Anesthesia Airway ManagementNCT01001078
Comparison Study of Two Supraglottic Airway Devices Used for Patients Under General Anesthesia
A study testing I-Gel supraglottic airway device, LMA Supreme supraglottic airway device and Standard endotracheal tube in people with Anesthesia Airway Management, run by Maisonneuve-Rosemont Hospital at 1 site in Canada.
Trial summary
This study will compare two supraglottic airway devices currently on the market. Supraglottic airway devices are used during general anesthesia to provide a patent airway. They are inserted blindly into the mouth once the patient has lost consciousness and they cover the laryngeal inlet. Read more about this study
- Who is it for?
- Ages 18 years and olderHealthy volunteers can take part
- Conditions the study is for
- Anesthesia Airway Management
- What is being tested?
- I-Gel supraglottic airway device, LMA Supreme supraglottic airway device and Standard endotracheal tube
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 100 participants
- Where
- 1 siteCanada
- Study dates
- Sep 2009to Mar 2011
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- Healthy volunteers accepted
- Adult, Older adult
Who may be able to join
- Weight more than 50 kg;
- Body Mass Index less than 30;
- ASA I to III.
Who cannot join
- Symptomatic gastro oesophageal reflux disease (GERD);
- Non fasted patients;
- Nasogastric tube in place;
- Intestinal obstruction;
- Ear, nose and throat surgery in the past or deformation of the airway;
- Known difficult airway (Cormack-Lehane grade 3-4);
- Oral cavity opening less than 3 cm;
- Pregnancy.
Trial locations and contacts
1 site in 1 country.
About this study
Official title: Prospective Randomized Trial Comparing the Performance of Two Supraglottic Airway Management Devices: I-Gel and Supreme LMA
This study will compare two supraglottic airway devices currently on the market. Supraglottic airway devices are used during general anesthesia to provide a patent airway. They are inserted blindly into the mouth once the patient has lost consciousness and they cover the laryngeal inlet. The I-Gel and the Supreme laryngeal mask airway (LMA) are the two devices to be compared. The I-Gel has no inflatable cuff, which makes it different from all other supraglottic airway devices currently in use. The Supreme shares some characteristics of the other LMA devices, but it is disposable. Our main goal will be to compare the airway leak pressure and the peak airway pressure of each devices. We will also measure the time needed for insertion, number of attempts needed to secure the airway and side effects related to the airway (cough, dysphagia, trauma, hoarseness of voice, sore throat).
Read the detailed description
First of all, we will record the following demographic variables: age, sex, height, weight, BMI, surgery planned, American Society of Anesthesiology (ASA) status, dental state (good, bad, edentulous), tobacco use (none, ceased, active) and packs year, Mallampati score, thyromental distance and whether or not there is trauma on the tongue, teeth or lips prior to surgery.
Induction of anesthesia will be standardized with Fentanyl 1-3 mcg/kiloGram(kG) and Propofol 1-3,5 mg/kG. The use of myorelaxant will be mandatory, but either succinylcholine or rocuronium will be used as long as complete disappearance of T1 is observed before attempting supraglottic airway insertion.
Prior to insertion, ease of ventilation (yes/no) will be noted.
The two airway devices will be used according to manufacturers instructions. The sizes of the devices will be chosen according to weight:
I-Gel: 50-90 kG: #4 and 90 Kg and up: #5 LMA Supreme: 50-70 Kg: #4 and 70-100 Kg: #5.
Patients will be randomized in either I-Gel or LMA Supreme group. Two insertions with the same device will be allowed before a crossover to the other device. If the third attempt fails, a standard endotracheal tube will be put in the trachea and blood in the oral cavity will be sought. After each failed attempt (no thoracic expansion, no carbon dioxide (CO2) square wave or leak at standardized volume. Blood on the devices will also be recorded.
Since the LMA Supreme has a cuff, we will inflate it to 25 mL and then by intervals of 5 mL of air (max 45 mL) we will inflate it at a volume where no audible leak is heard at a standardized ventilation scheme of volume controlled 8 mL/kg X 10 breaths per minute.
Once fixed with tape, we will measure peak and mean airway pressure for three breaths after 30 seconds of volume controlled breathing. Then the fresh gas flow will be fixed at 3 L/min of 50-50 air-oxygen blend and the Adjustable Pressure Limiting valve will be closed. We will look either for a plateau pressure at which a leak equal to the fresh gas flow or a pressure of 40 cm of water.
We will sought adverse effect like bronchospasm, cough, regurgitation, desaturation, aspiration and deglutition movement.
Ease of insertion will be score on a scale of 1 to 4. 1 being very easy and 4 being very hard.
Endoscopic evaluation will take place afterwards. First in the ventilation tube. Scoring for laryngeal view:
- Vocal cords all visible;
- Partial view of the cords including arytenoids;
- View of the epiglottis only;
- Other (LMA or pharynx).
Downfolding of the epiglottis will be sought.
Then we will go through the oesophageal port and see if the oesophageal mucosa il visible.
During surgery, we will check if airway manipulation and/or withdrawal was necessary and why.
If traces of blood on the devices are present at the end of the intervention it will be noted.
Two interviews will take place one at the post anesthesia care unit and another the following day by phone if the patient was having ambulatory surgery. Sore throat, cough, dysphagia and dysphonia will be asked to be graded: absent, mild, moderate or severe.
Pulse and arterial pressure with the operating room's sphygmomanometer and oximeter will be written before induction, after induction before insertion and at 1 and 2 minutes post insertion.
- Conditions
- Anesthesia Airway Management
- Being tested
- I-Gel supraglottic airway device, LMA Supreme supraglottic airway device, Standard endotracheal tube
- Keywords
- supraglottic airway devices, I-Gel, Supreme Laryngeal mask airway, Airway leak pressure, peak airway pressure
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
I-Gel supraglottic airway deviceDevice
2 attempts to secure the airway will be tried. If they both fail, there will be a crossover with a Supreme LMA.
LMA Supreme supraglottic airway deviceDevice
In the LMA Supreme group, the same interventions will be done as in the I-Gel group. The cuff of the LMA Supreme will be inflated with 25 mL of air and then by intervals of 5 mL until no leak is heard (max 45 mL). Leak pressure will also be measure at 45 mL. Once the leak test is done, the cuff will be deflated to initial volume necessary to avoid audible leaks.
Standard endotracheal tubeDevice
In case both supraglottic airway devices fail, a standard endotracheal tube will be inserted
| Study group | Role | What they receive |
|---|---|---|
| I-Gel supraglottic airway device | Active comparator | Group in which the I-Gel will be used in the first and second attempts to secure the airway. If a third attempt is needed, the LMA Supreme will be used. If the third attempt fails, a standard endotracheal tube will be used. Receives: Device: I-Gel supraglottic airway device; Device: Standard endotracheal tube |
| LMA Supreme supraglottic airway device | Active comparator | Group in which the LMA Supreme will be used in the first and second attempts to secure the airway. If a third attempt is needed, the I-Gel will be used. If the third attempt fails, a standard endotracheal tube will be used. Receives: Device: LMA Supreme supraglottic airway device; Device: Standard endotracheal tube |
| Standard endotracheal tube | Active comparator | A Standard endotracheal tube will be inserted in case both other airway devices (LMA Supreme and iGel) devices fail to provide adequate ventilation. Receives: Device: Standard endotracheal tube |
What is being measured
The primary outcome is the main question the study is designed to answer.
Measure of the Airway Leak Pressure
Measured: After introduction of the supraglottic device before the beginning of the surgery.
Secondary outcomes (5)
Measure of the Peak Airway Pressure
Measured: After introduction of the SAD before the beginning of the surgery.
Time Needed to Secure the Airway
Measured: From opening of the mouth to thoracic expansion and presence of End Tidal CO2 up to five minutes.
Number of Participants With Successful Attempts to Introduce the Devices
Measured: At the beginning of anesthesia before the beginning of the surgery.
Adverse Effects After Anesthesia (Sore Throat, Cough, Dysphagia, Dysphonia)
Intent to treat assessment of adverse events, per intervention, was intended. The adverse effects were assessed by phone; or in person if patient in the hospital; on the following day.
Measured: On the day following surgery
Airway Manipulation and Blood on Device at Removal
Measured: During and after anesthesia when the device is removed.
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted23 Oct 2009 (estimated)
- Study startSep 2009
- Primary completionMar 2011
- Study completionMar 2011
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Publications (18)Papers the sponsor has linked to this study.
- Other trials for Anesthesia Airway ManagementStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Supportive care
- Allocation
- Randomised
- Intervention model
- Crossover
- Masking
- Single blind
- Who is masked
- Participant
- Enrolment
- 100 participants (actual)
Sponsor and investigators
- Lead sponsor
- Maisonneuve-Rosemont Hospital
- Sponsor type
- University, hospital or other
- Responsible party
- Principal investigator, Nikola Joly, Maisonneuve-Rosemont Hospital
- Investigators
- Pierre Drolet, MD, FRCPC, Principal investigator, Hôpital Maisonneuve-Rosemont, Université de Montréal
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT01001078
- Sponsor's protocol ID
- HMR-Anesth-Joly
- Status last verified
- Jan 2015
- First submitted
- 21 Oct 2009
- Last updated
- 22 Jan 2015
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
Publications
Theiler LG, Kleine-Brueggeney M, Kaiser D, Urwyler N, Luyet C, Vogt A, Greif R, Unibe MM. Crossover comparison of the laryngeal mask supreme and the i-gel in simulated difficult airway scenario in anesthetized patients. Anesthesiology. 2009 Jul;111(1):55-62. doi: 10.1097/ALN.0b013e3181a4c6b9.
PubMed 19512881Janakiraman C, Chethan DB, Wilkes AR, Stacey MR, Goodwin N. A randomised crossover trial comparing the i-gel supraglottic airway and classic laryngeal mask airway. Anaesthesia. 2009 Jun;64(6):674-8. doi: 10.1111/j.1365-2044.2009.05898.x.
PubMed 19453322Weber U, Oguz R, Potura LA, Kimberger O, Kober A, Tschernko E. Comparison of the i-gel and the LMA-Unique laryngeal mask airway in patients with mild to moderate obesity during elective short-term surgery. Anaesthesia. 2011 Jun;66(6):481-7. doi: 10.1111/j.1365-2044.2011.06682.x.
PubMed 21568982Gatward JJ, Cook TM, Seller C, Handel J, Simpson T, Vanek V, Kelly F. Evaluation of the size 4 i-gel airway in one hundred non-paralysed patients. Anaesthesia. 2008 Oct;63(10):1124-30. doi: 10.1111/j.1365-2044.2008.05561.x. Epub 2008 Jul 9.
PubMed 18616521Richez B, Saltel L, Banchereau F, Torrielli R, Cros AM. A new single use supraglottic airway device with a noninflatable cuff and an esophageal vent: an observational study of the i-gel. Anesth Analg. 2008 Apr;106(4):1137-9, table of contents. doi: 10.1213/ane.0b013e318164f062.
PubMed 18349185Uppal V, Fletcher G, Kinsella J. Comparison of the i-gel with the cuffed tracheal tube during pressure-controlled ventilation. Br J Anaesth. 2009 Feb;102(2):264-8. doi: 10.1093/bja/aen366.
PubMed 19151051Wong DT, Yang JJ, Jagannathan N. Brief review: The LMA Supreme supraglottic airway. Can J Anaesth. 2012 May;59(5):483-93. doi: 10.1007/s12630-012-9673-0. Epub 2012 Feb 9.
PubMed 22318376Keller C, Brimacombe JR, Keller K, Morris R. Comparison of four methods for assessing airway sealing pressure with the laryngeal mask airway in adult patients. Br J Anaesth. 1999 Feb;82(2):286-7. doi: 10.1093/bja/82.2.286.
PubMed 10365012Brimacombe J, Berry A. A proposed fiber-optic scoring system to standardize the assessment of laryngeal mask airway position. Anesth Analg. 1993 Feb;76(2):457. No abstract available.
PubMed 8424538Teoh WH, Lee KM, Suhitharan T, Yahaya Z, Teo MM, Sia AT. Comparison of the LMA Supreme vs the i-gel in paralysed patients undergoing gynaecological laparoscopic surgery with controlled ventilation. Anaesthesia. 2010 Dec;65(12):1173-9. doi: 10.1111/j.1365-2044.2010.06534.x. Epub 2010 Oct 19.
PubMed 20958278Russo SG, Cremer S, Eich C, Jipp M, Cohnen J, Strack M, Quintel M, Mohr A. Magnetic resonance imaging study of the in vivo position of the extraglottic airway devices i-gel and LMA-Supreme in anaesthetized human volunteers. Br J Anaesth. 2012 Dec;109(6):996-1004. doi: 10.1093/bja/aes314. Epub 2012 Sep 25.
PubMed 23015619Chew EE, Hashim NH, Wang CY. Randomised comparison of the LMA Supreme with the I-Gel in spontaneously breathing anaesthetised adult patients. Anaesth Intensive Care. 2010 Nov;38(6):1018-22. doi: 10.1177/0310057X1003800609.
PubMed 21226431Ragazzi R, Finessi L, Farinelli I, Alvisi R, Volta CA. LMA Supreme vs i-gel--a comparison of insertion success in novices. Anaesthesia. 2012 Apr;67(4):384-8. doi: 10.1111/j.1365-2044.2011.07002.x. Epub 2012 Feb 13.
PubMed 22329593Chen X, Jiao J, Cong X, Liu L, Wu X. A comparison of the performance of the I-gel vs. the LMA-Sduring anesthesia: a meta-analysis of randomized controlled trials. PLoS One. 2013 Aug 12;8(8):e71910. doi: 10.1371/journal.pone.0071910. eCollection 2013.
PubMed 23951266Halwagi AE, Massicotte N, Lallo A, Gauthier A, Boudreault D, Ruel M, Girard F. Tracheal intubation through the I-gel supraglottic airway versus the LMA Fastrach: a randomized controlled trial. Anesth Analg. 2012 Jan;114(1):152-6. doi: 10.1213/ANE.0b013e318236f438. Epub 2011 Nov 10.
PubMed 22075016
Study at a glance
From the registry record, last updated by the sponsor on 22 Jan 2015. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Prospective Randomized Trial Comparing the Performance of Two Supraglottic Airway Management Devices: I-Gel and Supreme LMA
- ClinicalTrials.gov ID
- NCT01001078
- Acronym / sponsor ID
- HMR-Anesth-Joly
- Conditions
- Anesthesia Airway Management
- Keywords
- supraglottic airway devices, I-Gel, Supreme Laryngeal mask airway, Airway leak pressure, peak airway pressure
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 100 participants (actual)
- Lead sponsor
- Maisonneuve-Rosemont HospitalUniversity, hospital or other
- Study start
- Sep 2009
- Primary completion
- Mar 2011 (actual)
- Study completion
- Mar 2011 (actual)
- First posted
- 23 Oct 2009 (estimated)
- Results
- Posted 22 Jan 2015
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
This study will compare two supraglottic airway devices currently on the market. Supraglottic airway devices are used during general anesthesia to provide a patent airway. They are inserted blindly into the mouth once the patient has lost consciousness and they cover the laryngeal inlet. The I-Gel and the Supreme laryngeal mask airway (LMA) are the two devices to be compared. The I-Gel has no inflatable cuff, which makes it different from all other supraglottic airway devices currently in use. The Supreme shares some characteristics of the other LMA devices, but it is disposable. Our main goal will be to compare the airway leak pressure and the peak airway pressure of each devices. We will also measure the time needed for insertion, number of attempts needed to secure the airway and side effects related to the airway (cough, dysphagia, trauma, hoarseness of voice, sore throat).
Detailed description
First of all, we will record the following demographic variables: age, sex, height, weight, BMI, surgery planned, American Society of Anesthesiology (ASA) status, dental state (good, bad, edentulous), tobacco use (none, ceased, active) and packs year, Mallampati score, thyromental distance and whether or not there is trauma on the tongue, teeth or lips prior to surgery.
Induction of anesthesia will be standardized with Fentanyl 1-3 mcg/kiloGram(kG) and Propofol 1-3,5 mg/kG. The use of myorelaxant will be mandatory, but either succinylcholine or rocuronium will be used as long as complete disappearance of T1 is observed before attempting supraglottic airway insertion.
Prior to insertion, ease of ventilation (yes/no) will be noted.
The two airway devices will be used according to manufacturers instructions. The sizes of the devices will be chosen according to weight:
I-Gel: 50-90 kG: #4 and 90 Kg and up: #5 LMA Supreme: 50-70 Kg: #4 and 70-100 Kg: #5.
Patients will be randomized in either I-Gel or LMA Supreme group. Two insertions with the same device will be allowed before a crossover to the other device. If the third attempt fails, a standard endotracheal tube will be put in the trachea and blood in the oral cavity will be sought. After each failed attempt (no thoracic expansion, no carbon dioxide (CO2) square wave or leak at standardized volume. Blood on the devices will also be recorded.
Since the LMA Supreme has a cuff, we will inflate it to 25 mL and then by intervals of 5 mL of air (max 45 mL) we will inflate it at a volume where no audible leak is heard at a standardized ventilation scheme of volume controlled 8 mL/kg X 10 breaths per minute.
Once fixed with tape, we will measure peak and mean airway pressure for three breaths after 30 seconds of volume controlled breathing. Then the fresh gas flow will be fixed at 3 L/min of 50-50 air-oxygen blend and the Adjustable Pressure Limiting valve will be closed. We will look either for a plateau pressure at which a leak equal to the fresh gas flow or a pressure of 40 cm of water.
We will sought adverse effect like bronchospasm, cough, regurgitation, desaturation, aspiration and deglutition movement.
Ease of insertion will be score on a scale of 1 to 4. 1 being very easy and 4 being very hard.
Endoscopic evaluation will take place afterwards. First in the ventilation tube. Scoring for laryngeal view:
- Vocal cords all visible;
- Partial view of the cords including arytenoids;
- View of the epiglottis only;
- Other (LMA or pharynx).
Downfolding of the epiglottis will be sought.
Then we will go through the oesophageal port and see if the oesophageal mucosa il visible.
During surgery, we will check if airway manipulation and/or withdrawal was necessary and why.
If traces of blood on the devices are present at the end of the intervention it will be noted.
Two interviews will take place one at the post anesthesia care unit and another the following day by phone if the patient was having ambulatory surgery. Sore throat, cough, dysphagia and dysphonia will be asked to be graded: absent, mild, moderate or severe.
Pulse and arterial pressure with the operating room's sphygmomanometer and oximeter will be written before induction, after induction before insertion and at 1 and 2 minutes post insertion.
Study design
- Primary purpose
- Supportive care
- Allocation
- Randomised
- Intervention model
- Crossover
- Masking
- Single blind (participant)
- Number of arms
- 3
| Arm | Type | Receives |
|---|---|---|
| I-Gel supraglottic airway device | Active comparator | Device: I-Gel supraglottic airway device, Device: Standard endotracheal tubeGroup in which the I-Gel will be used in the first and second attempts to secure the airway. If a third attempt is needed, the LMA Supreme will be used. If the third attempt fails, a standard endotracheal tube will be used. |
| LMA Supreme supraglottic airway device | Active comparator | Device: LMA Supreme supraglottic airway device, Device: Standard endotracheal tubeGroup in which the LMA Supreme will be used in the first and second attempts to secure the airway. If a third attempt is needed, the I-Gel will be used. If the third attempt fails, a standard endotracheal tube will be used. |
| Standard endotracheal tube | Active comparator | Device: Standard endotracheal tubeA Standard endotracheal tube will be inserted in case both other airway devices (LMA Supreme and iGel) devices fail to provide adequate ventilation. |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (3 listed)
- Weight more than 50 kg;
- Body Mass Index less than 30;
- ASA I to III.
Key exclusion criteria (8 listed)
- Symptomatic gastro oesophageal reflux disease (GERD);
- Non fasted patients;
- Nasogastric tube in place;
- Intestinal obstruction;
- Ear, nose and throat surgery in the past or deformation of the airway;
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| I-Gel supraglottic airway device2 attempts to secure the airway will be tried. If they both fail, there will be a crossover with a Supreme LMA. | Device | I-Gel supraglottic airway device |
| LMA Supreme supraglottic airway deviceIn the LMA Supreme group, the same interventions will be done as in the I-Gel group. The cuff of the LMA Supreme will be inflated with 25 mL of air and then by intervals of 5 mL until no leak is heard (max 45 mL). Leak pressure will also be measure at 45 mL. Once the leak test is done, the cuff will be deflated to initial volume necessary to avoid audible leaks. | Device | LMA Supreme supraglottic airway device |
| Standard endotracheal tubeIn case both supraglottic airway devices fail, a standard endotracheal tube will be inserted | Device | I-Gel supraglottic airway device, LMA Supreme supraglottic airway device, Standard endotracheal tube |
Primary outcome
Measure of the Airway Leak Pressure
Time frame: After introduction of the supraglottic device before the beginning of the surgery.
5 secondary outcomes
Measure of the Peak Airway Pressure
Time frame: After introduction of the SAD before the beginning of the surgery.
Time Needed to Secure the Airway
Time frame: From opening of the mouth to thoracic expansion and presence of End Tidal CO2 up to five minutes.
Number of Participants With Successful Attempts to Introduce the Devices
Time frame: At the beginning of anesthesia before the beginning of the surgery.
Adverse Effects After Anesthesia (Sore Throat, Cough, Dysphagia, Dysphonia)
Time frame: On the day following surgery
Airway Manipulation and Blood on Device at Removal
Time frame: During and after anesthesia when the device is removed.
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 22 Jan 2015.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
- Background: PubMed 19512881Theiler LG, Kleine-Brueggeney M, Kaiser D, Urwyler N, Luyet C, Vogt A, Greif R, Unibe MM. Crossover comparison of the laryngeal mask supreme and the i-gel in simulated difficult airway scenario in anesthetized patients. Anesthesiology. 2009 Jul;111(1):55-62. doi: 10.1097/ALN.0b013e3181a4c6b9.
- Background: PubMed 19453322Janakiraman C, Chethan DB, Wilkes AR, Stacey MR, Goodwin N. A randomised crossover trial comparing the i-gel supraglottic airway and classic laryngeal mask airway. Anaesthesia. 2009 Jun;64(6):674-8. doi: 10.1111/j.1365-2044.2009.05898.x.
- Background: PubMed 21568982Weber U, Oguz R, Potura LA, Kimberger O, Kober A, Tschernko E. Comparison of the i-gel and the LMA-Unique laryngeal mask airway in patients with mild to moderate obesity during elective short-term surgery. Anaesthesia. 2011 Jun;66(6):481-7. doi: 10.1111/j.1365-2044.2011.06682.x.
- Background: PubMed 18616521Gatward JJ, Cook TM, Seller C, Handel J, Simpson T, Vanek V, Kelly F. Evaluation of the size 4 i-gel airway in one hundred non-paralysed patients. Anaesthesia. 2008 Oct;63(10):1124-30. doi: 10.1111/j.1365-2044.2008.05561.x. Epub 2008 Jul 9.
- Background: PubMed 18349185Richez B, Saltel L, Banchereau F, Torrielli R, Cros AM. A new single use supraglottic airway device with a noninflatable cuff and an esophageal vent: an observational study of the i-gel. Anesth Analg. 2008 Apr;106(4):1137-9, table of contents. doi: 10.1213/ane.0b013e318164f062.
- Background: PubMed 19151051Uppal V, Fletcher G, Kinsella J. Comparison of the i-gel with the cuffed tracheal tube during pressure-controlled ventilation. Br J Anaesth. 2009 Feb;102(2):264-8. doi: 10.1093/bja/aen366.
- Background: PubMed 22318376Wong DT, Yang JJ, Jagannathan N. Brief review: The LMA Supreme supraglottic airway. Can J Anaesth. 2012 May;59(5):483-93. doi: 10.1007/s12630-012-9673-0. Epub 2012 Feb 9.
- Background: PubMed 10365012Keller C, Brimacombe JR, Keller K, Morris R. Comparison of four methods for assessing airway sealing pressure with the laryngeal mask airway in adult patients. Br J Anaesth. 1999 Feb;82(2):286-7. doi: 10.1093/bja/82.2.286.
- Background: PubMed 8424538Brimacombe J, Berry A. A proposed fiber-optic scoring system to standardize the assessment of laryngeal mask airway position. Anesth Analg. 1993 Feb;76(2):457. No abstract available.
- Background: PubMed 20958278Teoh WH, Lee KM, Suhitharan T, Yahaya Z, Teo MM, Sia AT. Comparison of the LMA Supreme vs the i-gel in paralysed patients undergoing gynaecological laparoscopic surgery with controlled ventilation. Anaesthesia. 2010 Dec;65(12):1173-9. doi: 10.1111/j.1365-2044.2010.06534.x. Epub 2010 Oct 19.
- Background: PubMed 23015619Russo SG, Cremer S, Eich C, Jipp M, Cohnen J, Strack M, Quintel M, Mohr A. Magnetic resonance imaging study of the in vivo position of the extraglottic airway devices i-gel and LMA-Supreme in anaesthetized human volunteers. Br J Anaesth. 2012 Dec;109(6):996-1004. doi: 10.1093/bja/aes314. Epub 2012 Sep 25.
- Background: PubMed 21226431Chew EE, Hashim NH, Wang CY. Randomised comparison of the LMA Supreme with the I-Gel in spontaneously breathing anaesthetised adult patients. Anaesth Intensive Care. 2010 Nov;38(6):1018-22. doi: 10.1177/0310057X1003800609.
Related trials for Anesthesia Airway Management
Not yet recruiting
Tizanidine Premedication Versus Placebo for Laryngeal Mask Airway Insertion and Recovery in Adults Undergoing Ambulatory Surgery
Airway Management
Not yet recruitingObservational study
Preoperative Frailty and Suboptimal Mask Ventilation in Cardiac Surgery Patients
Frailty, Suboptimal Mask Ventilation
Not yet recruiting
Intracuff Dexmedetomidine Versus Alkalinized Lidocaine for Peri-Extubation Cough and Hemodynamic Responses in Elderly Patients
Airway Complication of Anesthesia, Hemodynamic Instability
Not yet recruiting
Comparison of Second- and Third-Generation Laryngeal Mask Airway in Pediatric Patients
Airway Management, Anesthesia, General
Not yet recruiting
Title of Study: Anesthesiologists' Learning Curves for Ultrasound Assessment of Difficult Airway
Airway Ultrasound, Difficult Airway Management
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Pierre Drolet, MD, FRCPC principal investigator
Hôpital Maisonneuve-Rosemont, Université de Montréal
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing I-Gel supraglottic airway device, LMA Supreme supraglottic airway device and Standard endotracheal tube in people with Anesthesia Airway Management, run by Maisonneuve-Rosemont Hospital at 1 site in Canada.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 3 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 22 Jan 2015. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Canada. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 22 Jan 2015). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Measure of the Airway Leak Pressure (time frame: After introduction of the supraglottic device before the beginning of the surgery.). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT01001078, which also lists 18 linked publications.