AzTrials
AzTrials

DiabetesNCT01714232

Evaluation of Blood Glucose Meter Systems

A study testing Contour® PLUS BGMS, OneTouch® SelectSimple™ BGMS and Accu-Chek® Performa BGMS in people with Diabetes, run by Ascensia Diabetes Care at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study was to evaluate the performance of a Bayer Blood Glucose Monitoring System (BGMS) and four additional Blood Glucose Monitoring Systems (BGMS) from other manufacturers. Read more about this study

Who is it for?
Ages 18 years and olderHealthy volunteers can take part
Conditions the study is for
Diabetes
What is being tested?
Contour® PLUS BGMS, OneTouch® SelectSimple™ BGMS and Accu-Chek® Performa BGMS
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
106 participants
Where
1 siteUnited States
Study dates
Oct 2012to Nov 2012
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Males and females, 18 years of age or older
  • Willing to complete all study procedures

Who cannot join

  • Blood Borne infections like hepatitis or HIV or infections such as tuberculosis
  • Hemophilia or any other bleeding disorder
  • Pregnancy

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Bayer HealthCare LLC, Diabetes Care

    Mishawaka, Indiana, 46544

Other trials in United States

About this study

Official title: Evaluation of Blood Glucose Meter Systems - Contour® PLUS Study

The purpose of this study was to evaluate the performance of a Bayer Blood Glucose Monitoring System (BGMS) and four additional Blood Glucose Monitoring Systems (BGMS) from other manufacturers. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

Read the detailed description

The purpose of this study was to evaluate the performance of a Bayer Blood Glucose Monitoring System (BGMS) and four additional Blood Glucose Monitoring Systems (BGMS) from other manufacturers. All meter BG results were compared with plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer). Performance of the five systems were evaluated across the glucose range of the BGMSs using capillary blood. All testing and lancing were performed by study staff and some samples were tested from subject fingertips. Additionally, some blood samples were glycolyzed to lower the glucose concentration levels and glucose solution was added to other samples to raise glucose concentration levels.

Conditions
Diabetes
Being tested
Contour® PLUS BGMS, OneTouch® SelectSimple™ BGMS, Accu-Chek® Performa BGMS, Accu-Chek® Active BGMS, Freestyle Freedom® BGMS
Medical subject headings
Diabetes Mellitus

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Contour® PLUS BGMSDevice

    Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

  • OneTouch® SelectSimple™ BGMSDevice

    Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

  • Accu-Chek® Performa BGMSDevice

    Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

  • Accu-Chek® Active BGMSDevice

    Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

  • Freestyle Freedom® BGMSDevice

    Study staff performed Blood Glucose (BG) tests with capillary fingerstick blood collected from subjects with diabetes and without diabetes (up to 10% of subjects without diabetes were included). All meter BG results were compared with capillary plasma results obtained with a reference laboratory glucose method (YSI Glucose Analyzer).

Study groups in NCT01714232
Study groupRoleWhat they receive
Study Staff Test BGMSsExperimentalAll testing and lancing were performed by the study staff; subjects did not perform any lancing or self-testing in this study. Study Staff lanced the fingers of subjects and tested the blood samples using five Blood Glucose Monitoring Systems (BGMS): Contour® PLUS BGMS; OneTouch® SelectSimple™ BGMS; Accu-Chek® Performa BGMS; Accu-Chek® Active BGMS; Freestyle Freedom® BGMS.
Receives: Device: Contour® PLUS BGMS; Device: OneTouch® SelectSimple™ BGMS; Device: Accu-Chek® Performa BGMS; Device: Accu-Chek® Active BGMS; Device: Freestyle Freedom® BGMS

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range

    Using the overall Blood Glucose (BG) range (27 to 460 mg/dL), the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD was calculated from the sum of all |(BG meter)-(BG reference)|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all 5 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD values indicate smaller differences between meter value and the reference value. Higher MARD values indicate higher differences between meter value and the reference value.

    Measured: 8 hours

Secondary outcomes (2)
  1. MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) in the Low Glucose Range(<=80 mg/dL)

    Using fresh and glycolyzed samples with Blood Glucose (BG) <=80 mg/dL, the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD was calculated from the sum of all |(BG meter)-(BG reference)|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all 5 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD values indicate smaller differences between meter value and the reference value. Higher MARD values indicate higher differences between meter value and the reference value.

    Measured: 8 hours

  2. MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) in the High Glucose Range (>180 mg/dL)

    Using samples with Blood Glucose >180 mg/dL, the Mean Absolute Relative Differences (MARD) between the BGM System readings and the YSI laboratory reference values were compared. MARD was calculated from the sum of all |(BG meter)-(BG reference)|/(BG reference) assessments, divided by the number of assessments, then multiplied by 100(%). Each evaluable sample was tested on all 5 BGMS, thus the same number of BG test results was analyzed for each BGMS intervention. Lower MARD values indicate smaller differences between meter value and the reference value. Higher MARD values indicate higher differences between meter value and the reference value.

    Measured: 8 hours

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted25 Oct 2012 (estimated)
  2. Study startOct 2012
  3. Primary completionNov 2012
  4. Study completionNov 2012

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
106 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT01714232
Sponsor's protocol ID
CTD-2012-008-01
Status last verified
Jan 2016
First submitted
23 Oct 2012
Last updated
29 Feb 2016
Data monitoring committee
No

Publications

  • Pardo S, Dunne N, Simmons DA. A comparison of bolus insulin dose errors based on results of a clinical trial of five blood glucose monitoring systems. Ther Deliv. 2019 Dec;10(12):793-799. doi: 10.4155/tde-2019-0047. Epub 2019 Dec 2.

    PubMed 31789119
  • Dunne N, Viggiani MT, Pardo S, Robinson C, Parkes JL. Accuracy Evaluation of CONTOUR((R))PLUS Compared With Four Blood Glucose Monitoring Systems. Diabetes Ther. 2015 Sep;6(3):377-88. doi: 10.1007/s13300-015-0121-3. Epub 2015 Jul 14.

    PubMed 26169192

Questions about this trial

What is this clinical trial studying?

A study testing Contour® PLUS BGMS, OneTouch® SelectSimple™ BGMS and Accu-Chek® Performa BGMS in people with Diabetes, run by Ascensia Diabetes Care at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 2 inclusion and 3 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 29 Feb 2016. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 27 Dec 2013). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

MARD (Mean Absolute Relative Difference Between BGMS Results and Reference Method Results) Across the Overall Tested Glucose Range (time frame: 8 hours). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT01714232, which also lists 2 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 29 Feb 2016.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in