AzTrials
AzTrials

Accuracy of Measurement of Respiration RateNCT02740504

Reassure Spot Respiration Rate Accuracy Compared to SOMNOScreen

An observational study of Accuracy of Measurement of Respiration Rate, run by ResMed.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

A trial conducted within the company with a new non-contact respiration rate (RR) sensor. The device was compared for accuracy in RR measurement with a cleared device, the SOMNOScreen Easy, which uses chest effort bands. Read more about this study

Who is it for?
Ages 18 to 75 yearsHealthy volunteers can take part
How many people
20 participants
Where
No sites listed yet
Study dates
Jan 2015to Mar 2015
Phase
Observational
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 75 years
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Study population: 20 persons selected to cover the range of age and BMI of 18 to 45

Who may be able to join

  • Greater or equal to 18 years old and less than or equal to 75 years old
  • BMI within the range of greater or equal to 18 and less than or equal to 45

Who cannot join

None listed.

About this study

Official title: A Test of the Spot Respiration Rate Accuracy of the New Reassure Device on Human Subjects

A trial conducted within the company with a new non-contact respiration rate (RR) sensor. The device was compared for accuracy in RR measurement with a cleared device, the SOMNOScreen Easy, which uses chest effort bands.

The RR was measured over a short duration on subjects sitting upright in a chair, with the non-contact sensor at various distances from the subject. A total of 20 subjects were tested, from both genders, selected to represent a wide range of age and body mass index

Read the detailed description

The Reassure uses Radio Frequency (RF) waves to detect the small movements of the human body due to respiration. The reference device, the SOMNOScreen, uses chest effort bands. The Reassure is non-contact and so more comfortable and convenient for quick RR measurement, in a similar way that an Infra-Red thermometer is compared to a traditional under-the-tongue mercury thermometer.

The aim is to explore the variations in the human body that could have an effect on the radio reflection from the human torso, and thus affect the potential detection and accuracy of breathing rate. Factors that could affect non-contact RF performance are:

  • High Body Mass Index (BMI) could potentially reduce the chest wall movement during breathing, or alter the nature of the breathing.
  • Elderly, thin skin (ie pensioner with low BMI) may have less RF reflection.
  • Female breast tissue may have an effect on detection or accuracy.

This was tested at a number of distances from the non-contact sensor and also with the subject having a thick comforter (duvet) laid across them.

The test conditions were

  • Distances of 0.5 m, 1.0m and 1.5m between the subject and the RF sensor.
  • Additional test condition of comforter (duvet) on the subject when the distance from subject to sensor is 1.0 m.
  • Each test condition carried out twice.

The use of the comforter in one test condition is to confirm that the RR measurement by the Reassure is unaffected by the thickness of clothing or blankets being worn.

Keywords
Respiration rate

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

Study groups in NCT02740504
Study groupRoleWhat they receive
SubjectsAll 20 subjects. Selected to have a range of ages (18 to less than 75) and BMI from 18.5 to 45

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Respiration Rate Accuracy

    The respiration rate measured contemporaneously by the Reassure and the reference device, SomnoScreen, are compared on a second-by-second basis, over the recording duration. The difference between the two respiration rate measurements (the delta) for all 20 subjects' recordings, in all positions, are plotted on a normal distribution chart, giving 8,800 data points. The distribution chart reveals the mean difference between the two devices' measurements and the Standard Deviation (SD). The measure of success is if the 95% (ie 2SD) of the difference in the respiration rate indicated by the two devices (the delta) is less than 5 breaths per minute. This measure is derived from the peer review paper by Lim et al, cited in the References section

    Measured: Recording duration of 2 minutes; measurement made after 45 seconds settling time, for at least 60 seconds

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted15 Apr 2016 (estimated)
  2. Study startJan 2015
  3. Primary completionMar 2015
  4. Study completionMar 2015

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Case only
Time perspective
Cross-sectional
Enrolment
20 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT02740504
Sponsor's protocol ID
D22105-106
Status last verified
Mar 2021
First submitted
12 Apr 2016
Last updated
2 Apr 2021
Data monitoring committee
No
Individual data sharing
No: Trial is for device performance validation and only recorded with Clinical Trials.gov as it involved human volunteers. No intent to publish

Publications

  • Lim WS, Carty SM, Macfarlane JT, Anthony RE, Christian J, Dakin KS, Dennis PM. Respiratory rate measurement in adults--how reliable is it? Respir Med. 2002 Jan;96(1):31-3. doi: 10.1053/rmed.2001.1203.

    PubMed 11863207

Questions about this trial

What is this clinical trial studying?

An observational study of Accuracy of Measurement of Respiration Rate, run by ResMed.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 2 inclusion and 0 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 2 Apr 2021. Status can change between updates, so confirm with the study contact before making plans.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 2 Apr 2021). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Respiration Rate Accuracy (time frame: Recording duration of 2 minutes; measurement made after 45 seconds settling time, for at least 60 seconds). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT02740504, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 2 Apr 2021.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in