DiabetesNCT02531386
System Accuracy Evaluation of 6 CE-marked Blood Glucose Monitoring System Following ISO 15197:2013
A study testing 6 blood glucose monitoring systems for self-testing (blood glucose meters) in people with Diabetes, run by Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm at 1 site in Germany.
Trial summary
The objective of this study is the evaluation of system accuracy following ISO 15197:2013 (E), clause 6.3 in which accuracy requirements and applicable test procedures for blood glucose monitoring systems intended for self-monitoring of blood glucose by patients are stipulated. Read more about this study
- Who is it for?
- Ages 18 years and olderHealthy volunteers can take part
- Conditions the study is for
- Diabetes
- What is being tested?
- 6 blood glucose monitoring systems for self-testing (blood glucose meters)
- Type of treatment
- Medical device
- How many people
- 120 participants
- Where
- 1 siteGermany
- Study dates
- Aug 2015to Nov 2015
- Phase
- Interventional
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- Healthy volunteers accepted
- Adult, Older adult
Who may be able to join
- Male or female, with type 1 diabetes, type 2 diabetes or no diabetes
- For provoked blood glucose excursions due to insulin dose adjustment:
- Male or female with type 1 diabetes or
- Type 2 diabetes and intensified insulin therapy or insulin pump therapy.
- Minimum age of 18 years
- Signed informed consent form
- Legally competent and capable to understand character, meaning and consequences of the study
Who cannot join
- Pregnancy or lactation period
- Severe acute disease (at the study physician's discretion)
- Severe chronic disease with potential risk during the test procedures (at the study physician's discretion)
- Physical or mental constitution that compromises the subject's capability to participate in the study (at the study physician's discretion)
- Incapability of giving informed consent
- < 18 years
Show all 16Show fewer
- Legally incompetent
- Accommodated in an institution (e.g. psychiatric clinic)
- Language barriers potentially compromising an adequate compliance with study procedures
- Dependent from investigator or sponsor
- For provoked blood glucose excursions 50 - 80 mg/dl (concentration category 2):
- Coronary heart disease
- Condition after myocardial infarction
- Cerebral incidence
- Peripheral arterial occlusive disease
- Hypoglycemia unawareness
Trial locations and contacts
1 site in 1 country.
Germany1 site
Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm
Ulm, 89081
About this study
Official title: System Accuracy Evaluation of Accu-Chek Aviva Nano (Roche Diagnostics), Accu-Chek Performa Nano (Roche Diagnostics), FreeStyle Lite (Abbott Diabetes Care), OneTouch Verio IQ (LifeScan), Contour Next Link 2.4 (Bayer Consumer Care), One Additional CE-marked Blood Glucose Monitoring System Following ISO 15197:2013
The objective of this study is the evaluation of system accuracy following ISO 15197:2013 (E), clause 6.3 in which accuracy requirements and applicable test procedures for blood glucose monitoring systems intended for self-monitoring of blood glucose by patients are stipulated. In this study, system accuracy evaluation will be performed with CE-marked BGMS (BGMS = Blood Glucose Monitoring System) common among insulin pump users - Accu-Chek Aviva Nano and Accu-Chek Performa Nano (Roche Diagnostics), FreeStyle Lite (Abbott Diabetes Care), OneTouch Verio IQ (LifeScan) Contour Next Link 2.4 (Bayer Healthcare Diabetes Care) and Accu-Chek Mobile with three different reagent system lots for each BGMS.
The study will be performed in 2 parts. Each part will comprise testing of 3 BGMS.
- Conditions
- Diabetes
- Being tested
- 6 blood glucose monitoring systems for self-testing (blood glucose meters)
- Medical subject headings
- Diabetes Mellitus
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
6 blood glucose monitoring systems for self-testing (blood glucose meters)Device
Also called Accu-Chek Aviva Nano, Accu-Chek Performa Nano, FreeStyle Lite, OneTouch Verio IQ, Contour Next Link 2.4, Accu-Chek Mobile
What is being measured
The primary outcome is the main question the study is designed to answer.
System accuracy criteria (see description)
System accuracy criteria
Following ISO 15197:2013 (E), the BGMS shall meet both of the following criteria:
Criterion A: 95 % of the measured glucose values shall fall within either ± 15 mg/dl (0.83 mmol/l) of the average comparison measurement result at glucose concentrations < 100 mg/dl (5.55 mmol/l) or within ± 15 % at glucose concentrations ≥ 100 mg/dl (5.55 mmol/l).
Each lot shall pass acceptability criterion A.
Criterion B: 99 % of individual glucose measured values shall fall within zones A and B of the Consensus Error Grid (CEG) for type 1 diabetes.
Applying ISO 15197:2003 accuracy criteria:
95 % of the measured glucose values shall fall within either ± 15 mg/dl (0.83 mmol/l) of the average comparison measurement result at glucose concentrations < 75 mg/dl (4.2 mmol/l) or within ± 20 % at glucose concentrations ≥ 75 mg/dl (4.2 mmol/l).
Measured: For each subject, the experimental phase has an expected duration of up to 6 hours
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted24 Aug 2015 (estimated)
- Study startAug 2015
- Primary completionSep 2015
- Study completionNov 2015
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Other trials for DiabetesStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Enrolment
- 120 participants (estimated)
Sponsor and investigators
- Lead sponsor
- Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm
- Sponsor type
- University, hospital or other
- Collaborators
- Bayer
- Responsible party
- Sponsor
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT02531386
- Sponsor's protocol ID
- IDT-1525-BL
- Status last verified
- Sep 2015
- First submitted
- 7 Aug 2015
- Last updated
- 9 Nov 2015
- Results posted
- No
Study at a glance
From the registry record, last updated by the sponsor on 9 Nov 2015. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- System Accuracy Evaluation of Accu-Chek Aviva Nano (Roche Diagnostics), Accu-Chek Performa Nano (Roche Diagnostics), FreeStyle Lite (Abbott Diabetes Care), OneTouch Verio IQ (LifeScan), Contour Next Link 2.4 (Bayer Consumer Care), One Additional CE-marked Blood Glucose Monitoring System Following ISO 15197:2013
- ClinicalTrials.gov ID
- NCT02531386
- Acronym / sponsor ID
- IDT-1525-BL
- Conditions
- Diabetes
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 120 participants (estimated)
- Lead sponsor
- Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität UlmUniversity, hospital or other
- Collaborators
- Bayer
- Study start
- Aug 2015
- Primary completion
- Sep 2015 (actual)
- Study completion
- Nov 2015 (actual)
- First posted
- 24 Aug 2015 (estimated)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The objective of this study is the evaluation of system accuracy following ISO 15197:2013 (E), clause 6.3 in which accuracy requirements and applicable test procedures for blood glucose monitoring systems intended for self-monitoring of blood glucose by patients are stipulated. In this study, system accuracy evaluation will be performed with CE-marked BGMS (BGMS = Blood Glucose Monitoring System) common among insulin pump users - Accu-Chek Aviva Nano and Accu-Chek Performa Nano (Roche Diagnostics), FreeStyle Lite (Abbott Diabetes Care), OneTouch Verio IQ (LifeScan) Contour Next Link 2.4 (Bayer Healthcare Diabetes Care) and Accu-Chek Mobile with three different reagent system lots for each BGMS.
The study will be performed in 2 parts. Each part will comprise testing of 3 BGMS.
Study design
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (7 listed)
- Male or female, with type 1 diabetes, type 2 diabetes or no diabetes
- For provoked blood glucose excursions due to insulin dose adjustment:
- Minimum age of 18 years
- Signed informed consent form
- Legally competent and capable to understand character, meaning and consequences of the study
Key exclusion criteria (16 listed)
- Pregnancy or lactation period
- Severe acute disease (at the study physician's discretion)
- Severe chronic disease with potential risk during the test procedures (at the study physician's discretion)
- Physical or mental constitution that compromises the subject's capability to participate in the study (at the study physician's discretion)
- Incapability of giving informed consent
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| 6 blood glucose monitoring systems for self-testing (blood glucose meters)Also: Accu-Chek Aviva Nano, Accu-Chek Performa Nano, FreeStyle Lite, OneTouch Verio IQ, Contour Next Link 2.4, Accu-Chek Mobile | Device | – |
Primary outcome
System accuracy criteria (see description)
Time frame: For each subject, the experimental phase has an expected duration of up to 6 hours
System accuracy criteria Following ISO 15197:2013 (E), the BGMS shall meet both of the following criteria: Criterion A: 95 % of the measured glucose values shall fall within either ± 15 mg/dl (0.83 mmol/l) of the average comparison measurement result at glucose concentrations < 100 mg/dl (5.55 mmol/l) or within ± 15 % at glucose concentrations ≥ 100 mg/dl (5.55 mmol/l). Each lot shall pass acceptability criterion A. Criterion B: 99 % of individual glucose measured values shall fall within zones A and B of the Consensus Error Grid (CEG) for type 1 diabetes. Applying ISO 15197:2003 accuracy criteria: 95 % of the measured glucose values shall fall within either ± 15 mg/dl (0.83 mmol/l) of the average comparison measurement result at glucose concentrations < 75 mg/dl (4.2 mmol/l) or within ± 20 % at glucose concentrations ≥ 75 mg/dl (4.2 mmol/l).
Results and safety
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 9 Nov 2015.
Related trials for Diabetes
Not yet recruitingPhase 3
Efficacy and Safety of Once-Weekly SHR-3167 Compared With Once-Daily Insulin Glargine in Adults With Insulin-Naive Type 2 Diabetes
Type 2 Diabetes
Active, not recruiting
Digital Reinforcement of Motivational Interviewing Micro-Skills Among Primary Care Healthcare Professionals
Hypertension (Without Type 2 Diabetes Mellitus)
Recruiting
Effects of Lean Pork Loin Intake on Protein Homeostasis and Glucose Regulation in Prediabetic Adults
Hyperglycaemia (Non Diabetic), Prediabetes
Recruiting
PAW-Hydrogel for Chronic Wound Healing
Diabetic Foot Ulcer (DFU), Pressure Ulcer (PU)
Recruiting
Investigating Vascular Properties of HEMI and SPG Signals in Individuals With or at Risk for Chronic Kidney Disease
Chronic Kidney Disease, Hypertension (HTN)
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
The record lists no contact. Use the official record to reach the sponsor.
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing 6 blood glucose monitoring systems for self-testing (blood glucose meters) in people with Diabetes, run by Institut für Diabetes-Technologie Forschungs- und Entwicklungsgesellschaft mbH an der Universität Ulm at 1 site in Germany.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 7 inclusion and 16 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 9 Nov 2015. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in Germany. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
System accuracy criteria (see description) (time frame: For each subject, the experimental phase has an expected duration of up to 6 hours). It is what the study was designed to measure, not a finding.