Comparison of the Freiburg Visual Acuity Test (FrACT) With the EDTRS- and Landoltring Charts

An observational study of Ophthalmologically Normal Subjects, run by Schiefer Ulrich at 1 site in Germany.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

A) For the investigation of Visual acuity, inter-test agreement and the test-retest reliability were used three different visual acuity tests: ETDRS charts, Landoltring chart and Freiburg Visual Acuity Test (FrACt). Read more about this study

Who is it for?
Ages 18 years and olderHealthy volunteers can take part
Conditions the study is for
Ophthalmologically Normal Subjects
What is being tested?
Methodology of visual acuity assessment
Type of treatment
Medical device
How many people
25 participants
Where
1 siteGermany
Study dates
Jun 2014to Jun 2015
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Study population: Ophthalmologically normal subjects

Who may be able to join

  • age ≥ 18 years
  • Minimum distant visual acuity (without/with correction) 0.2 (Visucat, single letter optotypes [numbers], initial session). As long as minimum distant visual acuity limits are exceeded, test runs without optical corrections are preferred in order to enhance the spectrum/variety of tested visual acuity values
  • Ametropia:
  • maximum myopia sph -2.00 dpt maximum hyperopia sph +5.50 dpt maximum astigmatism cyl 2.5 dpt
  • informed consent

Who cannot join

  • Epilepsy / psychiatric disorders
  • Drugs affect the reaction time and / or determining the visual acuity, or the refractive power of the eye
  • Ophthalmic Education
  • Amblyopia
  • strabismus
  • Glaucoma IOP> 22 mmHg
Show all 16Show fewer
  • Nystagmus
  • Diabetic retinopathy (blood sugar-related retinal disease)
  • Very dry eyes (sicca symptoms)
  • eye surgery back are less than 3 months
  • Infectious diseases (conjunctivitis, corneal inflammation, choroidal inflammation)
  • miosis drugs
  • Pronounced eye injuries
  • Other eye disorders that affect the image quality
  • Macular diseases
  • albinism

Trial locations and contacts

1 site in 1 country.

Germany1 site
  • University of Applied Sciences, Study Course Ophthalmic Optics

    Aalen, Baden-Wurttemberg, 73430

Other trials in Germany

About this study

Official title: Comparison of the Freiburg Visual Acuity Test (FrACT) With the EDTRS- and Landoltring Charts: a) Assessment of the Inter-test Agreement and the Test-retest Reliability b) Analysis of Examinees' / Examiner's Subjective Rating and of Test Durations

  1. For the investigation of Visual acuity, inter-test agreement and the test-retest reliability were used three different visual acuity tests: ETDRS charts, Landoltring chart and Freiburg Visual Acuity Test (FrACt). The examinees passed each visual acuity test twice but the sequence of the visual acuity tests were randomize and so different for each examinee. The examinees were 24 adult Aalen university students (except students of ophthalmic optics in terms 2 - 7) with a minimal distant visual acuity without or with correction of 0.2. Another inclusion criteria was a maximum myopia of sph -2,00 dpt, a maximum hyperopia of sph +5,50 dpt and a maximum astigmatism of cyl 2,50 dpt. The Landolt C was projected in eight-orientations. For the FrACT Test we used two different devices for response input: a modified keypad with Landoltring symbols (FrACTk) and a haptic Landolt device with an adjustable Landoltring (FrACTh). The primary objective was the comparison between FrACTk and FrACTh. The secondary Objective contains the comparison between all visual acuity tests.
  2. Background: Landolt Cs and letters are commonly used optotypes. In the explorative study we compare four different ways of applying them: The Freiburg visual acuity test (FrACT) using a keypad and a haptic Landolt device, a Landolt C chart in accordance to ISO 8596 and DIN 58220 with eight-orientated Landolt C and ETDRS charts. The aims of the survey were to determine examinee´s/examiner´s subjective ratings using Visual analogue scales (VAS), the inter-test agreement between the tests and test-retest reliability of the tests, concerning the test duration.

Subjects and Methods: Twenty-four adult examinees with a visual acuity of ≥ 0,2 were included in the explorative study. The visual acuity of one eye was measured monocularly, randomized with regard to the leading/non-leading eye. The subjects passed every visual acuity test twice in a randomized sequence of the visual acuity tests, which is maintained by a pre-specified randomization list, making use of forced choice.

Read the detailed description
  1. FrACT (internet version 3.8.1) 1.1 haptic Landolt Ring device (FrACTh) 1.2 modified keypad with Landolt Ring symbols (FrACTk)
  2. ETDRS Visual Acuity Tester (Steinbeis Transferzentrum Biomedizinische Optik, Tübingen/Germany)
  3. Landoltring chart (Visus GmbH, Stuttgart/Germany)
  4. VISUCAT monitor with single optotype display(numbers; Argus Individuell Optic GmbG, Putzbrunn/Germany)
  5. Room illuminance level 250 lx, assessed with digital luxmeter (Peak Tech 5025, Peak Tech Prüf - und Messtechnik GmbH, Ahrensburg/Germany)
  6. Laptopdisplay (TravelMate, Acer, 8372 Taipeh/China): luminance level 151 cd/m2 , surround luminance level 75 cd/m2 , assessed with luminance meter (IS-110 Laptop Minolta, Tokyo/Japan)
  7. ETDRS Visual Acuity Tester: luminance level: 237 cd/m², surround luminance level 85 cd/m²
  8. Landoltring Chart: luminance level: 236 cd/m², surround luminance level 84 cd/m²
  9. VISUCAT: luminance level: 380 cd/m², surround luminance level 70 cd/m²
  10. Stopwatch (Apple iPhone 4S, Cupertino CA/USA)
  11. Laser Rangefinder (GLM 80 Professional, Bosch GmbH, Stuttgart/Germany)
Being tested
Methodology of visual acuity assessment
Keywords
visual acuity, Landolt C, ETDRS charts, Freiburg Visual Acuity Test, FrACT, Comparison, acceptance, Test retest reliability, duration, questionnaire

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Methodology of visual acuity assessmentDevice

    Also called VISUCAT, FrACT, ETDRS Chart, Landolt C

Study groups in NCT02198534
Study groupRoleWhat they receive
ophthalomogically normal subjectsNot described
Receives: Device: Methodology of visual acuity assessment

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. A) 1.Visual acuity (in terms of spatial resolution)

    Comparison of visual acuity measurements, visualized by Bland-Altman plots

    Measured: July - August, 2014 (1 month)

  2. Inter-test agreement between FrACTh and FrACTk with regard to visual acuity results

    Assessment of inter-test agreement between FrACTh and FrACTk by comparison between these two methods, visualized by Bland-Altman plots

    Measured: July - August, 2014 (1 month)

  3. Test-retest reliability of FrACTh and FrACTk with regard to visual acuity results

    Assessment of test-retest reliability of FrACTh and FrACTk by repeated measurements, visualized by Bland-Altman plots

    Measured: July - August, 2014 (1 month)

Secondary outcomes (2)
  1. Examinees´ / examer's subjective ratings regarding the test procedures

    Assessment of examinees´ / examiner´s subjective ratings, using box \& whisker plots for analysis of the visual analogue scale (VAS) results

    Measured: July - August, 2014 (1 month)

  2. Test duration (in seconds)

    Assessment of test duration and its comparison, using box \& whisker plots and/or Bland-Altman plots

    Measured: July - August, 2014 (1 month)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted23 Jul 2014 (estimated)
  2. Study startJun 2014
  3. Primary completionOct 2014
  4. Study completionJun 2015

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Cohort
Time perspective
Prospective
Enrolment
25 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT02198534
Sponsor's protocol ID
F-2014-049
Acronym
FEL
Status last verified
Nov 2016
First submitted
21 Jul 2014
Last updated
23 Nov 2016
Results posted
No
Data monitoring committee
No

Questions about this trial

What is this clinical trial studying?

An observational study of Ophthalmologically Normal Subjects, run by Schiefer Ulrich at 1 site in Germany.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 5 inclusion and 16 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 23 Nov 2016. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Germany. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

A) 1.Visual acuity (in terms of spatial resolution) (time frame: July - August, 2014 (1 month)). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 23 Nov 2016.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in