Diabetes Mellitus, Type 1NCT02614768
Standardized Evaluation of Subcutaneous Glucose Monitoring Systems Under Routine Environmental Conditions
A study testing Glucose Sensor in people with Diabetes Mellitus, Type 1, run by Medical University of Graz.
Trial summary
The purpose of this study is to assess accuracy and reliability of the SPIDIMAN continuous glucose monitoring (CGM) sensor system with regard to values as measured by Super GL and compare these results with similar evaluations of the Medtronic MiniMed 640G system, the Abbott FreeStyle Libre Flash Sensor and the DexCom… Read more about this study
- Who is it for?
- Ages 18 years and older
- Conditions the study is for
- Diabetes Mellitus, Type 1
- What is being tested?
- Glucose Sensor
- Type of treatment
- Injection or infusion, Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 12 participants
- Where
- No sites listed yet
- Study dates
- Jul 2015to Sep 2015
- Phase
- Interventional
- Are there results?
- Yes, posted in the registrySummary in the doctors' view
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 years and older
- All sexes
- No healthy volunteers
- Adult, Older adult
Who may be able to join
- Informed consent obtained after being advised of the nature of the study
- Male or female aged ≥18 years
- Type 1 diabetes for at least 6 months according to the World Health Organization (WHO) definition
- Treatment with multiple daily injections (MDI) or continuous subcutaneous insulin infusion (CSII) for at least 3 months
- Body Mass Index (BMI) < 35 kg/m²
- Willing and able to wear 5 CGM devices for the duration of the study and undergo all study procedures.
- HbA1c ≤ 86 mmol/mol
Who cannot join
- Any disease or condition which the investigator or treating physician feels would interfere with the trial or the safety of the subject
- Female of childbearing potential who is pregnant, breast-feeding or intend to become pregnant or is not using adequate contraceptive methods
- Any mental condition rendering the subject incapable of giving his consent
- Subject is using a medication that significantly impacts glucose metabolism (oral steroids) except if stable for at least the last three months and expected to remain stable for the study duration
- Subject may not use acetaminophen (paracetamol) while participating in the study
- Has severe medical or psychological condition(s) or chronic conditions/infections that in the opinion of the Investigator would compromise the subject's safety or successful participation in the study.
Show all 11Show fewer
- Subject is actively enrolled in another clinical trial
- Known adrenal gland problem, pancreatic tumour, or insulinoma
- Inability of the subject to comply with all study procedures
- Inability of the subject to understand the patient information.
- Subject donated blood in the last 3 months
About this study
The purpose of this study is to assess accuracy and reliability of the SPIDIMAN continuous glucose monitoring (CGM) sensor system with regard to values as measured by Super GL and compare these results with similar evaluations of the Medtronic MiniMed 640G system, the Abbott FreeStyle Libre Flash Sensor and the DexCom G4 Platinum Sensor in patients with type 1 Diabetes.
Read the detailed description
The study is a single-center open-label study in patients with type 1 diabetes treated with continuous subcutaneous insulin infusion (CSII) or multiple daily injections (MDI) therapy. The study will include a total of 12 subjects.
- Conditions
- Diabetes Mellitus, Type 1
- Being tested
- Glucose Sensor
- Keywords
- Glucose Monitoring System
- Medical subject headings
- Diabetes Mellitus, Type 1
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Glucose SensorDevice
The investigational intervention is CGM monitoring using four different CGM systems. Three of the four CGM devices which are used will be CGM systems which have received conformité européenne (CE) certification (Dexcom G4 Platinum, Medtronic MiniMed 640G system, Abbott FreeStyle Libre Flash). The SPIDIMAN sensor will be used for the first time in human subjects, does not yet have received CE certification and will be given an identifying label in addition to being labelled "for investigative use only".
| Study group | Role | What they receive |
|---|---|---|
| Glucose Sensor | Experimental | Continuous subcutaneous glucose monitoring using four different CGM systems in parallel will be performed throughout the study. Insulin therapy will be performed by the subjects themselves, as under daily life conditions. For the hypoglycaemia experiment an increased insulin bolus will be administered with meals (180% of the subject's calculated mealtime dose). Receives: Device: Glucose Sensor |
What is being measured
The primary outcome is the main question the study is designed to answer.
Percentage of Test-to-Reference Measurement Pairs With an Absolute Relative Difference (ARD) ≤15%
The primary outcome measure was overall sensor accuracy determined using ISO 15197:2013, which is percentage (%) of sensor values that are within ±0.8 mmol/L of the reference value at glucose concentrations <5.6 mmol/L and within ±15% at glucose concentrations ≥5.6 mmol/L. The sensor values were obtained using three different glucose monitoring systems, i.e. Abbott, Dexcom and Medtronic, in patients with T1D over a 12-hour period in standardized conditions. The reference values were obtained by analysis of venous plasma glucose samples.
Measured: 36 hours
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted25 Nov 2015 (estimated)
- Study startJul 2015
- Primary completionAug 2015
- Study completionSep 2015
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Posted resultsParticipant flow, outcome measures and adverse events reported by the sponsor.
- Other trials for Diabetes Mellitus, Type 1Studies of the same condition, open and under way.
- More trials run by Medical University of GrazThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Basic science
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 12 participants (actual)
Sponsor and investigators
- Lead sponsor
- Medical University of Graz
- Sponsor type
- University, hospital or other
- Responsible party
- Sponsor
- Investigators
- Julia Mader, Prof. Dr., Principal investigator, Medical University of Graz
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT02614768
- Sponsor's protocol ID
- SPIDIMAN_01
- Status last verified
- May 2023
- First submitted
- 24 Nov 2015
- Last updated
- 6 May 2023
- Results posted
- Yes, on ClinicalTrials.gov
- Data monitoring committee
- No
Study at a glance
From the registry record, last updated by the sponsor on 6 May 2023. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Standardized Evaluation of Subcutaneous Glucose Monitoring Systems Under Routine Environmental Conditions
- ClinicalTrials.gov ID
- NCT02614768
- Acronym / sponsor ID
- SPIDIMAN_01
- Conditions
- Diabetes Mellitus, Type 1
- Keywords
- Glucose Monitoring System
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 12 participants (actual)
- Lead sponsor
- Medical University of GrazUniversity, hospital or other
- Study start
- Jul 2015 (actual)
- Primary completion
- Aug 2015 (actual)
- Study completion
- Sep 2015 (actual)
- First posted
- 25 Nov 2015 (estimated)
- Results
- Posted 6 May 2023
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
The purpose of this study is to assess accuracy and reliability of the SPIDIMAN continuous glucose monitoring (CGM) sensor system with regard to values as measured by Super GL and compare these results with similar evaluations of the Medtronic MiniMed 640G system, the Abbott FreeStyle Libre Flash Sensor and the DexCom G4 Platinum Sensor in patients with type 1 Diabetes.
Detailed description
The study is a single-center open-label study in patients with type 1 diabetes treated with continuous subcutaneous insulin infusion (CSII) or multiple daily injections (MDI) therapy. The study will include a total of 12 subjects.
Study design
- Primary purpose
- Basic science
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
| Arm | Type | Receives |
|---|---|---|
| Glucose Sensor | Experimental | Device: Glucose SensorContinuous subcutaneous glucose monitoring using four different CGM systems in parallel will be performed throughout the study. Insulin therapy will be performed by the subjects themselves, as under daily life conditions. For the hypoglycaemia experiment an increased insulin bolus will be administered with meals (180% of the subject's calculated mealtime dose). |
Who may qualify
- Ages
- 18 years and older
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Adult, Older adult
Key inclusion criteria (7 listed)
- Informed consent obtained after being advised of the nature of the study
- Male or female aged ≥18 years
- Type 1 diabetes for at least 6 months according to the World Health Organization (WHO) definition
- Treatment with multiple daily injections (MDI) or continuous subcutaneous insulin infusion (CSII) for at least 3 months
- Body Mass Index (BMI) < 35 kg/m²
Key exclusion criteria (11 listed)
- Any disease or condition which the investigator or treating physician feels would interfere with the trial or the safety of the subject
- Female of childbearing potential who is pregnant, breast-feeding or intend to become pregnant or is not using adequate contraceptive methods
- Any mental condition rendering the subject incapable of giving his consent
- Subject is using a medication that significantly impacts glucose metabolism (oral steroids) except if stable for at least the last three months and expected to remain stable for the study duration
- Subject may not use acetaminophen (paracetamol) while participating in the study
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Glucose SensorThe investigational intervention is CGM monitoring using four different CGM systems. Three of the four CGM devices which are used will be CGM systems which have received conformité européenne (CE) certification (Dexcom G4 Platinum, Medtronic MiniMed 640G system, Abbott FreeStyle Libre Flash). The SPIDIMAN sensor will be used for the first time in human subjects, does not yet have received CE certification and will be given an identifying label in addition to being labelled "for investigative use only". | Device | Glucose Sensor |
Primary outcome
Percentage of Test-to-Reference Measurement Pairs With an Absolute Relative Difference (ARD) ≤15%
Time frame: 36 hours
The primary outcome measure was overall sensor accuracy determined using ISO 15197:2013, which is percentage (%) of sensor values that are within ±0.8 mmol/L of the reference value at glucose concentrations <5.6 mmol/L and within ±15% at glucose concentrations ≥5.6 mmol/L. The sensor values were obtained using three different glucose monitoring systems, i.e. Abbott, Dexcom and Medtronic, in patients with T1D over a 12-hour period in standardized conditions. The reference values were obtained by analysis of venous plasma glucose samples.
Results and safety
Loading the posted results from ClinicalTrials.gov…
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 6 May 2023.
- Posted resultsParticipant flow, baseline, all outcome measures and adverse events.
Related trials for Diabetes Mellitus, Type 1
Not yet recruitingPhase 3
A Study to Evaluate the Effectiveness and Safety of the Drug AZD6234 for Managing Weight in Adults Who Are Obese or Overweight. These Participants May or May Not Also Have Type 2 Diabetes
Obesity, Overweight
RecruitingPhase 1
Pharmacokinetics, Safety, and Immunogenicity Comparison of Bmab1700 and Opdivo® as Adjuvant Monotherapy in Participants With Melanoma
Melanoma
Active, not recruitingPhase 2
A Study of Botensilimab and Balstilimab for the Treatment of Colorectal Cancer
Metastatic Colorectal Cancer
Not yet recruiting
Inter-Arm Differences in Continuous Glucose Monitor Readings in Adults With Diabetes
Continuous Glucose Monitoring, Type 1 Diabetes
RecruitingObservational study
Estimated Risk of Preeclampsia in the First Trimester and Its Association With Subclinical Atherosclerosis and Cardiovascular Risk in Women With Type 1 Diabetes
Type 1 Diabetes, Preeclampsia
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Julia Mader, Prof. Dr. principal investigator
Medical University of Graz
Questions about this trial
What is this clinical trial studying?
A study testing Glucose Sensor in people with Diabetes Mellitus, Type 1, run by Medical University of Graz.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 7 inclusion and 11 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 6 May 2023. Status can change between updates, so confirm with the study contact before making plans.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
Yes. Results are posted in the registry (first posted 6 May 2023). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Percentage of Test-to-Reference Measurement Pairs With an Absolute Relative Difference (ARD) ≤15% (time frame: 36 hours). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 6 May 2023.