Clinical Variations and Pain Assessment in Newborns Submitted to Intratracheal Aspiration With Open and Closed System

A study testing Tracheal aspiration close system in people with Adverse Reaction to Systemic Agents, run by Federal University of São Paulo at 2 sites in Brazil.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

To compare clinical effects identified in infants undergoing intratracheal aspiration with open and closed systems and verify the presence and intensity of pain in newborns during intratracheal suction, according to the system applied.Clinical, randomized, controlled, and crossover study, performed at two neonatal… Read more about this study

Who is it for?
Ages 1 hour to 28 days
Conditions the study is for
Adverse Reaction to Systemic Agents
What is being tested?
Tracheal aspiration close system
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
13 participants
Where
2 sitesBrazil
Study dates
Jan 2010to Jul 2010
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 1 hour to 28 days
  • All sexes
  • No healthy volunteers
  • Child

Who may be able to join

  • infants from zero to seven days of age.
  • 26 weeks of gestational age.
  • responsible concordance in participation

Who cannot join

  • use of mechanical ventilation with high-frequency oscillatory.
  • use of inhaled nitric oxide.
  • acute phase of circulatory failure.
  • central nervous system disorders.
  • without reactivity to stimulus.
  • severe asphyxia.
  • Apgar score less than three in the fifth minute.
  • Presence of chromosomal abnormalities and congenital malformations

Trial locations and contacts

2 sites in 1 country.

Brazil2 sites
  • Federtal university of São Paulo

    São Paulo, São Paulo, 04024002

  • Universidade Federal de São Paulo

    São Paulo, São Paulo, 04024002

Other trials in Brazil

About this study

Official title: Clinical Variations and Pain Assessment in Newborns Submitted to Intratracheal Aspiration With Open and Closed System.

To compare clinical effects identified in infants undergoing intratracheal aspiration with open and closed systems and verify the presence and intensity of pain in newborns during intratracheal suction, according to the system applied.Clinical, randomized, controlled, and crossover study, performed at two neonatal intensive care units after approval by the ethics committee. The sample consisted of infants from zero to seven days of age, 26 weeks of gestational age, and the exclusion criteria were use of mechanical ventilation with high-frequency oscillatory, in use of inhaled nitric oxide, in the acute phase of circulatory failure, central nervous system disorders, without reactivity to stimulus, severe asphyxia, Apgar score less than three in the fifth minute, as well as, presence of chromosomal abnormalities and congenital malformations. After obtaining the informed consent from the responsible, children were randomly allocated into the experimental (CS) and control (OS) groups. The dependent variables of this study respiratory rate (RR), peripheral oxygen saturation (SatpO2), heart rate (HR), blood pressure (BP) and pain, as well as, of the complementary variables, were collected through observation, the Premature Infant Pain Profile (PIPP) scale application and medical records consultation. During the data collection the period of variables observation were: T1 (immediately before suctioning), T2 (during suctioning), T3 (immediately after) and T4 (10 minutes after). To the statistical analysis were applied Fisher exact test, McNemar, Binomial, Student t test, Wilcoxon and Mann Whitney, settling at 0.05 level of significance.

Read the detailed description

Background: Patients who underwent tracheal intubation require intratracheal tube secretions sucking, and open system (OS) or closed system (CS) can be used, in order to preserve the tracheal tube permeability and an adequate ventilation and blood oxygenation. Respiratory and hemodynamic complications, as well as, presence of pain can be possibly experienced in newborns submitted to such procedure. Objectives: To compare clinical effects identified in infants undergoing intratracheal aspiration with OS or CS and verify the presence and intensity of pain in newborns during intratracheal suction, according to the system applied. Methods: Clinical, randomized, controlled, and crossover study, performed at two neonatal intensive care units after approval by the ethics committee. The sample consisted of infants from zero to seven days of age, 26 weeks of gestational age, and the exclusion criteria were use of mechanical ventilation with high-frequency oscillatory, in use of inhaled nitric oxide, in the acute phase of circulatory failure, central nervous system disorders, without reactivity to stimulus, severe asphyxia, Apgar score less than three in the fifth minute, as well as, presence of chromosomal abnormalities and congenital malformations. After obtaining the informed consent from the responsible, children were randomly allocated into the experimental (CS) and control (OS) gruops. The dependent variables of this study respiratory rate (RR), peripheral oxygen saturation (SatpO2), heart rate (HR), blood pressure (BP) and pain, as well as, of the complementary variables, were collected through observation, the Premature Infant Pain Profile (PIPP) scale application and medical records consultation. During the data collection the period of variables observation were: T1 (immediately before suctioning), T2 (during suctioning), T3 (immediately after) and T4 (10 minutes after). To the statistical analysis were applied Fisher exact test, McNemar, Binomial, Student t test, Wilcoxon and Mann Whitney, settling at 0.05 level of significance. Results: The studied groups did not show significant differences concerning the complementary variables. There were no statistically significant influences regarding the use of OS and CS of intratracheal suction in any of the dependent variables studied. Clinical consequences were observed with the use of both systems. It was demonstrated that immediately after intratracheal aspiration, the newborns had higher RR average in OS (58,4±9,6) group when compared to CS (56,2±5,9). Average values of SatpO2 were lower in OS (90,5±4,8) group compared to CS (91,6±4,6). Bradycardia was not verified in any group, however, there was a trend toward higher mean values of HR immediately after intratracheal aspiration, especially with OS (149,5±17,4) compared with the values identified before the procedure (143,2±18,1). There was low variation in the mean BP value in the two groups during the period of assessment, using both techniques. Regarding pain it was found that, with OS the average score was 7,9±4,4 and 6,5±3,8 with CS, demonstrating presence of mild to moderate pain. During the use of OS intratracheal aspiration higher pain scores (p=0,038) where verified in newborn females compared to males, this result was not observed when using CS. Conclusion: There was no statistically significant difference regarding the clinical implications, the presence and intensity of pain observed in ventilated newborns, according to the use of OS or CS to intratracheal aspiration.

Being tested
Tracheal aspiration close system
Keywords
Artificial Respiration, Suction, Pain, Nursing, Neonatology
Medical subject headings
Pain

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Tracheal aspiration close systemDevice

    Also called aspiration system, suctioning system, tracheal suctioning system

    use of closed system in comparision to open system to tracheal aspiration of nweborns, analyzing pain and climical variations associated.

Study groups in NCT01276873
Study groupRoleWhat they receive
Closed SystemExperimentalApplication of Tracheal aspiration closed system, controlled by the use of Open system to tracheal aspiraiton.
Receives: Device: Tracheal aspiration close system

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Pain response to tracheal suction

    There were no statistically significant influences regarding the use of open (OS) or close system (CS) to intratracheal suction in this dependent variable. With OS the average score was 7,9±4,4 and 6,5±3,8 with CS, demonstrating presence of mild to moderate pain.

    Measured: 48 hours

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted13 Jan 2011 (estimated)
  2. Study startJan 2010
  3. Primary completionJul 2010
  4. Study completionJul 2010

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Crossover
Masking
None (open label)
Enrolment
13 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT01276873
Sponsor's protocol ID
1094/09
Status last verified
Dec 2010
First submitted
11 Jan 2011
Last updated
13 Jan 2011
Results posted
No
Data monitoring committee
Yes

Questions about this trial

What is this clinical trial studying?

A study testing Tracheal aspiration close system in people with Adverse Reaction to Systemic Agents, run by Federal University of São Paulo at 2 sites in Brazil.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 13 Jan 2011. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 2 sites in Brazil. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Pain response to tracheal suction (time frame: 48 hours). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 13 Jan 2011.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in