Stress HyperglycaemiaPhase 2CLASSICNCT01440842

Closed Loop Glucose Control in Intensive Care Unit

A Phase 2 study testing Closed-loop insulin delivery and Standard IV insulin infusion sliding scale in people with Stress Hyperglycaemia, run by University of Cambridge at 1 site in United Kingdom.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to assess the efficacy, safety and feasibility of a computer-based control algorithm to control glucose levels in adults in intensive care unit, in comparison to standard care. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Stress Hyperglycaemia
What is being tested?
Closed-loop insulin delivery and Standard IV insulin infusion sliding scale
Type of treatment
Injection or infusion
Could I receive a placebo?
No, everyone receives an active treatment
How many people
24 participants
Where
1 siteUnited Kingdom
Study dates
May 2012to Oct 2012
Phase
Phase 2
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Aged 18 years and older
  • Admitted to Neurosciences critical care unit; (all patients are eligible regardless of admitting diagnosis except where specified under exclusion criteria)
  • Stay in intensive care unit expected to be at least 48 hours
  • At least one of the following conditions applies:
  • Not on insulin infusion and single confirmed reference blood glucose level > 10.0 mmol/l
  • Already on insulin infusion including those subjects with pre-existing diabetes.

Who cannot join

  • Patients with diabetic ketoacidosis or hyperosmolar hyperglycaemic non-ketotic coma (HONK)
  • Patients who are receiving therapeutic hypothermia
  • Known or suspected allergy to insulin
  • Any disease or condition which the investigator or treating physician feels would interfere with the trial or the safety of the patient (i.e. liver failure, other fatal organ failures)
  • Patients with significant abnormalities of blood clotting.
  • Moribund patients likely to die within 48 hours
  • Pregnancy

Trial locations and contacts

1 site in 1 country.

United Kingdom1 site
  • Cambridge University Hospitals NHS Foundation Trust, Addenbrookes Hospital

    Cambridge, Cambridgeshire, CB2 0QQ

Other trials in United Kingdom

About this study

Official title: An Open-label, Single-centre, Randomised Controlled Trial to Assess the Safety and Efficacy of Automated Closed-loop Blood Glucose Control in Comparison With Standard Care in Adults in Intensive Care Unit

The purpose of this study is to assess the efficacy, safety and feasibility of a computer-based control algorithm to control glucose levels in adults in intensive care unit, in comparison to standard care.

Read the detailed description

This will be a single centre, open-label, randomised, parallel design, feasibility study conducted at Neurosciences Critical Care Unit (NCCU), Addenbrooke's hospital, Cambridge, UK. Study will aim for 24 adult subjects (12 participants in each arm of the trial) and study will last for up to 49 hours in each subject. Subjects will have a commercially available Conformité Européenne(CE) marked subcutaneous glucose sensor inserted at the start of the study. Glucose data from the sensor will be transmitted to a small bedside tablet computer, containing the algorithm which will determine insulin infusion rates aimed at maintaining glucose level between 6.0 - 8.0 mmol/L. The advice from the algorithm will be sent to the infusion pump via USB cable and insulin will be delivered intravenously. The system will also deliver intravenous glucose via a second infusion pump at times of low glucose levels. Samples for reference glucose values will be obtained either from an arterial line or central venous cannula and will be analysed using standard blood gas analyser in real time. Subjects randomised to standard care will receive intravenous insulin based on current treatment guidelines at the Neurosciences critical care unit (NCCU), Addenbrooke's Hospital, Cambridge, UK.

The primary outcome is time spent in target glucose range between 6.0 to 8.0 mmol/L as recorded by reference glucose. Secondary outcomes are the time spent with glucose levels above and below target, as recorded by reference glucose, and sensor accuracy. Safety includes evaluation of significant hypoglycaemia and hyperglycaemia and other adverse events.

Being tested
Closed-loop insulin delivery, Standard IV insulin infusion sliding scale
Keywords
Stress hyperglycaemia

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Closed-loop insulin deliveryOther

    Intravenous infusion delivery of Actrapid insulin and dextrose, dose calculated by Model Predictive Control (MPC) algorithm, based on continuous glucose sensor readings.

  • Standard IV insulin infusion sliding scaleOther

    Standard intravenous insulin infusion sliding scale as per intensive care unit protocol.

Study groups in NCT01440842
Study groupRoleWhat they receive
Closed-loop (Model Predictive Control Algorithm)ExperimentalNot described
Receives: Other: Closed-loop insulin delivery
Open loop (Standard treatment)Active comparatorNot described
Receives: Other: Standard IV insulin infusion sliding scale

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Percentage of glucose values in target range (6.0 - 8.0 mmol/l) as recorded by reference glucose measurements.

    Definition of Reference glucose = Blood Glucose level measured by arterial blood gas analyzer.

    During the study subjects will have the reference glucose measured every 1 hour and this is used for the calculation percentage glucose values.

    Measured: Up to 49 hours in each subject.

Secondary outcomes (8)
  1. Percentage of glucose values in range (4.0 - 10.0 mmol/l) as recorded by reference glucose measurements.

    Definition of Reference glucose = Blood Glucose level measured by arterial blood gas analyzer.

    During the study subjects will have the reference glucose measured every 1 hour and this is used for the calculation percentage glucose values.

    Measured: Up to 49 hours in each subject

  2. Percentage of glucose values <4.0 mmol/l and >8.0 mmol/l as recorded by reference glucose measurements.

    Definition of Reference glucose = Blood Glucose level measured by arterial blood gas analyzer.

    During the study subjects will have the reference glucose measured every 1 hour and this is used for the calculation percentage glucose values.

    Measured: Up to 49 hours in each subject.

  3. Percentage of glucose values > 10.0 mmol/l as recorded by reference glucose measurements.

    Definition of Reference glucose = Blood Glucose level measured by arterial blood gas analyzer.

    During the study subjects will have the reference glucose measured every 1 hour and this is used for the calculation percentage glucose values.

    Measured: Up to 49 hours in each subject,

  4. Mean and standard deviation of reference glucose measured using arterial blood gas analyzer

    Definition of Reference glucose = Blood Glucose level measured by arterial blood gas analyzer. This is reported in mmol/L

    During the study subjects will have the reference glucose measured every 1 hour and this is used for the calculation of mean (mmol/L) and standard deviation which will reflect the efficasy of closed loop insulin delivery.

    Measured: Up to 49 hours in each subject.

  5. Mean and median absolute and relative difference between matched pairs of subcutaneous glucose sensor and reference plasma glucose.

    For a given subcutaneous glucose sensor value, difference between the sensor and the reference glucose will be calculated. Example - Reference glucose 10 mmol/L and sensor glucose 12 mmol/L, therefore absolute difference will be 2 mmol/L. Mean and median of these deviations will be reported in mmol/L. The term relative implies that data has been converted to a percentage deviation from reference glucose.

    Measured: Up to 49 hours in each subject.

  6. Time to reach target glucose

    Measured: Up to 49 hours in each subject.

  7. Insulin infusion rates

    Measured: Up to 49 hours in each subject.

  8. Frequency and magnitude of significant hypoglycaemic (< 3.0 mmol/L), severe hypoglycaemic (<2.0 mmol/L) and significant hyperglycaemic (> 15mmol/l) episodes.

    Measured: Up to 49 hours in each subject.

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted27 Sep 2011 (estimated)
  2. Study startMay 2012
  3. Primary completionOct 2012
  4. Study completionOct 2012

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase 2
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Parallel groups
Masking
None (open label)
Enrolment
24 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT01440842
Sponsor's protocol ID
CLASSIC 01
Acronym
CLASSIC
Status last verified
Dec 2012
First submitted
22 Sep 2011
Last updated
3 Jan 2013
Results posted
No
Data monitoring committee
No

Publications

  • Leelarathna L, English SW, Thabit H, Caldwell K, Allen JM, Kumareswaran K, Wilinska ME, Nodale M, Mangat J, Evans ML, Burnstein R, Hovorka R. Feasibility of fully automated closed-loop glucose control using continuous subcutaneous glucose measurements in critical illness: a randomized controlled trial. Crit Care. 2013 Jul 24;17(4):R159. doi: 10.1186/cc12838.

    PubMed 23883613

Questions about this trial

What is this clinical trial studying?

A Phase 2 study testing Closed-loop insulin delivery and Standard IV insulin infusion sliding scale in people with Stress Hyperglycaemia, run by University of Cambridge at 1 site in United Kingdom.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 3 Jan 2013. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United Kingdom. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

What does this phase mean?

Phase 2 studies give the treatment to a larger group to see whether it has the intended effect and to keep checking safety.

Not every study follows the same path, and some studies (observational studies, device studies) have no phase.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Percentage of glucose values in target range (6.0 - 8.0 mmol/l) as recorded by reference glucose measurements. (time frame: Up to 49 hours in each subject.). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT01440842, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 3 Jan 2013.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in