AzTrials
AzTrials

Healthy NormalsNCT03257189

Evaluation of Performance, Usability, and Reliability of a Novel Device for Continuous Collection of Physiological Data

A study testing Physiological signal monitor, Heart rate and heart rate variability comparison device and Respiration rate comparison device in people with Healthy Normals, run by MC10 Inc. at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Study objective: The primary objective of the clinical investigation is to evaluate the accuracy of BioStamp nPoint system algorithm measurements. Study design: A single-site, non-significant risk, open-label, prospective non-randomized clinical investigation designed to validate the accuracy of the various… Read more about this study

Who is it for?
Ages 18 to 99 yearsHealthy volunteers can take part
Conditions the study is for
Healthy Normals
What is being tested?
Physiological signal monitor, Heart rate and heart rate variability comparison device and Respiration rate comparison device
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
30 participants
Where
1 siteUnited States
Study dates
31 Jul 2017to 30 Aug 2017
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 99 years
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Male or female volunteers, at least 18 years of age at the time of screening visit;
  • Fluent in English;
  • The subject is willing to comply with the protocol specified evaluations;
  • Subject is willing and cognitively able to sign informed consent

Who cannot join

  • Pregnancy;
  • Positive urine drug screen
  • History of active (clinically significant) skin disorders;
  • History of allergic response to silicones or adhesives;
  • Subjects with electronic implants of any kind (e.g. pacemaker)
  • History of sleep disorders or self-reported insomnia or other sleep conditions;
Show all 9Show fewer
  • Broken, damaged or irritated skin or rashes near the sensor application sites;
  • Subjects that are MC10 employees or shareholders, or a spouse or child of an MC10 employee or shareholder;
  • Subjects who are physically or cognitively unable to normally perform activities of daily living, assessed at the discretion of the investigator.

Trial locations and contacts

1 site in 1 country.

United States1 site
  • CTI Clinical Research Center

    Cincinnati, Ohio, 45212

Other trials in United States

About this study

Official title: A Study to Evaluate the Performance, Usability, and Reliability of a Novel Device for Continuous Collection of Physiological Data in Healthcare and Remote Settings

Study objective: The primary objective of the clinical investigation is to evaluate the accuracy of BioStamp nPoint system algorithm measurements.

Study design: A single-site, non-significant risk, open-label, prospective non-randomized clinical investigation designed to validate the accuracy of the various physiological parameters that the Wearable Sensor Patches acquire and the system processes.

Conditions
Healthy Normals
Being tested
Physiological signal monitor, Heart rate and heart rate variability comparison device, Respiration rate comparison device, Activity classification

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Physiological signal monitorDevice

    Wireless remote monitoring system intended for use by researchers and healthcare professionals for continuous collection of physiological data in home and healthcare settings.

  • Heart rate and heart rate variability comparison deviceDevice

    FDA cleared reference device that monitors heart rate and heart rate variability in subjects

  • Respiration rate comparison deviceDevice

    FDA cleared reference device that monitors respiration rate in subjects

  • Activity classificationOther

    Visual annotation of subject posture and other activities used for reference

Study groups in NCT03257189
Study groupRoleWhat they receive
Single ArmOtherThis single arm consists of all subjects which will interact with the device under investigation as well as comparator devices. This includes the Physiological signal monitor intervention, Heart rate and heart rate variability comparison device intervention, Respiration rate comparison device intervention, and Activity classification intervention.
Receives: Device: Physiological signal monitor; Device: Heart rate and heart rate variability comparison device; Device: Respiration rate comparison device; Other: Activity classification

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Accuracy of Heart Rate Measured in Beats Per Minute Compared to Reference Device

    Heart rate as measured by the subject device in beats per minute will be compared to the reference device. The Mean Absolute Error (MAE) between measurements from each device is presented.

    Measured: 2 days after informed consent

  2. Accuracy of Heart Rate Variability Measured by Root Mean Square of Successive Differences (HRV RMSSD) Compared to the Reference Device

    Heart Rate Variability RMSSD is measured by the subject device determined by root mean square of successive differences in neighboring RR intervals and will be compared to the reference device. The Mean Absolute Error (MAE) between measurements from each device is presented.

    Measured: 2 days after informed consent

  3. Accuracy of Heart Rate Variability Measured by Low Frequency Content to High Frequency Content Ratio (HRV Ratio) Compared to the Reference Device

    HRV Ratio as measured by the subject device determined by low frequency content to high frequency content ratio (HRV Ratio) in beats per minute will be compared to the reference device. The Mean Absolute Error (MAE) between measurements from each device is presented.

    Measured: 2 days after informed consent

  4. Accuracy of Respiration Rate Measured in Breaths Per Minute Compared to the Reference Device

    Respiration rate as measured by the subject device in breaths per minute will be compared to the reference device. The Mean Absolute Error (MAE) between measurements from each device is presented.

    Measured: 2 days after informed consent

  5. Accuracy of Activity Classification as Compared to Visual Annotation

    The device under test will classify a subjects activity into sleeping, standing, sitting, lying, walking and other activities as the subject performs various predefined activities, which are annotated by an observer. The device's Activity Classification will be compared to the observer's annotation. The percentage of correct classifications by the device against indicated visual observation is presented. The percent correct is not a per subject average, but is the percentage of all activity classifications which were correct when compared to ground truth observation.

    Measured: 2 days after informed consent

  6. Accuracy of Step Count Compared to an Observer's Manual Count

    The number of steps reported by the device under test during a 6 minute walk test will be compared to an observer's manual count of the number of steps taken.

    Measured: 2 days after informed consent

  7. Accuracy of Sleep Onset Time (Hours, Minutes, and Seconds), Sleep Wake Time (Hours, Minutes, and Seconds) as Compared to an Observer's Visual Annotation

    Sleep onset and wake times as reported by the device under test will be compared to an observer's visual annotation of the sleep onset and wake times. The Mean Absolute Error (MAE) between times as indicated by the device and observed times is presented.

    Measured: 2 days after informed consent

  8. Accuracy of Posture Classification as Compared to Visual Annotation

    The device under test will classify a subjects activity into sleep posture, standing posture, and sitting posture as the subject performs various predefined activities, which are annotated by an observer. The device's Posture Classification will be compared to the observer's annotation. The percentage of correct classifications by the device against indicated visual observation is presented. The percent correct is not a per subject average, but is the percentage of all posture classifications which were correct when compared to ground truth observation.

    Measured: 2 days after informed consent

  9. Number of Sensors That Remained Sufficiently Adhered to Subjects for 24 Hours as Assessed by a 5 Point Scale

    The level of sensor adhesion will be assessed by a 5 point scale (0- sensor is greater than or equal to 90% of sensor adhered, 4-sensor completely detached from subject). Sensors receiving scores of 0 and 1 will be determined to have acceptable adhesion, while scores of 2, 3, and 4 will be considered unacceptable adhesion. The percentage of sensors which had acceptable adhesion at sensor removal is presented.

    Measured: 2 days after informed consent

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted22 Aug 2017
  2. Study start31 Jul 2017
  3. Primary completion25 Aug 2017
  4. Study completion30 Aug 2017

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
30 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03257189
Sponsor's protocol ID
MC10-PTL-103
Status last verified
Jan 2019
First submitted
16 Aug 2017
Last updated
25 Jan 2019
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing Physiological signal monitor, Heart rate and heart rate variability comparison device and Respiration rate comparison device in people with Healthy Normals, run by MC10 Inc. at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 9 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 25 Jan 2019. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 25 Jan 2019). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Accuracy of Heart Rate Measured in Beats Per Minute Compared to Reference Device (time frame: 2 days after informed consent). The study lists 9 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 25 Jan 2019.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in