Comparison Evaluation of Blood Glucose Meter Systems

A study testing CONTOUR® NEXT EZ BGMS, FreeStyle Freedom Lite® BGMS and ACCU-CHEK® Aviva BGMS in people with Diabetes, run by Ascensia Diabetes Care at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study was to evaluate the performance of one investigational Bayer Blood Glucose meter (with investigational Bayer test strips) and four Blood Glucose Monitoring Systems from other companies. Read more about this study

Who is it for?
Ages 18 years and olderHealthy volunteers can take part
Conditions the study is for
Diabetes
What is being tested?
CONTOUR® NEXT EZ BGMS, FreeStyle Freedom Lite® BGMS and ACCU-CHEK® Aviva BGMS
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
146 participants
Where
1 siteUnited States
Study dates
Mar 2012to Mar 2012
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Males and females
  • 18 years of age or older
  • Willing to complete all study procedures

Who cannot join

  • Blood Borne infections like hepatitis or HIV or infections such as tuberculosis
  • Hemophilia or any other bleeding disorder
  • Pregnancy

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Bayer HealthCare LLC, Diabetes Care

    Mishawaka, Indiana, 46544

Other trials in United States

About this study

Official title: Evaluation of Blood Glucose Meter Systems - NACT Study

The purpose of this study was to evaluate the performance of one investigational Bayer Blood Glucose meter (with investigational Bayer test strips) and four Blood Glucose Monitoring Systems from other companies. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer). Performance of the five systems was evaluated across the glucose range of the BGMSs using capillary blood. All testing and lancing were performed by study staff and some blood samples were glycolyzed to lower the glucose levels.

Conditions
Diabetes
Being tested
CONTOUR® NEXT EZ BGMS, FreeStyle Freedom Lite® BGMS, ACCU-CHEK® Aviva BGMS, TRUEtrack® BGMS, OneTouch® Ultra®2 BGMS
Medical subject headings
Diabetes Mellitus

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • CONTOUR® NEXT EZ BGMSDevice

    Study staff performed Blood Glucose (BG) testing with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). Performance of CONTOUR® NEXT EZ BGMS was evaluated across the glucose range of the BGMs. All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

  • FreeStyle Freedom Lite® BGMSDevice

    Study staff performed Blood Glucose (BG) testing with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). Performance of FreeStyle Freedom Lite® BGMS was evaluated across the glucose range of the BGMs. All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

  • ACCU-CHEK® Aviva BGMSDevice

    Study staff performed Blood Glucose (BG) testing with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). Performance of ACCU-CHEK® Aviva BGMS was evaluated across the glucose range of the BGMs. All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

  • TRUEtrack® BGMSDevice

    Study staff performed Blood Glucose (BG) testing with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). Performance of TRUEtrack® BGMS was evaluated across the glucose range of the BGMs. All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

  • OneTouch® Ultra®2 BGMSDevice

    Study staff performed Blood Glucose (BG) testing with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). Performance of OneTouch® Ultra®2 BGMS was evaluated across the glucose range of the BGMs. All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

Study groups in NCT01548885
Study groupRoleWhat they receive
Study Staff Test BGMSsExperimentalAll testing and lancings were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems(BGMS): FreeStyle Freedom Lite® BGMS; ACCU-CHEK® Aviva BGMS; TRUEtrack® BGMS; OneTouch® Ultra®2 BGMS; CONTOUR® NEXT EZ BGMS.
Receives: Device: CONTOUR® NEXT EZ BGMS; Device: FreeStyle Freedom Lite® BGMS; Device: ACCU-CHEK® Aviva BGMS; Device: TRUEtrack® BGMS; Device: OneTouch® Ultra®2 BGMS

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range

    Using the overall Blood Glucose(BG) range (24 to 386mg/dL), the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD is calculated from the sum of all |(BG meter)-(BG reference)| / (BG Reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all five BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value . Higher MARD value indicates larger difference between meter value and the reference value.

    Measured: 10 hours

Secondary outcomes (1)
  1. MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Low Glucose Range (<70mg/dL)

    Using fresh and glycolyzed samples with Blood Glucose(BG) below 70 mg/dL, the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD is calculated from the sum of all |(BG meter)-(BG reference)| / (BG Reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all five BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD value indicates smaller difference between meter value and the reference value . Higher MARD value indicates larger difference between meter value and the reference value.

    Measured: 10 hours

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted8 Mar 2012 (estimated)
  2. Study startMar 2012
  3. Primary completionMar 2012
  4. Study completionMar 2012

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
146 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT01548885
Sponsor's protocol ID
CTD-PRO-2012-001-01
Status last verified
Jan 2016
First submitted
6 Mar 2012
Last updated
29 Feb 2016
Data monitoring committee
No

Questions about this trial

What is this clinical trial studying?

A study testing CONTOUR® NEXT EZ BGMS, FreeStyle Freedom Lite® BGMS and ACCU-CHEK® Aviva BGMS in people with Diabetes, run by Ascensia Diabetes Care at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 29 Feb 2016. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 6 Aug 2013). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range (time frame: 10 hours). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 29 Feb 2016.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in