CEP249 Pediatric Accuracy Study

A study testing Enlite sensors in people with Insulin Requiring Diabetes, run by Medtronic MiniMed, Inc. at 11 sites in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to evaluate the Enlite Sensor™ and Enlite 3 Sensor performance in pediatric patients with insulin requiring diabetes at the beginning, middle and end of sensor wear. Read more about this study

Who is it for?
Ages 2 to 18 years
Conditions the study is for
Insulin Requiring Diabetes
What is being tested?
Enlite sensors
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
186 participants
Where
11 sitesUnited States
Study dates
Mar 2015to Dec 2016
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 2 to 18 years
  • All sexes
  • No healthy volunteers
  • Child, Adult

Who may be able to join

  • Subject is 2-18 years of age at time of screening
  • Subject has been diagnosed with insulin requiring diabetes mellitus for at least one year.
  • Subject is willing to perform greater than or equal to 4 finger stick blood glucose measurements daily
  • Subject is willing to perform required sensor calibrations
  • Subject is willing to wear the system (Guardian Mobile application, pumps, sensors, meter) continuously throughout the study
  • Adequate venous access as assessed by investigator or appropriate staff

Who cannot join

  • Subject is unable to tolerate tape adhesive in the area of sensor placement
  • Subject has any unresolved adverse skin condition in the area of sensor or device placement (e.g., psoriasis, rash, Staphylococcus infection)
  • Subject is actively participating in an investigational study (drug or device) wherein they have received treatment from an investigational study (drug or device) in the last 2 weeks
  • Females of child-bearing potential who have a positive pregnancy test at screening or plans to become pregnant during the course of the study
  • Subjects with hematocrit lower than the normal age specific reference range per central or local lab testing

Trial locations and contacts

11 sites in 1 country.

11 sites in 1 country

United States11 sites
  • "Atlanta Diabetes Associates

    Atlanta, Georgia, 30318

  • Barbara Davis Center

    Aurora, Colorado, 80045

  • AMCR Institute

    Escondido, California, 92025

  • Indiana University, Riley Hospital for Children

    Indianapolis, Indiana, 46202

  • University of Iowa Children Hospital

    Iowa City, Iowa, 52242

  • Nemours Children's Clinic

    Jacksonville, Florida, 32207

  • Yale University, Yale Diabetes Research Program

    New Haven, Connecticut, 06511

  • OUHSC Department of Pediatrics

    Oklahoma City, Oklahoma, 73104

  • Rainier Clinical Research Center

    Renton, Washington, 98057

  • SoCal Diabetes

    Torrance, California, 90505

  • Diablo Clinical Research

    Walnut Creek, California, 94598

Other trials in United States

About this study

Official title: A Performance Evaluation of the Enlite™ and Enlite 3 Glucose Sensor to Support Use in Children

The purpose of this study is to evaluate the Enlite Sensor™ and Enlite 3 Sensor performance in pediatric patients with insulin requiring diabetes at the beginning, middle and end of sensor wear.

Read the detailed description

This study is a prospective, single arm, multi-center, in-clinic study.. There will be two consecutive phases of the study:

A total of up to 260 subjects (2-18 years old) will be enrolled at up to 15 investigational centers

Subjects will come for clinic visit for YSI reference glucose testing and for Enlite Sensor™ and Enlite 3 Sensor™ testing. SMBG (Self-Monitoring of Blood Glucose) reference measurement is for subjects 2-6 years of age and cannot tolerate YSI.

Study has a run-in period and a study period with a total of 10 visits

Being tested
Enlite sensors
Keywords
Insulin requiring diabetes, Children, Sensor, Accuracy

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Enlite sensorsDevice

    Subject wears enlite sensor first and followed by enlite 3 sensor.

Study groups in NCT01991470
Study groupRoleWhat they receive
Enlite sensorsExperimentalEach subject will first wear enlite sensors and will come for clinic visit for YSI (Yellow Spring Instruments). Then each subject will wear enlite 3 sensors and will come for clinic visit for YSI (Yellow Spring Instruments) or SMBG (Self-Monitoring of Blood Glucose) testing. SMBG testings are for subjects 2-6 years of age and cannot tolerate YSI. In addition. In addition, subjects aged 2-6 years old will not participate in Enlite phase. They only participate in Enlite 3 phase.
Receives: Device: Enlite sensors

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Lower Serial Number 530G Pump

    Enlite Sensor accuracy relative to the YSI (Yellow Spring Instruments) reference. Each subject wore two 530G pumps. The results for the lower serial number pump worn by each subject were arbitrarily selected and pooled across subjects for the primary dataset. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.

    Measured: 6 days

  2. Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With the Guardian Mobile System

    Enlite 3 Sensor accuracy using Guardian Mobile System compared to YSI or SMBG (Self-Monitoring of Blood Glucose) reference values. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) * 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.

    Measured: 7 days

Secondary outcomes (6)
  1. Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With the GST3C Transmitter

    Enlite 3 Sensor accuracy using GST3C transmitter used as a recorder compared to YSI or SMBG reference values. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) * 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.

    Measured: 7 days

  2. Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Higher Serial Number 530G Pump

    Enlite Sensor accuracy relative to the YSI reference. Each subject wore two 530G pumps. The results for the higher serial number pump worn by each subject were arbitrarily selected and pooled across subjects for the secondary dataset. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.

    Measured: 6 Days

  3. Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Lower Serial Number 530G Pump, Using One Additional Calibration During FST

    Enlite Sensor accuracy relative to the YSI reference, using one additional calibration during FST. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.

    Measured: 6 Days

  4. Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Higher Serial Number 530G Pump, Using One Additional Calibration During FST

    Enlite Sensor accuracy relative to the YSI reference, using one additional calibration during FST. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days were pooled together for reporting purpose.

    Measured: 6 Days

  5. Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With the Guardian Mobile System, Using One Additional Calibration During FST

    Enlite 3 Sensor accuracy using Guardian Mobile System compared to YSI or SMBG reference values, with one additional calibration during FST. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) * 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.

    Measured: 7 Days

  6. Enlite 3 Sensor Accuracy Mean Absolute Relative Difference (MARD), With GST3C Transmitter, Using One Additional Calibration During FST

    Enlite 3 Sensor accuracy using GST3C transmitter used as a recorder compared to YSI or SMBG reference values, with one additional calibration during FST. For subjects 2-6 years of age, SMBG was used as a reference value if YSI cannot be tolerated. MARD = Mean of ((Absolute difference of reference and Sensor glucose values / reference glucose values) * 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple testing days will be pooled together for reporting purpose.

    Measured: 7 Days

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted25 Nov 2013 (estimated)
  2. Study startMar 2015
  3. Primary completionDec 2016
  4. Study completionDec 2016

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Intervention model
Sequential
Masking
None (open label)
Enrolment
186 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT01991470
Sponsor's protocol ID
CEP249
Status last verified
Aug 2018
First submitted
18 Nov 2013
Last updated
27 Sep 2018
Data monitoring committee
No

Publications

  • Slover RH, Tryggestad JB, DiMeglio LA, Fox LA, Bode BW, Bailey TS, Brazg R, Christiansen MP, Sherr JL, Tsalikian E, Kaiserman KB, Sullivan A, Huang S, Shin J, Lee SW, Kaufman FR. Accuracy of a Fourth-Generation Continuous Glucose Monitoring System in Children and Adolescents with Type 1 Diabetes. Diabetes Technol Ther. 2018 Sep;20(9):576-584. doi: 10.1089/dia.2018.0109. Epub 2018 Jul 31.

    PubMed 30063162

Questions about this trial

What is this clinical trial studying?

A study testing Enlite sensors in people with Insulin Requiring Diabetes, run by Medtronic MiniMed, Inc. at 11 sites in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 5 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 27 Sep 2018. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 11 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 27 Sep 2018). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Enlite Sensor Accuracy Mean Absolute Relative Difference (MARD), With Lower Serial Number 530G Pump (time frame: 6 days). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT01991470, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 27 Sep 2018.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in