Intubation; Difficult or FailedNCT03247803

Air-Q® SP Versus Williams Intubating Airway for Single-Operator FOI

A study testing air-Q® SP and Williams Airway Intubator (WA) in people with Intubation; Difficult or Failed and Laryngeal Masks, run by University of Wisconsin, Madison at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The air-Q and air-Q SP are commercially-available supraglottic airways (SGAs) that can be used for primary airway maintenance or as a conduit for FOI, as can be the Williams, Ovassapian, and Berman Intubating Airways. Read more about this study

Who is it for?
Ages 18 years and older
What is being tested?
air-Q® SP and Williams Airway Intubator (WA)
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
60 participants
Where
1 siteUnited States
Study dates
11 Jun 2018to 8 Mar 2019
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Adult patients (≥18 years);
  • The patient is scheduled for an elective surgery or procedure that will take place in the UWHC main operating room, outpatient surgical center, or ambulatory procedure center; and
  • The patient's primary anesthesia care team has planned for general anesthesia with orotracheal intubation for airway maintenance.

Who cannot join

  • The patient is < 18 years of age;
  • The patient is non-English speaking;
  • The patient is known or believed to be pregnant;
  • The patient is a prisoner;
  • The patient has impaired decision-making capacity;
  • The patient is at increased risk for aspiration, including non-fasted or emergency surgery, and cases of uncontrolled gastroesophageal reflux disease, hiatal hernia, gastroparesis, esophageal dysmotility, prior esophagectomy, and/or emesis within twenty-four hours of the surgery or procedure;
Show all 11Show fewer
  • The patient has limited mouth opening or oropharyngeal anatomy making successful placement of the study airway conduit unlikely;
  • The patient has airway exam features, a medical condition, or a past airway management history, which prompts the attending anesthesiologist to plan to maintain spontaneous ventilation during tracheal intubation;
  • The patient pre-operatively is requiring supplemental oxygen;
  • The patient has moderate to severe cardiac disease, as evidenced by a left ventricular ejection fraction of ≤ 50% on the most recent echocardiogram, severe aortic valve stenosis, atrial fibrillation with a resting pre-operative heart rate of ≥ 90 beats per minute, and/or known, multi-vessel coronary artery, managed conservatively with medical management or by intervention with multiple coronary artery stents or coronary bypass grafting; and
  • The patient has an allergy to glycopyrrolate.

Trial locations and contacts

1 site in 1 country.

United States1 site
  • University of Wisconsin Hopital and Clinics

    Madison, Wisconsin, 53792

Other trials in United States

About this study

Official title: Comparison of Use of a Shortened Air-Q® Self-Pressurizing Intubating Laryngeal Airway Versus the Williams Intubating Airway for Single-Operator Flexible Bronchoscopic Intubation.

The air-Q and air-Q SP are commercially-available supraglottic airways (SGAs) that can be used for primary airway maintenance or as a conduit for FOI, as can be the Williams, Ovassapian, and Berman Intubating Airways. The air-Q products have the same overall geometrical design, but differ in their cuff inflating mechanisms, where the air-Q uses a traditional pilot balloon valve, while the air-Q SP utilizes a self-inflating cuff that dynamically adjusts according to the relative resistances to airflow between the patient's lungs and the device's cuff. The purpose of this study is to compare the single-operator intubation success using the air-Q SP versus the Williams Intubating Airway as a conduit for flexible bronchoscopic intubation (FBI) and to compare the single-operator intubation success using the air-Q SP versus the Williams Intubating Airway as a conduit for flexible bronchoscopic intubation. Both devices are commercially-available and being used in the study in accordance with their labeled intended use. The investigator envisage that the endpoint of the study will be single operator intubation success rate.

Read the detailed description

Despite the introduction of videolaryngoscopes, flexible bronchoscopic intubation (FBI), also known as fiberoptic intubation (FOI), remains a critical method for achieving tracheal intubation in patients with difficult airways. Unfortunately, however, FOI often requires two-operators for success. One operator is required to manipulate the flexible bronchoscope through the patient's upper airway and into their trachea, while the second operator applies one or more maneuvers (e.g., jaw thrust, anterior jaw lift, or lingual retraction) with or without the use of a commercially-available fiberoptic intubating airway.

Previous studies have demonstrated the effectiveness of the currently available FIAs, but also illustrate the frequent need for an assistant for successful FOI. In this study, an assistant performed a jaw lift and a chin lift in all subjects as well.

The air-Q and air-Q SP are commercially-available supraglottic airways (SGAs) that can be used for primary airway maintenance or as a conduit for FOI, as can be the Williams, Ovassapian, and Berman Intubating Airways. The air-Q products have the same overall geometrical design, but differ in their cuff inflating mechanisms, where the air-Q uses a traditional pilot balloon valve, while the air-Q SP utilizes a self-inflating cuff that dynamically adjusts according to the relative resistances to airflow between the patient's lungs and the device's cuff. In prior trials comparing the air-Q and air-Q SP with the two other commercially-available SGAs, the air-Q and air-Q SP bowl design provided a partial or full view of the vocal cords in 80-92% of subjects without requiring any additional airway maneuvers applied by an assistant (e.g., jaw thrust, anterior jaw lift, or lingual retraction) (Galgon, 2011; Galgon, 2015). Presumably, in these situations, successful tracheal intubation using the air-Q or air-Q SP could have been achieved by a single operator versus requiring one or more assistants. However, this potential performance advantage has not yet been tested. Therefore, the purpose of this study is to compare use of the air-Q SP against the WA for FBI performed by a single operator.

The air-Q SP, rather than the air-Q, is chosen for this study because its lack of a pilot balloon valve allows its stem to be easily shortened to ease device removal over the coaxially placed tracheal tube. The WA is chosen for this study because it has performed best in past trials and is the airway routinely used by the study principal investigator (PI) for FBI. Both devices are commercially-available and being used in the study in accordance with their labeled intended use.

Being tested
air-Q® SP, Williams Airway Intubator (WA)
Keywords
Air-Q® SP, Williams Airway Intubator (WA), Flexible Bronchoscopic intubation (FBI), Fiberoptic Intubation (FOI), Fiberoptic Intubating Airway(FIA), Laryngeal Mask Airway

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • air-Q® SPDevice

    Also called air-Q SP placement for airway maintenance

    air-Q SP placement for airway maintenance,

  • Williams Airway Intubator (WA)Device

    Also called WA is placed in subject's oropharynx.

    WA is placed in subject's oropharynx. If the vocal cords are not in view at the end of the WA, the operator may ask an assistant to perform an airway maneuver (e.g., jaw thrust, anterior jaw lift, or lingual traction) to attempt to bring the vocal cords into view

Study groups in NCT03247803
Study groupRoleWhat they receive
Air-Q SPExperimentalair-Q Self-Pressurizing intubating Laryngeal Airway, sizes 3.5 and 4.5 (Mecury Medical, Clearwater, FL, USA)
Receives: Device: air-Q® SP
Williams Intubating AirwayExperimentalAirway Intubator, Williams, Adult Female, Single Use, Molded Surlyn Plastic, 9 cm or Airway Intubator, Williams, Adult Male, Single Use, Molded Surlyn Plastic, 10 cm
Receives: Device: Williams Airway Intubator (WA)

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. The primary endpoint is single-operator intubation success.

    The primary purpose of this study is to compare the single-operator intubation success using the air-Q SP versus the Williams Intubating Airway as a conduit for flexible bronchoscopic intubation.

    Measured: 15 minutes

Secondary outcomes (10)
  1. Time required for conduit placement

    The secondary purposes of this study are to compare device performance characteristics and patient responses to use of the air-Q SP versus the Williams Intubating Airway as a conduit for flexible bronchoscopic intubation

    Measured: 15 minutes

  2. Time required for each intubation attempt

    The secondary purposes of this study are to compare device performance characteristics and patient responses to use of the air-Q SP versus the Williams Intubating Airway as a conduit for flexible bronchoscopic intubation

    Measured: 15 minutes

  3. Overall intubation time

    The secondary purposes of this study are to compare device performance characteristics and patient responses to use of the air-Q SP versus the Williams Intubating Airway as a conduit for flexible bronchoscopic intubation

    Measured: 15 minutes

  4. Number of attempts required for conduit placement

    The secondary purposes of this study are to compare device performance characteristics and patient responses to use of the air-Q SP versus the Williams Intubating Airway as a conduit for flexible bronchoscopic intubation

    Measured: 15 minutes

  5. Number of attempts for successful intubation

    The secondary purposes of this study are to compare device performance characteristics and patient responses to use of the air-Q SP versus the Williams Intubating Airway as a conduit for flexible bronchoscopic intubation

    Measured: 15 minutes

  6. Number and types of assistive airway maneuvers (e.g., jaw thrust, lingual retraction, etc.) for successful intubation

    The secondary purposes of this study are to compare device performance characteristics and patient responses to use of the air-Q SP versus the Williams Intubating Airway as a conduit for flexible bronchoscopic intubation

    Measured: 15 minutes

  7. Successful intubation method

    The secondary purposes of this study are to compare device performance characteristics and patient responses to use of the air-Q SP versus the Williams Intubating Airway as a conduit for flexible bronchoscopic intubation

    Measured: 15 minutes

  8. Tracheal tube dislodgment rate during conduit removal

    The secondary purposes of this study are to compare device performance characteristics and patient responses to use of the air-Q SP versus the Williams Intubating Airway as a conduit for flexible bronchoscopic intubation

    Measured: 15 minutes

  9. Oxygen desaturation rate

    The secondary purposes of this study are to compare device performance characteristics and patient responses to use of the air-Q SP versus the Williams Intubating Airway as a conduit for flexible bronchoscopic intubation

    Measured: 15 minutes

  10. Post operative sore throat rate.

    The secondary purposes of this study are to compare device performance characteristics and patient responses to use of the air-Q SP versus the Williams Intubating Airway as a conduit for flexible bronchoscopic intubation

    Measured: 4 hours

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted14 Aug 2017
  2. Study start11 Jun 2018
  3. Primary completion8 Mar 2019
  4. Study completion8 Mar 2019

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Parallel groups
Model description
Adult patients presenting for a surgical or interventional procedure requiring general anesthesia and tracheal intubation, who do not meet any of the study exclusion criteria, and who give their informed consent, will be randomly allocated to receive either an air-Q SP or Williams Intubating Airway as a conduit to aid flexible bronchoscopic intubation.
Masking
Double blind
Who is masked
Participant, Outcomes assessor
Enrolment
60 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03247803
Sponsor's protocol ID
2017-0542
Status last verified
Mar 2019
First submitted
28 Jul 2017
Last updated
22 Mar 2019
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No: There is no plan to make individual participant data available to other researchers.

Publications

  • Galgon RE, Schroeder KM, Han S, Andrei A, Joffe AM. The air-Q((R)) intubating laryngeal airway vs the LMA-ProSeal(TM) : a prospective, randomised trial of airway seal pressure. Anaesthesia. 2011 Dec;66(12):1093-100. doi: 10.1111/j.1365-2044.2011.06863.x. Epub 2011 Aug 22.

    PubMed 21880031

Questions about this trial

What is this clinical trial studying?

A study testing air-Q® SP and Williams Airway Intubator (WA) in people with Intubation; Difficult or Failed and Laryngeal Masks, run by University of Wisconsin, Madison at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 11 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 22 Mar 2019. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

The primary endpoint is single-operator intubation success. (time frame: 15 minutes). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT03247803, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 22 Mar 2019.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in