Diabetes MellitusAPCam09NCT02129868

Automated Closed-loop in Children and Adolescents With Type 1 Diabetes

A study testing Closed-loop (Android Closed-Loop platform) in people with Diabetes Mellitus and Diabetes Mellitus, Type 1, run by University of Cambridge at 1 site in United Kingdom.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

People with type 1 diabetes need regular insulin injections or continuous delivery of insulin using a pump. Keeping blood sugars in the normal range is known to reduce long term complications. However, achieving treatment goals can be very difficult due to the risk of low glucose levels (hypoglycaemia). Read more about this study

Who is it for?
Ages 6 to 18 years
What is being tested?
Closed-loop (Android Closed-Loop platform)
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
13 participants
Where
1 siteUnited Kingdom
Study dates
Apr 2014to Apr 2015
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 6 to 18 years
  • All sexes
  • No healthy volunteers
  • Child, Adult

Who may be able to join

  • The subject is between 6 and 18 years of age.
  • The subject has had type 1 diabetes, as defined by WHO criteria for at least 1 year or is confirmed C-peptide negative.
  • The subject will have been an insulin pump user for at least 3 months, with a good knowledge of insulin dose adjustment.
  • HbA1c between below 11 % based on analysis from central laboratory.
  • The subject is literate in English.
  • The subject is willing to undertake all study related activities.

Who cannot join

  • Non-type 1 diabetes mellitus including those secondary to chronic disease.
  • Any other physical or psychological disease likely to interfere with the normal conduct of the study and interpretation of the study results.
  • Current treatment with drugs known to interfere with glucose metabolism such as systemic corticosteroids, non-selective beta-blockers and MAO inhibitors.
  • Known or suspected allergy against insulin.
  • Subjects with clinically significant nephropathy, neuropathy or proliferative retinopathy as judged by the investigator
  • Patient is pregnant, or breast feeding during the period of the study.
Show all 11Show fewer
  • Total daily insulin dose ≥ 2 Units/kg/day
  • Total daily insulin dose < 10 Units/day
  • Severe visual impairment
  • Severe hearing impairment
  • Subjects using implanted internal pacemaker

Trial locations and contacts

1 site in 1 country.

United Kingdom1 site
  • University of Cambridge

    Cambridge, CB2 0QQ

Other trials in United Kingdom

About this study

Official title: An Open-label, Single-centre, Randomised, 2-period Cross-over Study to Assess the Efficacy and Safety of a Novel Automated Overnight Closed-loop Glucose Control System on Day 1 of Continuous Glucose Monitoring Sensor Insertion in Comparison to Day 3 to 4 After Sensor Insertion in Children and Adolescents With Type 1 Diabetes

People with type 1 diabetes need regular insulin injections or continuous delivery of insulin using a pump. Keeping blood sugars in the normal range is known to reduce long term complications. However, achieving treatment goals can be very difficult due to the risk of low glucose levels (hypoglycaemia). One solution is to use a system where the amount of insulin injected closely matches the blood sugar levels on a continuous basis. This can be achieved by what is known as a "closed loop system" where a small glucose sensor placed under the skin communicates with a computer containing an algorithm that drives a subcutaneous insulin pump. Previous studies conducted under carefully controlled clinical research facility environment, in Cambridge, United Kingdom, as well as several other centres have shown that closed-loop glucose control is superior to usual insulin pump therapy. The next logical step in the development pathway is to test closed loop systems in the home environment. An essential requirement for conducting closed-loop studies outside clinical research facility is an automated system where wireless data transmission takes place between the glucose sensor and insulin pump.

The purpose of the present study is to evaluate the efficacy and safety of automated overnight closed-loop, in children and adolescents with type 1 diabetes, using a novel system which has greatest potential for use in the home setting. The study will take place at a clinical research facility on two occasions, using a standardised protocol. The performance of the closed-loop system will be evaluated on day 1 of continuous glucose monitoring (CGM) sensor life as compared to on days 3 to 4 of sensor life. Data and experience gained from this study will be used for further refinements and development of the system for future home use.

Read the detailed description

The main objective of the current study is to evaluate the efficacy, safety and utility/reliability of a novel automated overnight closed-loop glucose control system in children and adolescents with type 1 diabetes. This will inform the future development of closed-loop systems in children and adolescents aged 6 to 18 years with type 1 diabetes.

This is an open-label, randomised, 2-period cross-over study, using the Medtronic Android closed loop platform or similar, comparing the safety and efficacy of overnight closed loop glucose control on day 1 compared with day 3 to 4 of CGM sensor life. Participants will be randomised to undergo two overnight studies in a clinical research facility, during which glucose levels will be controlled by the computer-based closed-loop algorithm on day 1 of CGM sensor insertion or day 3 to 4 after sensor insertion.

A total of 16 children and adolescents with T1D between 6 and 18 years of age will be recruited through the Outpatient Diabetes Clinic in the Department of Paediatrics, Addenbrooke's Hospital and Paediatric Diabetes Centres in London, Norwich, and Ipswich, to allow for 12 available for assessment. All participants will be on insulin pump treatment with a good knowledge of insulin dose adjustments. The study will take place at the Wellcome Trust Clinical Research Facility (WTCRF), Cambridge. Participants will be admitted at the clinical research facility in the evening and stay until the following morning. During the two closed-loop visits, automated closed-loop glucose control with wireless data transmission will be performed. A Model-Predictive-Control algorithm will determine the insulin infusion rate based on interstitial glucose measured by a continuous subcutaneous glucose monitoring (CGM) system. Intravenous sampling for glucose and insulin levels will also be carried out on both study visits for post hoc validation of glucose sensor data and for future modelling purposes.

The primary efficacy endpoint is the time spent in the target plasma glucose range from 3.9 to 8 mmol/L. Safety evaluation is focused on the frequency and duration of episodes of hypoglycaemia.

Being tested
Closed-loop (Android Closed-Loop platform)
Keywords
Type 1 diabetes, Closed-loop glucose control, Artificial Pancreas, Continuous subcutaneous insulin infusion, Continuous glucose monitoring
Medical subject headings
Diabetes Mellitus, Diabetes Mellitus, Type 1, Glucose Metabolism Disorders, Endocrine System Diseases, Autoimmune Diseases

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Closed-loop (Android Closed-Loop platform)Device

    The closed-loop system is purpose-built and comprises a hand-held computer containing a model predictive control (MPC) based glucose control algorithm and communicating with the CGM device and the insulin pump.

    The Android Closed-Loop platform employs the Medtronic MiniMed Paradigm® Veo™ insulin pump system with CGM capability (use of MiniLink™Transmitter and sensor). A Radiofrequency (RF) translator module translates the RF protocol to Bluetooth® technology. An Android device containing an algorithm will use sensor glucose data to calculate pump strokes.

Study groups in NCT02129868
Study groupRoleWhat they receive
Closed-loop on day 1 of CGM sensor lifeExperimentalGlucose level is controlled by the automated closed-loop glucose control system on day 1 after CGM sensor insertion.
Receives: Device: Closed-loop (Android Closed-Loop platform)
Closed-loop on day 3 of CGM sensor lifeActive comparatorGlucose level is controlled by the automated closed-loop glucose control system on day 3 or 4 after CGM sensor insertion.
Receives: Device: Closed-loop (Android Closed-Loop platform)

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Time spent with plasma glucose concentration in the target glucose range from 3.9 to 8.0 mmol/l.

    The primary outcome measure is time spent with plasma glucose concentration in the target range (3.9-8.0mmol/L) between 21:30 on Day 1 and 07:30 on Day 2 (10 hours), as obtained with closed-loop insulin delivery on day 1 of CGM sensor insertion as compared with closed-loop insulin delivery on day 3 to 4 of CGM sensor insertion.

    Measured: 10 hours

Secondary outcomes (7)
  1. Time spent with plasma glucose concentration below the target range (<3.9mmol/L) between 21:30 on Day 1 and 07:30 on Day 2 (10 hours).

    Measured: 10 hours

  2. Time spent with plasma glucose concentration above the target range (>8mmol/L) between 21:30 on Day 1 and 07:30 on Day 2 (10 hours).

    Measured: 10 hours

  3. Mean and standard deviation of plasma glucose levels between 21:30 on Day 1 and 07:30 on Day 2 (10 hours).

    Measured: 10 hours

  4. Total insulin delivery between 21:30 on Day 1 and 07:30 on Day 2 (10 hours).

    Measured: 10 hours

  5. Total basal insulin delivery between 21:30 on Day 1 and 07:30 on Day 2 (10 hours).

    Measured: 10 hours

  6. Plasma insulin concentrations between 21:30 on Day 1 and 07:30 on Day 2 (10 hours).

    Measured: 10 hours

  7. Absolute relative difference between matched pairs of subcutaneous glucose sensor and reference plasma glucose.

    Measured: 10 hours

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted2 May 2014 (estimated)
  2. Study startApr 2014
  3. Primary completionApr 2015
  4. Study completionApr 2015

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Crossover
Masking
None (open label)
Enrolment
13 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT02129868
Sponsor's protocol ID
APCam09
Acronym
APCam09
Status last verified
Feb 2017
First submitted
23 Apr 2014
Last updated
3 Mar 2017
Results posted
No
Data monitoring committee
Yes

Publications

  • Hovorka R, Allen JM, Elleri D, Chassin LJ, Harris J, Xing D, Kollman C, Hovorka T, Larsen AM, Nodale M, De Palma A, Wilinska ME, Acerini CL, Dunger DB. Manual closed-loop insulin delivery in children and adolescents with type 1 diabetes: a phase 2 randomised crossover trial. Lancet. 2010 Feb 27;375(9716):743-51. doi: 10.1016/S0140-6736(09)61998-X. Epub 2010 Feb 4.

    PubMed 20138357
  • Hovorka R, Kumareswaran K, Harris J, Allen JM, Elleri D, Xing D, Kollman C, Nodale M, Murphy HR, Dunger DB, Amiel SA, Heller SR, Wilinska ME, Evans ML. Overnight closed loop insulin delivery (artificial pancreas) in adults with type 1 diabetes: crossover randomised controlled studies. BMJ. 2011 Apr 13;342:d1855. doi: 10.1136/bmj.d1855.

    PubMed 21493665
  • Elleri D, Allen JM, Biagioni M, Kumareswaran K, Leelarathna L, Caldwell K, Nodale M, Wilinska ME, Acerini CL, Dunger DB, Hovorka R. Evaluation of a portable ambulatory prototype for automated overnight closed-loop insulin delivery in young people with type 1 diabetes. Pediatr Diabetes. 2012 Sep;13(6):449-53. doi: 10.1111/j.1399-5448.2012.00903.x. Epub 2012 Jul 23.

    PubMed 22817340
  • Kumareswaran K, Elleri D, Allen JM, Harris J, Xing D, Kollman C, Nodale M, Murphy HR, Amiel SA, Heller SR, Wilinska ME, Acerini CL, Evans ML, Dunger DB, Hovorka R. Meta-analysis of overnight closed-loop randomized studies in children and adults with type 1 diabetes: the Cambridge cohort. J Diabetes Sci Technol. 2011 Nov 1;5(6):1352-62. doi: 10.1177/193229681100500606.

    PubMed 22226252
  • Tauschmann M, Allen JM, Wilinska ME, Ruan Y, Thabit H, Acerini CL, Dunger DB, Hovorka R. Sensor Life and Overnight Closed Loop: A Randomized Clinical Trial. J Diabetes Sci Technol. 2017 May;11(3):513-521. doi: 10.1177/1932296816678631. Epub 2016 Nov 11.

    PubMed 27837162

Questions about this trial

What is this clinical trial studying?

A study testing Closed-loop (Android Closed-Loop platform) in people with Diabetes Mellitus and Diabetes Mellitus, Type 1, run by University of Cambridge at 1 site in United Kingdom.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 11 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 3 Mar 2017. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United Kingdom. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Time spent with plasma glucose concentration in the target glucose range from 3.9 to 8.0 mmol/l. (time frame: 10 hours). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT02129868, which also lists 5 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 3 Mar 2017.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in