The Medtronic Harmony™ Transcatheter Pulmonary Valve Clinical Study

A study testing Harmony TPV System in people with Congenital Heart Disease and Tetrology of Fallot, run by Medtronic Heart Valves at 12 sites in 3 countries.

Active, not recruiting

Under way, not looking for participants. Everyone needed has enrolled. The study is still running.

Official record

Search all trials

Trial summary

The purpose of this study is to further evaluate the safety and effectiveness of the Harmony™ TPV system. The Pivotal/CAS phases of the study have transitioned into a post-approval study to confirm the long-term functionality of transcatheter implantation of the Medtronic Harmony TPV. Read more about this study

Who is it for?
Any age
What is being tested?
Harmony TPV System
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
86 participants
Where
12 sitesin 3 countries
Study dates
Jan 2017to Feb 2031
Phase
Interventional

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages Any age
  • All sexes
  • No healthy volunteers
  • Child, Adult, Older adult

Who may be able to join

  • Subject has severe pulmonary regurgitation as assessed via echocardiography or CMR determined PR fraction >/= 30%
  • Subject has clinical indication for surgical placement of an RV-PA conduit or bioprosthetic pulmonary valve
  • Subject is willing to consent to participate

Who cannot join

  • Patients with right ventricular outflow tract obstruction (RVOTO) lesions surgically treated with an RV-to-PA conduit implant
  • RVOT anatomy or morphology that is unfavorable for device anchoring
  • Positive pregnancy test
  • Life expectancy of less than 1 year

Trial locations and contacts

12 sites in 3 countries.

12 sites in 3 countries

United States9 sites
  • Nationwide Children's Hospital

    Columbus, Ohio, 43215

  • Texas Children's Hospital

    Houston, Texas, 77030

  • Ronald Regan UCLA Medical Center

    Los Angeles, California, 90095

  • Yale University

    New Haven, Connecticut, 06510

  • Stanford University Medical Center

    Palo Alto, California, 94304

  • The Children's Hospital of Philadelphia

    Philadelphia, Pennsylvania, 19104

  • Mayo Clinic

    Rochester, Minnesota, 55902

  • Primary Children's Hospital

    Salt Lake City, Utah, 84113

  • Seattle Children's Hospital

    Seattle, Washington, 98105

Other trials in United States

Canada2 sites
  • The Hospital for Sick Children (SickKids)

    Toronto, Ontario, M5G 1X8

  • Toronto General Hospital

    Toronto, Ontario, M5G 2C4

Other trials in Canada

Japan1 site
  • National Cerebral and Cardiovascular Center

    Suita, Osaka, 565-8565

Other trials in Japan

About this study

The purpose of this study is to further evaluate the safety and effectiveness of the Harmony™ TPV system. The Pivotal/CAS phases of the study have transitioned into a post-approval study to confirm the long-term functionality of transcatheter implantation of the Medtronic Harmony TPV.

Read the detailed description

The continued clinical experience addendum is a prospective, multi-center, non-randomized, interventional study to evaluate the safety and effectiveness of the Harmony TPV system. All implanted subjects will receive the Harmony TPV 22 or Harmony mTPV 25 device. This phase allows up to 45 subjects implanted with TPV 22 in the United States and Canada, and up to 84 subjects implanted with mTPV 25 in the United States.

The Post Approval Phase (PAS) addendum is a prospective, multi-center, non-randomized, post-market study to evaluate the safety and effectiveness of the Harmony TPV system in the United States. All implanted subjects have been implanted with the Harmony TPV 22 or Harmony mTPV 25 device during the pivotal or CAS phase of the study. The PAS addendum extends follow-up from five years to ten years for consenting subjects.

Being tested
Harmony TPV System
Medical subject headings
Heart Defects, Congenital, Tetralogy of Fallot, Pulmonary Valve Insufficiency

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Harmony TPV SystemDevice

    The Harmony™ TPV (Model NTPV0022) is comprised of a 22mm porcine pericardium valve, sewn to a polyester-covered nitinol frame.

    The Harmony TPV 25 and mTPV 25 (Model HTPV254952 and HTPV254252, respectively) is comprised of a 25mm porcine pericardium valve, sewn to a polyester covered asymmetrical hourglass nitinol frame with larger diameter inflow and outflow as well as shorter length compared to TPV 22 (Harmony TPV 25 Model HTPV254952 is not applicable for Continued Clinical Experience).

    The Harmony Delivery System (DS) for the TPV 22, TPV 25, and mTPV 25 (NTPVDS0022 and HTPVDS0025) are all 25 Fr delivery systems using a coil loading catheter.

Study groups in NCT02979587
Study groupRoleWhat they receive
Harmony TPV SystemOtherIntervention Device: Harmony Transcatheter Pulmonary Valves and Delivery Systems
Receives: Device: Harmony TPV System

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Freedom From Procedure- or Device-related Mortality at 30 Days.

    The primary safety endpoint is point estimate of freedom from procedure or device-related mortality rate at 30 days post procedure.

    Measured: 30 days

  2. Number of Participants With Acceptable Hemodynamic Function Composite at 6 Months

    Defined as:

    • Mean RVOT gradient as measured by continuous-wave Doppler ≤40 mmHg -AND-
    • Pulmonary regurgitant fraction as measured by magnetic resonance imaging <20%

    Measured: 6 months

Secondary outcomes (9)
  1. Number of Participants With Technical Success at Exit From Catheterization Lab/Operating Room (OR)

    Technical success at exit from catheterization lab/operating room (OR), as defined as:

    • No device- or procedural-related mortality, with
    • Successful access, delivery and retrieval of the delivery system, and
    • Deployment and correct positioning (including minor repositioning if needed) of the single intended device, and
    • No need for additional unplanned or emergency surgery or re-intervention related to the device or access procedure

    Measured: At exit from catheterization lab/operating room (OR)

  2. Device Success Out to 5 Years

    Device success is defined as:

    • No device- or procedural-related mortality, with
    • Original intended device in place, and
    • No additional surgical or interventional procedures related to access or the device since completion of the original procedure (i.e., exit from the catheterization lab), and
    • Intended performance of the device, as defined as:
    • Structural performance: No migration, embolization, detachment, major stent fracture, hemolysis, thrombosis, endocarditis, and
    • Hemodynamic performance: Relief of insufficiency (PR < moderate) without producing the opposite (mean RVOT gradient > 40 mmHg) as measured by continuous wave Doppler, and
    • Absence of para-device complications, as defined by:
    • PVL = moderate, or
    • Erosion, or
    • RVOT or PA rupture

    Measured: 5 years

  3. Number of Participants With Procedural Success at 30 Days

    Procedural success is defined as:

    • Device success at 30 days, and
    • None of the following device- or procedure-related serious adverse events:
    • Life-threatening major bleed
    • Major vascular or cardiac structural complications required unplanned reintervention or surgery
    • Stage 2 or 3 acute kidney injury (AKI) (includes new dialysis)
    • Pulmonary embolism
    • Severe heart failure (HF) or hypotension requiring IV inotrope, ultrafiltration, or mechanical circulatory support
    • Prolonged intubation >48 hours

    Measured: 30 days

  4. Freedom From TPV Dysfunction Out to 5 Years

    TPV dysfunction is defined as any one of the following:

    • RVOT reoperation for device-related reasons
    • Catheter re-intervention of TPV
    • Hemodynamic dysfunction of the TPV (moderate or greater pulmonary regurgitation, and/or a mean RVOT gradient >40 mmHg)

    Measured: 5 years (5 year data has an anticipated reporting date of April 2026)

  5. Assessment of Safety

    All procedure-related serious adverse events. All device-related serious adverse events. Death (all-cause, procedural, and device-related).

    Measured: Ongoing

  6. Characterization of Quality of Life Scores Out to 5 Years

    Quality of life score over time will be assessed by the SF-36 at pre-implant, 1 month post-implant, 6 months post-implant, 1 year post-implant, 2 years post-implant, and 3 years. The analysis cohort will be the implanted > 24 hours cohort. Minimum score of 0 and maximum score of 100 is possible. Higher values are considered to be a better outcome.

    Measured: Baseline, 30 Day, 6 Month, 1 Year, 2 Year, 3 Year

  7. Characterization of Right Ventricle Remodeling Following TPV Implant

    Right ventricle remodeling will be assessed via CMR at pre-implant and 6 months post-implant. The analysis cohort will be implanted longer than 24 hours cohort. The characterization will be made using right ventricular end diastolic volume (ml).

    Measured: Baseline, 6 Month

  8. Characterization of Quality of Life Scores Out to 5 Years

    Quality of life score over time will be assessed by the SF-36 at 4 \& 5 years. The analysis cohort will be the implanted > 24 hours cohort. Minimum score of 0 and maximum score of 100 is possible. Higher values are considered to be a better outcome.

    Measured: 4 Year, 5 Year (4 & 5 year data has an anticipated reporting date of April 2026)

  9. Characterization of Right Ventricle Remodeling Following TPV Implant

    Right ventricle remodeling will be assessed via CMR at 2 years post- implant and 5 years post-implant. The analysis cohort will be implanted longer than 24 hours cohort. The characterization will be made using right ventricular end diastolic volume (ml).

    Measured: 2 Year, 5 Year (2 & 5 year data has an anticipated reporting date of April 2026)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted1 Dec 2016 (estimated)
  2. Study startJan 2017
  3. Primary completion17 Jul 2020
  4. Study completionFeb 2031 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
86 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT02979587
Sponsor's protocol ID
MDT16004CON001
Status last verified
Sep 2025
First submitted
23 Nov 2016
Last updated
19 Sep 2025
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
Yes

Publications

  • Morray BH, Gillespie MJ, Cheatham JP, Salavitabar A, Peng L, Jones TK, Levi DS, Gray RG, Asnes J, Cabalka AK, Fujimoto K, Qureshi AM, Bergersen L, Benson LN, Haugan D, McElhinney DB. Midterm Outcomes in a Pooled Cohort of Harmony Transcatheter Pulmonary Valve Recipients. Circ Cardiovasc Interv. 2025 Sep;18(9):e015196. doi: 10.1161/CIRCINTERVENTIONS.125.015196. Epub 2025 Jul 16.

    PubMed 40665900
  • Gillespie MJ, Bergersen L, Benson LN, Weng S, Cheatham JP. 5-Year Outcomes From the Harmony Native Outflow Tract Early Feasibility Study. JACC Cardiovasc Interv. 2021 Apr 12;14(7):816-817. doi: 10.1016/j.jcin.2021.01.046. No abstract available.

    PubMed 33826508