DiabetesNCT02352298
Protocol for the Evaluation of Dario™ BGMS Performance at Elevated Altitude
A study testing Dario Blood Glucose Monitoring System and YSI Analyzer in people with Diabetes, run by LabStyle Innovations Ltd. at 1 site in United States.
Trial summary
The purpose of this study is to assess the performance of the Dario Blood Glucose Monitor (BGMS), at elevated altitudes, by obtaining whole blood fingerstick samples from individuals and analyzing the blood on the Dario Blood Glucose Monitor (BGMS) and comparing the readings to readings obtained from the YSI 2300 STAT… Read more about this study
- Who is it for?
- Ages 18 years and olderHealthy volunteers can take part
- Conditions the study is for
- Diabetes
- What is being tested?
- Dario Blood Glucose Monitoring System and YSI Analyzer
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 52 participants
- Where
- 1 siteUnited States
- Study dates
- Feb 2015to Feb 2015
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- Healthy volunteers accepted
- Adult, Older adult
Who may be able to join
- Age of subject is 18 years or older
- Able to speak and read English
Who cannot join
- Pregnant
- Has any condition that the principal investigator believes may interfere in the subject's participation in the trial.
Trial locations and contacts
1 site in 1 country.
About this study
Official title: Protocol for the Evaluation of Dario™ Blood Glucose Monitor (BGMS) Performance at Elevated Altitude
The purpose of this study is to assess the performance of the Dario Blood Glucose Monitor (BGMS), at elevated altitudes, by obtaining whole blood fingerstick samples from individuals and analyzing the blood on the Dario Blood Glucose Monitor (BGMS) and comparing the readings to readings obtained from the YSI 2300 STAT Plus glucose analyzer.
Read the detailed description
The study will be performed with 50 fresh capillary blood samples, collected by fingerstick using Dario single use lancet. Additional fingerstick with dedicated disposable lancets will be used to collect sufficient capillary fresh blood to be analyzed by the YSI 2300 STAT Plus analyzer. Additional fingersticks may be necessary if volume of blood from first fingerstick is insufficient.
- Conditions
- Diabetes
- Being tested
- Dario Blood Glucose Monitoring System, YSI Analyzer
- Medical subject headings
- Diabetes Mellitus
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Dario Blood Glucose Monitoring SystemDevice
Fingerstick to obtain blood sample for analysis with the Dario BGMS
YSI AnalyzerDevice
Fingerstick to obtain blood sample for analysis with YSI
| Study group | Role | What they receive |
|---|---|---|
| Dario Blood Glucose Monitoring System | Experimental | Blood obtained via fingerstick and blood glucose level is tested on the Dario Blood Glucose Monitoring System Receives: Device: Dario Blood Glucose Monitoring System |
| YSI STAT | Active comparator | Blood obtained via fingerstick and blood glucose level is tested on the YSI STAT for comparison to the results obtained with the Dario Blood Glucose Monitoring System Receives: Device: YSI Analyzer |
What is being measured
The primary outcome is the main question the study is designed to answer.
Dario Blood Glucose Monitoring System Accuracy at High Altitude (ISO 15197:2003)
Accuracy of Dario Blood Glucose Monitoring System when compared to Yellow Springs Instrument 2300 reference method at 10,152 feet above sea level. Acceptance criteria set per (International Organization for Standardization) ISO 15197:2003. Greater than or equal to 95% of the Dario results shall fall within plus or minus 15 mg/dL of the reference method value (for glucose concentrations <75 mg/dL). For glucose concentrations greater than or equal to 75 mg/dL, the Dario results shall fall within plus or minus 20% of the reference method value.
Measured: 6 seconds
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted2 Feb 2015 (estimated)
- Study startFeb 2015
- Primary completionFeb 2015
- Study completionFeb 2015
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Other trials for DiabetesStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Other
- Allocation
- Non-randomised
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 52 participants (actual)
Sponsor and investigators
- Lead sponsor
- LabStyle Innovations Ltd.
- Sponsor type
- Industry (pharma, biotech, devices)
- Responsible party
- Sponsor
- Investigators
- Christie Beatson, Study chair, Barbara Davis Center
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT02352298
- Sponsor's protocol ID
- CP-0003
- Status last verified
- May 2017
- First submitted
- 21 Jan 2015
- Last updated
- 9 Jun 2017
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
- Individual data sharing
- No
Study at a glance
From the registry record, last updated by the sponsor on 9 Jun 2017. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Protocol for the Evaluation of Dario™ Blood Glucose Monitor (BGMS) Performance at Elevated Altitude
- ClinicalTrials.gov ID
- NCT02352298
- Acronym / sponsor ID
- CP-0003
- Conditions
- Diabetes
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 52 participants (actual)
- Lead sponsor
- LabStyle Innovations Ltd.Industry (pharma, biotech, devices)
- Study start
- Feb 2015
- Primary completion
- Feb 2015 (actual)
- Study completion
- Feb 2015 (actual)
- First posted
- 2 Feb 2015 (estimated)
- Results
- Posted 9 Jun 2017
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The purpose of this study is to assess the performance of the Dario Blood Glucose Monitor (BGMS), at elevated altitudes, by obtaining whole blood fingerstick samples from individuals and analyzing the blood on the Dario Blood Glucose Monitor (BGMS) and comparing the readings to readings obtained from the YSI 2300 STAT Plus glucose analyzer.
Detailed description
The study will be performed with 50 fresh capillary blood samples, collected by fingerstick using Dario single use lancet. Additional fingerstick with dedicated disposable lancets will be used to collect sufficient capillary fresh blood to be analyzed by the YSI 2300 STAT Plus analyzer. Additional fingersticks may be necessary if volume of blood from first fingerstick is insufficient.
Study design
- Primary purpose
- Other
- Allocation
- Non-randomised
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 2
| Arm | Type | Receives |
|---|---|---|
| Dario Blood Glucose Monitoring System | Experimental | Device: Dario Blood Glucose Monitoring SystemBlood obtained via fingerstick and blood glucose level is tested on the Dario Blood Glucose Monitoring System |
| YSI STAT | Active comparator | Device: YSI AnalyzerBlood obtained via fingerstick and blood glucose level is tested on the YSI STAT for comparison to the results obtained with the Dario Blood Glucose Monitoring System |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (2 listed)
- Age of subject is 18 years or older
- Able to speak and read English
Key exclusion criteria (2 listed)
- Pregnant
- Has any condition that the principal investigator believes may interfere in the subject's participation in the trial.
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Dario Blood Glucose Monitoring SystemFingerstick to obtain blood sample for analysis with the Dario BGMS | Device | Dario Blood Glucose Monitoring System |
| YSI AnalyzerFingerstick to obtain blood sample for analysis with YSI | Device | YSI STAT |
Primary outcome
Dario Blood Glucose Monitoring System Accuracy at High Altitude (ISO 15197:2003)
Time frame: 6 seconds
Accuracy of Dario Blood Glucose Monitoring System when compared to Yellow Springs Instrument 2300 reference method at 10,152 feet above sea level. Acceptance criteria set per (International Organization for Standardization) ISO 15197:2003. Greater than or equal to 95% of the Dario results shall fall within plus or minus 15 mg/dL of the reference method value (for glucose concentrations <75 mg/dL). For glucose concentrations greater than or equal to 75 mg/dL, the Dario results shall fall within plus or minus 20% of the reference method value.
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 9 Jun 2017.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
- Individual participant dataSharing planned: No
Related trials for Diabetes
Not yet recruiting
Different Aerobic-to-Resistance Training Ratios in Adults Aged 50 Years and Older With Prediabetes
Prediabetes
Not yet recruiting
Efficacy, Safety, and Treatment Satisfaction of Once-Weekly Tirzepatide in Obese Type 2 Diabetic Patients
Type 2 Diabetes Mellitus (T2DM), Obesity, Overweight
Active, not recruiting
Helping Adults With Type 2 Diabetes Use Their Health Data for Healthy Diabetes Self-Management
Type 2 Diabetes Mellitus (T2DM)
Not yet recruiting
Care COMPASS Study
Diabetes Mellitus ( Type 1 and Type 2), Hypertension
Enrolling by invitationObservational study
Personalizing A1c Using Hematologic Indices and CGM for Accurate Diabetes Diagnosis
Type 2 Diabetes
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Christie Beatson study chair
Barbara Davis Center
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Dario Blood Glucose Monitoring System and YSI Analyzer in people with Diabetes, run by LabStyle Innovations Ltd. at 1 site in United States.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 2 inclusion and 2 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 9 Jun 2017. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 9 Jun 2017). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Dario Blood Glucose Monitoring System Accuracy at High Altitude (ISO 15197:2003) (time frame: 6 seconds). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 9 Jun 2017.