Overactive BladderNCT00437528
Safety and Applicability Study of a Novel Heat Flow Sensor Unit for Measuring Urinary Bladder Capacity
An observational study of Overactive Bladder and Urge Urinary Incontinence, run by Rambam Health Care Campus at 1 site in Israel.
Trial summary
A novel sensor (a device named sensurine)was developed in the aim of real- time measurement of bladder volume. The sensurine device is a wearable, fully passive, non-invasive and compact heat flow sensor (patent pending) and electronic control unit. Read more about this study
- Who is it for?
- Ages 18 to 50 yearsHealthy volunteers can take part
- Conditions the study is for
- Overactive Bladder, Urge Urinary Incontinence, Enuresis
- How many people
- 10 participants
- Where
- 1 siteIsrael
- Study dates
- Nov 2006to Dec 2007
- Phase
- Observational
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 to 50 years
- All sexes
- Healthy volunteers accepted
- Adult
Who may be able to join
- All subjects will be aged 18-50, active, in good general health and without urinary complaints or documented urinary tract dysfunctions
Who cannot join
- Pregnancy, breast feeding, known allergic sensitivity to medical adhesive tapes, cognitive incompatibility and mental incompetence
Trial locations and contacts
1 site in 1 country.
Israel1 site, 1 recruiting
- Recruiting
Rambam Healthcare Campus
Haifa, Northern District, 31096
Ilan Gruenwald, MD
About this study
Official title: Phase I Study for Evaluating the Safety and Applicability of a Novel Heat Flow Sensor Unit for Measuring Urinary Bladder Capacity
A novel sensor (a device named sensurine)was developed in the aim of real- time measurement of bladder volume. The sensurine device is a wearable, fully passive, non-invasive and compact heat flow sensor (patent pending) and electronic control unit. The device will serve as a tool for managing and treating bladder dysfunction (by behavioral treatment), such as overactive bladder without incontinence, urge incontinence and voiding difficulties. In this phase I of the clinical trial the sensor unit will be tested for technological feasibility demonstration and for collecting design data.
Read the detailed description
The Primary objectives are to asses the safety and the physical performance of the SenseUrine's sensor unit and to validate the concept of determining bladder volume by measuring heat flow. The Secondary objective is to optimize the device accuracy Primary endpoints: To show that the sensor has no negative effect on the subject, to determine bladder urine volume, based on the sensor's signals, in an accuracy of ±10% compared to the actual volume. Secondary endpoint - to determine bladder urine volume, in an accuracy of ±10%, in different ambient conditions (temperature and clothing).
Inclusion criteria - All subjects will be aged 18-50, active, in good general health and without urinary complaints or documented urinary tract dysfunctions. Exclusion criteria - Pregnancy, breast feeding, known allergic sensitivity to medical adhesive tapes, cognitive incompatibility and mental incompetence.
On the first visit, the subject will sign an informed consent form. A medical intake including demographic data will be collected. Stage 1 will serve for basic testing of the sensor unit and for measuring the maximal and minimal bladder volume. The sensor will be attached to the lower abdomen wall and checked for functionality. Each subject will fill up his bladder (by drinking water) until sensing a strong urge to void. The subject will then void until he is empty into a measurement cup. Ultrasound measurement (US) of bladder volume, and sensor signal will be read and recorded by a standard miliVoltmeter just before and immediately after emptying the bladder. Sensor, US and cup measurement readings will not be done by the same investigator. Continuing to stage 2 will depend on the preliminary results (sensor readings compared to the measured volume and US readings) determining feasibility.. Stage 2 will serve for measuring the dynamics of bladder heat conductivity during the filling and voiding phases. On the 2nd visit the subject will fill up his bladder by gradually drinking 500CC of water (over 60-90 minutes) until sensing a strong urge to void. The subject will drink an additional 50-100CC of water and wait around 10 minutes in the aim of achieving maximal bladder capacity. The sensor signals will be recorded from the beginning of this bladder filling stage every 5-10 minutes. Then, the subject will be requested to void in a controlled manner and to intentionally stop urinating after emptying 50-100CC of urine (voided volume will be determined by the measurement cup). The controlled voiding of 50-100CC will be repeated until the bladder is empty. The sensor signals will be read and recorded during both stages. Visit 3 will be performed in the same manner but in different positions during measurements (supine-on the back, sides and front, sitting, and standing). Visit 4 (final visit) will serve for demonstrating the sensor unit's performance in different ambient temperature and clothing. Two subjects will be randomly selected to examine the sensor in an ambient of 15ºC (this will be controlled by an air-conditioned room with an external thermometer) for measuring the dynamic range (maximal and minimal bladder capacity). The same will be done at room temperature twice- with light and heavy clothing. The other 2 subjects will be randomly selected to examine the sensor's performance in an ambient of 40ºC (hot temperature will be achieved by an electric heater in a small room controlled with an external thermometer) for measuring the dynamic range (maximal and minimal bladder capacity). The same will be done at room temperature twice- with light and heavy clothing.
All output signals will be logged and processed to show the correlation between the calculated heat conductance of the bladder and the volume. A table consisting the calculated indexes of conductivity vs. bladder volumes will be designed.
General: 10 subjects will be recruited by advertisement with notification of some compensation for their time and travel to the study site. Every session will last about 4 hours long. In each working day 3-5 subjects will participate and the total trial period is expected to be approximately 20 working days.
- Conditions
- Overactive Bladder, Urge Urinary Incontinence, Enuresis, Self Clean Intermittent Catheterization
- Keywords
- Heat flow, Sensor, Overactive bladder, Urinary incontinence
- Medical subject headings
- Urinary Bladder, Overactive, Urinary Incontinence, Urge, Enuresis, Urinary Incontinence
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted21 Feb 2007 (estimated)
- Study startNov 2006
- Study completionDec 2007
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Other trials for Overactive BladderStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Observational
- Time perspective
- Prospective
- Enrolment
- 10 participants (estimated)
Sponsor and investigators
- Lead sponsor
- Rambam Health Care Campus
- Sponsor type
- University, hospital or other
- Investigators
- Yoram Vardi, Prof, Principal investigator, Rambam Health CampusIlan Gruenwald, MD, Study director, Rambam Health CampusIrena Papier, RN, MPA, Study director, Rambam Health Campus
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT00437528
- Sponsor's protocol ID
- 2456_CTIL
- Status last verified
- Feb 2007
- First submitted
- 19 Feb 2007
- Last updated
- 17 May 2007
- Results posted
- No
- Data monitoring committee
- No
Study at a glance
From the registry record, last updated by the sponsor on 17 May 2007. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Phase I Study for Evaluating the Safety and Applicability of a Novel Heat Flow Sensor Unit for Measuring Urinary Bladder Capacity
- ClinicalTrials.gov ID
- NCT00437528
- Acronym / sponsor ID
- 2456_CTIL
- Conditions
- Overactive Bladder, Urge Urinary Incontinence, Enuresis, Self Clean Intermittent Catheterization
- Keywords
- Heat flow, Sensor, Overactive bladder, Urinary incontinence
- Study type
- Observational
- Recruitment status
- Unknown status
- Enrolment
- 10 participants (estimated)
- Lead sponsor
- Rambam Health Care CampusUniversity, hospital or other
- Study start
- Nov 2006
- Study completion
- Dec 2007
- First posted
- 21 Feb 2007 (estimated)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
A novel sensor (a device named sensurine)was developed in the aim of real- time measurement of bladder volume. The sensurine device is a wearable, fully passive, non-invasive and compact heat flow sensor (patent pending) and electronic control unit. The device will serve as a tool for managing and treating bladder dysfunction (by behavioral treatment), such as overactive bladder without incontinence, urge incontinence and voiding difficulties. In this phase I of the clinical trial the sensor unit will be tested for technological feasibility demonstration and for collecting design data.
Detailed description
The Primary objectives are to asses the safety and the physical performance of the SenseUrine's sensor unit and to validate the concept of determining bladder volume by measuring heat flow. The Secondary objective is to optimize the device accuracy Primary endpoints: To show that the sensor has no negative effect on the subject, to determine bladder urine volume, based on the sensor's signals, in an accuracy of ±10% compared to the actual volume. Secondary endpoint - to determine bladder urine volume, in an accuracy of ±10%, in different ambient conditions (temperature and clothing).
Inclusion criteria - All subjects will be aged 18-50, active, in good general health and without urinary complaints or documented urinary tract dysfunctions. Exclusion criteria - Pregnancy, breast feeding, known allergic sensitivity to medical adhesive tapes, cognitive incompatibility and mental incompetence.
On the first visit, the subject will sign an informed consent form. A medical intake including demographic data will be collected. Stage 1 will serve for basic testing of the sensor unit and for measuring the maximal and minimal bladder volume. The sensor will be attached to the lower abdomen wall and checked for functionality. Each subject will fill up his bladder (by drinking water) until sensing a strong urge to void. The subject will then void until he is empty into a measurement cup. Ultrasound measurement (US) of bladder volume, and sensor signal will be read and recorded by a standard miliVoltmeter just before and immediately after emptying the bladder. Sensor, US and cup measurement readings will not be done by the same investigator. Continuing to stage 2 will depend on the preliminary results (sensor readings compared to the measured volume and US readings) determining feasibility.. Stage 2 will serve for measuring the dynamics of bladder heat conductivity during the filling and voiding phases. On the 2nd visit the subject will fill up his bladder by gradually drinking 500CC of water (over 60-90 minutes) until sensing a strong urge to void. The subject will drink an additional 50-100CC of water and wait around 10 minutes in the aim of achieving maximal bladder capacity. The sensor signals will be recorded from the beginning of this bladder filling stage every 5-10 minutes. Then, the subject will be requested to void in a controlled manner and to intentionally stop urinating after emptying 50-100CC of urine (voided volume will be determined by the measurement cup). The controlled voiding of 50-100CC will be repeated until the bladder is empty. The sensor signals will be read and recorded during both stages. Visit 3 will be performed in the same manner but in different positions during measurements (supine-on the back, sides and front, sitting, and standing). Visit 4 (final visit) will serve for demonstrating the sensor unit's performance in different ambient temperature and clothing. Two subjects will be randomly selected to examine the sensor in an ambient of 15ºC (this will be controlled by an air-conditioned room with an external thermometer) for measuring the dynamic range (maximal and minimal bladder capacity). The same will be done at room temperature twice- with light and heavy clothing. The other 2 subjects will be randomly selected to examine the sensor's performance in an ambient of 40ºC (hot temperature will be achieved by an electric heater in a small room controlled with an external thermometer) for measuring the dynamic range (maximal and minimal bladder capacity). The same will be done at room temperature twice- with light and heavy clothing.
All output signals will be logged and processed to show the correlation between the calculated heat conductance of the bladder and the volume. A table consisting the calculated indexes of conductivity vs. bladder volumes will be designed.
General: 10 subjects will be recruited by advertisement with notification of some compensation for their time and travel to the study site. Every session will last about 4 hours long. In each working day 3-5 subjects will participate and the total trial period is expected to be approximately 20 working days.
Study design
- Time perspective
- Prospective
Who may qualify
- Ages
- 18 to 50 years
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Adult
Key inclusion criteria (1 listed)
- All subjects will be aged 18-50, active, in good general health and without urinary complaints or documented urinary tract dysfunctions
Key exclusion criteria (1 listed)
- Pregnancy, breast feeding, known allergic sensitivity to medical adhesive tapes, cognitive incompatibility and mental incompetence
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Primary outcome
Not reported in the source record.
Results and safety
Results are not available in the source record.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 17 May 2007.
Related trials for Overactive Bladder
Recruiting
Safety and Performance of UCon Patch Electrode
Urinary Incontinence, Urinary Incontinence, Urge
Active, not recruitingPhase 2/3
Examining the Effects of Intra-detrusor Botox at Time of HoLEP in Men With Overactive Bladder Symptoms
Urinary Incontinence, Overactive Bladder Syndrome
Recruiting
UCon Treatment of Overactive Bladder (OAB) in Males
Urinary Incontinence, Urge Incontinence
Active, not recruitingObservational study
Overactive Bladder Symptoms and Urinary Incontinence In Women With Rheumatoid Arthritis
Rheumatoid Arthritis, Rheumatoid Arthritis (RA)
RecruitingObservational study
Predictive Value of Inflammatory Biomarkers for Mirabegron Response in Women With Overactive Bladder
Overactive Bladder (OAB)
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Yoram Vardi, Prof principal investigator
Rambam Health Campus
Ilan Gruenwald, MD study director
Rambam Health Campus
Irena Papier, RN, MPA study director
Rambam Health Campus
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
An observational study of Overactive Bladder and Urge Urinary Incontinence, run by Rambam Health Care Campus at 1 site in Israel.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 1 inclusion and 1 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: status not confirmed, as of the sponsor's last update on 17 May 2007. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Israel. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Results are not available in the source record.
How can a doctor refer a patient or contact the research team?
Contact Ilan Gruenwald, MD (00972-54-4-474341, i_gruenwald@rambam.health.gov.il), or a site near the patient. AzTrials does not enrol participants or take referrals.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 17 May 2007.