GlaucomaNCT02246777
Alcon® Ex-PRESS® Glaucoma Filtration Device in Japanese Patients
A study testing Ex-PRESS® Glaucoma Filtration Device, Model P50PL in people with Glaucoma, run by Alcon Research at 1 site in Japan.
Trial summary
The purpose of this study is to assess the efficacy and safety of the Alcon Ex-PRESS® Glaucoma Filtration Device (Ex-PRESS) in Japanese subjects with normal tension glaucoma. Read more about this study
- Who is it for?
- Ages 20 years and older
- Conditions the study is for
- Glaucoma
- What is being tested?
- Ex-PRESS® Glaucoma Filtration Device, Model P50PL
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 32 participants
- Where
- 1 siteJapan
- Study dates
- 26 Feb 2014to 24 Oct 2016
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 20 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- Diagnosis of normal tension glaucoma.
- Indicated for filtration surgery using Ex-PRESS®.
- Understand and provide Informed Consent.
- Other protocol-defined inclusion criteria may apply.
Who cannot join
- Women who are pregnant, lactating, or planning to be pregnant during the study period.
- Diagnosis of angle closure glaucoma or secondary glaucoma.
- History of glaucoma surgery.
- Ophthalmologic surgery within the past 6 months.
- Difficulty with applanation tonometry measurement.
- Corneal dystrophy.
Show all 10Show fewer
- Infectious/non-infectious conjunctivitis, keratitis, or uveitis in either eye.
- Severe blepharitis or dry eye.
- History of metal allergy.
- Other protocol-defined exclusion criteria may apply.
Trial locations and contacts
1 site in 1 country.
About this study
Official title: Assessment of Efficacy and Safety of the EX-PRESS® Glaucoma Filtration Device in Patients With Normal Tension Glaucoma
The purpose of this study is to assess the efficacy and safety of the Alcon Ex-PRESS® Glaucoma Filtration Device (Ex-PRESS) in Japanese subjects with normal tension glaucoma.
- Conditions
- Glaucoma
- Being tested
- Ex-PRESS® Glaucoma Filtration Device, Model P50PL
- Keywords
- Glaucoma
- Medical subject headings
- Glaucoma
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Ex-PRESS® Glaucoma Filtration Device, Model P50PLDevice
| Study group | Role | What they receive |
|---|---|---|
| Ex-PRESS | Experimental | Ex-PRESS® Glaucoma Filtration Device, Model P50PL, implanted in the anterior chamber of the study eye during glaucoma surgery intended for the lifetime of the patient Receives: Device: Ex-PRESS® Glaucoma Filtration Device, Model P50PL |
What is being measured
The primary outcome is the main question the study is designed to answer.
Mean Intraocular Pressure (IOP)
IOP (fluid pressure inside the eye) was assessed using Goldmann applanation tonometry and reported in mmHg. A higher IOP can be a greater risk factor for developing glaucoma or glaucoma progression (leading to optic nerve damage). Only one eye contributed to the analysis.
Measured: Month 3, Month 6, Month 12 Post-Operative
Change From Baseline in IOP
IOP (fluid pressure inside the eye) was assessed using Goldmann applanation tonometry and reported in mmHg. A higher IOP can be a greater risk factor for developing glaucoma or glaucoma progression (leading to optic nerve damage). A more negative change from baseline indicates a greater improvement, i.e., a reduction of IOP. Only one eye contributed to the analysis.
Measured: Baseline (Pre-Operative), Month 3, Month 6, Month 12 Post-Operative
Percent Change From Baseline in IOP
IOP (fluid pressure inside the eye) was assessed using Goldmann applanation tonometry and reported in mmHg. A higher IOP can be a greater risk factor for developing glaucoma or glaucoma progression (leading to optic nerve damage). A more negative percent change from baseline indicates a greater improvement, i.e., a reduction of IOP. Only one eye contributed to the analysis.
Measured: Baseline (Pre-Operative), Month 3, Month 6, Month 12 Post-Operative
Percentage of Eyes With IOP Lowering Rate of 20% or More From Baseline up to Month 12
IOP (fluid pressure inside the eye) was assessed using Goldmann applanation tonometry and reported in mmHg. A higher IOP can be a greater risk factor for developing glaucoma or glaucoma progression (leading to optic nerve damage). Only one eye contributed to the analysis.
Measured: Baseline (Pre-Operative), Month 3, Month 6, Month 12 Post-Operative
Percentage of Eyes Receiving Drug Therapy for Glaucoma Necessary to Maintain the IOP
IOP (fluid pressure inside the eye) was assessed using Goldmann applanation tonometry and reported in mmHg. A higher IOP can be a greater risk factor for developing glaucoma or glaucoma progression (leading to optic nerve damage). For some subjects, both left and right eyes were targeted for the study, and data from both eyes were analyzed for this safety endpoint, as specified in the protocol.
Measured: Month 3, Month 6, Month 12 Post-Operative
Percentage of Eyes Receiving Secondary Surgical Treatment (Including Laser Therapy) Necessary to Maintain the IOP
IOP (fluid pressure inside the eye) was assessed using Goldmann applanation tonometry and reported in mmHg. A higher IOP can be a greater risk factor for developing glaucoma or glaucoma progression (leading to optic nerve damage). Secondary surgical treatment included needling, laser suture lysis, and conjunctival and scleral flap sutures. For some subjects, both left and right eyes were targeted for the study, and data from both eyes were analyzed for this safety endpoint, as specified in the protocol. An eye may have received more than one procedure.
Measured: Month 3, Month 6, Month 12 Post-Operative
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted23 Sep 2014 (estimated)
- Study start26 Feb 2014
- Primary completion24 Oct 2016
- Study completion24 Oct 2016
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Other trials for GlaucomaStudies of the same condition, open and under way.
- More trials run by Alcon ResearchThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Treatment
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 32 participants (actual)
Sponsor and investigators
- Lead sponsor
- Alcon Research
- Sponsor type
- Industry (pharma, biotech, devices)
- Responsible party
- Sponsor
- Investigators
- Alcon, A Novartis Division, Study director, Alcon Japan, Ltd.
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT02246777
- Sponsor's protocol ID
- ALJ-P2013-1
- Other IDs
- UMIN000013243
- Status last verified
- Jul 2018
- First submitted
- 18 Sep 2014
- Last updated
- 20 Jul 2018
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
Study at a glance
From the registry record, last updated by the sponsor on 20 Jul 2018. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Assessment of Efficacy and Safety of the EX-PRESS® Glaucoma Filtration Device in Patients With Normal Tension Glaucoma
- ClinicalTrials.gov ID
- NCT02246777
- Acronym / sponsor ID
- ALJ-P2013-1
- Conditions
- Glaucoma
- Keywords
- Glaucoma
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 32 participants (actual)
- Lead sponsor
- Alcon ResearchIndustry (pharma, biotech, devices)
- Study start
- 26 Feb 2014 (actual)
- Primary completion
- 24 Oct 2016 (actual)
- Study completion
- 24 Oct 2016 (actual)
- First posted
- 23 Sep 2014 (estimated)
- Results
- Posted 20 Jul 2018
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The purpose of this study is to assess the efficacy and safety of the Alcon Ex-PRESS® Glaucoma Filtration Device (Ex-PRESS) in Japanese subjects with normal tension glaucoma.
Study design
- Primary purpose
- Treatment
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
| Arm | Type | Receives |
|---|---|---|
| Ex-PRESS | Experimental | Device: Ex-PRESS® Glaucoma Filtration Device, Model P50PLEx-PRESS® Glaucoma Filtration Device, Model P50PL, implanted in the anterior chamber of the study eye during glaucoma surgery intended for the lifetime of the patient |
Who may qualify
- Ages
- 20 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (4 listed)
- Diagnosis of normal tension glaucoma.
- Indicated for filtration surgery using Ex-PRESS®.
- Understand and provide Informed Consent.
- Other protocol-defined inclusion criteria may apply.
Key exclusion criteria (10 listed)
- Women who are pregnant, lactating, or planning to be pregnant during the study period.
- Diagnosis of angle closure glaucoma or secondary glaucoma.
- History of glaucoma surgery.
- Ophthalmologic surgery within the past 6 months.
- Difficulty with applanation tonometry measurement.
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Ex-PRESS® Glaucoma Filtration Device, Model P50PL | Device | Ex-PRESS |
Primary outcome
Mean Intraocular Pressure (IOP)
Time frame: Month 3, Month 6, Month 12 Post-Operative
IOP (fluid pressure inside the eye) was assessed using Goldmann applanation tonometry and reported in mmHg. A higher IOP can be a greater risk factor for developing glaucoma or glaucoma progression (leading to optic nerve damage). Only one eye contributed to the analysis.
Change From Baseline in IOP
Time frame: Baseline (Pre-Operative), Month 3, Month 6, Month 12 Post-Operative
IOP (fluid pressure inside the eye) was assessed using Goldmann applanation tonometry and reported in mmHg. A higher IOP can be a greater risk factor for developing glaucoma or glaucoma progression (leading to optic nerve damage). A more negative change from baseline indicates a greater improvement, i.e., a reduction of IOP. Only one eye contributed to the analysis.
Percent Change From Baseline in IOP
Time frame: Baseline (Pre-Operative), Month 3, Month 6, Month 12 Post-Operative
IOP (fluid pressure inside the eye) was assessed using Goldmann applanation tonometry and reported in mmHg. A higher IOP can be a greater risk factor for developing glaucoma or glaucoma progression (leading to optic nerve damage). A more negative percent change from baseline indicates a greater improvement, i.e., a reduction of IOP. Only one eye contributed to the analysis.
Percentage of Eyes With IOP Lowering Rate of 20% or More From Baseline up to Month 12
Time frame: Baseline (Pre-Operative), Month 3, Month 6, Month 12 Post-Operative
IOP (fluid pressure inside the eye) was assessed using Goldmann applanation tonometry and reported in mmHg. A higher IOP can be a greater risk factor for developing glaucoma or glaucoma progression (leading to optic nerve damage). Only one eye contributed to the analysis.
Percentage of Eyes Receiving Drug Therapy for Glaucoma Necessary to Maintain the IOP
Time frame: Month 3, Month 6, Month 12 Post-Operative
IOP (fluid pressure inside the eye) was assessed using Goldmann applanation tonometry and reported in mmHg. A higher IOP can be a greater risk factor for developing glaucoma or glaucoma progression (leading to optic nerve damage). For some subjects, both left and right eyes were targeted for the study, and data from both eyes were analyzed for this safety endpoint, as specified in the protocol.
Percentage of Eyes Receiving Secondary Surgical Treatment (Including Laser Therapy) Necessary to Maintain the IOP
Time frame: Month 3, Month 6, Month 12 Post-Operative
IOP (fluid pressure inside the eye) was assessed using Goldmann applanation tonometry and reported in mmHg. A higher IOP can be a greater risk factor for developing glaucoma or glaucoma progression (leading to optic nerve damage). Secondary surgical treatment included needling, laser suture lysis, and conjunctival and scleral flap sutures. For some subjects, both left and right eyes were targeted for the study, and data from both eyes were analyzed for this safety endpoint, as specified in the protocol. An eye may have received more than one procedure.
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 20 Jul 2018.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
Related trials for Glaucoma
Recruiting
Implementing AI-based Glaucoma Screening Within Federally Qualified Health Centers
Glaucoma, Ocular Hypertension
Not yet recruiting
AI-Assisted vs Conventional Teleglaucoma Referral
Glaucoma
Recruiting
Brain Stimulation Effects on Orientation and Mobility Skills in Adults With Vision Impairment
Retinitis Pigmentosa (RP), Rod Cone Dystrophy
Recruiting
Automated Applanation Tonometry - Updated
Glaucoma, Glaucoma, Suspect
RecruitingPhase 2
A Phase 2a Study Comparing the Efficacy, Safety and Tolerability of Pregabalin Eye Drops Comparing With Timolol Eye Drops in Adults With Glaucoma or High Eye Pressure.
Primary Open - Angle Glaucoma (POAG), Ocular Hypertension (OHT)
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Alcon, A Novartis Division study director
Alcon Japan, Ltd.
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Ex-PRESS® Glaucoma Filtration Device, Model P50PL in people with Glaucoma, run by Alcon Research at 1 site in Japan.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 4 inclusion and 10 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 20 Jul 2018. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Japan. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 20 Jul 2018). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Mean Intraocular Pressure (IOP) (time frame: Month 3, Month 6, Month 12 Post-Operative). The study lists 6 primary outcomes in all. It is what the study was designed to measure, not a finding.