Study of Carbon Dioxide Exchange Patterns During Robotic Prostatectomies

A study testing AirSeal System and Standard Endopath in people with Prostate Cancer and Pneumoperitoneum, run by OhioHealth at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The AirSeal® System will allow for lower carbon dioxide (CO2) absorption rates than the standard Endopath System. Read more about this study

Who is it for?
Ages 18 years and older, men only
Conditions the study is for
Prostate Cancer, Pneumoperitoneum
What is being tested?
AirSeal System and Standard Endopath
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
109 participants
Where
1 siteUnited States
Study dates
Aug 2016to 15 Jul 2019
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • Men only
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Age>18
  • Eligible and elect to undergo robotic prostatectomy surgery @ OhioHealth Dublin Methodist Hospital

Who cannot join

  • Age < 18
  • Emergency surgery
  • Ascites
  • BMI>44 or <18

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Dublin Methodist Hospital

    Columbus, Ohio, 43016

Other trials in United States

About this study

Official title: A Comparative Study of CO2 Exchange Patterns Between Valve-free Trocar (AirSeal®) Versus Standard Trocar (Endopath®) During Robotic Prostatectomies

The AirSeal® System will allow for lower carbon dioxide (CO2) absorption rates than the standard Endopath System.

Read the detailed description

The primary observation to be made is the level of carbon dioxide in the arterial blood, specifically when the patient is insufflated and deflated during the surgery. As a result of the consistent pressure that the AirSeal® system maintains, carbon dioxide will be absorbed in lower levels when compared to the standard of care systems in which fluctuations in intra-peritoneum pressure are seen.

Being tested
AirSeal System, Standard Endopath
Keywords
AirSeal, Trocar, Carbon Dioxide Exchange and Absorption
Medical subject headings
Prostatic Neoplasms, Pneumoperitoneum

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • AirSeal SystemDevice

    Also called Valve-free Trocar

    To analyze the results of patients randomized to either the AirSeal system or Standard Endopath trocar.

  • Standard EndopathDevice

    Also called Endopath Trocar

    This group receives the Standard Endopath Trocar for intraoperative insufflation.

Study groups in NCT02114164
Study groupRoleWhat they receive
AirSeal SystemActive comparatorThis group receives the AirSeal System for intraoperative insufflation.
Receives: Device: AirSeal System
Standard EndopathActive comparatorThis group receives the Standard Endopath Trocar for intraoperative insufflation.
Receives: Device: Standard Endopath

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Intraoperative Pneumoperitoneal Pressure (mmHg)

    Pneumoperitoneal pressure is the pressure exerted by air or gas in the abdominal (peritoneal) cavity. As the amount of air/gas increases, the pressure (mmHg) increases. The highest pressure that occurred during the procedure was compared between groups.

    Measured: Through end of procedure, an average of 156 minutes

Secondary outcomes (7)
  1. Intraoperative Arterial Blood Gas Carbon Dioxide (mmHg).

    The amount of dissolved carbon dioxide (PaCO2) in arterial blood (mmHg) during the operative procedure

    Measured: Baseline, 60 minutes, and at end of procedure, an average of 156 minutes

  2. End-tidal Carbon Dioxide Pressure (mmHg)

    End-tidal carbon dioxide pressure is the partial pressure of carbon dioxide at the end of exhalation. Normal value is 35-45 mmHg.

    Measured: Baseline, 60 minutes, and at end of procedure, an average of 156 minutes

  3. Carbon Dioxide Elimination (mmHg)

    Carbon dioxide elimination was calculated using end-tidal carbon dioxide pressure (EtCO2), tidal volume, respiratory rate, barometric pressure (pB = 760mm Hg), partial pressure of water vapor (PH2O = 13 mm Hg), and patient weight (kg)

    Measured: Baseline, 60 minutes, and at end of procedure, an average of 156 minutes

  4. Intraoperative Arterial Blood Gas Oxygen (mmHg).

    The amount of dissolved oxygen (PaO2) in arterial blood (mmHg).

    Measured: Baseline, 60 minutes, and end of procedure, an average of 156 minutes

  5. Number of Ventilator Interventions

    The number of times the anesthesiologist had to adjust ventilator settings

    Measured: Through end of procedure, an average of 156 minutes

  6. Procedure Time (Minutes)

    The duration of the procedure

    Measured: Through end of procedure, an average of 156 minutes

  7. Smoke Evacuation Quality

    Smoke evacuation quality throughout the procedure was assessed by the surgeon as below average, average, or above average

    Measured: Through end of procedure, an average of 156 minutes

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted15 Apr 2014 (estimated)
  2. Study startAug 2016
  3. Primary completion3 May 2018
  4. Study completion15 Jul 2019

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Supportive care
Allocation
Randomised
Intervention model
Parallel groups
Masking
Single blind
Who is masked
Participant
Enrolment
109 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT02114164
Sponsor's protocol ID
OH1-13-00489
Status last verified
Oct 2021
First submitted
10 Jan 2014
Last updated
26 Nov 2021
Data monitoring committee
No

Publications

  • Abaza R, Martinez O, Murphy C. Randomized Controlled Comparison of Valveless Trocar (AirSeal) vs Standard Insufflator with Ultralow Pneuomoperitoneum During Robotic Prostatectomy. J Endourol. 2021 Jul;35(7):1020-1024. doi: 10.1089/end.2020.1025. Epub 2021 Jan 21.

    PubMed 33349135

Questions about this trial

What is this clinical trial studying?

A study testing AirSeal System and Standard Endopath in people with Prostate Cancer and Pneumoperitoneum, run by OhioHealth at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 2 inclusion and 4 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 26 Nov 2021. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 26 Nov 2021). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Intraoperative Pneumoperitoneal Pressure (mmHg) (time frame: Through end of procedure, an average of 156 minutes). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT02114164, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.