Lung DiseaseNCT02699268

A Clinical Study Aiming to Validate the Equipment: VoluSense Pediatrics (VSP)

An observational study of Lung Disease, run by Haukeland University Hospital at 1 site in Norway.

Unknown status

Status not confirmed. The sponsor has not verified this record for over two years, so it may be out of date.

Official record
Search all trials

Trial summary

This is a cross-sectional observational study, aiming to validate the accuracy of tidal breathing measurements in infants made by VoluSense Pediatrics, a lung function method based on electromagnetic inductance plethysmography, compared to an ultrasonic flowmeter. Read more about this study

Who is it for?
Ages 2 days to 4 weeksHealthy volunteers can take part
Conditions the study is for
Lung Disease
What is being tested?
Lung function measurement
How many people
50 participants
Where
1 siteNorway
Study dates
Dec 2015to Mar 2016
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 2 days to 4 weeks
  • All sexes
  • Healthy volunteers accepted
  • Child

Study population: Term born and preterm born babies ≥ 2000 grams.

Who may be able to join

  • Healthy term born or preterm born babies ≥ 2000 grams.

Who cannot join

  • Being in an unstable clinical condition.
  • Weight < 2000 grams.

Trial locations and contacts

1 site in 1 country.

Mariann H Bentsen, MD contact

+47 55970871

bens@helse-bergen.no

Norway1 site, 1 recruiting
  • Haukeland University Hospital, Pediatric department, NICU

    Bergen, Western Norway, 5021

    Mariann H Bentsen, MD, +47 55970871, bens@helse-bergen.no

    Thomas Halvorsen, MD, +47 55975300, thomas.halvorsen@helse-bergen.no

    Recruiting

Other trials in Norway

About this study

Official title: Validation of VoluSense Pediatrics, a New Method of Infant Spirometry Based on Electromagnetic Inductance Plethysmography

This is a cross-sectional observational study, aiming to validate the accuracy of tidal breathing measurements in infants made by VoluSense Pediatrics, a lung function method based on electromagnetic inductance plethysmography, compared to an ultrasonic flowmeter. VoluSense Pediatrics consists of a flexible vest placed around the torso of the baby, and changes in tidal volume and flow is measured. The study will enable a better understanding of the usefulness of this equipment. The equipment is made by VoluSense Norway AS, and is owned by Haukeland University Hospital.

Read the detailed description

All infants will be assessed at the pediatric department at Haukeland University Hospital, and all data obtained by the same two investigators (Bentsen and Olsen). The infants will be dressed in the appropriate-sized vest selected according to the length from the armpit to the hip of the child. Care will be taken to ensure that the vest fits snugly around the torso and that the width of the copper wire zig-zag coveres the entire thorax and abdomen including the pubic region. The vest will be applied directly to the skin; neither body nor diaper will be worn underneath. The infants will then be fed before they will be placed supine in a cot and encouraged to sleep. No sedation will be used. Once the infant has fallen asleep, data will be collected with VSP for 2-3 minutes before the Exhalyzer D® neonatal facemask connected in series with the ultrasonic flow sensor will be placed gently, but firmly over the infant's mouth and nose, ensuring a good seal. Once the facemask is in place, a marker will be inserted in the VSP recording. The Exhalyzer D® measurement will be started about thirty seconds after the facemask has been put on, to allow adaptation to the facemask. Simultaneous recording using both devices will then be done for about 1 minute before the mask will be removed. New markers will be inserted in the VSP recording as the Exhalyzer D® measurement is started and stopped facilitating analysis of the same breathing cycles.

Data will be analysed and graphs will be created with SPSS version 22 (IBM SPSS Statistics, New York, US) and MedCalc version 13.1 (MedCalc Software, Mariakerke, Belgium). Bland-Altman plots will be used to visualize agreement between the two measurement techniques. For each pair of measurements, the pairwise mean of the breathing parameters will be plotted on the x-axis, and the pairwise difference expressed as a percentage of their mean value on the y-axis. The mean difference and the 95% limits of agreement and their confidence intervals will be represented by horizontal lines in the graphs.

Conditions
Lung Disease
Being tested
Lung function measurement
Medical subject headings
Lung Diseases

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Lung function measurementOther

    The infants will be exposed to lung function measurements performed by a new metod (Volusense Pediatrics) and a validated method (EcoMedics Exhalyzer D). The two Methods will be compared. This is not an intervential study, but a study aiming to validate a New Method versus an old method

Study groups in NCT02699268
Study groupRoleWhat they receive
Infants (term borns and preterm borns)Intervention: simultaneous lung function measurement using VoluSense Pediatrics and a mask-based method with an ultrasonic flowmeter (EcoMedics Exhalyzer)
Receives: Other: Lung function measurement

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Tidal volume. Unit is milliliter

    Tidal volume will be measured using the two Methods and compared using Bland-Altman statistics

    Measured: 1 hour

Secondary outcomes (5)
  1. Respiratory Rate. Unit is counts per minute

    Respiratory rate will be measured using the two Methods and compared using Bland-Altman statistics

    Measured: 1 hour

  2. Peak tidal expiratory flow (TPTEF). Unit is ml/sec

    TPTEF will be measured using the two Methods and compared using Bland-Altman statistics

    Measured: 1 hour

  3. the ratio of inspiratory to expiratory time (Ti/Te). This is a ratio, and thus no unit

    Ti/Te will be measured using the two Methods and compared using Bland-Altman statistics

    Measured: 1 hour

  4. The ratio of tidal expiratory flow at 50% of expired volume to peak tidal expiratory flow (TEF 50/PTEF).

    TEF 50/PTEF will be measured using the two Methods and compared using Bland-Altman statistics

    Measured: 1 hour

  5. Total expiratory time (Tptef/Te). This is a ratio, and thus no unit

    Tptef/Te will be measured using the two Methods and compared using Bland-Altman statistics

    Measured: 1 hour

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted4 Mar 2016 (estimated)
  2. Study startDec 2015
  3. Primary completionMar 2016 (estimated)
  4. Study completionMar 2016 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Cohort
Time perspective
Prospective
Enrolment
50 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT02699268
Sponsor's protocol ID
REK Vest 2010/496-8
Status last verified
Feb 2016
First submitted
1 Dec 2015
Last updated
4 Mar 2016
Results posted
No
Data monitoring committee
No
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

An observational study of Lung Disease, run by Haukeland University Hospital at 1 site in Norway.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 1 inclusion and 2 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: status not confirmed, as of the sponsor's last update on 4 Mar 2016. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Norway. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Results are not available in the source record.

How can a doctor refer a patient or contact the research team?

Contact Mariann H Bentsen, MD (+47 55970871, bens@helse-bergen.no), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Tidal volume. Unit is milliliter (time frame: 1 hour). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 4 Mar 2016; AzTrials last refreshed its copy on 6 Nov 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in