Efficacy of Oropharyngeal Seal PressureGuardianNCT02063516
Comparison of Clinical Performance of the Guardian Laryngeal Mask With LMA Proseal
A study testing Guardian Laryngeal Mask and Proseal Laryngeal Mask Airway in people with Efficacy of Oropharyngeal Seal Pressure, run by Tianjin Medical University General Hospital at 1 site in China.
Trial summary
To compare insertion characteristics of 2 different supraglottic devices (Guardian LMA and Proseal LMA) and to observe any associated complications. Read more about this study
- Who is it for?
- Ages 18 to 65 years
- Conditions the study is for
- Efficacy of Oropharyngeal Seal Pressure
- What is being tested?
- Guardian Laryngeal Mask and Proseal Laryngeal Mask Airway
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 80 participants
- Where
- 1 siteChina
- Study dates
- Jun 2013to Nov 2014
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 to 65 years
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
None listed.
Who cannot join
- Body weight 50 Kg
- Pregnancy
- Difficult airway
- Recent history of upper respiratory tract infection and sore throats
- A history of esophageal reflux disease
- ENT surgery
- Laparoscopic surgery
- Inability to understand the Study Information Sheet and provide a written consent to take part in the study.
Trial locations and contacts
1 site in 1 country.
China1 site
Tianjin Medical University General Hospital
Tianjin, Tianjin Municipality, 300052
About this study
Official title: Comparison of Clinical Performance of the Guardian Laryngeal Mask With LMA Proseal - A Perspective, Randomized,Blind Study
To compare insertion characteristics of 2 different supraglottic devices (Guardian LMA and Proseal LMA) and to observe any associated complications.
Read the detailed description
Condition:
The study focuses on the comparison of the ease of insertion, oropharyngeal leak pressure, fiberoptic position and ease of gastric tube placement between the Guardian LMA and the LMA Proseal in paralyzed, anesthetized patients.
Device: Guardian Laryngeal Mask Device: LMA Proseal
Study Type:
Interventional
Study Design:
Allocation: Randomized Endpoint Classification: Efficacy Study Intervention Model: Parallel Assignment Masking: Single Blind (Subject)
- Conditions
- Efficacy of Oropharyngeal Seal Pressure
- Being tested
- Guardian Laryngeal Mask, Proseal Laryngeal Mask Airway
- Keywords
- Airway Cuff, Guardian, Proseal, Cuff Pressure, laryngeal Mask Airway
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Guardian Laryngeal MaskDevice
Guardian Laryngeal Mask: The Guardian Laryngeal Mask (Ultimate Medical Pty Ltd, Richmond, Vic, Australia) is a new silicone-based single-use extraglottic airway device that forms a seal with the glottis for ventilation and with the hypopharynx for airway protection, and provides a gastric drainage port. In addition, it has a port for suctioning material from the hypopharynx and a pilot balloon valve that constantly monitors intracuff pressure. In the following randomized, non-crossover study, the investigators test if oropharyngeal leak pressures differed between the Guardian Laryngeal Mask and the LMA proseal in paralysed, anaesthetised patients.
Proseal Laryngeal Mask AirwayDevice
Proseal Laryngeal Mask Airway: Silicon based, reusable, modified dorsal cuff and drainage tube
| Study group | Role | What they receive |
|---|---|---|
| Guardian | Experimental | Device: Guardian Laryngeal Mask Receives: Device: Guardian Laryngeal Mask |
| Proseal | Experimental | Device:Proseal Laryngeal Mask Airway Receives: Device: Proseal Laryngeal Mask Airway |
What is being measured
The primary outcome is the main question the study is designed to answer.
Oropharyngeal Seal Pressure
This will be measured over the full range of cuff volumes (0-40 ml) and at an intracuff pressure of 60 cm H2O.
Measured: 5 min
Secondary outcomes (1)
Anatomic Position
This will be determined fiberscopically via the airway tube over the full range of cuff volumes and at an intracuff pressure of 60 cm H2O.
Measured: 5 min
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted14 Feb 2014 (estimated)
- Study startJun 2013
- Primary completionAug 2014
- Study completionNov 2014
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Publications (1)Papers the sponsor has linked to this study.
- Other trials for Efficacy of Oropharyngeal Seal PressureStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Allocation
- Randomised
- Intervention model
- Parallel groups
- Masking
- Single blind
- Who is masked
- Participant
- Enrolment
- 80 participants (actual)
Sponsor and investigators
- Lead sponsor
- Tianjin Medical University General Hospital
- Sponsor type
- University, hospital or other
- Responsible party
- Principal investigator, Haiyun Wang, Tianjin Medical University General Hospital
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT02063516
- Sponsor's protocol ID
- Wang001
- Acronym
- Guardian
- Status last verified
- Dec 2014
- First submitted
- 13 Jan 2014
- Last updated
- 26 Jan 2015
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- Yes
Publications
Pajiyar AK, Wen Z, Wang H, Ma L, Miao L, Wang G. Comparisons of clinical performance of Guardian laryngeal mask with laryngeal mask airway ProSeal. BMC Anesthesiol. 2015 May 1;15:69. doi: 10.1186/s12871-015-0039-3.
PubMed 25929558
Study at a glance
From the registry record, last updated by the sponsor on 26 Jan 2015. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Comparison of Clinical Performance of the Guardian Laryngeal Mask With LMA Proseal - A Perspective, Randomized,Blind Study
- ClinicalTrials.gov ID
- NCT02063516
- Acronym / sponsor ID
- Guardian / Wang001
- Conditions
- Efficacy of Oropharyngeal Seal Pressure
- Keywords
- Airway Cuff, Guardian, Proseal, Cuff Pressure, laryngeal Mask Airway
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 80 participants (actual)
- Lead sponsor
- Tianjin Medical University General HospitalUniversity, hospital or other
- Study start
- Jun 2013
- Primary completion
- Aug 2014 (actual)
- Study completion
- Nov 2014 (actual)
- First posted
- 14 Feb 2014 (estimated)
- Results
- Posted 18 Dec 2014
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
To compare insertion characteristics of 2 different supraglottic devices (Guardian LMA and Proseal LMA) and to observe any associated complications.
Detailed description
Condition:
The study focuses on the comparison of the ease of insertion, oropharyngeal leak pressure, fiberoptic position and ease of gastric tube placement between the Guardian LMA and the LMA Proseal in paralyzed, anesthetized patients.
Device: Guardian Laryngeal Mask Device: LMA Proseal
Study Type:
Interventional
Study Design:
Allocation: Randomized Endpoint Classification: Efficacy Study Intervention Model: Parallel Assignment Masking: Single Blind (Subject)
Study design
- Allocation
- Randomised
- Intervention model
- Parallel groups
- Masking
- Single blind (participant)
- Number of arms
- 2
| Arm | Type | Receives |
|---|---|---|
| Guardian | Experimental | Device: Guardian Laryngeal MaskDevice: Guardian Laryngeal Mask |
| Proseal | Experimental | Device: Proseal Laryngeal Mask AirwayDevice:Proseal Laryngeal Mask Airway |
Who may qualify
- Ages
- 18 to 65 years
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (0 listed)
Key exclusion criteria (8 listed)
- Body weight 50 Kg
- Pregnancy
- Difficult airway
- Recent history of upper respiratory tract infection and sore throats
- A history of esophageal reflux disease
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Guardian Laryngeal MaskGuardian Laryngeal Mask: The Guardian Laryngeal Mask (Ultimate Medical Pty Ltd, Richmond, Vic, Australia) is a new silicone-based single-use extraglottic airway device that forms a seal with the glottis for ventilation and with the hypopharynx for airway protection, and provides a gastric drainage port. In addition, it has a port for suctioning material from the hypopharynx and a pilot balloon valve that constantly monitors intracuff pressure. In the following randomized, non-crossover study, the investigators test if oropharyngeal leak pressures differed between the Guardian Laryngeal Mask and the LMA proseal in paralysed, anaesthetised patients. | Device | Guardian |
| Proseal Laryngeal Mask AirwayProseal Laryngeal Mask Airway: Silicon based, reusable, modified dorsal cuff and drainage tube | Device | Proseal |
Primary outcome
Oropharyngeal Seal Pressure
Time frame: 5 min
This will be measured over the full range of cuff volumes (0-40 ml) and at an intracuff pressure of 60 cm H2O.
1 secondary outcome
Anatomic Position
Time frame: 5 min
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 26 Jan 2015.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
- Derived from this study: PubMed 25929558Pajiyar AK, Wen Z, Wang H, Ma L, Miao L, Wang G. Comparisons of clinical performance of Guardian laryngeal mask with laryngeal mask airway ProSeal. BMC Anesthesiol. 2015 May 1;15:69. doi: 10.1186/s12871-015-0039-3.
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
The record lists no contact. Use the official record to reach the sponsor.
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Guardian Laryngeal Mask and Proseal Laryngeal Mask Airway in people with Efficacy of Oropharyngeal Seal Pressure, run by Tianjin Medical University General Hospital at 1 site in China.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 0 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 26 Jan 2015. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in China. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 18 Dec 2014). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Oropharyngeal Seal Pressure (time frame: 5 min). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT02063516, which also lists 1 linked publication.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 26 Jan 2015; AzTrials last refreshed its copy on 8 Nov 2026.