Accuracy Assessment of an Automatic Blood Pressure Measurement Device in Adult Volunteers

A study testing Sphygmo: Automatic Blood Pressure Monitor and GE Dinamap ProCare Automatic Blood Pressure Monitor in people with Pre Eclampsia, run by William Marsh Rice University at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The research team has developed an automatic blood pressure monitor (Sphygmo) to be used for the monitoring and diagnosis of pre-eclampsia in pregnant women, particularly in low-resource settings where current monitoring is limited. 90 adult volunteers will be enrolled by researchers at Rice University. Read more about this study

Who is it for?
Ages 18 years and olderHealthy volunteers can take part
Conditions the study is for
Pre Eclampsia
What is being tested?
Sphygmo: Automatic Blood Pressure Monitor and GE Dinamap ProCare Automatic Blood Pressure Monitor
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
90 participants
Where
1 siteUnited States
Study dates
Jan 2015to 12 Oct 2017
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • 18 years of age or older
  • Willing and able to provide informed consent

Who cannot join

  • Unable to provide informed consent

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Rice University

    Houston, Texas, 77030

Other trials in United States

About this study

The research team has developed an automatic blood pressure monitor (Sphygmo) to be used for the monitoring and diagnosis of pre-eclampsia in pregnant women, particularly in low-resource settings where current monitoring is limited.

90 adult volunteers will be enrolled by researchers at Rice University. The participant will be seated in a comfortable chair with arm at heart level. Arm circumference will be measured and a blood pressure cuff will be placed on the arm. The cuff will be inflated and blood pressure measurements will be taken by a commercially available device and by the Sphygmo device. Blood pressure measurements from both devices will be recorded. The participant's blood pressure will be measured up to 9 times with a waiting period of 45-60 seconds between each measurement.

The results of this study will be used to optimize the blood pressure detection algorithm and thus further develop the device.

Conditions
Pre Eclampsia
Being tested
Sphygmo: Automatic Blood Pressure Monitor, GE Dinamap ProCare Automatic Blood Pressure Monitor
Medical subject headings
Pre-Eclampsia

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Sphygmo: Automatic Blood Pressure MonitorDevice

    A team of engineers from Rice University has recently developed Sphygmo, an ambulatory, low-cost blood pressure monitor for use in the diagnosis and management of pre-eclampsia in low-resource hospitals.

  • GE Dinamap ProCare Automatic Blood Pressure MonitorDevice

    This commercially available gold standard blood pressure monitor will be used as a control to assess the accuracy of the Sphygmo device.

Study groups in NCT02267577
Study groupRoleWhat they receive
Healthy Adults VolunteersExperimentalMen and women over the age of 18 will have their blood pressure measured with both the Sphygmo: Automatic Blood Pressure Monitor device and the GE Dinamap ProCare automatic blood pressure monitor.
Receives: Device: Sphygmo: Automatic Blood Pressure Monitor; Device: GE Dinamap ProCare Automatic Blood Pressure Monitor

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Accuracy of Blood Pressure Measurement by Sphygmo Relative to the Clinical Standard.

    Mean difference of systolic and diastolic blood pressure readings between Sphygmo device and the gold standard sphygmomanometer.

    Measured: Blood pressure was measured an average of 9 times for each participant during their single visit and the mean of these measurements was recorded and used for analysis. The measurement period lasted approximately 30-45 minutes.

  2. Percentage of Sphygmo Readings That Were With < 5 mmHg, < 10 mmHg, and < 15 mmHg of the Readings by the Gold Standard Sphygmomanometer.

    The British Hypertension society defines a specific criteria for the accuracy of a sphygmomanometer. Devices are graded according to the cumulative percentage of readings that have an absolute difference between the more favorable observer's mercury sphygmomanometer readings and the test device of < 5 mmHg, < 10 mmHg, and < 15 mmHg. A letter grade of "A" requires that over 60%, 85%, and 95% were achieved in the < 5 mmHg, < 10 mmHg, and <15 mmHg categories, respectively. A letter grade of "B" requires that over 50%, 75%, and 90% were achieved in the < 5 mmHg, < 10 mmHg, and <15 mmHg categories, respectively. The test device achieved a grade of (B/B) with the validation data from this study.

    Measured: Blood pressure was measured an average of 9 times for each participant during their single visit and the mean of these measurements was recorded and used for analysis. The measurement period lasted approximately 30-45 minutes.

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted17 Oct 2014 (estimated)
  2. Study startJan 2015
  3. Primary completionOct 2015
  4. Study completion12 Oct 2017

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Device feasibility
Allocation
Not applicable
Intervention model
Single group
Model description
Participants enrolled in the study will have their blood pressure measured multiple times with both the GE Dinamap Procare commercial blood pressure monitor and Sphygmo, the experimental blood pressure monitoring device.
Masking
None (open label)
Enrolment
90 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT02267577
Sponsor's protocol ID
14-082F-A
Other IDs
51262
Status last verified
Mar 2020
First submitted
9 Oct 2014
Last updated
6 Mar 2020
Data monitoring committee
No

Questions about this trial

What is this clinical trial studying?

A study testing Sphygmo: Automatic Blood Pressure Monitor and GE Dinamap ProCare Automatic Blood Pressure Monitor in people with Pre Eclampsia, run by William Marsh Rice University at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 2 inclusion and 1 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 6 Mar 2020. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 15 Feb 2019). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Accuracy of Blood Pressure Measurement by Sphygmo Relative to the Clinical Standard. (time frame: Blood pressure was measured an average of 9 times for each participant during their single visit and the mean of these measurements was recorded and used for analysis. The measurement period lasted approximately 30-45 minutes.). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 6 Mar 2020.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in