Respiratory InsufficiencyNCT00810498
Trilogy Comparison Study - Adults
A study testing Trilogy in people with Respiratory Insufficiency and Respiratory Failure, run by Philips Respironics at 1 site in Canada.
Trial summary
The objective of this study (Part II) will be to compare the performance of the Trilogy device and a conventional ventilator (participant's current device). Both devices will be set to the participants' current mode and prescription settings. Read more about this study
- Who is it for?
- Ages 21 to 85 years
- Conditions the study is for
- Respiratory Insufficiency, Respiratory Failure
- What is being tested?
- Trilogy
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 11 participants
- Where
- 1 siteCanada
- Study dates
- 12 Mar 2008to 5 Feb 2009
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 21 to 85 years
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- Age > 21 years of age; < 85 years of age
- Able to follow instructions
- Able to provide informed consent
- Considered clinically stable, with low flow oxygen requirements less than 15 LPM (or FIO2 less than 60 %) and PEEP no greater than 10 cm H2O.
- Requiring mechanical ventilation through nasal/facial mask,tracheostomy or endotracheal tube (ET Tube)
Who cannot join
- Clinically unstable, i.e.,
- Acute Respiratory Failure
- Participants with refractory hypotension defined as systolic blood pressure less than 90 mm Hg despite inotropic agents),
- Uncontrolled cardiac ischemia or arrhythmias,
- or as otherwise determined inappropriate for the study as determined by the investigator.
- Patients suffering from metastatic or terminal cancer and patients with do-not-resuscitate orders
- Pregnancy
Trial locations and contacts
1 site in 1 country.
About this study
Official title: Comparing Known Modes of Ventilation Delivered by Trilogy Versus Conventional Mechanical Ventilators
The objective of this study (Part II) will be to compare the performance of the Trilogy device and a conventional ventilator (participant's current device). Both devices will be set to the participants' current mode and prescription settings.
Read the detailed description
- Primary Hypothesis and end-point In patients with respiratory insufficiency, when compared to conventional ventilators, the Trilogy device achieves comparable level of gas exchange (measured as partial pressure of CO2 in arterial or venous blood)and SpO2 levels during various modes of ventilation.
- Secondary Hypothesis and end-points In patients with respiratory insufficiency, when compared to conventional ventilators, the Trilogy device achieves comparable breathing pattern (tidal volume, respiratory rate, minute ventilation), hemodynamics (heart rate), and subjective measures of breathing comfort (measured by modified Borg dyspnea scale).
- Conditions
- Respiratory Insufficiency, Respiratory Failure
- Being tested
- Trilogy, Standard of Care
- Medical subject headings
- Respiratory Insufficiency
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
TrilogyDevice
Exposure for one hour on the Trilogy ventilator
Standard of CareDevice
Exposure to participants current ventilator
| Study group | Role | What they receive |
|---|---|---|
| Trilogy | Experimental | Trilogy Device Receives: Device: Trilogy |
| Standard of Care | Active comparator | Participants prescribed ventilator Receives: Device: Standard of Care |
What is being measured
The primary outcome is the main question the study is designed to answer.
Comparable Level of Gas Exchange (Measured as Partial Pressure of CO2 in Arterial or Venous Blood).
Blood gas was obtained for this study one hour after use of each device.
Measured: After one hour of ventilator use
Comparable Level of Gas Exchange SpO2 Levels During Various Modes of Ventilation.
Pulse Oximetry was measured after one hour of use of each ventilator.
Measured: After one hour of ventilator use
Secondary outcomes (3)
Tidal Volume
Tidal volume is the amount of air that moves in or out of the lungs with each respiratory cycle. It measures around 500 mL in an average healthy adult male and approximately 400 mL in a healthy female. Average tidal volume was compare after a hour of use of each device.
Measured: One hour during ventilator use.
Minute Ventilation
Minute ventilation is the volume of gas inhaled or exhaled from a person's lungs per minute. This was measured after one hour of use of each device.
Measured: One hour during ventilator use.
Hemodynamics (Heart Rate)
Number of beats per minute. Measured after one hour of use of each device.
Measured: One hour during ventilator use.
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted18 Dec 2008 (estimated)
- Study start12 Mar 2008
- Primary completion5 Feb 2009
- Study completion5 Feb 2009
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Other trials for Respiratory InsufficiencyStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Treatment
- Allocation
- Randomised
- Intervention model
- Crossover
- Masking
- Single blind
- Who is masked
- Participant
- Enrolment
- 11 participants (actual)
Sponsor and investigators
- Lead sponsor
- Philips Respironics
- Sponsor type
- Industry (pharma, biotech, devices)
- Responsible party
- Sponsor
- Investigators
- Steven Mink, MD, Principal investigator, University of Manitoba
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT00810498
- Sponsor's protocol ID
- MR-0609-TRI2-SS
- Status last verified
- Aug 2026
- First submitted
- 17 Dec 2008
- Last updated
- 21 Sep 2026
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- Yes
Study at a glance
From the registry record, last updated by the sponsor on 21 Sep 2026. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Comparing Known Modes of Ventilation Delivered by Trilogy Versus Conventional Mechanical Ventilators
- ClinicalTrials.gov ID
- NCT00810498
- Acronym / sponsor ID
- MR-0609-TRI2-SS
- Conditions
- Respiratory Insufficiency, Respiratory Failure
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 11 participants (actual)
- Lead sponsor
- Philips RespironicsIndustry (pharma, biotech, devices)
- Study start
- 12 Mar 2008 (actual)
- Primary completion
- 5 Feb 2009 (actual)
- Study completion
- 5 Feb 2009 (actual)
- First posted
- 18 Dec 2008 (estimated)
- Results
- Posted 3 Feb 2021
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The objective of this study (Part II) will be to compare the performance of the Trilogy device and a conventional ventilator (participant's current device). Both devices will be set to the participants' current mode and prescription settings.
Detailed description
- Primary Hypothesis and end-point In patients with respiratory insufficiency, when compared to conventional ventilators, the Trilogy device achieves comparable level of gas exchange (measured as partial pressure of CO2 in arterial or venous blood)and SpO2 levels during various modes of ventilation.
- Secondary Hypothesis and end-points In patients with respiratory insufficiency, when compared to conventional ventilators, the Trilogy device achieves comparable breathing pattern (tidal volume, respiratory rate, minute ventilation), hemodynamics (heart rate), and subjective measures of breathing comfort (measured by modified Borg dyspnea scale).
Study design
- Primary purpose
- Treatment
- Allocation
- Randomised
- Intervention model
- Crossover
- Masking
- Single blind (participant)
- Number of arms
- 2
| Arm | Type | Receives |
|---|---|---|
| Trilogy | Experimental | Device: TrilogyTrilogy Device |
| Standard of Care | Active comparator | Device: Standard of CareParticipants prescribed ventilator |
Who may qualify
- Ages
- 21 to 85 years
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (5 listed)
- Age > 21 years of age; < 85 years of age
- Able to follow instructions
- Able to provide informed consent
- Considered clinically stable, with low flow oxygen requirements less than 15 LPM (or FIO2 less than 60 %) and PEEP no greater than 10 cm H2O.
- Requiring mechanical ventilation through nasal/facial mask,tracheostomy or endotracheal tube (ET Tube)
Key exclusion criteria (7 listed)
- Clinically unstable, i.e.,
- Patients suffering from metastatic or terminal cancer and patients with do-not-resuscitate orders
- Pregnancy
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| TrilogyExposure for one hour on the Trilogy ventilator | Device | Trilogy |
| Standard of CareExposure to participants current ventilator | Device | Standard of Care |
Primary outcome
Comparable Level of Gas Exchange (Measured as Partial Pressure of CO2 in Arterial or Venous Blood).
Time frame: After one hour of ventilator use
Blood gas was obtained for this study one hour after use of each device.
Comparable Level of Gas Exchange SpO2 Levels During Various Modes of Ventilation.
Time frame: After one hour of ventilator use
Pulse Oximetry was measured after one hour of use of each ventilator.
3 secondary outcomes
Tidal Volume
Time frame: One hour during ventilator use.
Minute Ventilation
Time frame: One hour during ventilator use.
Hemodynamics (Heart Rate)
Time frame: One hour during ventilator use.
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 21 Sep 2026.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
Related trials for Respiratory Insufficiency
Not yet recruiting
Modified Inhaled Nitric Oxide in Ventilated Patients With Acute Respiratory Failure
Acute Respiratory Failure (ARF), Nitric Oxide
Not yet recruitingObservational study
Impact of Inspiratory Effort and Patient-Ventilator Interaction During the Transition Phase
Mechanically Ventilated Patients, Acute Respiratory Failure (ARF)
Not yet recruitingObservational study
Development and Prospective Validation of the AutoEIT Foundation Model for ARDS
ARDS (Acute Respiratory Distress Syndrome), Respiratory Failure
Recruiting
Humidification Effects on Airway Complications and Clinical Outcomes in Trauma ICU Patients on Mechanical Ventilation
Respiratory Insufficiency, Wounds and Injuries
Not yet recruitingObservational study
Diaphragm Ultrasound and Rapid Shallow Breathing Index for Predicting Extubation Success
Ventilator Weaning, Respiratory Insufficiency
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Steven Mink, MD principal investigator
University of Manitoba
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Trilogy in people with Respiratory Insufficiency and Respiratory Failure, run by Philips Respironics at 1 site in Canada.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 5 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 21 Sep 2026. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Canada. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 3 Feb 2021). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Comparable Level of Gas Exchange (Measured as Partial Pressure of CO2 in Arterial or Venous Blood). (time frame: After one hour of ventilator use). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.