Performance of an Investigational Blood Glucose Monitoring System

A study testing Ninja 3 PLUS Investigational BG Monitoring System in people with Diabetes, run by Ascensia Diabetes Care at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study was to demonstrate that untrained subjects who have diabetes can operate the Investigational Blood Glucose Monitoring System (BGMS) and obtain valid glucose results. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Diabetes
What is being tested?
Ninja 3 PLUS Investigational BG Monitoring System
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
113 participants
Where
1 siteUnited States
Study dates
Mar 2013to Mar 2013
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Males and females, 18 years of age or older.
  • People with type 1 or type 2 diabetes.
  • Able to speak, read and understand English.
  • Willing to complete all study procedures.

Who cannot join

  • Hemophilia or any other bleeding disorder.
  • Pregnancy.
  • Physical, visual or neurological impairments that would make the person unable to perform testing with the BGM.
  • Previously participated in a blood glucose monitoring study using the Ninja 3 Plus BGMS.
  • Working for a medical laboratory, hospital or other clinical setting that involves training on or clinical use of blood glucose meters.
  • Working for a competitive medical device company, or having an immediate family member who works for such a company.
  • A condition, which in the opinion of the investigator or designee, would put the person or study conduct at risk.

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Consumer Product Testing Co.

    Fairfield, New Jersey, 07004

Other trials in United States

About this study

Official title: Performance of the Ninja 3 Plus Blood Glucose Monitoring System

The purpose of this study was to demonstrate that untrained subjects who have diabetes can operate the Investigational Blood Glucose Monitoring System (BGMS) and obtain valid glucose results.

Conditions
Diabetes
Being tested
Ninja 3 PLUS Investigational BG Monitoring System
Medical subject headings
Diabetes Mellitus

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Ninja 3 PLUS Investigational BG Monitoring SystemDevice

    Untrained subjects with diabetes performed self Blood Glucose (BG) tests with capillary fingerstick and palm blood using the Ninja 3 PLUS Investigational BG Monitoring System. Study staff tested subject fingerstick blood. All BG results were compared to a reference laboratory glucose method.

Study groups in NCT01824355
Study groupRoleWhat they receive
Intended Users of the Monitoring SystemExperimentalUntrained subjects with diabetes used the Ninja 3 PLUS Investigational BG Monitoring System. General enrollment criteria for the 'Intended Users' population: 1. At least 60% of subjects were younger than 65 years of age. 2. At least 20% had type 1 diabetes. 3. At least 50% with type 2 diabetes were insulin users.
Receives: Device: Ninja 3 PLUS Investigational BG Monitoring System

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Percent of Self Test Fingerstick Blood Glucose Results Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Glucose Method

    Untrained subjects with diabetes who self-tested fingerstick blood used the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of BGMS results within +/- 15mg/dL (for reference BG results <75mg/dL) and within +/- 20% (for reference BG results >=75mg/dL) of the YSI reference method results.

    Measured: 1 hour

Secondary outcomes (5)
  1. Percent of Self Test Fingerstick Blood Glucose Results Within ±15 mg/dL (< 100 mg/dL) and Within ±15% (≥ 100 mg/dL) of the Laboratory Glucose Method

    Untrained subjects with diabetes who self-tested fingerstick blood used the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of BGMS results within +/- 15mg/dL (for reference BG results <100mg/dL) and within +/- 15% (for reference BG results >=100mg/dL) of the YSI reference method results.

    Measured: 1 hour

  2. Percent of Self Test Fingerstick Blood Glucose Results Within ±12.5 mg/dL (< 100 mg/dL) and Within ±12.5% (≥ 100 mg/dL) of the Laboratory Glucose Method

    Untrained subjects with diabetes who self-tested fingerstick blood used the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of BGMS results within +/- 12.5mg/dL (for reference BG results <100mg/dL) and within +/- 12.5% (for reference BG results >=100mg/dL) of the YSI reference method results.

    Measured: 1 hour

  3. Percent of Blood Glucose Results From Alternative Site Testing (AST) of the Palm Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Method.

    Untrained subjects with diabetes self-tested Alternative Site (AST) Palm blood using the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS AST results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of AST BGMS results within +/- 15mg/dL (<75mg/dL YSI capillary plasma) and within +/- 15% (>=75mg/dL YSI capillary plasma).

    Measured: 1 hour

  4. Percent of Fingerstick Blood Glucose Results Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Glucose Method When Tested by Study Staff

    Study Staff tested subject fingerstick blood using the Ninja 3 PLUS investigational Blood Glucose Monitoring System (BGMS). BGMS results were compared with capillary plasma BG results obtained with a reference laboratory glucose method - YSI Life Sciences (YSI) Analyzer. BG meter results were used to calculate the number of BGMS results within +/- 15mg/dL (for reference BG results <75mg/dL) and within +/- 20% (for reference BG results >=75mg/dL) of the YSI reference method results.

    Measured: one hour

  5. Number of Subject Responses That 'Strongly Agree' or 'Agree' or Are 'Neutral' With Questionnaire Statements

    Staff obtained subject responses using short questionnaires to provide feedback on instructions for use and the basic operation of the BGMS. Subjects could respond 'Strongly Agree' 'Agree' 'Neutral' 'Disagree' or 'Strongly Disagree.'

    Measured: 1 hour

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted4 Apr 2013 (estimated)
  2. Study startMar 2013
  3. Primary completionMar 2013
  4. Study completionMar 2013

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
113 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT01824355
Sponsor's protocol ID
CTD-2012-010-01
Status last verified
Jan 2016
First submitted
1 Apr 2013
Last updated
29 Feb 2016
Data monitoring committee
No

Publications

  • Caswell M, Brown D, Frank J, Wallace JF, Pardo S. Accuracy and user performance evaluation of a blood glucose monitoring system which wirelessly transmits results to compatible insulin pumps. Curr Med Res Opin. 2020 May;36(5):757-764. doi: 10.1080/03007995.2020.1734919. Epub 2020 Mar 17.

    PubMed 32096654

Questions about this trial

What is this clinical trial studying?

A study testing Ninja 3 PLUS Investigational BG Monitoring System in people with Diabetes, run by Ascensia Diabetes Care at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 29 Feb 2016. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 6 Jan 2015). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Percent of Self Test Fingerstick Blood Glucose Results Within ±15 mg/dL (< 75 mg/dL) and Within ±20% (≥ 75 mg/dL) of the Laboratory Glucose Method (time frame: 1 hour). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT01824355, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 29 Feb 2016.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in