Atrial FibrillationNCT03095131

Assessment of the Clinical Efficacy of the Heart Spectrum Blood Pressure Monitor for Diagnosis of Atrial Fibrillation

A study testing Heart Spectrum Blood Pressure Monitor in people with Atrial Fibrillation, run by Ostar Meditech Corp.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Question: Is 12-lead electrocardiography (ECG) the most accurate and convenient diagnostic method for atrial fibrillation (AF)? Findings: The Heart Spectrum Blood Pressure Monitor can be used at home, and is capable of determining the occurrence of AF. Three measurement methods were used in this study. Read more about this study

Who is it for?
Ages 20 years and olderHealthy volunteers can take part
Conditions the study is for
Atrial Fibrillation
What is being tested?
Heart Spectrum Blood Pressure Monitor
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
62 participants
Where
No sites listed yet
Study dates
22 Feb 2016to 7 Dec 2016
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 20 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Subjects older than 20 years of age who had been diagnosed with AF using 12-lead ECG, or were healthy (without AF) were included.

Who cannot join

  • Subjects who had been exposed to high frequency surgical equipment during testing, those with cardiac pacemakers or implantable defibrillators, and pregnant women were excluded.

About this study

Official title: Assessment of the Clinical Efficacy of the Heart Spectrum Blood Pressure Monitor for Diagnosis of Atrial Fibrillation: An Unblinded Clinical Trial

Question: Is 12-lead electrocardiography (ECG) the most accurate and convenient diagnostic method for atrial fibrillation (AF)? Findings: The Heart Spectrum Blood Pressure Monitor can be used at home, and is capable of determining the occurrence of AF. Three measurement methods were used in this study. The sensitivity, specificity, PPV, and NPV of the methods used are between 90% and 100%, and the Heart Spectrum Blood Pressure Monitor can be effectively applied to AF detection.

Meaning: The Heart Spectrum Blood Pressure Monitor can be used at home and can replace the need for an in-office ECG, thereby accomplishing meaningful telecare.

Read the detailed description

Importance: Atrial fibrillation (AF) is the most common arrhythmia. The most common diagnostic method, 12-lead ECG, can record episodes of arrhythmia from which the type and severity can be determined. The Heart Spectrum Blood Pressure Monitor (P2, OSTAR Meditech Corp.) is used to measure cardiovascular pressure change, with fast Fourier transform (FFT) analysis, to obtain the heart rate frequency variability and accurate blood pressure data.

Objective: To compare the diagnostic efficacy of the Heart Spectrum Blood Pressure Monitor with a 12-lead ECG in patients with AF. Three measurement methods were used in this study: blood pressure, mean arterial pressure (MAP) which is calculated from individual blood pressure as constant pressure, and a constant pressure 60 mmHg, in order to analyze the heart index and compare it with simultaneous 12-lead ECG.

Design: The physician used a 12-lead ECG and the Heart Spectrum Blood Pressure Monitor simultaneously. The Heart Spectrum Blood Pressure Monitor used an FFT analysis for the diagnosis of AF; the findings were compared with the readings from the 12-lead ECG.

Setting: This unblinded clinical trial was conducted at the Emergency Room of Taipei Medical University Hospital.

Participants: Twenty-nine subjects with AF and 33 without AF, aged 25 to 97 years (mean: 63.5 years), were included.

Intervention(s) for clinical trials Subjects who had been exposed to high frequency surgical equipment during testing, those with cardiac pacemakers or implantable defibrillators, and pregnant women were excluded.

Trial Registration: TMU JIRB No.: N201510061

Being tested
Heart Spectrum Blood Pressure Monitor
Keywords
Heart Spectrum Blood Pressure MonitorHeart Spectrum
Medical subject headings
Atrial Fibrillation

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Heart Spectrum Blood Pressure MonitorDevice

    The Heart Spectrum Blood Pressure Monitor provides a convenient and effective filtering application compared with traditional methods that require a visit to a physician for a 12-lead ECG.

Study groups in NCT03095131
Study groupRoleWhat they receive
12-lead ECGExperimentalNot described
Receives: Device: Heart Spectrum Blood Pressure Monitor

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. The Heart Spectrum Blood Pressure Monitor used an FFT analysis for the diagnosis of AF; the findings were compared with the readings from the 12-lead ECG.

    In the emergency room, when medical staff found patients with AF, they informed the research physicians to study the case. The Heart Spectrum Blood Pressure Monitor was compared with 12-lead ECG (ELI 250) after written informed consent was obtained from the patient. The results were interpreted by the emergency physician. Finally, the sensitivity, specificity, positive predictive value (PPV), and negative predictive value (NPV) were analyzed to evaluate the clinical efficacy.

    Measured: monitoring for about 60 seconds per test

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted29 Mar 2017
  2. Study start22 Feb 2016
  3. Primary completion23 Jun 2016
  4. Study completion7 Dec 2016

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Device feasibility
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
62 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03095131
Sponsor's protocol ID
TMU JIRB No.: N201510061
Status last verified
Mar 2017
First submitted
17 Mar 2017
Last updated
29 Mar 2017
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Heart Spectrum Blood Pressure Monitor in people with Atrial Fibrillation, run by Ostar Meditech Corp.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 1 inclusion and 1 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 29 Mar 2017. Status can change between updates, so confirm with the study contact before making plans.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

The Heart Spectrum Blood Pressure Monitor used an FFT analysis for the diagnosis of AF; the findings were compared with the readings from the 12-lead ECG. (time frame: monitoring for about 60 seconds per test). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.