Healthy VolunteerPATCHSANTESpO2NCT02842255
Calibration of a New Reflectance Oximeter.
A study testing Calibrate the oximeter ALTITRAINER200® in people with Healthy Volunteer, run by University Hospital, Grenoble at 1 site in France.
Trial summary
This study is about the development of a new medical device for the monitoring of of sleep disordered breathing. This device is developed by the CEA-LETI-LE2S and will be composed of an oximeter, an actimeter in order to measure the activity level of the subject, his state of sleeplessness/sleep, and a CO2 partial… Read more about this study
- Who is it for?
- Ages 18 to 50 yearsHealthy volunteers can take part
- Conditions the study is for
- Healthy Volunteer
- What is being tested?
- Calibrate the oximeter ALTITRAINER200®
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 10 participants
- Where
- 1 siteFrance
- Study dates
- Oct 2015to Jan 2016
- Phase
- Interventional
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 18 to 50 years
- All sexes
- Healthy volunteers accepted
- Adult
Who may be able to join
- Subjects (both gender) aged between 18 to 50 years affiliated to a certain social security, that has given their free consent to participate to the study
- Subjets without any pathology known and not under medical treatment likely to suffer from controlled hypoxia at levels demanded in the protocol.
- Subjets having a regular practice of sport
Who cannot join
- Subjet non affiliated to a certain social security
- Smoker or person exposed to passive smoking
- History of acute mountain sickness or known sensitivity to hypoxia
- Any chronic pathology known, in particular pathologies behind high rates of methaemoglobin
- Pregnant, parturient or breastfeeding women, person deprived of liberty by judicial or administrative decision, person under a legally protected order and could'nt be included in clinical trials (persons mentioned in L1121-5 to L1121-8 of CSP).
- Subjets likely to not be cooperative or respectful of the inherent constraints to study participation.
Trial locations and contacts
1 site in 1 country.
About this study
This study is about the development of a new medical device for the monitoring of of sleep disordered breathing. This device is developed by the CEA-LETI-LE2S and will be composed of an oximeter, an actimeter in order to measure the activity level of the subject, his state of sleeplessness/sleep, and a CO2 partial pressure transcutaneous sensor (PtcCO2).
Developing this device is realized in several steps : a study (ALMOST) is already in progress and its goal is to acquire a database of polysomnography to create automatic algorithms in order to detect respiratory disorders with this device under development.
The main goal of the study is to calibrate the oxymeter under development.
- Conditions
- Healthy Volunteer
- Being tested
- Calibrate the oximeter ALTITRAINER200®
- Keywords
- Sleep apnea, Oximeter, Calibration, Oxygen partial pressure
- Medical subject headings
- Sleep Apnea Syndromes
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
Calibrate the oximeter ALTITRAINER200®Device
During normoxic condition and four situations of hypoxia.
| Study group | Role | What they receive |
|---|---|---|
| Healthy volunteers | Experimental | Not described Receives: Device: Calibrate the oximeter ALTITRAINER200® |
What is being measured
The primary outcome is the main question the study is designed to answer.
Calibration of the oximeter in normoxic situation
Measures with ALTITRAINER200® system Ambiant air : 98-96 % partial pressure in O2
Measured: 2 hours
Calibration of the oximeter in hypoxia situation
Measures with ALTITRAINER200® system Situation 1 : 90 % (92-88%) partial pressure in O2 Situation 2 : 85% (87-83%) partial pressure in O2 Situation 3 : 80% (82-78%) partial pressure in O2 Situation 4 : 75%(77%-73%) partial pressure in O2
Measured: 2 hours
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted22 Jul 2016 (estimated)
- Study startOct 2015
- Primary completionJan 2016
- Study completionJan 2016
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Publications (12)Papers the sponsor has linked to this study.
- Other trials for Healthy VolunteerStudies of the same condition, open and under way.
- More trials run by University Hospital, GrenobleThe sponsor's other registered studies.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Enrolment
- 10 participants (actual)
Sponsor and investigators
- Lead sponsor
- University Hospital, Grenoble
- Sponsor type
- University, hospital or other
- Responsible party
- Sponsor
- Investigators
- Bernard Wuyam, Doctor, Principal investigator, Grenoble Hospital University
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT02842255
- Sponsor's protocol ID
- 38RC15.098
- Acronym
- PATCHSANTESpO2
- Status last verified
- Jul 2016
- First submitted
- 20 Jul 2016
- Last updated
- 22 Jul 2016
- Results posted
- No
- Data monitoring committee
- No
Publications
Berry RB, Budhiraja R, Gottlieb DJ, Gozal D, Iber C, Kapur VK, Marcus CL, Mehra R, Parthasarathy S, Quan SF, Redline S, Strohl KP, Davidson Ward SL, Tangredi MM; American Academy of Sleep Medicine. Rules for scoring respiratory events in sleep: update of the 2007 AASM Manual for the Scoring of Sleep and Associated Events. Deliberations of the Sleep Apnea Definitions Task Force of the American Academy of Sleep Medicine. J Clin Sleep Med. 2012 Oct 15;8(5):597-619. doi: 10.5664/jcsm.2172.
PubMed 23066376Dempsey JA, Veasey SC, Morgan BJ, O'Donnell CP. Pathophysiology of sleep apnea. Physiol Rev. 2010 Jan;90(1):47-112. doi: 10.1152/physrev.00043.2008.
PubMed 20086074Chouri-Pontarollo N, Borel JC, Tamisier R, Wuyam B, Levy P, Pepin JL. Impaired objective daytime vigilance in obesity-hypoventilation syndrome: impact of noninvasive ventilation. Chest. 2007 Jan;131(1):148-55. doi: 10.1378/chest.06-1159.
PubMed 17218569Becker HF, Piper AJ, Flynn WE, McNamara SG, Grunstein RR, Peter JH, Sullivan CE. Breathing during sleep in patients with nocturnal desaturation. Am J Respir Crit Care Med. 1999 Jan;159(1):112-8. doi: 10.1164/ajrccm.159.1.9803037.
PubMed 9872827Peppard PE, Young T, Palta M, Skatrud J. Prospective study of the association between sleep-disordered breathing and hypertension. N Engl J Med. 2000 May 11;342(19):1378-84. doi: 10.1056/NEJM200005113421901.
PubMed 10805822Perrin C, Unterborn JN, Ambrosio CD, Hill NS. Pulmonary complications of chronic neuromuscular diseases and their management. Muscle Nerve. 2004 Jan;29(1):5-27. doi: 10.1002/mus.10487.
PubMed 14694494Flemons WW, Reimer MA. Development of a disease-specific health-related quality of life questionnaire for sleep apnea. Am J Respir Crit Care Med. 1998 Aug;158(2):494-503. doi: 10.1164/ajrccm.158.2.9712036.
PubMed 9700127Nitzan M, Romem A, Koppel R. Pulse oximetry: fundamentals and technology update. Med Devices (Auckl). 2014 Jul 8;7:231-9. doi: 10.2147/MDER.S47319. eCollection 2014.
PubMed 25031547Mollard P, Bourdillon N, Letournel M, Herman H, Gibert S, Pichon A, Woorons X, Richalet JP. Validity of arterialized earlobe blood gases at rest and exercise in normoxia and hypoxia. Respir Physiol Neurobiol. 2010 Jul 31;172(3):179-83. doi: 10.1016/j.resp.2010.05.017. Epub 2010 May 21.
PubMed 20493971Richalet JP, Larmignat P, Poitrine E, Letournel M, Canoui-Poitrine F. Physiological risk factors for severe high-altitude illness: a prospective cohort study. Am J Respir Crit Care Med. 2012 Jan 15;185(2):192-8. doi: 10.1164/rccm.201108-1396OC. Epub 2011 Oct 27.
PubMed 22071330Savourey G, Launay JC, Besnard Y, Guinet-Lebreton A, Alonso A, Sauvet F, Bourrilhon C. Normo or hypobaric hypoxic tests: propositions for the determination of the individual susceptibility to altitude illnesses. Eur J Appl Physiol. 2007 May;100(2):193-205. doi: 10.1007/s00421-007-0417-8. Epub 2007 Feb 24.
PubMed 17323073Aguilaniu B, Maitre J, Diab S, Perrault H, Peronnet F. Detection of disturbances in pulmonary gas exchanges during exercise from arterialized earlobe PO2. Respir Physiol Neurobiol. 2011 Jun 30;177(1):30-5. doi: 10.1016/j.resp.2011.03.005. Epub 2011 Mar 17.
PubMed 21397053
Study at a glance
From the registry record, last updated by the sponsor on 22 Jul 2016. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Calibration of a New Reflectance Oximeter.
- ClinicalTrials.gov ID
- NCT02842255
- Acronym / sponsor ID
- PATCHSANTESpO2 / 38RC15.098
- Conditions
- Healthy Volunteer
- Keywords
- Sleep apnea, Oximeter, Calibration, Oxygen partial pressure
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 10 participants (actual)
- Lead sponsor
- University Hospital, GrenobleUniversity, hospital or other
- Study start
- Oct 2015
- Primary completion
- Jan 2016 (actual)
- Study completion
- Jan 2016 (actual)
- First posted
- 22 Jul 2016 (estimated)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
This study is about the development of a new medical device for the monitoring of of sleep disordered breathing. This device is developed by the CEA-LETI-LE2S and will be composed of an oximeter, an actimeter in order to measure the activity level of the subject, his state of sleeplessness/sleep, and a CO2 partial pressure transcutaneous sensor (PtcCO2).
Developing this device is realized in several steps : a study (ALMOST) is already in progress and its goal is to acquire a database of polysomnography to create automatic algorithms in order to detect respiratory disorders with this device under development.
The main goal of the study is to calibrate the oxymeter under development.
Study design
- Allocation
- Not applicable
- Intervention model
- Single group
- Masking
- None (open label)
- Number of arms
- 1
| Arm | Type | Receives |
|---|---|---|
| Healthy volunteers | Experimental | Device: Calibrate the oximeter ALTITRAINER200® |
Who may qualify
- Ages
- 18 to 50 years
- Sex
- All
- Healthy volunteers
- Accepted
- Age groups
- Adult
Key inclusion criteria (3 listed)
- Subjects (both gender) aged between 18 to 50 years affiliated to a certain social security, that has given their free consent to participate to the study
- Subjets without any pathology known and not under medical treatment likely to suffer from controlled hypoxia at levels demanded in the protocol.
- Subjets having a regular practice of sport
Key exclusion criteria (6 listed)
- Subjet non affiliated to a certain social security
- Smoker or person exposed to passive smoking
- History of acute mountain sickness or known sensitivity to hypoxia
- Any chronic pathology known, in particular pathologies behind high rates of methaemoglobin
- Pregnant, parturient or breastfeeding women, person deprived of liberty by judicial or administrative decision, person under a legally protected order and could'nt be included in clinical trials (persons mentioned in L1121-5 to L1121-8 of CSP).
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| Calibrate the oximeter ALTITRAINER200®During normoxic condition and four situations of hypoxia. | Device | Healthy volunteers |
Primary outcome
Calibration of the oximeter in normoxic situation
Time frame: 2 hours
Measures with ALTITRAINER200® system Ambiant air : 98-96 % partial pressure in O2
Calibration of the oximeter in hypoxia situation
Time frame: 2 hours
Measures with ALTITRAINER200® system Situation 1 : 90 % (92-88%) partial pressure in O2 Situation 2 : 85% (87-83%) partial pressure in O2 Situation 3 : 80% (82-78%) partial pressure in O2 Situation 4 : 75%(77%-73%) partial pressure in O2
Results and safety
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 22 Jul 2016.
- Background: PubMed 23066376Berry RB, Budhiraja R, Gottlieb DJ, Gozal D, Iber C, Kapur VK, Marcus CL, Mehra R, Parthasarathy S, Quan SF, Redline S, Strohl KP, Davidson Ward SL, Tangredi MM; American Academy of Sleep Medicine. Rules for scoring respiratory events in sleep: update of the 2007 AASM Manual for the Scoring of Sleep and Associated Events. Deliberations of the Sleep Apnea Definitions Task Force of the American Academy of Sleep Medicine. J Clin Sleep Med. 2012 Oct 15;8(5):597-619. doi: 10.5664/jcsm.2172.
- Background: PubMed 20086074Dempsey JA, Veasey SC, Morgan BJ, O'Donnell CP. Pathophysiology of sleep apnea. Physiol Rev. 2010 Jan;90(1):47-112. doi: 10.1152/physrev.00043.2008.
- Background: PubMed 17218569Chouri-Pontarollo N, Borel JC, Tamisier R, Wuyam B, Levy P, Pepin JL. Impaired objective daytime vigilance in obesity-hypoventilation syndrome: impact of noninvasive ventilation. Chest. 2007 Jan;131(1):148-55. doi: 10.1378/chest.06-1159.
- Background: PubMed 9872827Becker HF, Piper AJ, Flynn WE, McNamara SG, Grunstein RR, Peter JH, Sullivan CE. Breathing during sleep in patients with nocturnal desaturation. Am J Respir Crit Care Med. 1999 Jan;159(1):112-8. doi: 10.1164/ajrccm.159.1.9803037.
- Background: PubMed 10805822Peppard PE, Young T, Palta M, Skatrud J. Prospective study of the association between sleep-disordered breathing and hypertension. N Engl J Med. 2000 May 11;342(19):1378-84. doi: 10.1056/NEJM200005113421901.
- Background: PubMed 14694494Perrin C, Unterborn JN, Ambrosio CD, Hill NS. Pulmonary complications of chronic neuromuscular diseases and their management. Muscle Nerve. 2004 Jan;29(1):5-27. doi: 10.1002/mus.10487.
- Background: PubMed 9700127Flemons WW, Reimer MA. Development of a disease-specific health-related quality of life questionnaire for sleep apnea. Am J Respir Crit Care Med. 1998 Aug;158(2):494-503. doi: 10.1164/ajrccm.158.2.9712036.
- Background: PubMed 25031547Nitzan M, Romem A, Koppel R. Pulse oximetry: fundamentals and technology update. Med Devices (Auckl). 2014 Jul 8;7:231-9. doi: 10.2147/MDER.S47319. eCollection 2014.
- Background: PubMed 20493971Mollard P, Bourdillon N, Letournel M, Herman H, Gibert S, Pichon A, Woorons X, Richalet JP. Validity of arterialized earlobe blood gases at rest and exercise in normoxia and hypoxia. Respir Physiol Neurobiol. 2010 Jul 31;172(3):179-83. doi: 10.1016/j.resp.2010.05.017. Epub 2010 May 21.
- Background: PubMed 22071330Richalet JP, Larmignat P, Poitrine E, Letournel M, Canoui-Poitrine F. Physiological risk factors for severe high-altitude illness: a prospective cohort study. Am J Respir Crit Care Med. 2012 Jan 15;185(2):192-8. doi: 10.1164/rccm.201108-1396OC. Epub 2011 Oct 27.
- Background: PubMed 17323073Savourey G, Launay JC, Besnard Y, Guinet-Lebreton A, Alonso A, Sauvet F, Bourrilhon C. Normo or hypobaric hypoxic tests: propositions for the determination of the individual susceptibility to altitude illnesses. Eur J Appl Physiol. 2007 May;100(2):193-205. doi: 10.1007/s00421-007-0417-8. Epub 2007 Feb 24.
- Background: PubMed 21397053Aguilaniu B, Maitre J, Diab S, Perrault H, Peronnet F. Detection of disturbances in pulmonary gas exchanges during exercise from arterialized earlobe PO2. Respir Physiol Neurobiol. 2011 Jun 30;177(1):30-5. doi: 10.1016/j.resp.2011.03.005. Epub 2011 Mar 17.
Related trials for Healthy Volunteer
Not yet recruiting
Neuroergonomic Analysis of a Brain-Computer Interface Device (BCI)
Healthy Volunteers - Male and Female
RecruitingPhase 1
First-in-Human Trial to Assess REGN20423 in Healthy Adult Participants and Adult Participants With Atopic Dermatitis
Healthy Volunteer, Atopic Dermatitis
Not yet recruitingObservational study
Establishment of Biobank of Nasal Epithelium Samples From Healthy Volunteers
Healthy
RecruitingObservational study
Improved Methods of Cell Selection for Bone Marrow Transplant Alternatives
Graft-Versus-Host Disease, Graft-versus-leukemia
Not yet recruitingObservational study
Screening Protocol
Pain
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Bernard Wuyam, Doctor principal investigator
Grenoble Hospital University
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing Calibrate the oximeter ALTITRAINER200® in people with Healthy Volunteer, run by University Hospital, Grenoble at 1 site in France.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 3 inclusion and 6 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 22 Jul 2016. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in France. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
Calibration of the oximeter in normoxic situation (time frame: 2 hours). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT02842255, which also lists 12 linked publications.
Does a listing mean the treatment is approved or proven to work?
No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.
How often is this information updated?
AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 22 Jul 2016.