Atrial FibrillationPhase 4Avoid FFSNCT00512915
Avoid FFS - Use of the Atrial Pacemaker Lead 1699 With Very Short Tip Ring Spacing to Avoid Far Field Sensing
A Phase 4 study testing pacemaker implantation in people with Atrial Fibrillation and Bradycardia, run by Abbott Medical Devices at 10 sites in Germany.
Trial summary
The purpose of this study is to evaluate the possibility to program the shortest possible Post Ventricular Atrial Blanking Period (PVAB) and high sensitivity without getting inappropriate Mode Switch due to Far Field R-Wave sensing when using the new Tendril 1699T lead. Read more about this study
- Who is it for?
- Ages 18 years and older
- Conditions the study is for
- Atrial Fibrillation, Bradycardia
- What is being tested?
- pacemaker implantation
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 204 participants
- Where
- 10 sitesGermany
- Study dates
- Dec 2006to Aug 2008
- Phase
- Phase 4
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- Indication for dual chamber pacing
- Implantation of an Identity (ADX) DR or Victory or later pacemaker (St. Jude Medical)
- Bipolar atrial pacing electrode
- Age >= 18 years
Who cannot join
- Persistent / permanent Atrial Arrhythmia
- Pacemaker Revision
- Pregnancy
- Participation in another Study involving active implantable medical devices
- Unable to complete follow up
- Missing patient informed consent
- Cardiac surgery or myocardial infarction within the last 4 weeks
- Planned cardiac surgery within 3 months after enrollment
Trial locations and contacts
10 sites in 1 country.
10 sites in 1 country
Germany10 sites
Klinikum Coburg
Coburg, 96450
Klinik Fränkische Schweiz
Ebermannstadt, 91320
Kreisklinik Ebersberg
Ebersberg, 85560
Krankenhaus Waltershausen-Friedrichroda
Friedrichroda, 99894
Universitäres Herzzentrum Hamburg
Hamburg, 20246
Klinikum Memmingen
Memmingen, 87700
Kardiologische Praxis Dres. med. Bödigheimer / Mühling / Prof. Dr. med. Silber
München, 80331
Deutsches Herzzentrum Muenchen
Munich, 80636
Kreiskrankenhaus Ottweiler
Ottweiler, 66564
Universitätsklinikum Ulm
Ulm, 89081
About this study
Official title: Atriale Vorhofsonde 1699 Mit Sehr Kurzem Bipolabstand Zur Vermeidung Von Far Field Sensing
The purpose of this study is to evaluate the possibility to program the shortest possible Post Ventricular Atrial Blanking Period (PVAB) and high sensitivity without getting inappropriate Mode Switch due to Far Field R-Wave sensing when using the new Tendril 1699T lead. Comparison with Tendril 1688T or 1388T with optimized Post Ventricular Atrial Blanking (PVAB).
Read the detailed description
Far-field R-wave sensing (FFS) in the atrial channel of dual chamber pacemakers is a relevant source for inappropriate Mode Switch from the DDD mode to the DDI or VDI mode. Inappropriate loss of atrioventricular synchrony due to false positive Mode Switch is hemodynamically disadvantageous, may induce atrial tachyarrhythmias, can lead to pacemaker syndrome, and impairs the reliability of pacemaker Holter data. The aim of the study is to determine whether the use of a new atrial screw in lead with ultra short tip-ring distance of 1.1 mm (Study Group) eliminates the need for individual adjustment of the postventricular atrial blanking period (PVAB) based on an additional test in the Control Group with standard leads.
- Conditions
- Atrial Fibrillation, Bradycardia
- Being tested
- pacemaker implantation
- Keywords
- Farfield Sensing, Pacing lead, Postventricular atrial blanking, Bradycardia
- Medical subject headings
- Atrial Fibrillation, Bradycardia
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
pacemaker implantationDevice
Also called Active Arm: 1699T Tendril Optisense lead, Standard lead: 1688T Tendril SDX, 1788T Tendril ST, 1888T Tendril ST OPTIM
St. Jude Medical Dual Chamber Pacemaker model Identity (ADx) DR, Victory DR, or newer
| Study group | Role | What they receive |
|---|---|---|
| 1699T (Optisense) | Active comparator | Implantation of the Optisense Lead 1699T, programming of the shortest possible postventricular atrial blanking period (PVAB) Receives: Device: pacemaker implantation |
| Standard lead | Active comparator | Implantation of a standard bipolar atrial pacing lead. Optimization of the postventricular atrial blanking period (PVAB) after implantation. Receives: Device: pacemaker implantation |
What is being measured
The primary outcome is the main question the study is designed to answer.
Incidence of inappropriate mode switch (documented by stored EGM's) using the atrial lead Tendril 1699T with short post ventricular atrial blanking period (PVAB) compared to standard atrial lead (Tentril 1388 or 1688) with optimised PVAB
Measured: 1 and 3 months post implantation
Secondary outcomes (3)
Incidence of inappropriate mode switch in respect to lead position and ventricular stimulation frequency
Measured: 1 and 3 months post implantation
Atrial Flutter in stored EGM's
Measured: 1 and 3 months post implantation
2:1 lock in of Atrial Flutter
Measured: 1 and 3 months post implantation
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted8 Aug 2007 (estimated)
- Study startDec 2006
- Primary completionMay 2008
- Study completionAug 2008
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Publications (1)Papers the sponsor has linked to this study.
- Other trials for Atrial FibrillationStudies of the same condition, open and under way.
- More trials run by Abbott Medical DevicesThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase 4
- Primary purpose
- Prevention
- Allocation
- Randomised
- Intervention model
- Parallel groups
- Masking
- Single blind
- Who is masked
- Participant
- Enrolment
- 204 participants (actual)
Sponsor and investigators
- Lead sponsor
- Abbott Medical Devices
- Sponsor type
- Industry (pharma, biotech, devices)
- Investigators
- Christof Kolb, MD, Principal investigator, Deutsches Herzzentrum Muenchen
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT00512915
- Sponsor's protocol ID
- B84
- Acronym
- Avoid FFS
- Status last verified
- Feb 2019
- First submitted
- 7 Aug 2007
- Last updated
- 4 Feb 2019
- Results posted
- No
- Data monitoring committee
- No
Publications
Kolb C, Nolker G, Lennerz C, Jetter H, Semmler V, Purner K, Gutleben KJ, Reents T, Lang K, Lotze U; AVOID-FFS Investigators. Use of an atrial lead with very short tip-to-ring spacing avoids oversensing of far-field R-wave. PLoS One. 2012;7(6):e38277. doi: 10.1371/journal.pone.0038277. Epub 2012 Jun 22.
PubMed 22745661
Study at a glance
From the registry record, last updated by the sponsor on 4 Feb 2019. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Atriale Vorhofsonde 1699 Mit Sehr Kurzem Bipolabstand Zur Vermeidung Von Far Field Sensing
- ClinicalTrials.gov ID
- NCT00512915
- Acronym / sponsor ID
- Avoid FFS / B84
- Conditions
- Atrial Fibrillation, Bradycardia
- Keywords
- Farfield Sensing, Pacing lead, Postventricular atrial blanking, Bradycardia
- Phase
- Phase 4
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 204 participants (actual)
- Lead sponsor
- Abbott Medical DevicesIndustry (pharma, biotech, devices)
- Study start
- Dec 2006
- Primary completion
- May 2008 (actual)
- Study completion
- Aug 2008 (actual)
- First posted
- 8 Aug 2007 (estimated)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The purpose of this study is to evaluate the possibility to program the shortest possible Post Ventricular Atrial Blanking Period (PVAB) and high sensitivity without getting inappropriate Mode Switch due to Far Field R-Wave sensing when using the new Tendril 1699T lead. Comparison with Tendril 1688T or 1388T with optimized Post Ventricular Atrial Blanking (PVAB).
Detailed description
Far-field R-wave sensing (FFS) in the atrial channel of dual chamber pacemakers is a relevant source for inappropriate Mode Switch from the DDD mode to the DDI or VDI mode. Inappropriate loss of atrioventricular synchrony due to false positive Mode Switch is hemodynamically disadvantageous, may induce atrial tachyarrhythmias, can lead to pacemaker syndrome, and impairs the reliability of pacemaker Holter data. The aim of the study is to determine whether the use of a new atrial screw in lead with ultra short tip-ring distance of 1.1 mm (Study Group) eliminates the need for individual adjustment of the postventricular atrial blanking period (PVAB) based on an additional test in the Control Group with standard leads.
Study design
- Primary purpose
- Prevention
- Allocation
- Randomised
- Intervention model
- Parallel groups
- Masking
- Single blind (participant)
- Number of arms
- 2
| Arm | Type | Receives |
|---|---|---|
| 1699T (Optisense) | Active comparator | Device: pacemaker implantationImplantation of the Optisense Lead 1699T, programming of the shortest possible postventricular atrial blanking period (PVAB) |
| Standard lead | Active comparator | Device: pacemaker implantationImplantation of a standard bipolar atrial pacing lead. Optimization of the postventricular atrial blanking period (PVAB) after implantation. |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (4 listed)
- Indication for dual chamber pacing
- Implantation of an Identity (ADX) DR or Victory or later pacemaker (St. Jude Medical)
- Bipolar atrial pacing electrode
- Age >= 18 years
Key exclusion criteria (8 listed)
- Persistent / permanent Atrial Arrhythmia
- Pacemaker Revision
- Pregnancy
- Participation in another Study involving active implantable medical devices
- Unable to complete follow up
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| pacemaker implantationAlso: Active Arm: 1699T Tendril Optisense lead, Standard lead: 1688T Tendril SDX, 1788T Tendril ST, 1888T Tendril ST OPTIMSt. Jude Medical Dual Chamber Pacemaker model Identity (ADx) DR, Victory DR, or newer | Device | 1699T (Optisense), Standard lead |
Primary outcome
Incidence of inappropriate mode switch (documented by stored EGM's) using the atrial lead Tendril 1699T with short post ventricular atrial blanking period (PVAB) compared to standard atrial lead (Tentril 1388 or 1688) with optimised PVAB
Time frame: 1 and 3 months post implantation
3 secondary outcomes
Incidence of inappropriate mode switch in respect to lead position and ventricular stimulation frequency
Time frame: 1 and 3 months post implantation
Atrial Flutter in stored EGM's
Time frame: 1 and 3 months post implantation
2:1 lock in of Atrial Flutter
Time frame: 1 and 3 months post implantation
Results and safety
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 4 Feb 2019.
- Derived from this study: PubMed 22745661Kolb C, Nolker G, Lennerz C, Jetter H, Semmler V, Purner K, Gutleben KJ, Reents T, Lang K, Lotze U; AVOID-FFS Investigators. Use of an atrial lead with very short tip-to-ring spacing avoids oversensing of far-field R-wave. PLoS One. 2012;7(6):e38277. doi: 10.1371/journal.pone.0038277. Epub 2012 Jun 22.
Related trials for Atrial Fibrillation
Active, not recruiting
VOLT CE Mark Study
Atrial Arrhythmia, Atrial Fibrillation
RecruitingObservational study
AI-Based Cardiac Analysis Using 12-Lead and 1-Lead ECG
Heart Failure, Atrial Fibrillation (AF)
RecruitingObservational study
Observational Study of Cardiac Arrhythmias During Treatment With BTK Inhibitors or Venetoclax
Chronic Lymphocytic Leukemia (CLL), Waldenstr(SqrRoot)(Delta)m s Macroglobulinemia
Not yet recruiting
CATCH AF - Atrial Fibrillation Detection in Cryptogenic Stroke
Atrial Fibrillation (AF)
Active, not recruiting
ARIADNE Clinical Trial
Persistent Atrial Fibrillation
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Christof Kolb, MD principal investigator
Deutsches Herzzentrum Muenchen
10 sites in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A Phase 4 study testing pacemaker implantation in people with Atrial Fibrillation and Bradycardia, run by Abbott Medical Devices at 10 sites in Germany.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 4 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 4 Feb 2019. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 10 sites in Germany. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
What does this phase mean?
Phase 4 studies take place after a treatment is approved, to learn more about its long-term benefits and risks in everyday use.
Not every study follows the same path, and some studies (observational studies, device studies) have no phase.
Are results available for this study?
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Incidence of inappropriate mode switch (documented by stored EGM's) using the atrial lead Tendril 1699T with short post ventricular atrial blanking period (PVAB) compared to standard atrial lead (Tentril 1388 or 1688) with optimised PVAB (time frame: 1 and 3 months post implantation). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT00512915, which also lists 1 linked publication.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 4 Feb 2019.