique d'Ophthalmologie Hopitaux, Universitaires de Geneve 22 rue Alcide Jentzer 1205

    Geneva

Other trials in Switzerland

About this study

Investigational Phase of the Study:

The objective of this feasibility study is to evaluate the safety and utility of the Argus II Retinal Stimulation System in providing visual function to blind subjects with retinitis pigmentosa.

Post-Approval Phase of the Study:

To collect post-approval data in order to monitor the ongoing safety and reliability of the Argus II System

Read the detailed description

During the post-approval phase, subjects will undergo annual eye exams, assessments of medical status and adverse events, and measurement of stimulation thresholds. In addition, at the mutual agreement of the investigator and the subject, subjects will have the option of participating in psychophysical research which could occur as frequently as monthly. Functional tests at 10 years: photographic flash, square localization, direction of motion and grating visual acuity.

Being tested
Implant of Argus II Retinal Prosthesis
Keywords
retinitis pigmentosa, device, retinal implant, retinal prosthesis, RP, outer retinal degeneration
Medical subject headings
Retinitis Pigmentosa

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Implant of Argus II Retinal ProsthesisDevice

    epiretinal implantation of device

Study groups in NCT00407602
Study groupRoleWhat they receive
Implant of Argus II Retinal ProsthesisOtherThis is a single group study where the status and performance of the implanted eye prior to surgery serves as the comparator.
Receives: Device: Implant of Argus II Retinal Prosthesis

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Visual Acuity

    Grating Visual Acuity is performed at Baseline and at various timepoints throughout the study. As of the 10 year final testing, the total number of subjects that were able to score on the logMAR scale are indicated below. The device has a scale from 1.6 to 2.9 logMAR. Maximum likelihood estimation of visual acuity is based on BEST-PEST algorithm.

    Measured: 10 years

  2. Number of Participants With Adverse Events

    Evaluate ocular adverse events reported for all subjects from day of implant through 10 years of study participation

    Measured: 10 years

Secondary outcomes (6)
  1. Massof Activity Inventory

    The Massof Activity Inventory measures changes in ability to perform activities of daily living in terms of Goals and Tasks in logits (log-odds units). A logit scale is an equal interval linear scale that represents probability values in terms of real numbers. There is no minimum or maximum scale used in this testing. In this case, a Rasch distribution (using an Andrich rating model) was used to find the maximum likelihood of difficulty of achieving Goals and Tasks, represented in logits. The change in mean logit value between baseline and the 36Month visit for each subject are provided below along with the 95% confidence interval (CI) for the change. A positive change in the Goals or Tasks score which is greater than 0.2 (with a CI less than the mean logit value), is interpreted as a clinically significant increase in functional ability (better outcome).

    Measured: 3 years

  2. Quality of Life - Tasks for Daily Living

    Quality of life is being measured using the VisQOL survey. This instrument, developed by researchers at the University of Melbourne, Australia, is a vision and quality of life-related utility measure intended to evaluate healthcare interventions for the visually impaired. The utility score is a preference-based measure of vision-related quality of life (Dimension 7 score (Health Dimension Score (1=Best, 0=Worst)). It measures subjective quality of life on a scale of 0 to 1, where 0 represents worst possible quality of life and 1 represents perfect quality of life.

    Measured: 3 years

  3. Orientation and Mobility

    Argus II System for orientation and mobility, each subject completed two tests (walking to a high-contrast "door" on the wall and following a straight line on the floor) at various timepoints throughout the study. Accuracy (percent of trials where the subject successfully reached the target) was measured with the device system ON and system OFF. Average % success rates across the analysis population are provided for each modality and for each test at the 60 month time point.

    Measured: 5 years

  4. Visual Function

    Square localization (SL) and Direction of Motion (DOM) testing performed at specific protocol intervals. SL tests ability of subjects to locate high-contrast square on black screen. DOM tests ability of subjects to determine the direction a white bar is moving over a black screen. Each test is performed with the system ON and then compared to performance with the system OFF. Number of subjects who performed significantly better with the system ON is indicated below.

    Measured: 10 years

  5. Stability of Implant - 5 Years Post Implant

    Stability of electrode array on the retina. Qualitative review of imaging: retinal and array landmarks compared from all time points post implant.

    Measured: 5 years

  6. Device Reliability

    Number of participants with functional devices 10 years post implant

    Measured: 10 years

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted5 Dec 2006 (estimated)
  2. Study startSep 2006
  3. Primary completion31 Dec 2019
  4. Study completion31 Dec 2019

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
30 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT00407602
Sponsor's protocol ID
CP-003-001
Other IDs
National Eye Institute (NEI)
Status last verified
Jan 2022
First submitted
1 Dec 2006
Last updated
20 Jan 2022
Data monitoring committee
Yes
Individual data sharing
No

Publications

  • Weiland JD, Faraji B, Greenberg RJ, Humayun MS, Shellock FG. Assessment of MRI issues for the Argus II retinal prosthesis. Magn Reson Imaging. 2012 Apr;30(3):382-9. doi: 10.1016/j.mri.2011.12.005. Epub 2012 Jan 20.

    PubMed 22260934
  • Humayun MS, Dorn JD, da Cruz L, Dagnelie G, Sahel JA, Stanga PE, Cideciyan AV, Duncan JL, Eliott D, Filley E, Ho AC, Santos A, Safran AB, Arditi A, Del Priore LV, Greenberg RJ; Argus II Study Group. Interim results from the international trial of Second Sight's visual prosthesis. Ophthalmology. 2012 Apr;119(4):779-88. doi: 10.1016/j.ophtha.2011.09.028. Epub 2012 Jan 11.

    PubMed 22244176
  • Barry MP, Dagnelie G; Argus II Study Group. Use of the Argus II retinal prosthesis to improve visual guidance of fine hand movements. Invest Ophthalmol Vis Sci. 2012 Aug 1;53(9):5095-101. doi: 10.1167/iovs.12-9536.

    PubMed 22661464
  • Ahuja AK, Dorn JD, Caspi A, McMahon MJ, Dagnelie G, Dacruz L, Stanga P, Humayun MS, Greenberg RJ; Argus II Study Group. Blind subjects implanted with the Argus II retinal prosthesis are able to improve performance in a spatial-motor task. Br J Ophthalmol. 2011 Apr;95(4):539-43. doi: 10.1136/bjo.2010.179622. Epub 2010 Sep 29.

    PubMed 20881025
  • Arevalo JF, Al Rashaed S, Alhamad TA, Al Kahtani E, Al-Dhibi HA; KKESH Collaborative Retina Study Group. Argus II retinal prosthesis for retinitis pigmentosa in the Middle East: The 2015 Pan-American Association of Ophthalmology Gradle Lecture. Int J Retina Vitreous. 2021 Oct 27;7(1):65. doi: 10.1186/s40942-021-00324-6.

    PubMed 34706764
  • Sabbah N, Authie CN, Sanda N, Mohand-Said S, Sahel JA, Safran AB. Importance of eye position on spatial localization in blind subjects wearing an Argus II retinal prosthesis. Invest Ophthalmol Vis Sci. 2014 Nov 20;55(12):8259-66. doi: 10.1167/iovs.14-15392.

    PubMed 25414187
  • Ahuja AK, Yeoh J, Dorn JD, Caspi A, Wuyyuru V, McMahon MJ, Humayun MS, Greenberg RJ, Dacruz L. Factors Affecting Perceptual Threshold in Argus II Retinal Prosthesis Subjects. Transl Vis Sci Technol. 2013 Apr;2(4):1. doi: 10.1167/tvst.2.4.1. Epub 2013 Apr 12.

    PubMed 24049718
  • de Juan E Jr, Spencer R, Barale PO, da Cruz L, Neysmith J. Extraction of retinal tacks from subjects implanted with an epiretinal visual prosthesis. Graefes Arch Clin Exp Ophthalmol. 2013 Oct;251(10):2471-6. doi: 10.1007/s00417-013-2452-y. Epub 2013 Sep 8.

    PubMed 24013578
  • Stronks HC, Barry MP, Dagnelie G. Electrically elicited visual evoked potentials in Argus II retinal implant wearers. Invest Ophthalmol Vis Sci. 2013 Jun 6;54(6):3891-901. doi: 10.1167/iovs.13-11594.

    PubMed 23611993
  • Dorn JD, Ahuja AK, Caspi A, da Cruz L, Dagnelie G, Sahel JA, Greenberg RJ, McMahon MJ; gus II Study Group. The Detection of Motion by Blind Subjects With the Epiretinal 60-Electrode (Argus II) Retinal Prosthesis. JAMA Ophthalmol. 2013 Feb;131(2):183-9. doi: 10.1001/2013.jamaophthalmol.221.

    PubMed 23544203
  • Ahuja AK, Behrend MR. The Argus II retinal prosthesis: factors affecting patient selection for implantation. Prog Retin Eye Res. 2013 Sep;36:1-23. doi: 10.1016/j.preteyeres.2013.01.002. Epub 2013 Mar 14.

    PubMed 23500412
  • Olson JL, Velez-Montoya R, Mandava N, Stoldt CR. Intravitreal silicon-based quantum dots as neuroprotective factors in a model of retinal photoreceptor degeneration. Invest Ophthalmol Vis Sci. 2012 Aug 17;53(9):5713-21. doi: 10.1167/iovs.12-9745.

    PubMed 22761263
  • Perez Fornos A, Sommerhalder J, da Cruz L, Sahel JA, Mohand-Said S, Hafezi F, Pelizzone M. Temporal properties of visual perception on electrical stimulation of the retina. Invest Ophthalmol Vis Sci. 2012 May 4;53(6):2720-31. doi: 10.1167/iovs.11-9344.

    PubMed 22447863

Questions about this trial

What is this clinical trial studying?

A study testing Implant of Argus II Retinal Prosthesis in people with Retinitis Pigmentosa, run by Second Sight Medical Products at 12 sites in 5 countries.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 28 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 20 Jan 2022. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 12 sites in United States, United Kingdom, France, Mexico, Switzerland. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 30 Dec 2021). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Visual Acuity (time frame: 10 years). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT00407602, which also lists 13 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 20 Jan 2022.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in