Pre EclampsiaNCT02319174
Accuracy Assessment of an Automatic Blood Pressure Measurement Device in Pregnant Women
A study testing Sphygmo in people with Pre Eclampsia, run by William Marsh Rice University at 1 site in United States.
Trial summary
The purpose of the study is to help make a lower cost automatic blood pressure monitor device for diagnosis and monitoring of pre-eclampsia in pregnant women, where automatic blood pressure monitoring is limited or not available. Read more about this study
- Who is it for?
- Ages 18 years and older, women only
- Conditions the study is for
- Pre Eclampsia
- What is being tested?
- Sphygmo
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 43 participants
- Where
- 1 siteUnited States
- Study dates
- Jun 2015to 4 Nov 2016
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- Women only
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- Pregnant women receiving antenatal care at the UT OB/GYN clinic.
- Over the age of 18
Who cannot join
- Women under the age of 18
- Women unable to provide informed consent
Trial locations and contacts
1 site in 1 country.
United States1 site
University of Texas Health Science Center Houston
Houston, Texas, 77030
About this study
The purpose of the study is to help make a lower cost automatic blood pressure monitor device for diagnosis and monitoring of pre-eclampsia in pregnant women, where automatic blood pressure monitoring is limited or not available. The study will compare this low cost device to a commercially available system used for pre-eclamptic women in many United States hospitals that the investigators will be bringing to Malawi as a part of this study.
The team hopes to show that this lower cost blood pressure machine works well and can help women with pre-eclampsia. The study also aims to see if this machine is easy for the nurse to use.
Read more
70 pregnant women who are either at-risk or diagnosed with pre-eclampsia will be enrolled at University of Texas Health Science Center Houston. Patient arm circumference will be measured with measurement tape.
They will be seated upright in a comfortable chair with arm at heart level and an arm blood pressure cuff from either the automatic blood pressure monitor or a manual sphygmomanometer will be placed on the left arm.
The cuff will be inflated and then deflated until measurement concludes. Heart rate will be measured with tactile arterial palpation.The process will be repeated for a total of up to nine measurements, alternating between measurements with the automatic blood pressure monitor and the manual sphygmomanometer. There will be a waiting period of 45-60 seconds between each measurement.
The results of this study will help researchers understand the performance and usability of this device in Malawi and help decide if any design changes are needed.
Read the detailed description
The research team has developed an automatic blood pressure monitor to be used for the monitoring and diagnosis of pre-eclampsia in pregnant women, particularly in low-resource settings where current monitoring is limited. This study aims to evaluate the accuracy of the device's blood pressure measurements in pregnant and pre-eclamptic women. This device must be evaluated with the above described population because blood pressure measurement devices are known to perform differently in pregnant and pre-eclamptic women than in normal healthy adults.
This study will take place at the University of Texas Health Science Center (UTHSC) where up to 70 subjects will be recruited to participate during their regularly scheduled antenatal care checkups.
- The Patient arm circumference will be measured with measurement tape.
- The Patient will be seated upright in a comfortable chair with arm at heart level.
- An arm blood pressure cuff from either the automatic blood pressure monitor or a manual sphygmomanometer will be placed on the left arm of the patient over the brachial artery.
- Cuff will be inflated to ~200 mmHg and then deflated until measurement concludes.
- Heart rate will be measured with Tactile Arterial Palpation.
- Steps 3-4 will be repeated for a total of up to nine measurements, alternating between measurements with the automatic blood pressure monitor and the manual sphygmomanometer. There will be a waiting period of 45-60 seconds between each measurement.
- Conditions
- Pre Eclampsia
- Being tested
- Sphygmo
- Medical subject headings
- Pre-Eclampsia
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
SphygmoDevice
A team of engineers from Rice University has recently developed Sphygmo, an ambulatory, low-cost blood pressure monitor for use in the diagnosis and management of pre-eclampsia in low-resource hospitals
| Study group | Role | What they receive |
|---|---|---|
| Sphygmo | Experimental | All subjects will have blood pressure measured by both the research device (Sphygmo) and the commercially available device. All clinical decisions will be made using measurements from the commercially available device. Receives: Device: Sphygmo |
What is being measured
The primary outcome is the main question the study is designed to answer.
Accuracy of Blood Pressure Readings by Sphygmo
Mean difference of systolic and diastolic blood pressures between Sphygmo blood pressure measurements and measurements from the gold standard sphygmomanometer.
Measured: Blood pressure was measured an average of 9 times for each participant and the average of these measurements was recorded. Each measurement period lasted approximately 30-45 minutes
Percent of Sphygmo Readings That Were Within 5mmHg, 10mmHg, and 15mmHg of the Readings by the Gold Standard Sphygmomanometer.
The British Hypertension society defines a specific criteria for the accuracy of a sphygmomanometer. Devices are graded according to the cumulative percentage of readings that have an absolute difference between the more favorable observer's mercury sphygmomanometer readings and the test device of < 5 mmHg, < 10 mmHg, and < 15 mmHg. A letter grade of "A" requires that over 60%, 85%, and 95% were achieved in the < 5 mmHg, < 10 mmHg, and <15 mmHg categories, respectively. A letter grade of "B" requires that over 50%, 75%, and 90% were achieved in the < 5 mmHg, < 10 mmHg, and <15 mmHg categories, respectively. The test device achieved a grade of (A/A) with the validation data from this study.
Measured: Blood pressure was measured an average of 9 times for each participant during their single measurement period, and the average of those measurements was recorded. Each measurement period lasted approximately 30-45 minutes
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted18 Dec 2014 (estimated)
- Study startJun 2015
- Primary completionJun 2015
- Study completion4 Nov 2016
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Publications (6)Papers the sponsor has linked to this study.
- Other trials for Pre EclampsiaStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Device feasibility
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 43 participants (actual)
Sponsor and investigators
- Lead sponsor
- William Marsh Rice University
- Sponsor type
- University, hospital or other
- Responsible party
- Principal investigator, Rebecca Richards-Kortum, William Marsh Rice University
- Investigators
- Rebecca R Kortum, PhD, Principal investigator, William Marsh Rice University
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT02319174
- Sponsor's protocol ID
- 14-082F-B
- Status last verified
- Dec 2018
- First submitted
- 10 Dec 2014
- Last updated
- 19 Dec 2018
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
Publications
Natarajan P, Shennan AH, Penny J, Halligan AW, de Swiet M, Anthony J. Comparison of auscultatory and oscillometric automated blood pressure monitors in the setting of preeclampsia. Am J Obstet Gynecol. 1999 Nov;181(5 Pt 1):1203-10. doi: 10.1016/s0002-9378(99)70109-2.
PubMed 10561646Turner JA. Diagnosis and management of pre-eclampsia: an update. Int J Womens Health. 2010 Sep 30;2:327-37. doi: 10.2147/IJWH.S8550.
PubMed 21151680Dekker G, Sibai B. Primary, secondary, and tertiary prevention of pre-eclampsia. Lancet. 2001 Jan 20;357(9251):209-15. doi: 10.1016/S0140-6736(00)03599-6.
PubMed 11213110Magee LA, Abalos E, von Dadelszen P, Sibai B, Easterling T, Walkinshaw S; CHIPS Study Group. How to manage hypertension in pregnancy effectively. Br J Clin Pharmacol. 2011 Sep;72(3):394-401. doi: 10.1111/j.1365-2125.2011.04002.x.
PubMed 21545480De Greeff A, Ghosh D, Anthony J, Shennan A. Accuracy assessment of the Dinamap ProCare 400 in pregnancy and preeclampsia. Hypertens Pregnancy. 2010 Jan;29(2):198-205. doi: 10.3109/10641950902968650.
PubMed 20367507O'Brien E, Petrie J, Littler W, de Swiet M, Padfield PL, Altman DG, Bland M, Coats A, Atkins N. An outline of the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens. 1993 Jun;11(6):677-9. doi: 10.1097/00004872-199306000-00013. No abstract available.
PubMed 8397248
Study at a glance
From the registry record, last updated by the sponsor on 19 Dec 2018. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Accuracy Assessment of an Automatic Blood Pressure Measurement Device in Pregnant Women
- ClinicalTrials.gov ID
- NCT02319174
- Acronym / sponsor ID
- 14-082F-B
- Conditions
- Pre Eclampsia
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 43 participants (actual)
- Lead sponsor
- William Marsh Rice UniversityUniversity, hospital or other
- Collaborators
- The University of Texas Health Science Center, Houston
- Study start
- Jun 2015
- Primary completion
- Jun 2015 (actual)
- Study completion
- 4 Nov 2016 (actual)
- First posted
- 18 Dec 2014 (estimated)
- Results
- Posted 19 Dec 2018
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The purpose of the study is to help make a lower cost automatic blood pressure monitor device for diagnosis and monitoring of pre-eclampsia in pregnant women, where automatic blood pressure monitoring is limited or not available. The study will compare this low cost device to a commercially available system used for pre-eclamptic women in many United States hospitals that the investigators will be bringing to Malawi as a part of this study.
The team hopes to show that this lower cost blood pressure machine works well and can help women with pre-eclampsia. The study also aims to see if this machine is easy for the nurse to use.
70 pregnant women who are either at-risk or diagnosed with pre-eclampsia will be enrolled at University of Texas Health Science Center Houston. Patient arm circumference will be measured with measurement tape.
They will be seated upright in a comfortable chair with arm at heart level and an arm blood pressure cuff from either the automatic blood pressure monitor or a manual sphygmomanometer will be placed on the left arm.
The cuff will be inflated and then deflated until measurement concludes. Heart rate will be measured with tactile arterial palpation.The process will be repeated for a total of up to nine measurements, alternating between measurements with the automatic blood pressure monitor and the manual sphygmomanometer. There will be a waiting period of 45-60 seconds between each measurement.
The results of this study will help researchers understand the performance and usability of this device in Malawi and help decide if any design changes are needed.
Detailed description
The research team has developed an automatic blood pressure monitor to be used for the monitoring and diagnosis of pre-eclampsia in pregnant women, particularly in low-resource settings where current monitoring is limited. This study aims to evaluate the accuracy of the device's blood pressure measurements in pregnant and pre-eclamptic women. This device must be evaluated with the above described population because blood pressure measurement devices are known to perform differently in pregnant and pre-eclamptic women than in normal healthy adults.
This study will take place at the University of Texas Health Science Center (UTHSC) where up to 70 subjects will be recruited to participate during their regularly scheduled antenatal care checkups.
- The Patient arm circumference will be measured with measurement tape.
- The Patient will be seated upright in a comfortable chair with arm at heart level.
- An arm blood pressure cuff from either the automatic blood pressure monitor or a manual sphygmomanometer will be placed on the left arm of the patient over the brachial artery.
- Cuff will be inflated to ~200 mmHg and then deflated until measurement concludes.
- Heart rate will be measured with Tactile Arterial Palpation.
- Steps 3-4 will be repeated for a total of up to nine measurements, alternating between measurements with the automatic blood pressure monitor and the manual sphygmomanometer. There will be a waiting period of 45-60 seconds between each measurement.
Study design
- Primary purpose
- Device feasibility
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
| Arm | Type | Receives |
|---|---|---|
| Sphygmo | Experimental | Device: SphygmoAll subjects will have blood pressure measured by both the research device (Sphygmo) and the commercially available device. All clinical decisions will be made using measurements from the commercially available device. |
Who may qualify
- Ages
- 18 years and older
- Sex
- Female only
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (2 listed)
- Pregnant women receiving antenatal care at the UT OB/GYN clinic.
- Over the age of 18
Key exclusion criteria (2 listed)
- Women under the age of 18
- Women unable to provide informed consent
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| SphygmoA team of engineers from Rice University has recently developed Sphygmo, an ambulatory, low-cost blood pressure monitor for use in the diagnosis and management of pre-eclampsia in low-resource hospitals | Device | Sphygmo |
Primary outcome
Accuracy of Blood Pressure Readings by Sphygmo
Time frame: Blood pressure was measured an average of 9 times for each participant and the average of these measurements was recorded. Each measurement period lasted approximately 30-45 minutes
Mean difference of systolic and diastolic blood pressures between Sphygmo blood pressure measurements and measurements from the gold standard sphygmomanometer.
Percent of Sphygmo Readings That Were Within 5mmHg, 10mmHg, and 15mmHg of the Readings by the Gold Standard Sphygmomanometer.
Time frame: Blood pressure was measured an average of 9 times for each participant during their single measurement period, and the average of those measurements was recorded. Each measurement period lasted approximately 30-45 minutes
The British Hypertension society defines a specific criteria for the accuracy of a sphygmomanometer. Devices are graded according to the cumulative percentage of readings that have an absolute difference between the more favorable observer's mercury sphygmomanometer readings and the test device of < 5 mmHg, < 10 mmHg, and < 15 mmHg. A letter grade of "A" requires that over 60%, 85%, and 95% were achieved in the < 5 mmHg, < 10 mmHg, and <15 mmHg categories, respectively. A letter grade of "B" requires that over 50%, 75%, and 90% were achieved in the < 5 mmHg, < 10 mmHg, and <15 mmHg categories, respectively. The test device achieved a grade of (A/A) with the validation data from this study.
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 19 Dec 2018.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
- Background: PubMed 10561646Natarajan P, Shennan AH, Penny J, Halligan AW, de Swiet M, Anthony J. Comparison of auscultatory and oscillometric automated blood pressure monitors in the setting of preeclampsia. Am J Obstet Gynecol. 1999 Nov;181(5 Pt 1):1203-10. doi: 10.1016/s0002-9378(99)70109-2.
- Background: PubMed 21151680Turner JA. Diagnosis and management of pre-eclampsia: an update. Int J Womens Health. 2010 Sep 30;2:327-37. doi: 10.2147/IJWH.S8550.
- Background: PubMed 11213110Dekker G, Sibai B. Primary, secondary, and tertiary prevention of pre-eclampsia. Lancet. 2001 Jan 20;357(9251):209-15. doi: 10.1016/S0140-6736(00)03599-6.
- Background: PubMed 21545480Magee LA, Abalos E, von Dadelszen P, Sibai B, Easterling T, Walkinshaw S; CHIPS Study Group. How to manage hypertension in pregnancy effectively. Br J Clin Pharmacol. 2011 Sep;72(3):394-401. doi: 10.1111/j.1365-2125.2011.04002.x.
- Background: PubMed 20367507De Greeff A, Ghosh D, Anthony J, Shennan A. Accuracy assessment of the Dinamap ProCare 400 in pregnancy and preeclampsia. Hypertens Pregnancy. 2010 Jan;29(2):198-205. doi: 10.3109/10641950902968650.
- Background: PubMed 8397248O'Brien E, Petrie J, Littler W, de Swiet M, Padfield PL, Altman DG, Bland M, Coats A, Atkins N. An outline of the revised British Hypertension Society protocol for the evaluation of blood pressure measuring devices. J Hypertens. 1993 Jun;11(6):677-9. doi: 10.1097/00004872-199306000-00013. No abstract available.
Related trials for Pre Eclampsia
Recruiting
Exercise to Improve Heart and Metabolic Health After Childbirth in Women at High Risk for Preeclampsia
Postpartum Exercise, Cardiometabolic Health
RecruitingObservational study
Molecular Study of the Maternal-fetal Interface in Preeclampsia.
Preeclampsia
Active, not recruitingObservational study
The CoPenHagen PREeClampsia and cardIOvascUlar diSease Study
Pre-Eclampsia, Atherosclerosis
RecruitingObservational study
Estimated Risk of Preeclampsia in the First Trimester and Its Association With Subclinical Atherosclerosis and Cardiovascular Risk in Women With Type 1 Diabetes
Type 1 Diabetes, Preeclampsia
Not yet recruitingPhase 2
Furosemide vs Placebo in Severe Preeclampsia Postpartum
Severe Preeclampsia
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Rebecca R Kortum, PhD principal investigator
William Marsh Rice University
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Sphygmo in people with Pre Eclampsia, run by William Marsh Rice University at 1 site in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 2 inclusion and 2 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 19 Dec 2018. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 19 Dec 2018). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Accuracy of Blood Pressure Readings by Sphygmo (time frame: Blood pressure was measured an average of 9 times for each participant and the average of these measurements was recorded. Each measurement period lasted approximately 30-45 minutes). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT02319174, which also lists 6 linked publications.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 19 Dec 2018.