Diabetes Mellitus, Type 1NCT00928057

Comparison of Glycemic Control Among Diabetics Using Three Different Pen Needles

A study testing 4mm x 32G pen needle, 8mm x 31G pen needle and 5mm x 31G pen needle in people with Diabetes Mellitus, Type 1 and Diabetes Mellitus, Type 2, run by Becton, Dickinson and Company at 4 sites in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Anxiety about needles is a commonly expressed concern by diabetics about beginning insulin therapy. A shorter, thinner pen needle that delivers insulin with the safety and efficacy profile of currently marketed pen needles may appeal to many diabetic patients as the new needle may be perceived as less intimidating and… Read more about this study

Who is it for?
Ages 18 to 75 years
What is being tested?
4mm x 32G pen needle, 8mm x 31G pen needle and 5mm x 31G pen needle
Could I receive a placebo?
No, everyone receives an active treatment
How many people
173 participants
Where
4 sitesUnited States
Study dates
Jun 2009to Nov 2009
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 75 years
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Insulin requiring diabetics (type 1 or type 2)
  • Using a pen device for injections at least once per day, for at least two months prior to baseline (Study Visit Two).
  • 18 to 75 years of age, inclusive.
  • Body Mass Index from 18 to 50 kg/m², inclusive.
  • Hemoglobin A1c from 5.5 to 9.5 percent (%), inclusive.
  • Largest single dose of insulin a day less than or equal to 40 units

Who cannot join

  • Current status or history of a medical condition that would contraindicate treatment with the study product or other conditions which, in the opinion of the Investigator, would place the subject at risk or have the potential to confound interpretation of the study results (i.e. recent history of ketoacidosis, hypoglycemic unawareness, etc.).
  • Hemophilia or any other disorder which causes excessive bleeding or requires frequent transfusions.
  • History of intravenous drug abuse (self-reported).
  • Bleeding disorder, hemophilia or subject is currently taking anti-coagulant medications (e.g., heparin, Coumadin®, warfarin).
  • Self-reported blood borne infection.
  • Pregnant (from medical history only).

Trial locations and contacts

4 sites in 1 country.

United States4 sites
  • Atlanta Diabetes Associates

    Atlanta, Georgia, 30309

  • Radiant Research

    Birmingham, Alabama, 35209

  • AMCR Institute, Inc.

    Escondido, California, 92026

  • Rainier Clinical Research Center, Inc.

    Renton, Washington, 98057

Other trials in United States

About this study

Official title: Comparison of Glycemic Control Among Diabetics Using the 4mm x 32G BD Pen Needle Versus the 8mm x 31G BD Pen Needle and the 5mm x 31G BD Pen Needle

Anxiety about needles is a commonly expressed concern by diabetics about beginning insulin therapy. A shorter, thinner pen needle that delivers insulin with the safety and efficacy profile of currently marketed pen needles may appeal to many diabetic patients as the new needle may be perceived as less intimidating and more comfortable. Currently marketed pen needles range in length from 5 to 12.7 millimeters (mm).

The primary purpose of this study was to evaluate whether the investigational 4mm x 32 Gauge(G) pen needle manufactured by Becton, Dickinson and Company (BD) provides equivalent glucose control (as measured by fructosamine levels) as the currently marketed BD 5mm x 31 G and BD 8mm x 31 G pen needles (PN)in diabetic subjects with varying insulin dosage regimens.

Read the detailed description

Within each dose group, subjects were randomly assigned at baseline to one of two study arms. One study arm (4/5 mm PN) compared the BD 4mm PN to the 5mm PN, and the other arm (4/8 mm PN) compared the BD 4mm PN to the 8mm PN, in a crossover fashion. Based on the randomization, subjects used either the 4mm or the other assigned PN (5mm or 8mm) for the first 3 weeks, then switched to the alternate PN for the next 3 weeks. Glycemic control was assessed by serum fructosamine (FRU) levels at the end of each 3 week period.

Based on their baseline insulin regimen, subjects were also assigned to an insulin dosing group (Low or Regular) to help ensure balance among treatment arms. Requirements for each group were as follows: Low Dose: largest single dose of insulin each day with a pen device must be less than or equal to 20 units. Regular Dose: largest single dose of insulin each day, with a pen device, must be 21-40 units.

Explanation of Visits and Timing of Assessments:

Visit 1:

  • Screening
  • Informed Consent
  • Demographics
  • Inclusion /Exclusion
  • Hemoglobin A1c measurement (HbA1c)

Visit 2 (Baseline:)

  • Insulin dose group assignment (Low dose or Regular dose)
  • Randomization
  • Fructosamine blood sample collected
  • Dispense pen needle (PN) assigned first

Visit 3 (approximately 3 weeks after starting first PN)

  • Pain rated by subject relative to baseline PN
  • Review Adverse Events (AEs) and reported events of injection site leakage
  • Fructosamine blood sample collected
  • Dispense second assigned PN

Visit 4 (approximately 3 weeks after starting second PN):

  • Review AEs and reported events of injection site leakage
  • Fructosamine blood sample collected
  • Pain rated relative to previously used PN
  • Study Completion
Being tested
4mm x 32G pen needle, 8mm x 31G pen needle, 5mm x 31G pen needle
Keywords
Pen Needle, Diabetes, Insulin, Glycemic, Pain, Fructosamine, Leakage, Hypoglycemia, Hyperglycemia
Medical subject headings
Diabetes Mellitus, Type 1, Diabetes Mellitus, Type 2, Diabetes Mellitus, Insulin Resistance, Pain, Hypoglycemia, Hyperglycemia

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • 4mm x 32G pen needleDevice

    Also called BD Ultra-Fine™ Nano pen needle, Catalog number 320122

    For 3 weeks, subjects used this pen needle with their own pen device for all daily insulin injections they usually administered themselves with a pen device. Subjects followed their usual insulin regimen and there was no upper limit on total daily insulin dosage or number of injections. Subjects were advised to inject straight in when using the 4mm PN, with no pinch up.

  • 8mm x 31G pen needleDevice

    Also called BD Ultra-Fine™ Short pen needle, Catalog number 320109

    For 3 weeks, subjects used this pen needle with their own pen device for all daily insulin injections they usually administered themselves with a pen device. Subjects followed their usual insulin regimen and there was no upper limit on total daily insulin dosage or number of injections. Subjects were directed to use pinch-up when injecting in the abdomen or thigh with the 8mm PN, and no pinch-up at other injection sites.

  • 5mm x 31G pen needleDevice

    Also called BD Ultra-Fine™ Mini pen needle, Catalog Number 320119

    For 3 weeks, subjects used this pen needle with their own pen device for all daily insulin injections they usually administered themselves with a pen device. Subjects followed their usual insulin regimen and there was no upper limit on total daily insulin dosage or number of injections. Subjects were advised to inject straight in when using the 5mm PN, with no pinch up.

Study groups in NCT00928057
Study groupRoleWhat they receive
4 mm / 8 mm PNExperimentalSubjects randomized to this study arm used either the 4mm PN or the 8mm PN for 3 weeks, then switched to the alternate PN for another 3 weeks. Order of PN use was randomly determined.
Receives: Device: 4mm x 32G pen needle; Device: 8mm x 31G pen needle
4 mm / 5 mm PNExperimentalSubjects randomized to this study arm used either the 4mm PN or the 5mm PN for 3 weeks, then switched to the alternate PN for another 3 weeks. Order of PN use was randomly determined.
Receives: Device: 4mm x 32G pen needle; Device: 5mm x 31G pen needle

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Percent (%) Absolute Change in Fructosamine

    Within each study arm (4mm / 5mm PN, or 4mm / 8mm PN) subjects used one pen needle (PN) for 3 weeks then switched to the alternate PN for the next three weeks, for a total of 6 weeks. The principal endpoint measure of glycemic control is the percent absolute change in serum fructosamine concentration, based on the fructosamine (FRU) concentration measured in micromoles per liter (umol/L) at the end of each three week period. The change was calculated according to the following formula:

    Percent absolute change in FRU, or %|∆ FRU|= 100*[FRU(4mm)-FRU(5 or 8mm)]/[FRU(5 or 8mm)].

    Measured: 3 weeks per pen needle, from visit 2-3 and visit 3-4.

Secondary outcomes (4)
  1. Percent Absolute Change in Fructosamine, by Dose Group

    Glycemic control was determined separately for each insulin dose group in the same manner as described for the primary outcome measure, according to the following formula:

    %|∆ FRU|= 100*[FRU(4mm)-FRU(5 or 8mm)]/[FRU(5 or 8mm)].

    Measured: 3 weeks per pen needle, from visit 2-3 and visit 3-4.

  2. Number of Subjects With Severe Unexplained Hypoglycemic Events

    Severe Unexplained Hypoglycemia is defined as an event in which the subject's blood glucose is below 50 milligrams per deciliter (mg/dL) and/or they required assistance from another person for treatment and there is no identified cause (for example, the subject skipped a meal). These events were also reported as Adverse Events.

    Measured: During 3 weeks using each pen needle

  3. Number of Subjects With Severe Unexplained Hyperglycemic Events

    Severe Unexplained Hyperglycemia is defined as requiring a visit to the emergency room or hospitalization and/or having a blood glucose value above 450mg/dL without an identified cause (for example, the subject missed an insulin dose). These events were also reported as Adverse Events.

    Measured: During 3 weeks using each pen needle

  4. Relative Injection Pain Score Assessed by Subject

    After using the second assigned pen needle (PN) for 3 weeks, subjects used a 150mm Visual Analog Scale (VAS) to rate the pain of the second PN relative to the first PN. The VAS was anchored at the center (0mm) with "as painful", and at the extreme ends with "much less painful" (-75mm) and "much more painful" (+75mm). VAS scores were adjusted for the order of PN use, such that a positive score means that the 4mm PN was scored as more painful than the reference PN (5mm or 8mm), and a negative score indicates that the 4mm PN was less painful than the reference.

    Measured: Visit 4: 18-24 days after starting 2nd pen needle

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted25 Jun 2009 (estimated)
  2. Study startJun 2009
  3. Primary completionNov 2009
  4. Study completionNov 2009

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Allocation
Randomised
Intervention model
Crossover
Masking
None (open label)
Enrolment
173 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT00928057
Sponsor's protocol ID
BDDC-08-011
Status last verified
Sep 2010
First submitted
24 Jun 2009
Last updated
20 Oct 2010
Data monitoring committee
No

Publications

  • Hirsch LJ, Gibney MA, Albanese J, Qu S, Kassler-Taub K, Klaff LJ, Bailey TS. Comparative glycemic control, safety and patient ratings for a new 4 mm x 32G insulin pen needle in adults with diabetes. Curr Med Res Opin. 2010 Jun;26(6):1531-41. doi: 10.1185/03007995.2010.482499.

    PubMed 20429832
  • Hirsch LJ, Gibney MA, Li L, Berube J. Glycemic control, reported pain and leakage with a 4 mm x 32 G pen needle in obese and non-obese adults with diabetes: a post hoc analysis. Curr Med Res Opin. 2012 Aug;28(8):1305-11. doi: 10.1185/03007995.2012.709181. Epub 2012 Jul 18.

    PubMed 22762347

Questions about this trial

What is this clinical trial studying?

A study testing 4mm x 32G pen needle, 8mm x 31G pen needle and 5mm x 31G pen needle in people with Diabetes Mellitus, Type 1 and Diabetes Mellitus, Type 2, run by Becton, Dickinson and Company at 4 sites in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 6 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 20 Oct 2010. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 4 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 20 Oct 2010). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Percent (%) Absolute Change in Fructosamine (time frame: 3 weeks per pen needle, from visit 2-3 and visit 3-4.). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT00928057, which also lists 2 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 20 Oct 2010.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in