Safety and Effectiveness of the MED-EL Electric-Acoustic System

A study testing Electric Acoustic System in people with Hearing Loss, run by Med-El Corporation at 14 sites in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this investigation is to demonstrate the safety and effectiveness of the MED-EL Electric-Acoustic System (EAS), a medical device that combines the use of a cochlear implant with an external electric-acoustic processor designed to provide benefit in speech perception and sound quality to individuals with… Read more about this study

Who is it for?
Ages 18 to 70 years
Conditions the study is for
Hearing Loss
What is being tested?
Electric Acoustic System
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
73 participants
Where
14 sitesUnited States
Study dates
Feb 2007to Feb 2016
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 70 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Audiologic tests suggest a moderate sloping to severe/profound sensorineural hearing loss in the ear to be implanted. The non-implanted ear may fall outside of these criteria; however, threshold levels cannot be better than the indication criteria (for example, the non-implanted ear must be 60 dB or poorer at 1000 Hz, 70 dB or poorer at 2000-8000 Hz).
  • Pure-tone air-conduction threshold levels for the ear to be implanted shall fall at or within the following levels:
  • 250 Hz - 500 Hz hearing loss less than or equal to 65 dB HL 750 Hz hearing loss less than or equal to 75 dB HL 1000-1500 Hz hearing loss less than or equal to 60 dBHL (ie. 60-110+ dB HL) 2000 Hz - 8000 Hz hearing loss less than or equal to 70 dB HL
  • Pure-tone air-conduction threshold levels shall fall at or within the levels listed in the following chart Frequency (Hz): 250 500 750 1000 1500 2000 4000 8000 Lower Limit: 10 10 10 60 60 70 70 70 Upper Limit: 65 65 75 *110+ *110+ *110+ *110+ *90+
  • Pure-tone air-conduction thresholds for both ears are within 20 dB of each other at 250, 500 and 1000 Hz.
  • Air-bone gap at 500, 1000, 2000 and 4000 Hz should be <10 dB at two or more of these frequencies.
Show all 17Show fewer
  • Minimal benefit from optimally fit hearing aid/s, preferably bilateral, with monosyllabic word scores in quiet of ≤50% in the best-aided condition.
  • Normal middle ear anatomy and function (based on clinical assessment of tympanometry and acoustic reflex results). No prior middle ear surgery or history of postadolescent, chronic middle ear infections or inner ear disorders (i.e. vertigo or Meniere's syndrome).
  • No evidence that hearing loss origin is retrocochlear.
  • Current user of bilateral acoustic hearing aids for at least 3 months. Note: If the subject has not been a successful hearing aid user (i.e. improper fit, feedback and/or discomfort from high frequency amplification), he/she must complete a hearing aid trial in the ear to be implanted for at least 1 month. If the ear to be implanted is poorer than the contralateral ear, then the subject must complete a hearing aid trial in both ears for at least 1 month to ensure that he/she is tested in the best-aided condition.
  • Adults 18-70 years of age at time of implantation.
  • Persons who currently exhibit sufficient understanding and communicative skills to comprehend general conversation in an "oral/aural mode" through normal conversation channels.
  • English as primary language.
  • Appropriate motivation and expectation levels.
  • Expanded Criteria:
  • Study Arm 2 candidates are those who meet the inclusion criteria listed above with the exception that:
  • In the ear to be implanted, the pure-tone air conduction threshold(s) at 250, 500, and/or 750 Hz will be greater than or equal to 0 dB HL and less than 10 dB HL, and/or In the ear to be implanted, the pure-tone air conduction threshold(s) at 1000 and/or 1500 Hz will be greater than or equal to 0 dB HL and less than 60 dB HL and/or In their best-aided condition, they score 51-60% on monosyllabic word scores.

Who cannot join

  • Conductive, retrocochlear or central auditory disorders.
  • Hearing loss in the ear to be implanted that has demonstrated a recent fluctuation at two or more frequencies of 15 dB in either direction in the last 2 years as demonstrated by serial audiograms. Note: In cases where 2-year documentation is missing, documentation of at least 18 months is necessary along with patient report of no fluctuation in the past 2 years. If 18-month data is unavailable, the patient must be monitored until 18-month data is available.
  • Any physical, psychological, or emotional disorder that would interfere with surgery or the ability to perform on test and rehabilitation procedures.
  • Developmental delays or organic brain dysfunction.
  • Physical or geographic limitations that may interfere with the completion of scheduled follow-up evaluations.
  • Skin or scalp conditions that could preclude magnetic attachment of the speech processor or use of the acoustic hearing aid.

Trial locations and contacts

14 sites in 1 country.

14 sites in 1 country

United States14 sites
  • University of Michigan

    Ann Arbor, Michigan, 48108

  • University of North Carolina Hospital

    Chapel Hill, North Carolina, 27514

  • University of Texas Southwestern Medical Center

    Dallas, Texas, 75390

  • Duke University Medical Center

    Durham, North Carolina, 27710

  • Indiana University School of Medicine

    Indianapolis, Indiana, 46202

  • University of Kansas Medical Center

    Kansas City, Kansas, 66160

  • University of Miami

    Miami, Florida, 33136

  • Medical College of Wisconsin

    Milwaukee, Wisconsin, 53226

  • New York Eye & Ear

    New York, New York, 10003

  • Boys Town

    Omaha, Nebraska, 68131

  • Hospital of the University of Pennsylvania

    Philadelphia, Pennsylvania, 19104

  • Oregon Health Sciences Center

    Portland, Oregon, 97239

  • Swedish Medical Center

    Seattle, Washington, 98122

  • Stanford University Medical Center

    Stanford, California, 94035

Other trials in United States

About this study

Official title: The MED-EL EAS (Electric-Acoustic System) Using the PULSARCI100 FLEXeas / SONATATI100 FLEXeas and the DUET Speech Processor

The purpose of this investigation is to demonstrate the safety and effectiveness of the MED-EL Electric-Acoustic System (EAS), a medical device that combines the use of a cochlear implant with an external electric-acoustic processor designed to provide benefit in speech perception and sound quality to individuals with a sensorineural hearing loss with minimal changes in residual hearing. The acoustic component of the processor will aid residual acoustic hearing in low frequency ranges, while the cochlear implant and electric component of the processor will be used to electrically stimulate the auditory nerve across a wide range of frequencies necessary for speech perception.

Conditions
Hearing Loss
Being tested
Electric Acoustic System
Keywords
sensorineural hearing loss, ski slope audiogram, electric acoustic stimulation
Medical subject headings
Hearing Loss, Hearing Loss, Sensorineural

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Electric Acoustic SystemDevice

    Also called EAS, Electric Acoustic Systems

    Combination of a cochlear implant and a hearing aid

Study groups in NCT00747435
Study groupRoleWhat they receive
Original Audiological CriteriaExperimentalInclusion Criteria: Pure-tone air-conduction threshold levels shall fall at or within the levels listed in the following chart. Frequency (Hz): 250 500 750 1000 1500 2000 4000 8000 Lower Limit: 10 10 10 60 60 70 70 70 Upper Limit: 65 65 75 *110+ *110+*110+ *110+ *90+ Minimal benefit from optimally fit hearing aid/s, preferably bilateral, with monosyllabic word scores in quiet of ≤50% in the best-aided condition.
Receives: Device: Electric Acoustic System
Expanded Audiological CriteriaExperimentalPure-tone air-conduction threshold levels shall fall at or within the levels listed in the following chart. Frequency (Hz): 250 500 750 1000 1500 2000 4000 8000 Lower Limit: 10 10 10 60 60 70 70 70 Upper Limit: 65 65 75 *110+ *110+*110+ *110+ *90+ Minimal benefit from optimally fit hearing aid/s, preferably bilateral, with monosyllabic word scores in quiet of ≤50% in the best-aided condition. Study Arm 2 candidates are those who meet the inclusion criteria listed above with the exception that: In the ear to be implanted, the pure-tone air conduction threshold(s) at 250, 500, and/or 750 Hz will be greater than or equal to 0 dB HL and less than 10 dB HL, and/or In the ear to be implanted, the pure-tone air conduction threshold(s) at 1000 and/or 1500 Hz will be greater than or equal to 0 dB HL and less than 60 dB HL and/or In their best-aided condition, they score 51-60% on monosyllabic word scores.
Receives: Device: Electric Acoustic System

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Improvement in Speech Perception in Noise With EAS When Compared to the Preoperative Hearing Aid Alone Condition.

    CUNY sentences in noise are scored as the percent correct of words in each sentence. The total percent correct for preoperative hearing aid use was subtracted from the total percent correct for EAS at 12 months giving a percentage point improvement in speech perception with EAS.

    Measured: 12 months post initial activation

Secondary outcomes (4)
  1. Improvement in Speech Perception in Noise With EAS When Compared to the Cochlear Implant Alone Condition

    CUNY sentences in noise are scored as the percent correct of words in each sentence. The total percent correct for the CI Alone condition was subtracted from the total percent correct for the EAS condition at 12 months giving an improvement in speech perception with EAS compared to CI Alone.

    Measured: 12 months initial activation

  2. Improvement on Speech Perception in the CI-only Condition as Compared to the Preoperative With Hearing Aid Condition.

    CNC words are scored as the percent correct out of 50 words. The percent correct for preoperative hearing aid use was subtracted from the percent correct for CI Alone at 12 months giving an improvement on speech perception for CI Alone compared to preoperative hearing aids.

    Measured: 12 months post initial activation

  3. Increased Benefit With EAS as Compared to Their Preoperative Hearing Aid Condition as Measured by the APHAB Questionnaire.

    A lower score on the APHAB indicates a better performance. The global score for preoperative hearing aid use was subtracted from the global score for EAS at 12 months giving an improvement on the APHAB with EAS.

    Measured: 12 months post initial activation

  4. Increase in Satisfaction With EAS Compared to Preoperative Hearing Aid Condition as Measured by the HDSS.

    Satisfaction is ranked from 1-5, with 1 being very dissatisfied and 5 being very satisfied. Data reported is the percentage of subjects who experienced an increase in satisfaction when using EAS compared to preoperative hearing aids.

    Measured: 12 months post initial activation

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted5 Sep 2008 (estimated)
  2. Study startFeb 2007
  3. Primary completionFeb 2016
  4. Study completionFeb 2016

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Non-randomised
Intervention model
Single group
Masking
None (open label)
Enrolment
73 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT00747435
Sponsor's protocol ID
G040002
Acronym
EAS
Status last verified
Apr 2017
First submitted
4 Sep 2008
Last updated
17 May 2017
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes

Publications

  • Skarzynski H, Lorens A, Piotrowska A, Anderson I. Preservation of low frequency hearing in partial deafness cochlear implantation (PDCI) using the round window surgical approach. Acta Otolaryngol. 2007 Jan;127(1):41-8. doi: 10.1080/00016480500488917.

    PubMed 17364328
  • Skarzynski H, Lorens A, Piotrowska A, Anderson I. Partial deafness cochlear implantation provides benefit to a new population of individuals with hearing loss. Acta Otolaryngol. 2006 Sep;126(9):934-40. doi: 10.1080/00016480600606632.

    PubMed 16864490
  • Skarzynski H, Lorens A, Piotrowska A. [New method of partial deafness treatment]. Otolaryngol Pol. 2004;58(4):811-6. Polish.

    PubMed 15603395
  • Skarzynski H, Lorens A, Piotrowska A. A new method of partial deafness treatment. Med Sci Monit. 2003 Apr;9(4):CS20-4.

    PubMed 12709676
  • Kiefer J, Pok M, Adunka O, Sturzebecher E, Baumgartner W, Schmidt M, Tillein J, Ye Q, Gstoettner W. Combined electric and acoustic stimulation of the auditory system: results of a clinical study. Audiol Neurootol. 2005 May-Jun;10(3):134-44. doi: 10.1159/000084023. Epub 2005 Feb 17.

    PubMed 15724084
  • Kiefer J, Gstoettner W, Baumgartner W, Pok SM, Tillein J, Ye Q, von Ilberg C. Conservation of low-frequency hearing in cochlear implantation. Acta Otolaryngol. 2004 Apr;124(3):272-80. doi: 10.1080/00016480310000755a.

    PubMed 15141755
  • von Ilberg C, Kiefer J, Tillein J, Pfenningdorff T, Hartmann R, Sturzebecher E, Klinke R. Electric-acoustic stimulation of the auditory system. New technology for severe hearing loss. ORL J Otorhinolaryngol Relat Spec. 1999 Nov-Dec;61(6):334-40. doi: 10.1159/000027695.

    PubMed 10545807
  • Gstoettner WK, Helbig S, Maier N, Kiefer J, Radeloff A, Adunka OF. Ipsilateral electric acoustic stimulation of the auditory system: results of long-term hearing preservation. Audiol Neurootol. 2006;11 Suppl 1:49-56. doi: 10.1159/000095614. Epub 2006 Oct 6.

    PubMed 17063011
  • Gstoettner W, Kiefer J, Baumgartner WD, Pok S, Peters S, Adunka O. Hearing preservation in cochlear implantation for electric acoustic stimulation. Acta Otolaryngol. 2004 May;124(4):348-52. doi: 10.1080/00016480410016432.

    PubMed 15224851
  • Gstoettner WK, van de Heyning P, O'Connor AF, Morera C, Sainz M, Vermeire K, Mcdonald S, Cavalle L, Helbig S, Valdecasas JG, Anderson I, Adunka OF. Electric acoustic stimulation of the auditory system: results of a multi-centre investigation. Acta Otolaryngol. 2008 Sep;128(9):968-75. doi: 10.1080/00016480701805471.

    PubMed 19086194

Questions about this trial

What is this clinical trial studying?

A study testing Electric Acoustic System in people with Hearing Loss, run by Med-El Corporation at 14 sites in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 17 inclusion and 6 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 17 May 2017. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 14 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 17 May 2017). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Improvement in Speech Perception in Noise With EAS When Compared to the Preoperative Hearing Aid Alone Condition. (time frame: 12 months post initial activation). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT00747435, which also lists 10 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 17 May 2017.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in