Trial summary
Pacemakers and ICD's have traditionally been accepted as contraindications to MRI due to safety concerns. Recent work in our laboratory as well as our pilot study of approximately 500 patients has shown the safety of most contemporary implanted rhythm management devices in the MRI environment given appropriate… Read more about this study
- Who is it for?
- Ages 9 to 100 years
- Conditions the study is for
- Conditions Requiring Diagnostic Evaluation
- What is being tested?
- Perform MRI in patients with pacemaker or ICD
- How many people
- 1,275 participants
- Where
- 1 siteUnited States
- Study dates
- 30 Dec 2011to 22 Jan 2015
- Phase
- Observational
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 9 to 100 years
- All sexes
- No healthy volunteers
- Child, Adult, Older adult
Study population: Males and females 9 years old or greater weighing more than 80 lbs who require a clinically indicated MRI
Who may be able to join
- Patients with an absolute clinical need for MR imaging and PPM (year 1996 model or later) or ICD's (year 2000 or later)
- -
Who cannot join
- Inclusion: Patients with an absolute clinical need for MR imaging and PPM (year 1996 model or later) or ICD's (year 2000 or later)
- Patients with pacemaker models before 1996 and ICD models before year 2000.
- Pacemaker defendant patients with ICD.
- Patients who weighing less than 80 lbs.
- Patients less than 9 years old
- Patients who complete the MRI standard screening form and are deemed inappropriate for MRI for any reason
- Pregnant patients will be excluded in their first trimester and will not get gadolinium at any time during their pregnancy
- Patient with abandoned/capped leads or leads implanted <4 weeks unless the procedure is deemed critical by the referring physician and approved by the PI
Trial locations and contacts
1 site in 1 country.
United States1 site
Johns Hopkins Hospital, Blalcok 5 MRI
Baltimore, Maryland, 21205
About this study
Official title: Safety of Clinically Indicated Magnetic Resonance Imaging in Patients With Permanent Pacemakers (PPM) and Implanted Cardioverter Defibrillators
Pacemakers and ICD's have traditionally been accepted as contraindications to MRI due to safety concerns. Recent work in our laboratory as well as our pilot study of approximately 500 patients has shown the safety of most contemporary implanted rhythm management devices in the MRI environment given appropriate precautions. Because the investigators have completed testing in animals weighing as little as 80 lbs. without incident, the investigators feel they can safely image children at this weight or greater (Circulation 2004; 110:475-482).
The goal of this study is to expand on our pilot study and to develop a protocol, consistent with the Department of Health and Human Services (DHHS) requirements, that will increase our safety experience with clinically indicated MR imaging in patients with implanted cardiac pacemakers and ICD's.
Read more
This is a cohort study. One thousand seven hundred patients with a clinical need for MRI will be recruited. Risks and Benefits associated with Magnetic Resonance (MR) imaging in the setting of implantable devices will be discussed and informed consent obtained. The type of device and leads will be ascertained. Patients will be monitored by a physician or a Registered Nurse (RN) with ACLS certification and familiarity with ICD/Pacemaker programming during the study. When imaging children who meet age and weight criteria, a physician will be present during the entire study. Fellow and attending radiology and electrophysiology support will be available in house in all cases.
Standard device testing including battery voltages, thresholds, sensing and lead impedances will be measured.
Read the detailed description
The primary objective of this study is to develop a protocol, consistent with DHHS/ Center for Medicare and Medicaid Services (CMS) guidelines and requirements to further document safety in clinically indicated MR imaging in patients with implanted cardioverter defibrillators (ICD's) and pacemakers.
MRI's will be conducted on patients who require a clinically indicated MRI only. All studies will be done in the 1.5 Tesla MRI unit. ECG monitoring pads will be placed on the patients for the duration of the study. An external defibrillator and Advance Cardiovascular Life Support (ACLS) drugs will be on hand. Heart rate, blood pressure, O2 saturation will also be monitored non-invasively throughout the study. All devices will undergo a complete interrogation prior to imaging. Parameters such as atrial and ventricular pacing thresholds, R and P wave amplitudes, lead impedance, and battery status will be measured and recorded. PPMs will be programmed to an asynchronous mode if dependent and to an inhibited mode in patients without pacemaker dependence.
The pacemaker function of ICD's in pacemaker independent patients will be programmed to ventricular inhibited (VVI) mode at 50 bpm. Pacemaker dependant patients with ICDs will be excluded.
In addition, there is a theoretical concern that patients who have capped or abandoned leads in their chest may be at risk for heating. While there is little or no objective data to support this concern, given this theoretical concern, patients in this category will only be admitted to the study if the referring physician deems the MRI critical and after the PI has carefully considered the risk/benefit ration for each patient who falls in to this category.
Similarly, there historically has also been a concern that newly implanted leads (<4 weeks) are at risk for lead to dislodgement d/t possible torque from the MRI environment. This too is only a theoretical concern and there is little or no scientific evidence to support this claim. Nevertheless, patients with newly implanted pacemaker/ICD leads will be considered by the PI for protocol inclusion only if the referring physician deems the MRI to be critical and after considering the risk/benefit to each individual patient.
After the appropriate MRI protocol for each MRI patient's unique condition has been completed, the device will be re-programmed to its original settings and completely interrogated to detect any changes in device performance.
- Being tested
- Perform MRI in patients with pacemaker or ICD
- Keywords
- MRI, Magnetic Resonance Imaging, pacemakers, ICD
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Perform MRI in patients with pacemaker or ICDDiagnostic test
Perform MRI in patients with pacemaker or ICD
| Study group | Role | What they receive |
|---|---|---|
| MRI in patients with Pacemaker or ICD | Perform MRI in patients with pacemaker or ICD and evaluate patient and device safety Receives: Diagnostic Test: Perform MRI in patients with pacemaker or ICD |
What is being measured
The primary outcome is the main question the study is designed to answer.
Patient safety and device malfunction during or after MRI
Device function will be assessed by comparing pre-vs. post MRI battery voltage, impedance, sensing and capture thresholds
Measured: post MRI
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted27 Aug 2014 (estimated)
- Study start30 Dec 2011
- Primary completion22 Jan 2015
- Study completion22 Jan 2015
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Publications (1)Papers the sponsor has linked to this study.
- Other trials for Conditions Requiring Diagnostic EvaluationStudies of the same condition, open and under way.
- More trials run by Johns Hopkins UniversityThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Observational
- Observational model
- Cohort
- Time perspective
- Prospective
- Enrolment
- 1,275 participants (actual)
Sponsor and investigators
- Lead sponsor
- Johns Hopkins University
- Sponsor type
- University, hospital or other
- Responsible party
- Sponsor
- Investigators
- Henry R. Halperin, M.D.,M.A., Principal investigator, Johns Hopkins University
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT02227004
- Sponsor's protocol ID
- NA_00051707
- Status last verified
- Apr 2021
- First submitted
- 20 Aug 2014
- Last updated
- 20 Apr 2021
- Results posted
- No
- Data monitoring committee
- No
- Individual data sharing
- No
Publications
Nazarian S, Hansford R, Rahsepar AA, Weltin V, McVeigh D, Gucuk Ipek E, Kwan A, Berger RD, Calkins H, Lardo AC, Kraut MA, Kamel IR, Zimmerman SL, Halperin HR. Safety of Magnetic Resonance Imaging in Patients with Cardiac Devices. N Engl J Med. 2017 Dec 28;377(26):2555-2564. doi: 10.1056/NEJMoa1604267.
PubMed 29281579
Study at a glance
From the registry record, last updated by the sponsor on 20 Apr 2021. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Safety of Clinically Indicated Magnetic Resonance Imaging in Patients With Permanent Pacemakers (PPM) and Implanted Cardioverter Defibrillators
- ClinicalTrials.gov ID
- NCT02227004
- Acronym / sponsor ID
- NA_00051707
- Keywords
- MRI, Magnetic Resonance Imaging, pacemakers, ICD
- Study type
- Observational
- Recruitment status
- Completed
- Enrolment
- 1,275 participants (actual)
- Lead sponsor
- Johns Hopkins UniversityUniversity, hospital or other
- Study start
- 30 Dec 2011 (actual)
- Primary completion
- 22 Jan 2015 (actual)
- Study completion
- 22 Jan 2015 (actual)
- First posted
- 27 Aug 2014 (estimated)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
Pacemakers and ICD's have traditionally been accepted as contraindications to MRI due to safety concerns. Recent work in our laboratory as well as our pilot study of approximately 500 patients has shown the safety of most contemporary implanted rhythm management devices in the MRI environment given appropriate precautions. Because the investigators have completed testing in animals weighing as little as 80 lbs. without incident, the investigators feel they can safely image children at this weight or greater (Circulation 2004; 110:475-482).
The goal of this study is to expand on our pilot study and to develop a protocol, consistent with the Department of Health and Human Services (DHHS) requirements, that will increase our safety experience with clinically indicated MR imaging in patients with implanted cardiac pacemakers and ICD's.
This is a cohort study. One thousand seven hundred patients with a clinical need for MRI will be recruited. Risks and Benefits associated with Magnetic Resonance (MR) imaging in the setting of implantable devices will be discussed and informed consent obtained. The type of device and leads will be ascertained. Patients will be monitored by a physician or a Registered Nurse (RN) with ACLS certification and familiarity with ICD/Pacemaker programming during the study. When imaging children who meet age and weight criteria, a physician will be present during the entire study. Fellow and attending radiology and electrophysiology support will be available in house in all cases.
Standard device testing including battery voltages, thresholds, sensing and lead impedances will be measured.
Detailed description
The primary objective of this study is to develop a protocol, consistent with DHHS/ Center for Medicare and Medicaid Services (CMS) guidelines and requirements to further document safety in clinically indicated MR imaging in patients with implanted cardioverter defibrillators (ICD's) and pacemakers.
MRI's will be conducted on patients who require a clinically indicated MRI only. All studies will be done in the 1.5 Tesla MRI unit. ECG monitoring pads will be placed on the patients for the duration of the study. An external defibrillator and Advance Cardiovascular Life Support (ACLS) drugs will be on hand. Heart rate, blood pressure, O2 saturation will also be monitored non-invasively throughout the study. All devices will undergo a complete interrogation prior to imaging. Parameters such as atrial and ventricular pacing thresholds, R and P wave amplitudes, lead impedance, and battery status will be measured and recorded. PPMs will be programmed to an asynchronous mode if dependent and to an inhibited mode in patients without pacemaker dependence.
The pacemaker function of ICD's in pacemaker independent patients will be programmed to ventricular inhibited (VVI) mode at 50 bpm. Pacemaker dependant patients with ICDs will be excluded.
In addition, there is a theoretical concern that patients who have capped or abandoned leads in their chest may be at risk for heating. While there is little or no objective data to support this concern, given this theoretical concern, patients in this category will only be admitted to the study if the referring physician deems the MRI critical and after the PI has carefully considered the risk/benefit ration for each patient who falls in to this category.
Similarly, there historically has also been a concern that newly implanted leads (<4 weeks) are at risk for lead to dislodgement d/t possible torque from the MRI environment. This too is only a theoretical concern and there is little or no scientific evidence to support this claim. Nevertheless, patients with newly implanted pacemaker/ICD leads will be considered by the PI for protocol inclusion only if the referring physician deems the MRI to be critical and after considering the risk/benefit to each individual patient.
After the appropriate MRI protocol for each MRI patient's unique condition has been completed, the device will be re-programmed to its original settings and completely interrogated to detect any changes in device performance.
Study design
- Observational model
- Cohort
- Time perspective
- Prospective
- Number of arms
- 1
| Arm | Type | Receives |
|---|---|---|
| MRI in patients with Pacemaker or ICD | – | Diagnostic Test: Perform MRI in patients with pacemaker or ICDPerform MRI in patients with pacemaker or ICD and evaluate patient and device safety |
Who may qualify
- Ages
- 9 to 100 years
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Child, Adult, Older adult
- Study population
- Males and females 9 years old or greater weighing more than 80 lbs who require a clinically indicated MRI
Key inclusion criteria (2 listed)
- Patients with an absolute clinical need for MR imaging and PPM (year 1996 model or later) or ICD's (year 2000 or later)
- -
Key exclusion criteria (8 listed)
- Inclusion: Patients with an absolute clinical need for MR imaging and PPM (year 1996 model or later) or ICD's (year 2000 or later)
- Patients with pacemaker models before 1996 and ICD models before year 2000.
- Pacemaker defendant patients with ICD.
- Patients who weighing less than 80 lbs.
- Patients less than 9 years old
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Perform MRI in patients with pacemaker or ICDPerform MRI in patients with pacemaker or ICD | Diagnostic test | MRI in patients with Pacemaker or ICD |
Primary outcome
Patient safety and device malfunction during or after MRI
Time frame: post MRI
Device function will be assessed by comparing pre-vs. post MRI battery voltage, impedance, sensing and capture thresholds
Results and safety
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 20 Apr 2021.
- Results publication: PubMed 29281579Nazarian S, Hansford R, Rahsepar AA, Weltin V, McVeigh D, Gucuk Ipek E, Kwan A, Berger RD, Calkins H, Lardo AC, Kraut MA, Kamel IR, Zimmerman SL, Halperin HR. Safety of Magnetic Resonance Imaging in Patients with Cardiac Devices. N Engl J Med. 2017 Dec 28;377(26):2555-2564. doi: 10.1056/NEJMoa1604267.
- PubMed Citation #1Link provided by the sponsor.
- Individual participant dataSharing planned: No
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Henry R. Halperin, M.D.,M.A. principal investigator
Johns Hopkins University
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
An observational study of Conditions Requiring Diagnostic Evaluation, run by Johns Hopkins University at 1 site in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 2 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 20 Apr 2021. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Patient safety and device malfunction during or after MRI (time frame: post MRI). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT02227004, which also lists 1 linked publication.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 20 Apr 2021.