Comparison Evaluation of Blood Glucose Meter Systems in a Clinical Setting

A study testing Bayer G3 / Tatsu System, Accu-Chek® Aviva Nano Meter/Accu-Chek® Aviva Test Strips and Freestyle Lite® Meter and Test Strips with ZipwikTM tabs in people with Diabetes, run by Ascensia Diabetes Care at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study was to test the performance of an investigational Bayer Blood Glucose (BG) meter and five additional Blood Glucose Monitoring Systems (BGMSs). The goal was to obtain glucose samples from subjects with diabetes across the meter operating ranges. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Diabetes
What is being tested?
Bayer G3 / Tatsu System, Accu-Chek® Aviva Nano Meter/Accu-Chek® Aviva Test Strips and Freestyle Lite® Meter and Test Strips with ZipwikTM tabs
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
146 participants
Where
1 siteUnited States
Study dates
May 2012to Jun 2012
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Diagnosis of type 1 or type 2 diabetes mellitus
  • 18 years of age or older
  • Willing to follow all study procedures

Who cannot join

  • Received an investigational drug or device within the last 30 days
  • Current use or recent exposure to any prescription medication within 24 hours prior to screening that in the opinion of the investigator could have an influence on the subject's ability to participate in the study or on the performance on the test device
  • Positive pregnancy test obtained at screening (test performed for females only, except if surgically sterile or three years past menopause)
  • Hematocrit <32% or >55%
  • Inadequate veins (in the opinion of the investigator or designated study staff) for venous blood withdrawal or intravenous catheter insertion
  • Symptomatic coronary artery disease with a history of angina, or history of a myocardial infarction or coronary intervention (e.g. percutaneous transluminal coronary angioplasty [PTCA], stent placement) or coronary artery bypass graft (CABG) within the past 6 months
Show all 14Show fewer
  • Cerebrovascular accident within the past 6 months
  • Diagnosis of the following diabetic autonomic neuropathies: symptomatic orthostatic hypotension, heart rate anomalies, gastroparesis
  • Being treated for malignancy, except basal cell or squamous cell skin cancers
  • Major surgical operation within 30 days prior to screening
  • Active alcohol abuse or substance abuse (as judged by the principal investigator)
  • Severe mental illness, mental incapacity, unwillingness or language barriers precluding adequate understanding or cooperation (as judged by principal investigator)
  • Other medical conditions that would pose safety concerns, interfere with study conduct or seriously compromise study integrity, as judged by principal investigator (reason for exclusion will be clearly documented by investigator or designee)
  • Working for a competitive medical device company, or having an immediate family member or someone who is not a family member but is living within the household of someone who works for such a company

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Rainier Clinical Research Center Inc

    Renton, Washington, 98057

Other trials in United States

About this study

Official title: Clinical Evaluation of Bayer G3 / Tatsu System and BGMs From Other Companies

The purpose of this study was to test the performance of an investigational Bayer Blood Glucose (BG) meter and five additional Blood Glucose Monitoring Systems (BGMSs). The goal was to obtain glucose samples from subjects with diabetes across the meter operating ranges. A capillary sample was collected from each subject at 3 different times during the visit to obtain natural capillary blood samples with a range of glucose concentrations. Some blood samples were glycolyzed to lower the glucose concentration levels and glucose solution was added to other samples to raise glucose concentration levels. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer). All lancings and testing were performed by study staff.

Conditions
Diabetes
Being tested
Bayer G3 / Tatsu System, Accu-Chek® Aviva Nano Meter/Accu-Chek® Aviva Test Strips, Freestyle Lite® Meter and Test Strips with ZipwikTM tabs, OneTouch® Ultra®2 / OneTouch® Ultra® Blue Test Strips, One Touch® VerioTM Pro/One Touch® VerioTM Test Strips
Medical subject headings
Diabetes Mellitus

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Bayer G3 / Tatsu SystemDevice

    Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

  • Accu-Chek® Aviva Nano Meter/Accu-Chek® Aviva Test StripsDevice

    Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

  • Freestyle Lite® Meter and Test Strips with ZipwikTM tabsDevice

    Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

  • OneTouch® Ultra®2 / OneTouch® Ultra® Blue Test StripsDevice

    Study staff performed Blood Glucose (BG) tested with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

  • One Touch® VerioTM Pro/One Touch® VerioTM Test StripsDevice

    Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

  • Truetrack® Meter/Truetrack® Test StripsDevice

    Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

Study groups in NCT01614613
Study groupRoleWhat they receive
Intended BGM UsersExperimentalSubjects were assigned to subgroups according to their initial glucose levels. SUBGROUP 1 goal: safely decrease subject glucose levels during the visit. SUBGROUP 2 goal: safely raise subject glucose levels. Also, blood samples were modified to obtain glucose concentrations across the needed glucose ranges while maintaining subject safety. Staff tested the blood samples using the 6 Blood Glucose Monitoring Systems. Bayer G3/Tatsu System;Accu-Chek® Aviva Nano Meter/Accu-Chek® Aviva Test Strips; Freestyle Lite® Meter and Test Strips with ZipwikTM tabs;OneTouch® Ultra®2 / OneTouch® Ultra® Blue Test Strips;One Touch® VerioTM Pro/One Touch® VerioTM Test Strips;Truetrack® Meter/Truetrack® Test Strips
Receives: Device: Bayer G3 / Tatsu System; Device: Accu-Chek® Aviva Nano Meter/Accu-Chek® Aviva Test Strips; Device: Freestyle Lite® Meter and Test Strips with ZipwikTM tabs; Device: OneTouch® Ultra®2 / OneTouch® Ultra® Blue Test Strips; Device: One Touch® VerioTM Pro/One Touch® VerioTM Test Strips; Device: Truetrack® Meter/Truetrack® Test Strips

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. MAD Mean Absolute Value of the Difference Between Blood Glucose Meter (BGMS) Results and Corresponding YSI Blood Glucose (BG) Results in the Low BG Range(<70 mg/dL)

    Using samples with BG < 70 mg/dL, the Mean Absolute Value of the Differences Between BGM system readings and the YSI laboratory reference values were compared. MAD was calculated from the sum of all |(BG meter)-(BG reference)| assessments, divided by the number of assessments. Each evaluable sample was tested on all 6 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MAD value indicates smaller difference between meter value and the reference value. Higher MAD value indicates larger difference between meter value and the reference value.

    Measured: 6 hours

Secondary outcomes (4)
  1. MAD Mean Absolute Value of the Difference Between Blood Glucose Meter (BGMS) Results and Corresponding YSI Blood Glucose (BG) Results in the BG Range 70 to 180 mg/dL

    Using samples with BG 70 mg/dL to 180 mg/dL, the Mean Absolute Value of the Differences Between BGM system readings and the YSI laboratory reference values were compared. MAD was calculated from the sum of all|(BG meter)-(BG reference)| assessments, divided by the number of assessments. Each evaluable sample was tested on all 6 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MAD value indicates smaller difference between meter value and the reference value. Higher MAD value indicates larger difference between meter value and the reference value.

    Measured: 6 hours

  2. MAD Mean Absolute Value of the Difference Between Blood Glucose Meter (BGMS) Results and Corresponding YSI Blood Glucose (BG) Results in the BG Range >180 mg/dL

    Using samples with BG >180 mg/dL, the Mean Absolute Value of the Differences Between BGM system readings and the YSI laboratory reference values was compared. MAD was calculated from the sum of all |(BG meter)-(BG reference)| assessments, divided by the number of assessments. Each evaluable sample was tested on all 6 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MAD value indicates smaller difference between meter value and the reference value. Higher MAD value indicates larger difference between meter value and the reference value.

    Measured: 6 hours

  3. Standard Deviations of BGMS Differences (Between BGM Meter Readings and the YSI Laboratory Reference Values Across the BG Range of All Evaluable Samples 21 mg/dL to 496 mg/dL)

    Variability of each meter system was determined by calculating the Standard Deviation derived from each meter's differences between Blood Glucose Meter (BGMS)results and corresponding YSI Blood Glucose (BG) results.

    Measured: 6 hours

  4. Number of Subject Responses 'Strongly Agree' 'Agree' or 'Neutral' With Questionnaire Statements

    Subjects responded to statements about meter accuracy and diabetes management. Subject responses: 'Strongly Agree' 'Agree' 'Neutral' 'Disagree' or 'Strongly Disagree'or No Response. □1ACCURACY HELPS 1A-with my ability to talk with my HCP. 1B-my satisfaction with my self monitoring of diabetes. 1C-with my ability to manage my diabetes. 1D-prevent low BG. 1E-understand how food/exercise affects low BG. □2 I WOULD USE ONLY the meter and strips my insurance company pays for, even if a more accurate meter was available. □3 I USE MY CURRENT METER BECAUSE 3A-my HCP gave it to me. 3B-my insurance company covers the strips. 3C-I think it is the most accurate meter. □4 I WOULD SWITCH meters for a more accurate meter. □5BEING ABLE TO APPLY MORE BLOOD IS IMPORTANT 5A-as I have wasted test strips by not having enough blood to fill the strip 5B-as this would save test strips 5C-I would switch to a meter with this feature

    Measured: 1 hour

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted8 Jun 2012 (estimated)
  2. Study startMay 2012
  3. Primary completionJun 2012
  4. Study completionJun 2012

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
146 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT01614613
Sponsor's protocol ID
CTD-2010-009-02
Status last verified
Jan 2016
First submitted
13 May 2012
Last updated
29 Feb 2016
Data monitoring committee
No

Questions about this trial

What is this clinical trial studying?

A study testing Bayer G3 / Tatsu System, Accu-Chek® Aviva Nano Meter/Accu-Chek® Aviva Test Strips and Freestyle Lite® Meter and Test Strips with ZipwikTM tabs in people with Diabetes, run by Ascensia Diabetes Care at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 14 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 29 Feb 2016. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 9 Sep 2013). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

MAD Mean Absolute Value of the Difference Between Blood Glucose Meter (BGMS) Results and Corresponding YSI Blood Glucose (BG) Results in the Low BG Range(<70 mg/dL) (time frame: 6 hours). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 29 Feb 2016.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in