Cardiovascular DiseaseNCT01599741

X-ray Dose Reduction Study for Endovascular Interventional Radiology

A study testing Low-dose DSA (83% reduction compared to normal dose) with novel X-ray imaging technology. and Normal dose DSA with conventional X-ray imaging technology in people with Cardiovascular Disease, run by Philips Clinical & Medical Affairs Global at 1 site in Netherlands.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

ClarityIQ is a novel X-ray imaging technology, that combines advanced real-time image noise reduction algorithms, with state-of-the-art hardware to reduce patient entrance dose significantly. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Cardiovascular Disease
What is being tested?
Low-dose DSA (83% reduction compared to normal dose) with novel X-ray imaging technology. and Normal dose DSA with conventional X-ray imaging technology
Type of treatment
Surgery or procedure
Could I receive a placebo?
No, everyone receives an active treatment
How many people
51 participants
Where
1 siteNetherlands
Study dates
Jun 2012to May 2013
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Patients older than 18 years of age undergoing digital subtraction angiography for diagnosis or treatment of iliac artery disease.

Who cannot join

  • Patients not willing or unable to give consent to participate
  • Patients already involved in a clinical trial
  • Patients under the age of 18
  • Pregnant or breastfeeding women
  • Patients with kidney disease (eGFR < 60)

Trial locations and contacts

1 site in 1 country.

Netherlands1 site
  • St. Antonius Ziekenhuis

    Nieuwegein, 3430 EM

Other trials in Netherlands

About this study

Official title: Clinical Validation and Evaluation of X-ray Image Processing for Endovascular Digital Subtraction Angiography and Intervention

ClarityIQ is a novel X-ray imaging technology, that combines advanced real-time image noise reduction algorithms, with state-of-the-art hardware to reduce patient entrance dose significantly. This is realized by anatomy-specific optimization of the full acquisition chain (grid switch, beam filtering, pulse width, spot size, detector and image processing engine) for every clinical task individually. Furthermore, smaller focal spot sizes, shorter pulses and the introduction of automatic real-time motion compensation in subtraction imaging are used, which are known to positively influence image quality . The final effect on the clinical image quality in peripheral angiography is investigated in this study.

Read the detailed description

X-ray dose and image quality are related by laws of physics. Low dose and high image quality cannot be achieved at the same time. However, image processing algorithms can help an x-ray system to acquire images with lower dose without influencing image quality or achieving higher image quality with equal dose. The primary aim of this study is to verify if the diagnostic image quality is equal or better when using a significant reduction in X-ray dose for endovascular digital subtraction angiography in combination with a novel imaging technology compared to using normal dose with conventional image technology.

Being tested
Low-dose DSA (83% reduction compared to normal dose) with novel X-ray imaging technology., Normal dose DSA with conventional X-ray imaging technology
Keywords
X-rays, Endovascular angiography, Image processing
Medical subject headings
Cardiovascular Diseases

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Low-dose DSA (83% reduction compared to normal dose) with novel X-ray imaging technology.Radiation

    Also called ClarityIQ

    Digital Subtraction Angiography (DSA) of iliac artery with reduced dose settings (83% dose reduction expected) in combination with novel X-ray imaging technology.

  • Normal dose DSA with conventional X-ray imaging technologyRadiation

    Also called AlluraXper

    Digital Subtraction Angiography (DSA) of iliac artery with normal dose settings in combination with conventional X-ray imaging technology.

Study groups in NCT01599741
Study groupRoleWhat they receive
ClarityIQExperimentalLow-dose DSA (83% reducation compared to normal dose) with novel X-ray imaging technology.
Receives: Radiation: Low-dose DSA (83% reduction compared to normal dose) with novel X-ray imaging technology.
AlluraXperActive comparatorNormal dose DSA with conventional X-ray imaging technology.
Receives: Radiation: Normal dose DSA with conventional X-ray imaging technology

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Image Quality

    Overall proportion where diagnostic image quality of ClarityIQ is scored equal or better compared to AlluraXper by the blinded reviewers. Reading is performed by simultaneous visual comparison of image quality of AlluraXper and ClarityIQ by multiple blinded readers. All blinded readers have rated all side-by-side presented images. The images are presented in a random order and blinded to the reader. The hypothesis is that the overall proportion where diagnostic image quality of ClarityIQ is scored equal or better is ≥ than 0.80. Combined for all raters, the lower bound of the one-sided 95% CI (lower bound of the two-sided 90% CI) is used.

    Measured: 1 Day

Secondary outcomes (2)
  1. Radiation Dose Measurements: Dose Area Product (DAP)

    Percentage dose change of ClarityIQ vs. AlluraXper in DAP calculated by DAP/frame for DSA. Negative change means a reduction in dose for ClarityIQ vs. AlluraXper.

    DAP was measured during the ClarityIQ and AlluraXper runs during the procedure.

    Measured: Participants were followed for the duration of the procedure

  2. Radiation Dose Measurements: Air Kerma (AK)

    Percentage dose change of ClarityIQ vs. AlluraXper in AK calculated by AK/frame for DSA. Negative change means a reduction in dose for ClarityIQ vs. AlluraXper.

    AK was measured during the ClarityIQ and AlluraXper runs during the procedure.

    Measured: Participants were followed for the duration of the procedure

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted16 May 2012 (estimated)
  2. Study startJun 2012
  3. Primary completionMay 2013
  4. Study completionMay 2013

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Non-randomised
Masking
None (open label)
Enrolment
51 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT01599741
Sponsor's protocol ID
XCY607-100093
Status last verified
Mar 2022
First submitted
8 May 2012
Last updated
8 Mar 2022
Data monitoring committee
No

Questions about this trial

What is this clinical trial studying?

A study testing Low-dose DSA (83% reduction compared to normal dose) with novel X-ray imaging technology. and Normal dose DSA with conventional X-ray imaging technology in people with Cardiovascular Disease, run by Philips Clinical & Medical Affairs Global at 1 site in Netherlands.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 1 inclusion and 5 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 8 Mar 2022. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Netherlands. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 5 Mar 2014). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Image Quality (time frame: 1 Day). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 8 Mar 2022.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in