Diabetes MellitusPhase 4PolarisNCT01504620
Performance Evaluation of Blood Glucose Monitoring Systems
A Phase 4 study testing Sugar infusion and insulin infusion in people with Diabetes Mellitus, run by IKFE Institute for Clinical Research and Development at 1 site in Germany.
Trial summary
The goal of this single-center, comparative, open label, in-vitro diagnostic device performance evaluation was to investigate the performance of two sanofi-aventis BGMSs (BGStar® and iBGStar®) with regard to system accuracy (Polaris), precision and Hct influence (Helios). Read more about this study
- Who is it for?
- Ages 18 to 75 yearsHealthy volunteers can take part
- Conditions the study is for
- Diabetes Mellitus
- What is being tested?
- Sugar infusion and insulin infusion
- Type of treatment
- Injection or infusion
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 106 participants
- Where
- 1 siteGermany
- Study dates
- Jan 2011to Jun 2011
- Phase
- Phase 4
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 to 75 years
- All sexes
- Healthy volunteers accepted
- Adult, Older adult
Who may be able to join
- healthy subjects or patients with type 1 or type 2 diabetes
- screening glucose levels fall into required glucose ranges
- screening hematocrit values between 36 % to 55 %
Who cannot join
- history of hypotension during blood draws
- intake of drugs known to interfere with blood glucose readings
- biochemical safety parameters outside of reference ranges
- Hb < 11 g/dL
- lack of compliance
- history of frequent hypoglycemia
Trial locations and contacts
1 site in 1 country.
Germany1 site
IKFE - Institute for Clinical Research and Development
Mainz, 55116
About this study
Official title: Performance Evaluation of BGStar and iBGStar in Terms of Accuracy (Polaris), Intra-assay Precision, and Hematocrit Interference (Helios Substudy) Following ISO15197 and TNO Guidelines
The goal of this single-center, comparative, open label, in-vitro diagnostic device performance evaluation was to investigate the performance of two sanofi-aventis BGMSs (BGStar® and iBGStar®) with regard to system accuracy (Polaris), precision and Hct influence (Helios).
Read the detailed description
The main study (Polaris) investigated blood glucose measuring accuracy and intra-assay precision while the substudy (Helios) investigated potential haematocrit interference of sanofi-aventis BGMSs BGStar® and iBGStar® and partly of competitor BGMSs according to ISO 15197 [1] and TNO 2001 Quality Guideline [2], respectively.
The primary objective of the system accuracy evaluation is to identify the number of blood glucose readings of both sanofi-aventis devices within the ISO 15197 area of acceptable accuracy of ± 20% [DEV%] for blood samples with glucose concentration > 75 mg/dL (4.2 mmol/L) and within the ISO 15197 area of acceptable accuracy of ± 15 mg/dL (± 0.83 mmol/L) for blood samples with glucose concentration < 75 mg/dL (4.2 mmol/L) as compared to the StatStrip® Connectivity reference method. The primary objective of the haematocrit interference evaluation is to demonstrate that the BGStar® and iBGStar® devices meet the Hct interference requirements according to TNO 2001 Quality Guideline when used to measure BG of venous whole blood as compared to a venous plasma reference method (YSI 2300 STAT Plus™) at Hct levels between 35 - 55% in terms of:(1) a mean deviation not exceeding ± 10% [DEV%] for hyperglycaemic glucose concentrations (150 - 180, 250 - 280, 320 - 350 mg/dL) and (2)a mean deviation not exceeding ± 18 mg/dL (1.0 mmol/L) [DEVtotal] for hypo- and normoglycaemic glucose values (50 - 60, 100 - 120 mg/dL) compared to the reference method at Hct levels between 35 - 55%.
- Conditions
- Diabetes Mellitus
- Being tested
- Sugar infusion, insulin infusion
- Keywords
- diabetes mellitus, blood glucose self-assessment, Accuracy, Intra-Assay precision, Hematocrit interference
- Medical subject headings
- Diabetes Mellitus
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Sugar infusionOther
Also called 5 % glucose solution
Infusion of glucose to achieve high blood glucose levels
insulin infusionDrug
Also called insuman rapid
i.v. infusion of insulin
| Study group | Role | What they receive |
|---|---|---|
| Sugar infusion | Other | Diagnostic assessment of blood glucose by means of different devices Receives: Other: Sugar infusion |
| insulin infusion | Experimental | infusion of insulin to achieve low glucose levels Receives: Drug: insulin infusion |
What is being measured
The primary outcome is the main question the study is designed to answer.
Accuracy of the blood glucose meter in comparison to a standard reference
After start of the experiment, blood glucose will be measured by means of the test devices and blood will be drawn for assessment of venous blood glucose concentration with a standard laboratory reference method
Measured: within 10 min up to 300 min after start of experiment
Secondary outcomes (2)
Intra-Assay precision
Blood samples from 5 individuals with different levels of blood glucose were each measured with 10 meters of both investigational devices, the BGStar® and iBGStar®, ten times resulting in a total number of 500 measurements.
Measured: within 10 min up to 300 min after start of experiment
Hematocrit Interference (Helios)
One patient provided blood and artificial samples are generated from venous heparinized blood. These blood samples were adjusted to five different BG levels (2.8 mmol/L, 5.6 mmol/L, 8.3 mmol/L, 13.9 mmol/L, 19.4 mmol/L) at five different Hct levels (35%, 40%, 45%, 50%, 55%) to fulfill the range required by the TNO 2001 and tested by all devices.
Measured: within 10 min up to 300 min after start of the experiment
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted5 Jan 2012 (estimated)
- Study startJan 2011
- Primary completionApr 2011
- Study completionJun 2011
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Other trials for Diabetes MellitusStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase 4
- Primary purpose
- Diagnosis
- Allocation
- Non-randomised
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 106 participants (actual)
Sponsor and investigators
- Lead sponsor
- IKFE Institute for Clinical Research and Development
- Sponsor type
- University, hospital or other
- Collaborators
- Sanofi-Synthelabo
- Responsible party
- Sponsor
- Investigators
- Andreas Pfützner, MD, PhD, Principal investigator, IKFE Institute for Clinical Research and Development
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT01504620
- Sponsor's protocol ID
- BGSTA_C_05445
- Other IDs
- SAN-HCT-002 Helios, SAN-BGM-002 Precision
- Acronym
- Polaris
- Status last verified
- Dec 2011
- First submitted
- 27 Dec 2011
- Last updated
- 5 Jan 2012
- Results posted
- No
- Data monitoring committee
- No
Study at a glance
From the registry record, last updated by the sponsor on 5 Jan 2012. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Performance Evaluation of BGStar and iBGStar in Terms of Accuracy (Polaris), Intra-assay Precision, and Hematocrit Interference (Helios Substudy) Following ISO15197 and TNO Guidelines
- ClinicalTrials.gov ID
- NCT01504620
- Acronym / sponsor ID
- Polaris / BGSTA_C_05445
- Conditions
- Diabetes Mellitus
- Keywords
- diabetes mellitus, blood glucose self-assessment, Accuracy, Intra-Assay precision, Hematocrit interference
- Phase
- Phase 4
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 106 participants (actual)
- Lead sponsor
- IKFE Institute for Clinical Research and DevelopmentUniversity, hospital or other
- Collaborators
- Sanofi-Synthelabo
- Study start
- Jan 2011
- Primary completion
- Apr 2011 (actual)
- Study completion
- Jun 2011 (actual)
- First posted
- 5 Jan 2012 (estimated)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The goal of this single-center, comparative, open label, in-vitro diagnostic device performance evaluation was to investigate the performance of two sanofi-aventis BGMSs (BGStar® and iBGStar®) with regard to system accuracy (Polaris), precision and Hct influence (Helios).
Detailed description
The main study (Polaris) investigated blood glucose measuring accuracy and intra-assay precision while the substudy (Helios) investigated potential haematocrit interference of sanofi-aventis BGMSs BGStar® and iBGStar® and partly of competitor BGMSs according to ISO 15197 [1] and TNO 2001 Quality Guideline [2], respectively.
The primary objective of the system accuracy evaluation is to identify the number of blood glucose readings of both sanofi-aventis devices within the ISO 15197 area of acceptable accuracy of ± 20% [DEV%] for blood samples with glucose concentration > 75 mg/dL (4.2 mmol/L) and within the ISO 15197 area of acceptable accuracy of ± 15 mg/dL (± 0.83 mmol/L) for blood samples with glucose concentration < 75 mg/dL (4.2 mmol/L) as compared to the StatStrip® Connectivity reference method. The primary objective of the haematocrit interference evaluation is to demonstrate that the BGStar® and iBGStar® devices meet the Hct interference requirements according to TNO 2001 Quality Guideline when used to measure BG of venous whole blood as compared to a venous plasma reference method (YSI 2300 STAT Plus™) at Hct levels between 35 - 55% in terms of:(1) a mean deviation not exceeding ± 10% [DEV%] for hyperglycaemic glucose concentrations (150 - 180, 250 - 280, 320 - 350 mg/dL) and (2)a mean deviation not exceeding ± 18 mg/dL (1.0 mmol/L) [DEVtotal] for hypo- and normoglycaemic glucose values (50 - 60, 100 - 120 mg/dL) compared to the reference method at Hct levels between 35 - 55%.
Study design
- Primary purpose
- Diagnosis
- Allocation
- Non-randomised
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 2
| Arm | Type | Receives |
|---|---|---|
| Sugar infusion | Other | Other: Sugar infusionDiagnostic assessment of blood glucose by means of different devices |
| insulin infusion | Experimental | Drug: insulin infusioninfusion of insulin to achieve low glucose levels |
Who may qualify
- Ages
- 18 to 75 years
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (3 listed)
- healthy subjects or patients with type 1 or type 2 diabetes
- screening glucose levels fall into required glucose ranges
- screening hematocrit values between 36 % to 55 %
Key exclusion criteria (6 listed)
- history of hypotension during blood draws
- intake of drugs known to interfere with blood glucose readings
- biochemical safety parameters outside of reference ranges
- Hb < 11 g/dL
- lack of compliance
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Sugar infusionAlso: 5 % glucose solutionInfusion of glucose to achieve high blood glucose levels | Other | Sugar infusion |
| insulin infusionAlso: insuman rapidi.v. infusion of insulin | Drug | insulin infusion |
Primary outcome
Accuracy of the blood glucose meter in comparison to a standard reference
Time frame: within 10 min up to 300 min after start of experiment
After start of the experiment, blood glucose will be measured by means of the test devices and blood will be drawn for assessment of venous blood glucose concentration with a standard laboratory reference method
2 secondary outcomes
Intra-Assay precision
Time frame: within 10 min up to 300 min after start of experiment
Hematocrit Interference (Helios)
Time frame: within 10 min up to 300 min after start of the experiment
Results and safety
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 5 Jan 2012.
Related trials for Diabetes Mellitus
Not yet recruitingPhase 3
Efficacy and Safety of Once-Weekly SHR-3167 Compared With Once-Daily Insulin Glargine in Adults With Insulin-Naive Type 2 Diabetes
Type 2 Diabetes
Active, not recruiting
Digital Reinforcement of Motivational Interviewing Micro-Skills Among Primary Care Healthcare Professionals
Hypertension (Without Type 2 Diabetes Mellitus)
Recruiting
Effects of Lean Pork Loin Intake on Protein Homeostasis and Glucose Regulation in Prediabetic Adults
Hyperglycaemia (Non Diabetic), Prediabetes
Recruiting
PAW-Hydrogel for Chronic Wound Healing
Diabetic Foot Ulcer (DFU), Pressure Ulcer (PU)
Recruiting
Investigating Vascular Properties of HEMI and SPG Signals in Individuals With or at Risk for Chronic Kidney Disease
Chronic Kidney Disease, Hypertension (HTN)
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Andreas Pfützner, MD, PhD principal investigator
IKFE Institute for Clinical Research and Development
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A Phase 4 study testing Sugar infusion and insulin infusion in people with Diabetes Mellitus, run by IKFE Institute for Clinical Research and Development at 1 site in Germany.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 3 inclusion and 6 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 5 Jan 2012. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Germany. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
What does this phase mean?
Phase 4 studies take place after a treatment is approved, to learn more about its long-term benefits and risks in everyday use.
Not every study follows the same path, and some studies (observational studies, device studies) have no phase.
Are results available for this study?
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Accuracy of the blood glucose meter in comparison to a standard reference (time frame: within 10 min up to 300 min after start of experiment). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 5 Jan 2012.