DiabetesNCT01824355
Performance of an Investigational Blood Glucose Monitoring System
A study testing Ninja 3 PLUS Investigational BG Monitoring System in people with Diabetes, run by Ascensia Diabetes Care at 1 site in United States.
Trial summary
The purpose of this study was to demonstrate that untrained subjects who have diabetes can operate the Investigational Blood Glucose Monitoring System (BGMS) and obtain valid glucose results. Read more about this study
- Who is it for?
- Ages 18 years and older
- Conditions the study is for
- Diabetes
- What is being tested?
- Ninja 3 PLUS Investigational BG Monitoring System
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 113 participants
- Where
- 1 siteUnited States
- Study dates
- Mar 2013to Mar 2013
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- Males and females, 18 years of age or older.
- People with type 1 or type 2 diabetes.
- Able to speak, read and understand English.
- Willing to complete all study procedures.
Who cannot join
- Hemophilia or any other bleeding disorder.
- Pregnancy.
- Physical, visual or neurological impairments that would make the person unable to perform testing with the BGM.
- Previously participated in a blood glucose monitoring study using the Ninja 3 Plus BGMS.
- Working for a medical laboratory, hospital or other clinical setting that involves training on or clinical use of blood glucose meters.
- Working for a competitive medical device company, or having an immediate family member who works for such a company.
- A condition, which in the opinion of the investigator or designee, would put the person or study conduct at risk.
Trial locations and contacts
1 site in 1 country.
United States1 site
Consumer Product Testing Co.
Fairfield, New Jersey, 07004
About this study
Official title: Performance of the Ninja 3 Plus Blood Glucose Monitoring System
The purpose of this study was to demonstrate that untrained subjects who have diabetes can operate the Investigational Blood Glucose Monitoring System (BGMS) and obtain valid glucose results.
- Conditions
- Diabetes
- Being tested
- Ninja 3 PLUS Investigational BG Monitoring System
- Medical subject headings
- Diabetes Mellitus
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Ninja 3 PLUS Investigational BG Monitoring SystemDevice
Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Ninja 3 PLUS Investigational BG Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to a reference laboratory glucose method.
| Study group | Role | What they receive |
|---|---|---|
| Intended Users of the Monitoring System | Experimental | Untrained subjects with diabetes used the Ninja 3 PLUS Investigational BG Monitoring System. General enrollment criteria for the 'Intended Users' population:
1. At least 60% of subjects were younger than 65 years of age.
2. At least 20% had type 1 diabetes.
3. At least 50% with type 2 diabetes were insulin users. Receives: Device: Ninja 3 PLUS Investigational BG Monitoring System |
What is being measured
The primary outcome is the main question the study is designed to answer.
Percent of Self Test Fingerstick Blood Glucose Results Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Glucose Method
Untrained subjects with diabetes who self-tested fingerstick blood used the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of BGMS results within +/- 15mg/dL (for reference BG results <75mg/dL) and within +/- 20% (for reference BG results >=75mg/dL) of the YSI reference method results.
Measured: 1 hour
Secondary outcomes (5)
Percent of Self Test Fingerstick Blood Glucose Results Within ±15 mg/dL (< 100 mg/dL) and Within ±15% (≥ 100 mg/dL) of the Laboratory Glucose Method
Untrained subjects with diabetes who self-tested fingerstick blood used the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of BGMS results within +/- 15mg/dL (for reference BG results <100mg/dL) and within +/- 15% (for reference BG results >=100mg/dL) of the YSI reference method results.
Measured: 1 hour
Percent of Self Test Fingerstick Blood Glucose Results Within ±12.5 mg/dL (< 100 mg/dL) and Within ±12.5% (≥ 100 mg/dL) of the Laboratory Glucose Method
Untrained subjects with diabetes who self-tested fingerstick blood used the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of BGMS results within +/- 12.5mg/dL (for reference BG results <100mg/dL) and within +/- 12.5% (for reference BG results >=100mg/dL) of the YSI reference method results.
Measured: 1 hour
Percent of Blood Glucose Results From Alternative Site Testing (AST) of the Palm Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Method.
Untrained subjects with diabetes self-tested Alternative Site (AST) Palm blood using the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS AST results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of AST BGMS results within +/- 15mg/dL (<75mg/dL YSI capillary plasma) and within +/- 15% (>=75mg/dL YSI capillary plasma).
Measured: 1 hour
Percent of Fingerstick Blood Glucose Results Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Glucose Method When Tested by Study Staff
Study Staff tested subject fingerstick blood using the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of BGMS results within +/- 15mg/dL (for reference BG results <75mg/dL) and within +/- 20% (for reference BG results >=75mg/dL) of the YSI reference method results.
Measured: one hour
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
Staff obtained subject responses using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' 'Agree' 'Neutral' 'Disagree' or 'Strongly Disagree.'
Measured: 1 hour
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted4 Apr 2013 (estimated)
- Study startMar 2013
- Primary completionMar 2013
- Study completionMar 2013
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Publications (1)Papers the sponsor has linked to this study.
- Other trials for DiabetesStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Diagnosis
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 113 participants (actual)
Sponsor and investigators
- Lead sponsor
- Ascensia Diabetes Care
- Sponsor type
- Industry (pharma, biotech, devices)
- Responsible party
- Sponsor
- Investigators
- Michael Caswell, PhD, Principal investigator, Consumer Product Testing Co.
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT01824355
- Sponsor's protocol ID
- CTD-2012-010-01
- Status last verified
- Jan 2016
- First submitted
- 1 Apr 2013
- Last updated
- 29 Feb 2016
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
Publications
Caswell M, Brown D, Frank J, Wallace JF, Pardo S. Accuracy and user performance evaluation of a blood glucose monitoring system which wirelessly transmits results to compatible insulin pumps. Curr Med Res Opin. 2020 May;36(5):757-764. doi: 10.1080/03007995.2020.1734919. Epub 2020 Mar 17.
PubMed 32096654
Study at a glance
From the registry record, last updated by the sponsor on 29 Feb 2016. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Performance of the Ninja 3 Plus Blood Glucose Monitoring System
- ClinicalTrials.gov ID
- NCT01824355
- Acronym / sponsor ID
- CTD-2012-010-01
- Conditions
- Diabetes
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 113 participants (actual)
- Lead sponsor
- Ascensia Diabetes CareIndustry (pharma, biotech, devices)
- Study start
- Mar 2013
- Primary completion
- Mar 2013 (actual)
- Study completion
- Mar 2013 (actual)
- First posted
- 4 Apr 2013 (estimated)
- Results
- Posted 6 Jan 2015
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The purpose of this study was to demonstrate that untrained subjects who have diabetes can operate the Investigational Blood Glucose Monitoring System (BGMS) and obtain valid glucose results.
Study design
- Primary purpose
- Diagnosis
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
| Arm | Type | Receives |
|---|---|---|
| Intended Users of the Monitoring System | Experimental | Device: Ninja 3 PLUS Investigational BG Monitoring SystemUntrained subjects with diabetes used the Ninja 3 PLUS Investigational BG Monitoring System. General enrollment criteria for the 'Intended Users' population: 1. At least 60% of subjects were younger than 65 years of age. 2. At least 20% had type 1 diabetes. 3. At least 50% with type 2 diabetes were insulin users. |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (4 listed)
Key exclusion criteria (7 listed)
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Ninja 3 PLUS Investigational BG Monitoring SystemUntrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Ninja 3 PLUS Investigational BG Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to a reference laboratory glucose method. | Device | Intended Users of the Monitoring System |
Primary outcome
Percent of Self Test Fingerstick Blood Glucose Results Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Glucose Method
Time frame: 1 hour
Untrained subjects with diabetes who self-tested fingerstick blood used the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of BGMS results within +/- 15mg/dL (for reference BG results <75mg/dL) and within +/- 20% (for reference BG results >=75mg/dL) of the YSI reference method results.
5 secondary outcomes
Percent of Self Test Fingerstick Blood Glucose Results Within ±15 mg/dL (< 100 mg/dL) and Within ±15% (≥ 100 mg/dL) of the Laboratory Glucose Method
Time frame: 1 hour
Percent of Self Test Fingerstick Blood Glucose Results Within ±12.5 mg/dL (< 100 mg/dL) and Within ±12.5% (≥ 100 mg/dL) of the Laboratory Glucose Method
Time frame: 1 hour
Percent of Blood Glucose Results From Alternative Site Testing (AST) of the Palm Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Method.
Time frame: 1 hour
Percent of Fingerstick Blood Glucose Results Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Glucose Method When Tested by Study Staff
Time frame: one hour
Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements
Time frame: 1 hour
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 29 Feb 2016.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
- Derived from this study: PubMed 32096654Caswell M, Brown D, Frank J, Wallace JF, Pardo S. Accuracy and user performance evaluation of a blood glucose monitoring system which wirelessly transmits results to compatible insulin pumps. Curr Med Res Opin. 2020 May;36(5):757-764. doi: 10.1080/03007995.2020.1734919. Epub 2020 Mar 17.
Related trials for Diabetes
Not yet recruitingPhase 3
Efficacy and Safety of Once-Weekly SHR-3167 Compared With Once-Daily Insulin Glargine in Adults With Insulin-Naive Type 2 Diabetes
Type 2 Diabetes
Active, not recruiting
Digital Reinforcement of Motivational Interviewing Micro-Skills Among Primary Care Healthcare Professionals
Hypertension (Without Type 2 Diabetes Mellitus)
Recruiting
Effects of Lean Pork Loin Intake on Protein Homeostasis and Glucose Regulation in Prediabetic Adults
Hyperglycaemia (Non Diabetic), Prediabetes
Recruiting
PAW-Hydrogel for Chronic Wound Healing
Diabetic Foot Ulcer (DFU), Pressure Ulcer (PU)
Recruiting
Investigating Vascular Properties of HEMI and SPG Signals in Individuals With or at Risk for Chronic Kidney Disease
Chronic Kidney Disease, Hypertension (HTN)
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Michael Caswell, PhD principal investigator
Consumer Product Testing Co.
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Ninja 3 PLUS Investigational BG Monitoring System in people with Diabetes, run by Ascensia Diabetes Care at 1 site in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 4 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 29 Feb 2016. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 6 Jan 2015). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Percent of Self Test Fingerstick Blood Glucose Results Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Glucose Method (time frame: 1 hour). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT01824355, which also lists 1 linked publication.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 29 Feb 2016.