Thora-3Di™ Respiratory Rate Validation

A study testing Structured Light Plethysmography (Thora-3Di) and BCI Capnograph 9004 (Smiths Medical) in people with Pulmonary Disease, run by Pneumacare Ltd at 2 sites in United Kingdom.

Unknown status

Status not confirmed. The sponsor has not verified this record for over two years, so it may be out of date.

Official record
Search all trials

Trial summary

Respiratory rate (RR) is a vital sign used to monitor clinical condition of a patient. Various devices using different techniques are available to measure RR. One technique, capnography, uses continuous monitoring of expired gases, via a nasal cannula to assess end tidal CO2 concentration. Read more about this study

Who is it for?
Ages 18 to 80 yearsHealthy volunteers can take part
Conditions the study is for
Pulmonary Disease
What is being tested?
Structured Light Plethysmography (Thora-3Di) and BCI Capnograph 9004 (Smiths Medical)
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
43 participants
Where
2 sitesUnited Kingdom
Study dates
Feb 2015to May 2015
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 80 years
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Male or female,
  • Age range 18-80 years,
  • BMI range 18-40 kg/m2
  • Patients with respiratory disease or breathing impairment (Patient Group) or normal subjects with no previous or current diagnosis of respiratory disease (Normals group)

Who cannot join

  • Patient unable to sit in an upright position for required period of time
  • Patients with significant co morbidities (assessed by the clinician at screening only):
  • Significant unilateral lung disease e.g. pneumonectomy
  • Chest wall or spinal deformity e.g. scoliosis
  • Obstructive Sleep Apnoea, Apnoea hypopnoea index > 30 (if known)
  • BMI>40
  • Inability to consent/comply with trial protocol
  • Presence of an acute disease process that might interfere with test performance (e.g. Nausea, vomiting, persistent coughing)

Trial locations and contacts

2 sites in 1 country.

Rachel C Wilson, PhD contact

07907 071978

rachel.wilson@pneumacare.com

Iles, MD contact

United Kingdom2 sites, 2 recruiting
  • Cambridge University Hospitals Foundation Trust

    Cambridge, Cambridgeshire, CB2 0QQ

    Angelique Laubscher, 01223 811882, alaubscher@cambridge-clinical-research.co.uk

    Shuaib Nasser, MD

    Recruiting
  • Stowhealth

    Stowmarket, Suffolk, IP14 1NL

    Karen Norcott, 01223 811882, knorcott@cambridge-clinical-research.co.uk

    Neil Macey, MD

    Recruiting

Other trials in United Kingdom

About this study

Respiratory rate (RR) is a vital sign used to monitor clinical condition of a patient. Various devices using different techniques are available to measure RR. One technique, capnography, uses continuous monitoring of expired gases, via a nasal cannula to assess end tidal CO2 concentration. This long-established technique is frequently used for patients in intensive care. Both RR and the shape of the capnogram waveform, which has a wave pattern marked by alternating inspiratory and expiratory phases, are used to help monitor the patient. Clinician Over-scored End Tidal C02 (COSC) waveform (whereby an expert identifies and scores each breath on the waveform and counts the number of breaths per minute) is considered to be a "gold standard" for measuring RR.

Being able to record RR, without contact or interference with the patient is appealing because it requires minimal patient co-operation, enables measurements even during acute respiratory conditions, and may be more representative of "real life" physiology. Based on a principle originally described in the 1980s, a novel instrument that uses a completely non-contact system based on structured light plethysmography (SLP) has recently been designed by Cambridge University Hospitals Foundation Trust. The device has been refined with the development of the Thora-3Di by Pneumacare Ltd, Cambridge. A grid of visible light is projected onto the thoraco-abdominal wall and two digital video cameras record changes in the grid pattern due to breathing motion. A waveform is produced by the anterior excursion of the thoraco-abdominal wall (SLPVol) over time, and a numerical output of RR is provided . The present study aims to simultaneously measure tidal breathing with the Thora-3Di and a BCI Capnograph 9004 device to compare the Thora-3Di RR output against that of the gold standard clinician over-scored end-tidal C02 (COSC) and to assess equivalence of the two devices with the aim to establishing the validity of SLP for RR measurement.

Read the detailed description

Objectives:

The primary objective is to demonstrate equivalence of the Thora-3Di™ SLP device and a Clinician Over-scored End Tidal C02 (COSC) derived from a Capnograph (BCI Capnograph 9004), Smiths Medical) for measurement of Respiratory Rate.

Endpoint:

The primary endpoint is the difference in the Respiratory Rate between the two devices.

Test Article:

Thora-3Di™, non-contact; measures Respiratory Rate by SLP from the SLPVol/time trace.

Comparator:

Clinician Over-scored End Tidal C02 derived from a Capnograph (BCI Capnograph 9004, Smiths Medical); measures RR from the end tidal CO2 (EtCO2) waveform.

Patient Enrolment and Procedures:

A maximum of 43 subjects will be enrolled across 2 sites to achieve 34 evaluable data sets of:

  • 24 Patients with Respiratory disease/impairment (with a minimum of 7 and maximum of 17 to be enrolled at each site)
  • 10 Normal subjects with no current or previous diagnosis of respiratory disease (minimum of 3 and a maximum of 7 to be enrolled at each site)

Sample Size:

A sample size of 34 subjects has been determined to yield at least 80% power when the MSD is 0.4. A previous study found that the MSD between the test device and the Embletta device was 0.31, thus we conservatively use the value of 0.4 to calculate power. In that study, the mean difference was -0.26 and the standard deviation of the differences was 0.49. We assume that 80% of the enrolled subjects will yield evaluable data. Specifically, an enrolled sample size of 43 subjects will yield 43*.80 = 34 evaluable subjects. This sample size yields at least 80% power when the true MSD is 0.4 or smaller.

The study will consist of one stage, which will consist of:

  • Each participant will be measured simultaneously with both devices on 1 occasion for a duration of 5-9 minutes.
  • The participant will be measured in the seated position.

Measurement Devices:

End Tidal C02 will be measured using a Capnograph (BCI Capnograph 9004, Smiths Medical). Expired gas will be collected via a nasal cannula with sampling line and moisture trap. Expired gas will be sampled at a flow rate of 150 mL/min ± 20 ml. The device has an average delay time of 1.8 seconds and a response time of 2.1 seconds which will be taken into account in the analysis. The BCI Capnograph 9004 has a display range of 0 - 100 mmHg and is accurate to ±2 mmHg, further information can be found in the device instruction manual. Capnograph waveforms will be recorded into a computer by utilising the analogue output of the capnograph. The analogue output will be sampled at 30 Hz using a 16 bit analogue to digital converter (Micro 1401-3, Cambridge Electronic Design, UK). The sampling rate of 30 Hz is selected since it is equal to the display update rate of the waveform on the device, and it is therefore closest to what clinicians commonly assess. Clinician over-scored capnographic RR will be derived from the capnogram CO2 waveform. The Capnography device will be calibrated according to manufacturer instructions. The range of Respiration rate of the Capnography devices are 0 -150 brpm.

Thoraco-abdominal movement will be measured using a Thora-3Di. The grid size used will be 14 x 10. Five minutes of continuous resting tidal breathing data will be acquired for each subject. The signals are sampled at 30 Hz and represent the anterior thoraco-abdominal breathing motion. The Thora-3Di software provides a numerical output of RR as mean brpm, calculated from the SLPVol/Time trace over a selected time frame.

The five minutes of continuous resting tidal breathing data will be collected simultaneously from both devices.

Data Transfer:

Capnographic outputs will be acquired and saved as data files for subsequent processing and review. The Thora-3Di™ software thoraco-abdominal signals will be exported to comma-separated value (CSV) files.

Before data analysis, all measurements will be assessed on their quality, and any protocol violators identified. The Thora-3Di™ device software detects the noise level of the output signal, and where the software flags a measurement as excessively noisy a measurement will not be obtained. For example, if a patient moves their upper body, this may prohibit a valid Thora-3Di™ measurement from being recorded. Likewise, interference with, or condensation within, the Capnography sampling tubing may produce artefacts or prevent a valid Capnography trace being recorded. Where paired samples cannot be obtained due to protocol violations, this will render the data unanalyzable so these subjects will be excluded from the analysis. Details of protocol violators but will be tabulated with reasons for exclusion listed. Any specific sub-population or disease state where the Thora-3Di™ did not work in as expected will be noted. Details of analyzable datasets (i.e. excluding protocol violators) will be tabulated

Data Analysis:

The primary endpoint analysis is to compare the output respiratory rate of Thora-3Di with that of a COSC measured over one randomly selected epoch of one minute for each subject. The 5 minute measurement period will span the range of intended use of the Thora-3Di™. A duration of 1 minute has been chosen as an appropriate comparison period for the analysis, as the number of breaths in one minute is the duration over which Respiratory Rate is clinically defined. Where the SLP tracing is orange or there is a significant change in baseline or a spike caused by body movement or such that would otherwise be excluded by labelling, those segments of the recording will not be used in the analysis. This is in accordance with intended use of the device.

To identify the randomly selected epoch for each subject, a uniform random number generator will be used to generate a random starting sample from which point a 1 minute epoch of data will be truncated for further analysis. This will be done using the rand function in MATLAB. The code will be:

starting sample = round((rand*4)*60*Fs), where rand is the uniform random number generator, it is multiplied by 4 to increase the default range of [0,1] to [0,4]. rand*4 essentially produces the time (in minutes) from which 1 minute of data will be truncated. The rest of the code translate the starting minute to a sample number, that is, multiplied by 60 to get the time in seconds, multiplied by Fs (sampling rate of the device) to get the sample and finally, since samples are always integers, round the number to get a the closest starting sample number.

This random number generation process will be repeated in order to allocate a random starting point to each of the paired analyzable datasets. To derive the SLP Respiratory Rate, the section corresponding to a 1 minute epoch for each sample (as identified on the 'Appraised Paired Sample' table) will be selected on SLPVol/Time trace of the Thora-3Di. The RR value for this period, which is a direct output from the software, will be recorded for this time section.

To derive the COSC RR, an expert clinician will be provided with the capnograph traces only and will verify each breath in the indicated 1 minute epoch of the capnograph output and calculate the RR for the time section for each subject.

To ensure no bias in the scoring, the clinician scoring the capnograph output, will have no reference to the Thora-3Di™ trace or SLP Respiratory Rate and likewise, the individual deriving SLP Respiratory Rate will have no reference to the capnograph output or the derived COSC Respiratory Rate.

To ensure no bias is introduced from the data collection stage to the data processing stage, the operator collecting the data have no involvement in the scoring of the outputs.

Hypotheses:

The following hypotheses will be tested with a significance level of =0.05:

  • Null hypothesis: The true root mean squared difference (RMSD) between test and reference device RR is greater than or equal to 0.9 brpm.
  • Alternative hypothesis: The true RMSD is less than 0.9 brpm.

The hypotheses above are equivalent to the following:

Null hypothesis: The true mean squared difference (MSD) between test and reference device RR is greater than or equal to 0.81 (breaths per minute)2.

Alternative hypothesis: The true MSD is less than 0.81 (breaths per minute)2. A RMSD< 0.90 brpm implies that the 95% limits of agreement are within 2 brpm (see Section 14.2). Pre-defined limits of agreement of 2 brpm were suggested by Smith et al. 2011 as clinically insignificant.

Analysis Populations:

The study population will comprise normal subjects, and subjects with other respiratory diseases. The full analysis population will comprise all patients who were entered into the study, excluding protocol violators, and provided one pair of measurements from the Thora-3Di device and the COSC. This population will be used as part of the primary analysis.

Poolability of Data Across Sites:

To assess the poolability of data across sites, the mean absolute differences between RR measured by Thora 3Di and the Capnograph will compared with a two-sample t-test. If the p-value of the test is greater than 0.10, poolability of data will be concluded.

Primary Endpoint Analysis and Acceptance Criteria:

Respiratory rate counts from the Thora 3Di and the COSC will be compared in a pair-wise manner. To test the hypothesis that the MSD is less than 0.81 brpm2, a Wald test will be done. The following test statistic will be calculated and compared to the threshold value of -1.645 from a standard normal distribution:

T=(MSD-0.81)/(υ(τ=0.81)) where υ(τ=0.81) is the estimated standard error of the MSD assuming the mean MSD is 0.81, as defined in Section 14.2. If T<-1.645, the null hypothesis will be rejected and we will conclude that the test device has sufficient accuracy.

Additionally, Bland-Altman plots will be produced and the 95% limits of agreement will be calculated. Specifically, the differences between test and reference device RR measurements will be plotted against the mean of the two RR measurements. Reference lines representing the 95% limits of agreement, defined by the mean difference ±1.96x standard deviation of the difference, will also be included on the plot. A scatter plot with the linear regression line and its slope and intercept will also be given.

The acceptance criteria for primary endpoint is +/- 2 brpm. This is equivalent to a MSD<0.81 brpm2.

Overall Analysis All subjects entered into the study will be accounted for by listing the numbers of patients entered, the numbers in each population, withdrawals and the reasons for withdrawal. Summary statistics for adverse incidents, demographics, clinical characteristics, Respiration Rates from the Thora 3Di device and the COSC will be produced. In general, categorical data will be summarized using frequency counts and percentages, and continuous data will be summarized using means, standard deviations, minimums, medians and maximums. These summaries will be inspected for any concerning patterns and further analysis may be performed as appropriate.

Being tested
Structured Light Plethysmography (Thora-3Di), BCI Capnograph 9004 (Smiths Medical)
Keywords
Respiratory Rate, Structured Light Plethysmography, Capnography
Medical subject headings
Lung Diseases

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Structured Light Plethysmography (Thora-3Di)Device

    Also called Thora-3Di Structured Light Plethysmography

    Respiratory Rate measured by Structured Light Plethysmography Thora-3Di and Clinician Over-scored End Tidal C02 (COSC) during 1 randomised minute of a 5 minute measurement period

  • BCI Capnograph 9004 (Smiths Medical)Device

    Also called Capnocheck (Smiths Medical)

    Respiratory Rate measured by Structured Light Plethysmography Thora-3Di and Clinician Over-scored End Tidal C02 (COSC) during 1 randomised minute of a 5 minute measurement period

Study groups in NCT02374554
Study groupRoleWhat they receive
Respiratory Rate Measurement ComparisionOtherTo demonstrate equivalence of the Thora-3Di Structured Light Plethysmography device and a Clinician Over-scored End Tidal C02 (COSC)derived from a BCI Capnograph 9004 (Smiths Medical) for measurement of Respiratory Rate
Receives: Device: Structured Light Plethysmography (Thora-3Di); Device: BCI Capnograph 9004 (Smiths Medical)

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. The root mean squared difference (RMSD) in Respiratory Rate between the test (Thora-3Di) and reference Device (COSC)

    The test device will be accepted as sufficiently accurate if the RMSD is significantly less than 0.9 brpm (significance level 0.05, power 0.8)

    Measured: 1 minute (randomly selected) of a 5 minute measurement period

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted2 Mar 2015 (estimated)
  2. Study startFeb 2015
  3. Primary completionMay 2015 (estimated)
  4. Study completionMay 2015 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Supportive care
Allocation
Not applicable
Intervention model
Single group
Masking
Single blind
Who is masked
Outcomes assessor
Enrolment
43 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT02374554
Sponsor's protocol ID
PMC-T3D-001
Status last verified
Mar 2015
First submitted
23 Feb 2015
Last updated
6 Mar 2015
Results posted
No
Data monitoring committee
No

Publications

  • Bland JM. Sample size in guidelines trials. Fam Pract. 2000 Feb;17 Suppl 1:S17-20. doi: 10.1093/fampra/17.suppl_1.s17.

    PubMed 10735263
  • Cretikos MA, Bellomo R, Hillman K, Chen J, Finfer S, Flabouris A. Respiratory rate: the neglected vital sign. Med J Aust. 2008 Jun 2;188(11):657-9. doi: 10.5694/j.1326-5377.2008.tb01825.x.

    PubMed 18513176
  • Gaucher A, Frasca D, Mimoz O, Debaene B. Accuracy of respiratory rate monitoring by capnometry using the Capnomask(R) in extubated patients receiving supplemental oxygen after surgery. Br J Anaesth. 2012 Feb;108(2):316-20. doi: 10.1093/bja/aer383. Epub 2011 Dec 11.

    PubMed 22157953
  • Jaffe MB. Infrared measurement of carbon dioxide in the human breath: "breathe-through" devices from Tyndall to the present day. Anesth Analg. 2008 Sep;107(3):890-904. doi: 10.1213/ane.0b013e31817ee3b3.

    PubMed 18713902
  • Kodali BS. Capnography outside the operating rooms. Anesthesiology. 2013 Jan;118(1):192-201. doi: 10.1097/ALN.0b013e318278c8b6. No abstract available.

    PubMed 23221867
  • Morgan MD, Gourlay AR, Denison DM. An optical method of studying the shape and movement of the chest wall in recumbent patients. Thorax. 1984 Feb;39(2):101-6. doi: 10.1136/thx.39.2.101.

    PubMed 6701820
  • Storm-Versloot MN, Verweij L, Lucas C, Ludikhuize J, Goslings JC, Legemate DA, Vermeulen H. Clinical relevance of routinely measured vital signs in hospitalized patients: a systematic review. J Nurs Scholarsh. 2014 Jan;46(1):39-49. doi: 10.1111/jnu.12048. Epub 2013 Oct 11.

    PubMed 24124656
  • Yanagidate F, Dohi S. Modified nasal cannula for simultaneous oxygen delivery and end-tidal CO2 monitoring during spontaneous breathing. Eur J Anaesthesiol. 2006 Mar;23(3):257-60. doi: 10.1017/S0265021505002279.

    PubMed 16430798
  • Box, Hunter and Hunter (1978). Statistics for experimenters. Wiley. p118. ISBN 0471093157

  • de Boer WH, Lasenby J, Cameron J, Wareham R, Ahmad S, Roach C, Hills W, Iles R. SLP: a zero-contact non-invasive method for pulmonary function testing. In: Labrosse F, Zwiggelaar R, Liu Y, Tiddeman B, eds, Proceedings of the British Machine Vision Conference. BMVA Press, 2010; pp 85.1-85.12

  • Johnson NL, Kotz S, Balakrishnan N. 1970. Continuous Univariate Distributions, Volume 2, Wiley.

  • Smith I, Mackay J, Fahrid N, Krucheck D. Respiratory rate measurement: a comparison of methods. British Journal of Healthcare Assistants 2011; 5: 18-23

Questions about this trial

What is this clinical trial studying?

A study testing Structured Light Plethysmography (Thora-3Di) and BCI Capnograph 9004 (Smiths Medical) in people with Pulmonary Disease, run by Pneumacare Ltd at 2 sites in United Kingdom.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: status not confirmed, as of the sponsor's last update on 6 Mar 2015. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 2 sites in United Kingdom. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Results are not available in the source record.

How can a doctor refer a patient or contact the research team?

Contact Rachel C Wilson, PhD (07907 071978, rachel.wilson@pneumacare.com), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

The root mean squared difference (RMSD) in Respiratory Rate between the test (Thora-3Di) and reference Device (COSC) (time frame: 1 minute (randomly selected) of a 5 minute measurement period). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT02374554, which also lists 12 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 6 Mar 2015; AzTrials last refreshed its copy on 27 Oct 2026.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in