AzTrials
AzTrials

Sudden Cardiac DeathPROMISENCT00874445

Comparison of 50% Tilt and Tuned Waveforms in Single-Coil Active Can Configuration

An observational study of Sudden Cardiac Death, run by Abbott Medical Devices at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to compare the defibrillation efficacy between the 50/50% tilt biphasic waveform and the Tuned biphasic waveform in a single coil active can configuration. Read more about this study

Who is it for?
Ages 18 years and olderHealthy volunteers can take part
Conditions the study is for
Sudden Cardiac Death
What is being tested?
ICD/ CRT-D
Type of treatment
Medical device
How many people
77 participants
Where
1 siteUnited States
Study dates
Mar 2009to Sep 2010
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Study population: Patients with ICD/CRT-D indications

Who may be able to join

  • Patient meets standard indication for an ICD/ CRT-D and will be implanted with an FDA approved SJM ICD/ CRT-D and compatible defibrillation lead system in the left pectoral region.
  • Patient is able to tolerate DFT testing.

Who cannot join

  • Patient has the pulse generator on the right side.
  • Patient is pregnant.
  • Patient is less than 18 years old.
  • SVC coil was turned ON during DFT testing

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Michael Gold

    Charleston, South Carolina, 29425

Other trials in United States

About this study

Official title: Prospective Comparison of 50/50% Tilt and Tuned Defibrillation Waveforms in a Single Coil Active Can Configuration

The purpose of this study is to compare the defibrillation efficacy between the 50/50% tilt biphasic waveform and the Tuned biphasic waveform in a single coil active can configuration.

Read the detailed description

There has been one prospective study that found some benefit of using tuned waveforms over 50/50% tilt waveforms.11 Other studies have failed to show much change in defibrillation efficacy with different duration biphasic waveforms. A vast majority of patients in these studies were tested with dual coil leads. Although dual coil leads can lower defibrillation thresholds, it is known that dual coil leads can make the lead extraction difficult, lead to venous obstruction and are more prone to failure. Results from a recent study showed that that a single coil configuration is adequate in a vast majority of patients when DFT testing was performed with tuned waveforms. This makes the use of single-coil leads during ICD/ CRT-D implants an attractive option.

Theoretical analysis has shown that difference between the tuned and 50% tilt waveforms is amplified at high impedance levels with tuned waveforms being superior (unpublished). Accordingly, this study has been designed to prospectively compare the DFT estimates with tuned and 50% tilt waveforms in left-sided, active pectoral defibrillation lead systems when the SVC coil has been turned OFF or is not part of the shocking circuit.

Being tested
ICD/ CRT-D
Medical subject headings
Death, Sudden, Cardiac

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • ICD/ CRT-DDevice

    Patients in experimental group will have their ICDs programmed to Tuned waveform and those in the control group will have their ICDs programmed to Fixed Tilt waveform.

Study groups in NCT00874445
Study groupRoleWhat they receive
ICD shocks programmed to Tuned WaveformICD shocks programmed to Tuned Waveform
Receives: Device: ICD/ CRT-D
ICD shocks programmed to Fixed Tilt WaveformICD shocks programmed to Fixed Tilt Waveform
Receives: Device: ICD/ CRT-D

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. True DFTs in fixed tilt and tuned waveforms obtained in volts (V)

    Measured: Implant

Secondary outcomes (1)
  1. True DFTs in fixed tilt and tuned waveforms obtained in joules (J)

    Measured: Implant

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted2 Apr 2009 (estimated)
  2. Study startMar 2009
  3. Primary completionSep 2010
  4. Study completionSep 2010

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Cohort
Time perspective
Prospective
Enrolment
77 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT00874445
Sponsor's protocol ID
CRD 475
Acronym
PROMISE
Status last verified
Feb 2019
First submitted
31 Mar 2009
Last updated
4 Feb 2019
Results posted
No
Data monitoring committee
No

Publications

  • Gold M, Val-Mejias J, Leman RB, Tummala R, Goyal S, Kluger J, Kroll M, Oza A. Optimization of superior vena cava coil position and usage for transvenous defibrillation. Heart Rhythm. 2008 Mar;5(3):394-9. doi: 10.1016/j.hrthm.2007.12.001. Epub 2007 Dec 5.

    PubMed 18313597
  • Natarajan S, Henthorn R, Burroughs J, Esberg D, Zweibel S, Ross T, Kroll M, Gianola D, Oza A. "Tuned" defibrillation waveforms outperform 50/50% tilt defibrillation waveforms: a randomized multi-center study. Pacing Clin Electrophysiol. 2007 Jan;30 Suppl 1:S139-42. doi: 10.1111/j.1540-8159.2007.00624.x.

    PubMed 17302691
  • Mouchawar G, Kroll M, Val-Mejias JE, Schwartzman D, McKenzie J, Fitzgerald D, Prater S, Katcher M, Fain E, Syed Z. ICD waveform optimization: a randomized, prospective, pair-sampled multicenter study. Pacing Clin Electrophysiol. 2000 Nov;23(11 Pt 2):1992-5. doi: 10.1111/j.1540-8159.2000.tb07070.x.

    PubMed 11139975
  • Gold MR, Val-Mejias J, Cuoco F, Siddiqui M. Comparison of fixed tilt and tuned defibrillation waveforms: the PROMISE study. J Cardiovasc Electrophysiol. 2013 Mar;24(3):323-7. doi: 10.1111/jce.12041. Epub 2012 Dec 4.

    PubMed 23210764
  • Gold MR, Yu Y, Singh JP, Birgersdotter-Green U, Stein KM, Wold N, Meyer TE, Ellenbogen KA. Effect of Interventricular Electrical Delay on Atrioventricular Optimization for Cardiac Resynchronization Therapy. Circ Arrhythm Electrophysiol. 2018 Aug;11(8):e006055. doi: 10.1161/CIRCEP.117.006055.

    PubMed 30354310

Questions about this trial

What is this clinical trial studying?

An observational study of Sudden Cardiac Death, run by Abbott Medical Devices at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 2 inclusion and 4 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 4 Feb 2019. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

True DFTs in fixed tilt and tuned waveforms obtained in volts (V) (time frame: Implant). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT00874445, which also lists 5 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 4 Feb 2019.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in