AzTrials
AzTrials

HealthyNCT03187665

Alarms Due to Loss of Signal Integrity in Partially Wireless Versus Traditional Pulse Oximetry in Pediatric Patients

An observational study of Healthy, run by Pamela Petersen at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

This research study will help the investigators to learn more about a device used when children are sick called a pulse oximeter. The pulse oximeter measures how much oxygen is inside a child's blood without taking blood from the child. It is non-invasive meaning it does not enter the body. Read more about this study

Who is it for?
Ages 6 months to 17 yearsHealthy volunteers can take part
Conditions the study is for
Healthy
What is being tested?
Radical 7 and Radius 7
Type of treatment
Medical device
How many people
36 participants
Where
1 siteUnited States
Study dates
26 Jul 2017to 1 Nov 2017
Phase
Observational
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 6 months to 17 years
  • All sexes
  • Healthy volunteers accepted
  • Child

Study population: The subjects will be a sampling from the communities served by the Children's Hospital of Wisconsin. Pediatric subjects will be recruited via fliers placed in pediatric critical care areas, in the Translational Research Control Unit (TRU) and on the Medical College of Wisconsin (MCW) research bulletin board outside the MCW office of research. Subjects may include those responding to the fliers directly as well as the healthy siblings of patients being evaluated or treated in the Children's Hospital of Wisconsin Emergency Room (EDTC), Clinics or Hospital.

Who may be able to join

  • Healthy subjects between the ages of 6 months and 17 years at the time of data acquisition.

Who cannot join

  • Subjects will be excluded if they have any pre-existing cardiac or pulmonary chronic medical diagnoses as determined by asking the subject's guardian.

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Children's Hospital of Wisconsin

    Milwaukee, Wisconsin, 53226

Other trials in United States

About this study

Official title: The Ties That Bind: Evaluation of a Patient-worn Pulse Oximeter Compared to Traditional Pulse Oximetry on Loss of Signal Integrity in Pediatric Patients

This research study will help the investigators to learn more about a device used when children are sick called a pulse oximeter. The pulse oximeter measures how much oxygen is inside a child's blood without taking blood from the child. It is non-invasive meaning it does not enter the body. The device has a cable attached to it. At the end of the cable is a wrap that looks like a Band-Aid with a red light on it. This wrap is placed around a finger or toe. The red light gives the investigators a reading of how much oxygen is in the child's blood and the child's heart rate. Having a pulse oximeter connected to a child is painless. This device is used in many places. Besides hospitals, it is used in doctors' offices and in fitness centers.

This study will help the investigators learn more about whether a partially wireless, more portable pulse oximeter that connects to a small device worn on the child's arm or leg will give the investigators a more reliable signal/reading while letting children move more easily. The investigators will compare this device with the traditional wall-connected unit. Movement of the cable or a child moving may give a false oxygen reading. The investigators will ask the child to do activities that create movement and will look at the readings when the child moves. The investigators think the partially wireless pulse oximeter will be more reliable during movement than the traditional wall-connected unit.

Subjects will have two continuous pulse oximeter probes placed on them. These soft probes will go on a finger, toe, foot or hand and will be attached to two different pulse oximeter monitors. The child will then be asked to do common childhood activities based on their age for about 20 minutes. These activities will be play activities the child already does such as grabbing a toy, drawing with crayons or kicking a ball. Continuous pulse oximetry data will be recorded during the testing and will be stored in a way that it cannot be linked to the subject after the testing is complete.

Read the detailed description

PURPOSE OF THE STUDY This study will evaluate the loss of signal integrity secondary to motion in pediatric subjects being monitored with a newer, partially wireless pulse oximeter (the Radius 7) which eliminates the cable connecting the subject to a wall unit by instead connecting to a small device worn on the subject's arm or leg. The Radius 7 will be compared to the traditional wall-connected pulse oximeter currently used at Children's Hospital of Wisconsin (CHW) which requires a cable connecting to a wall unit (the Radical 7).

HYPOTHESIS Continuously monitoring pulse oximetry with a partially wireless monitor without physical connection to a wall unit will decrease episodes of loss of signal integrity generated by motion artifact in healthy pediatric subjects completing age appropriate activities.

BACKGROUND, SIGNIFICANCE, AND RATIONALE Pulse oximetry is used in both inpatient and outpatient environments to measure the oxygen saturation in a person's blood and the pulse, or heart rate. Pulse oximeters use a painless Band-Aid-like probe that attaches to a fingertip, toe, hand or foot and shines a light through the skin. The monitor then determines how much oxygen is in the blood and the pulse rate based on the way the light passes through the skin. The probe is connected to the monitor by a cable of varying lengths. Movement can cause incorrect pulse oximeter measurements.

Continuous pulse oximetry is considered a standard of care in pediatric intensive care units; however pediatric patients have a high incidence of false alarms in part due to motion artifact. These false alarms contribute to the 94% false alarm incidence in pediatric ICU patients. (Lawless) Motion artifact is potentially exacerbated by the long cables connecting the patients to the continuous monitors. Currently the FDA has not approved any completely cable-free pulse oximeters for continuous use; however Masimo has produced a device (the Radius 7) that overcomes the previously required cable connection to a wall unit via wireless transmission from a patient-worn continuous monitor. (Masimo) There are currently no studies looking at this patient worn monitor in relation to the motion artifacts that frequently cause false alarms in pediatric patients.

The Joint Commission has listed alarm fatigue as a top patient safety concern and as of January 1, 2016 has required hospitals to establish policies and procedures for managing alarm fatigue. Reduction in false alarms is necessary to reduce alarm fatigue. This pilot study will use clinical data to measure the effect of eliminating this cable-to-wall connection on motion artifact. Additionally, early mobility in critically ill patients has been shown to improve outcomes. Freeing patients from one of the cables currently tying them to the room may decrease the false alarms from motion artifact while simultaneously improving mobility, therefore improving patient outcomes and safety.

DESIGN AND METHODS Subjects: Healthy pediatric subjects ranging from 6 months to 17 years of age. Pediatric subjects will be recruited via fliers placed in pediatric critical care areas, in the Translational Research Control Unit (TRU) and on the MCW research bulletin board outside the MCW office of research. Subjects may include those responding to the fliers directly as well as the healthy siblings of patients being evaluated or treated in the Children's Hospital of Wisconsin Emergency Room (EDTC), Clinics or Hospital.

Setting: The Children's Hospital of Wisconsin (CHW): primarily in the TRU designed to mimic a CHW outpatient clinic room; and also including available sites near the EDTC or clinic to facilitate making testing more convenient for the subject's family if the subject is a healthy sibling of a patient being evaluated in the EDTC or clinic.

Procedures:

While being continuously monitored on the current equipment used by CHW (the wall-connected Radical 7) and the partially wireless Radius 7 pulse oximeter, subjects will be asked to complete age appropriate activities as developed by CHW physical therapists, for approximately twenty minutes. These activities include common childhood activities such as kicking a ball or drawing with crayons. If a subject declines to participate in an activity, an alternate will be suggested. The investigators will analyze the pulse oximetry data for instances of alarm state corresponding to loss of signal integrity as defined by complete inability to pick up the subject signal or a change in the pulse oximetry saturation value by 4%. (Seixas, Seixas and Pereira)

Endpoints: Enrollment of and data acquisition on 48 pediatric subjects: 12 in each of the following developmental age groups: 6 month-1year, 1-6 years, 6-10 years, and 11-17 years.

Data analysis:

All data will be collected in a de-identified manner in which subjects will be given a study case number and all data will be collected under that case number.

The number of episodes of loss of signal integrity leading to an alarm state generated by the wall-connected Radical 7 unit compared to the partially wireless Radius 7 unit during completion of the developmentally appropriate activities will be compared. Analysis will be performed by the Quantitative Health Sciences department at MCW and a 20% reduction in episodes of loss of signal integrity will be considered clinically significant.

TOTAL NUMBER OF HUMAN RESEARCH PARTICIPANTS PROPOSED FOR THIS STUDY. 48 pediatric subjects: 12 in each of the following developmental age groups: 6 months-1year, 1-6 years, 6-10 years, and 11-17 years. This number of participants will provide representative data for a pilot study investigating the most common age groups seen in the pediatric ICU in order to identify the age groups most likely to benefit from further study.

Conditions
Healthy
Being tested
Radical 7, Radius 7
Keywords
Pediatric, Pulse Oximeter, Wireless, Motion, False Alarm

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Radical 7Device

    Also called Masimo Radical 7 Pulse Oximeter

    The Radical 7 is the traditional wall-connected pulse oximeter currently used at Children's Hospital of Wisconsin (CHW) which requires a 12 foot cable connecting the patient to a wall unit. While being continuously monitored with the Radical 7 pulse oximeter, subjects will be asked to complete age appropriate activities. The pulse oximetry data will be analyzed for instances of alarm state corresponding to loss of signal integrity as defined by complete inability to pick up the subject signal or a change in the pulse oximetry saturation value by 4%.

  • Radius 7Device

    Also called Masimo Radius 7 Pulse Oximeter

    The Radius 7 is a newer, partially wireless pulse oximeter which eliminates the cable connecting the patient to a wall unit by instead connecting to a small device worn on the patient's arm or leg. The small device then sends the continuous pulse oximeter data wirelessly to the central monitoring station. While being continuously monitored with the Radius 7 pulse oximeter, subjects will be asked to complete age appropriate activities. The pulse oximetry data will be analyzed for instances of alarm state corresponding to loss of signal integrity as defined by complete inability to pick up the subject signal or a change in the pulse oximetry saturation value by 4%.

Study groups in NCT03187665
Study groupRoleWhat they receive
1-6 years12 subjects in the age range of 1-6 years old.
Receives: Device: Radical 7; Device: Radius 7
6-10 years12 subjects in the age range of 6-10 years old.
Receives: Device: Radical 7; Device: Radius 7
11-17 years12 subjects in the age range of 11-17 years old.
Receives: Device: Radical 7; Device: Radius 7

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. instances of alarm state corresponding to loss of signal integrity

    Loss of signal integrity is defined as complete inability to pick up the patient signal or a change in the pulse oximetry saturation value by 4%.

    Measured: 20 minutes

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted15 Jun 2017
  2. Study start26 Jul 2017
  3. Primary completion1 Nov 2017
  4. Study completion1 Nov 2017

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Observational
Observational model
Case only
Time perspective
Prospective
Enrolment
36 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03187665
Sponsor's protocol ID
953341-1
Status last verified
Nov 2017
First submitted
13 Jun 2017
Last updated
24 Nov 2017
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
Yes: At the end of the study, Masimo, who makes both of the pulse oximeters, will be able to see de-identified information. They will not be allowed to see "Protected Health Information" (PHI). IPD that will be shared includes Subject Number, Age, Race, Gender, Probe Location (dominant vs. nondominant hand), Activity (Likert score), Start time of activity, Stop time of activity, Radius 7 Probe location, Radical 7 Probe location, Instance of alarm state Radius 7, Duration of alarm state Radius 7, Loss of signal integrity Radius 7-Change in pulse ox saturation of 4% or more, Loss of signal integrity Radius 7-no signal, Instance of alarm state Radical 7, Duration of alarm state Radical 7, Loss of signal integrity Radical 7-Change in pulse ox saturation of 4% or more, and Loss of signal integrity Radical 7-no signal. This de-identified data with no PHI will be shared via USB stick or e-mail electronically.

Publications

  • Lawless ST. Crying wolf: false alarms in a pediatric intensive care unit. Crit Care Med. 1994 Jun;22(6):981-5.

    PubMed 8205831
  • Seixas DM, Seixas DM, Pereira MC, Moreira MM, Paschoal IA. Oxygen desaturation in healthy subjects undergoing the incremental shuttle walk test. J Bras Pneumol. 2013 Jun-Aug;39(4):440-6. doi: 10.1590/S1806-37132013000400007.

    PubMed 24068265

Questions about this trial

What is this clinical trial studying?

An observational study of Healthy, run by Pamela Petersen at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 1 inclusion and 1 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 24 Nov 2017. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

instances of alarm state corresponding to loss of signal integrity (time frame: 20 minutes). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT03187665, which also lists 2 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 24 Nov 2017.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in