Clinical Evaluation of Visco-Assisted CyPass® Micro-Stent Implantation in Patients With Open Angle Glaucoma

A study testing CyPass Micro-Stent and Viscoelastic in people with Open Angle Glaucoma, run by Transcend Medical, Inc. at 6 sites in 4 countries.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to assess the safety and effectiveness of the use of visco-assisted CyPass® Micro-Stent implantation for the lowering of intraocular pressure (IOP) in subjects who have open angle glaucoma (OAG). Read more about this study

Who is it for?
Ages 21 years and older
Conditions the study is for
Open Angle Glaucoma
What is being tested?
CyPass Micro-Stent and Viscoelastic
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
192 participants
Where
6 sitesin 4 countries
Study dates
21 Jun 2013to 22 Jun 2017
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 21 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Diagnosis of open angle glaucoma;
  • Unmedicated IOP between 21 - 36 mmHg, inclusive;
  • Normal angle anatomy at site of intended CyPass Micro-Stent implantation.

Who cannot join

  • Advanced glaucoma;
  • Prior incisional glaucoma surgery;
  • Acute angle closure, traumatic, congenital, malignant, uveitic or neovascular glaucoma;
  • Clinically significant ocular pathology other than glaucoma.

Trial locations and contacts

6 sites in 4 countries.

6 sites in 4 countries

Germany3 sites
  • Transcend Medical Investigative Site

    Lübeck

  • Transcend Medical Investigative Site

    Mainz

  • Transcend Medical Investigative Site

    Neubrandenburg

Other trials in Germany

Panama1 site
  • Transcend Medical Investigative Site

    Panama City

Other trials in Panama

Poland1 site
  • Transcend Medical Investigative Site

    Warsaw

Other trials in Poland

Spain1 site
  • Transcend Medical Investigative Site

    Madrid

Other trials in Spain

About this study

Official title: Randomized, Prospective Clinical Evaluation of the Safety and Effectiveness of Visco-Assisted CyPass® Implantation in Patients With Open Angle Glaucoma

The purpose of this study is to assess the safety and effectiveness of the use of visco-assisted CyPass® Micro-Stent implantation for the lowering of intraocular pressure (IOP) in subjects who have open angle glaucoma (OAG).

Read the detailed description

The study was conducted in 2 phases. Only one eye per subject was treated for each independent phase.

Dose Selection Phase (Cohort 1): Subjects were randomized in a 1:1:1 ratio and implanted with either a CyPass Micro-Stent with targeted delivery of 30 microliters (μl) ophthalmic 5 viscoelastic, a CyPass Micro-Stent with targeted delivery of 60 μl of ophthalmic viscoelastic, or a CyPass Micro-Stent without adjunct viscoelastic in the study eye.

Expansion Phase (Cohort 2): Subjects were randomized in a 1:1 ratio and implanted with either the CyPass Micro-Stent without adjunct viscoelastic or the CyPass Micro-Stent with 60 μl of ophthalmic viscoelastic (based on Dose Selection results). Subjects were randomized to treatment on the day of their surgical procedure. A total of 9 scheduled visits were planned including Screening (Day -45 to -2), Baseline (Day -15 to -1), Surgery (Day 0), 1 Day (Day 1), 1 Week (Day 5-9), 1 Month (Day 21-35), 3 Month (Day 70-98), 6 Month (Day 150-210), and 12 Month (Day 330-420) visits. The total expected duration of participation for each subject was up to 13 months.

Being tested
CyPass Micro-Stent, Viscoelastic
Medical subject headings
Glaucoma, Open-Angle

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • CyPass Micro-StentDevice

    Also called Model 2FX

    Small tube with a through-lumen designed to redirect aqueous fluid from the front to the back of the eye in order to reduce intraocular pressure. Designed to be permanently implanted in the eye.

  • ViscoelasticDevice

    Healon 5 ophthalmic viscoelastic used to increase the size of the aqueous drainage area created by the CyPass Micro-Stent

Study groups in NCT02448875
Study groupRoleWhat they receive
CyPassActive comparatorCyPass Micro-Stent without adjunct viscoelastic implanted in the study eye
Receives: Device: CyPass Micro-Stent
CyPass30ExperimentalCyPass Micro-Stent implantation followed by targeted delivery of 30 μl ophthalmic viscoelastic
Receives: Device: CyPass Micro-Stent; Device: Viscoelastic
CyPass60ExperimentalCyPass Micro-Stent implantation followed by targeted delivery of 60 μl ophthalmic viscoelastic
Receives: Device: CyPass Micro-Stent; Device: Viscoelastic

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Percentage of Subjects With ≥ 20% Decrease From Baseline to 12 Months Postoperative in IOP Without Use of Ocular Hypotensive Medication

    IOP (fluid pressure inside the eye) was assessed using Goldmann applanation tonometry and reported in millimeters of mercury (mmHg). A higher IOP can be a greater risk factor for developing glaucoma or glaucoma progression (leading to optic nerve damage). One eye (study eye) contributed to the analysis. No formal statistical hypothesis testing was planned for the study.

    Measured: Baseline (Day -1), Month 12 PostOperative

Secondary outcomes (4)
  1. Percentage of Subjects With Device-related Ocular Adverse Events

    A device related adverse event (AE) was any AE that was considered to be possibly, probably, or definitely related to the device in the opinion of the investigator. Reported categorically as intraoperative (start date on the date of surgery) and postoperative (start date after surgery). One eye (study eye) contributed to the analysis. No formal statistical hypothesis testing was planned for the study.

    Measured: Up to Month 12 PostOperative

  2. Mean Change From Baseline to 12 Months Postoperative in Medicated IOP

    IOP (fluid pressure inside the eye) was measured by Goldmann applanation tonometry. A higher IOP can be a greater risk for developing glaucoma or glaucoma progression (leading to optic nerve damage). A higher negative change indicates improvement. One eye (study eye) contributed to the analysis. No formal statistical hypothesis testing was planned for the study.

    Measured: Baseline (Day -1), Month 12 PostOperative

  3. Percentage of Eyes Using Ocular Hypotensive Medication at 12 Months

    The use of ocular hypotensive medications was assessed in subjects with at least one IOP-lowering medication at 12 Months. One eye (study eye) contributed to the analysis. No formal statistical hypothesis testing was planned for the study.

    Measured: Month 12 PostOperative

  4. Mean Change From Screening to 12 Months Postoperative in Number of Topical IOP-lowering Medications Used

    The number of IOP lowering medications in subjects at 12 months was compared to medicated baseline. A higher negative change indicates improvement. One eye (study eye) contributed to the analysis. No formal statistical hypothesis testing was planned for the study.

    Measured: Screening (Day -2), Month 12 PostOperative

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted20 May 2015 (estimated)
  2. Study start21 Jun 2013
  3. Primary completion22 Jun 2017
  4. Study completion22 Jun 2017

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Parallel groups
Masking
None (open label)
Enrolment
192 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT02448875
Sponsor's protocol ID
TMI-13-01
Other IDs
GLD122-P001
Acronym
ViscoPass
Status last verified
May 2019
First submitted
12 May 2015
Last updated
12 Jun 2019
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing CyPass Micro-Stent and Viscoelastic in people with Open Angle Glaucoma, run by Transcend Medical, Inc. at 6 sites in 4 countries.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 4 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 12 Jun 2019. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 6 sites in Germany, Panama, Poland, Spain. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 30 May 2019). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Percentage of Subjects With ≥ 20% Decrease From Baseline to 12 Months Postoperative in IOP Without Use of Ocular Hypotensive Medication (time frame: Baseline (Day -1), Month 12 PostOperative). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 12 Jun 2019.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in