Standardized Procedure for the Assessment of New-to-market Continuous Glucose Monitoring Systems

A study testing CGM Monitoring in people with Diabetes Mellitus, run by Academisch Medisch Centrum - Universiteit van Amsterdam (AMC-UvA) at 1 site in Netherlands.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to assess the accuracy of two continuous glucose monitoring devices (the Dexcom G4 Platinum and Medtronic Enlite systems) in patients with type 1 diabetes mellitus. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Diabetes Mellitus
What is being tested?
CGM Monitoring
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
24 participants
Where
1 siteNetherlands
Study dates
Jan 2013to Sep 2013
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • aged 18 years or above
  • diagnosed with Type 1 diabetes mellitus at least 6 months according to the WHO definition
  • Body Mass Index (BMI) <35 kg/m²
  • willing and able to wear a CGM device for the duration of the study and undergo all study procedures
  • HbA1c <10%
  • signed informed consent form prior to study entry

Who cannot join

  • Patient is pregnant, or breast feeding during the period of the study.
  • Patient is using a medication that significantly impacts glucose metabolism (oral steroids) except if stable for at least the last three months and expected to remain stable for the study duration
  • Patient may not use acetaminophen (paracetamol) while participating in the study
  • Has severe medical or psychological condition(s) or chronic conditions/infections that in the opinion of the Investigator would compromise the patient's safety or successful participation in the study.
  • Patient is actively enrolled in another clinical trial or took part in a study within 30 days
  • Known adrenal gland problem, pancreatic tumour, or insulinoma
Show all 9Show fewer
  • Inability of the patient to comply with all study procedures
  • Inability of the patient to understand the patient information.
  • Patient donated blood in the last 3 months

Trial locations and contacts

1 site in 1 country.

Netherlands1 site
  • Academic Medical Center

    Amsterdam, North Holland, 1105AZ

Other trials in Netherlands

About this study

The purpose of this study is to assess the accuracy of two continuous glucose monitoring devices (the Dexcom G4 Platinum and Medtronic Enlite systems) in patients with type 1 diabetes mellitus.

Read the detailed description

The study will use two CE marked Continuous Glucose Monitoring (CGM) systems; Dexcom G4 Platinum CGM CE-marked 2012 (Dexcom, USA) and the Medtronic Paradigm Veo system with Enlite Sensor CE-marked 2011 (Medtronic, USA). Both the Dexcom G4 Platinum and the Medtronic Enlite sensors will be obtained through normal commercial channels rather than directly from the manufacturer. The two CGM sensors will be worn concomitantly by the participant during the visit to the Clinical Research Center (CRC). This visit will have a duration of 6 hours. Blood will be drawn for the determination of glucose levels. The patient will receive his usual breakfast and an increased insulin bolus (180% of the patient's calculated mealtime dose) will be administered to correct the breakfast glucose excursion with the aim of inducing a period of minor hypoglycaemia. Blood sampling will continue until the end of the admission. At the end of this CRC part, the patient will continue to wear the two sensors at home. In the case of sensor failure before the CRC session on the third day into the study, patients will be instructed to insert a new sensor per the manufacturer's instruction for use and to notify study coordinators of the event. Sensors will be worn until the end of the six day study duration. Patients will be asked to perform at least 6 fingersticks per day for blood glucose measurements with the study glucometer. The study will end on the 6th day after initial sensor insertion. Patients will return to the CRC to have the sensor removed and their CGM sensor and blood glucometer data downloaded from the devices. In case of failure of both sensors after the CRC session but before the scheduled six day study duration, patients will return to the CRC for sensor removal and for data download.

Being tested
CGM Monitoring
Medical subject headings
Diabetes Mellitus

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • CGM MonitoringDevice

    Also called Dexcom G4 Platinum and Medtronic Enlite CGM

    All patients will be fitted with the two study CGM systems. Paradigm Veo will be configured to use the CGM part only if another pump is used.

Study groups in NCT01751932
Study groupRoleWhat they receive
CGM MonitoringExperimentalFitting of Dexcom G4 Platinum CGM monitor and Medtronic Enlite CGM monitor
Receives: Device: CGM Monitoring

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Mean Absolute Relative Difference (MARD)

    MARD will be assessed as an average of the first 6 days of wear

    Measured: up to day 6 of use

Secondary outcomes (1)
  1. Accuracy of sensors per glycemic range and trial phase

    Accuracy of the sensor in the hypoglycaemic (defined as a blood glucose value ≤3.9 mmol/L) and hyperglycaemic area (defined as a blood glucose value ≥10.0 mmol/L) as an average of the first six days of use. Additionally, a separate analysis will be performed to assess sensor performance and accuracy per day of sensor life.

    Measured: up to day 6 of use

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted18 Dec 2012 (estimated)
  2. Study startJan 2013
  3. Primary completionSep 2013
  4. Study completionSep 2013

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
24 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT01751932
Sponsor's protocol ID
SPACE2
Acronym
SPACE2
Status last verified
Dec 2013
First submitted
13 Dec 2012
Last updated
16 Dec 2013
Results posted
No
Data monitoring committee
Yes

Publications

  • Kropff J, Bruttomesso D, Doll W, Farret A, Galasso S, Luijf YM, Mader JK, Place J, Boscari F, Pieber TR, Renard E, DeVries JH. Accuracy of two continuous glucose monitoring systems: a head-to-head comparison under clinical research centre and daily life conditions. Diabetes Obes Metab. 2015 Apr;17(4):343-9. doi: 10.1111/dom.12378. Epub 2014 Sep 10.

    PubMed 25132320

Questions about this trial

What is this clinical trial studying?

A study testing CGM Monitoring in people with Diabetes Mellitus, run by Academisch Medisch Centrum - Universiteit van Amsterdam (AMC-UvA) at 1 site in Netherlands.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 9 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 16 Dec 2013. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Netherlands. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Mean Absolute Relative Difference (MARD) (time frame: up to day 6 of use). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT01751932, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 16 Dec 2013.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in