Standardized Evaluation of Subcutaneous Glucose Monitoring Systems Under Routine Environmental Conditions

A study testing Glucose Sensor in people with Diabetes Mellitus, Type 1, run by Medical University of Graz.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to assess accuracy and reliability of the SPIDIMAN continuous glucose monitoring (CGM) sensor system with regard to values as measured by Super GL and compare these results with similar evaluations of the Medtronic MiniMed 640G system, the Abbott FreeStyle Libre Flash Sensor and the DexCom… Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Diabetes Mellitus, Type 1
What is being tested?
Glucose Sensor
Could I receive a placebo?
No, everyone receives an active treatment
How many people
12 participants
Where
No sites listed yet
Study dates
Jul 2015to Sep 2015
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Informed consent obtained after being advised of the nature of the study
  • Male or female aged ≥18 years
  • Type 1 diabetes for at least 6 months according to the World Health Organization (WHO) definition
  • Treatment with multiple daily injections (MDI) or continuous subcutaneous insulin infusion (CSII) for at least 3 months
  • Body Mass Index (BMI) < 35 kg/m²
  • Willing and able to wear 5 CGM devices for the duration of the study and undergo all study procedures.
  • HbA1c ≤ 86 mmol/mol

Who cannot join

  • Any disease or condition which the investigator or treating physician feels would interfere with the trial or the safety of the subject
  • Female of childbearing potential who is pregnant, breast-feeding or intend to become pregnant or is not using adequate contraceptive methods
  • Any mental condition rendering the subject incapable of giving his consent
  • Subject is using a medication that significantly impacts glucose metabolism (oral steroids) except if stable for at least the last three months and expected to remain stable for the study duration
  • Subject may not use acetaminophen (paracetamol) while participating in the study
  • Has severe medical or psychological condition(s) or chronic conditions/infections that in the opinion of the Investigator would compromise the subject's safety or successful participation in the study.
Show all 11Show fewer
  • Subject is actively enrolled in another clinical trial
  • Known adrenal gland problem, pancreatic tumour, or insulinoma
  • Inability of the subject to comply with all study procedures
  • Inability of the subject to understand the patient information.
  • Subject donated blood in the last 3 months

About this study

The purpose of this study is to assess accuracy and reliability of the SPIDIMAN continuous glucose monitoring (CGM) sensor system with regard to values as measured by Super GL and compare these results with similar evaluations of the Medtronic MiniMed 640G system, the Abbott FreeStyle Libre Flash Sensor and the DexCom G4 Platinum Sensor in patients with type 1 Diabetes.

Read the detailed description

The study is a single-center open-label study in patients with type 1 diabetes treated with continuous subcutaneous insulin infusion (CSII) or multiple daily injections (MDI) therapy. The study will include a total of 12 subjects.

Being tested
Glucose Sensor
Keywords
Glucose Monitoring System
Medical subject headings
Diabetes Mellitus, Type 1

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Glucose SensorDevice

    The investigational intervention is CGM monitoring using four different CGM systems. Three of the four CGM devices which are used will be CGM systems which have received conformité européenne (CE) certification (Dexcom G4 Platinum, Medtronic MiniMed 640G system, Abbott FreeStyle Libre Flash). The SPIDIMAN sensor will be used for the first time in human subjects, does not yet have received CE certification and will be given an identifying label in addition to being labelled "for investigative use only".

Study groups in NCT02614768
Study groupRoleWhat they receive
Glucose SensorExperimentalContinuous subcutaneous glucose monitoring using four different CGM systems in parallel will be performed throughout the study. Insulin therapy will be performed by the subjects themselves, as under daily life conditions. For the hypoglycaemia experiment an increased insulin bolus will be administered with meals (180% of the subject's calculated mealtime dose).
Receives: Device: Glucose Sensor

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Percentage of Test-to-Reference Measurement Pairs With an Absolute Relative Difference (ARD) ≤15%

    The primary outcome measure was overall sensor accuracy determined using ISO 15197:2013, which is percentage (%) of sensor values that are within ±0.8 mmol/L of the reference value at glucose concentrations <5.6 mmol/L and within ±15% at glucose concentrations ≥5.6 mmol/L. The sensor values were obtained using three different glucose monitoring systems, i.e. Abbott, Dexcom and Medtronic, in patients with T1D over a 12-hour period in standardized conditions. The reference values were obtained by analysis of venous plasma glucose samples.

    Measured: 36 hours

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted25 Nov 2015 (estimated)
  2. Study startJul 2015
  3. Primary completionAug 2015
  4. Study completionSep 2015

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Basic science
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
12 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT02614768
Sponsor's protocol ID
SPIDIMAN_01
Status last verified
May 2023
First submitted
24 Nov 2015
Last updated
6 May 2023
Data monitoring committee
No

Questions about this trial

What is this clinical trial studying?

A study testing Glucose Sensor in people with Diabetes Mellitus, Type 1, run by Medical University of Graz.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 7 inclusion and 11 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 6 May 2023. Status can change between updates, so confirm with the study contact before making plans.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 6 May 2023). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Percentage of Test-to-Reference Measurement Pairs With an Absolute Relative Difference (ARD) ≤15% (time frame: 36 hours). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 6 May 2023.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in