Respiratory Distress Syndrome in Premature InfantE-FlowNCT03306524
The Role of Circuit Flow During Mechanical Ventilation of Neonates
A study testing SIPPV_VG_0_08, SIPPV_VG_0_16 and SIPPV_VG_0_24 in people with Respiratory Distress Syndrome in Premature Infant, run by Cambridge University Hospitals NHS Foundation Trust at 1 site in United Kingdom.
Trial summary
During neonatal mechanical ventilation inflating pressures, tidal volumes, and inflation and expiration times need to be set and adjusted to optimise oxygenation and carbon dioxide removal. Read more about this study
- Who is it for?
- Ages 1 day to 2 months
- Conditions the study is for
- Respiratory Distress Syndrome in Premature Infant
- What is being tested?
- SIPPV_VG_0_08, SIPPV_VG_0_16 and SIPPV_VG_0_24
- Type of treatment
- Medical device
- Could I receive a placebo?
- No, everyone receives an active treatment
- How many people
- 12 participants
- Where
- 1 siteUnited Kingdom
- Study dates
- 1 Jul 2017to 30 Jun 2018
- Phase
- Interventional
- Are there results?
- None posted in the registry
Key requirements
Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.
- Ages 1 day to 2 months
- All sexes
- No healthy volunteers
- Child
Who may be able to join
- Birth weight < 2 kg;
- Ventilated with SIPPV-VG modes,
- Informed parental consent,
- Clinician assent.
Who cannot join
- Baby's respiratory condition unstable (Inspired oxygen (FiO2) > 50%, PaCO2 > 8.5kPa or <5kPa in the last 12 hours)
- Extubation planned in the next 12 hours;
- Neonatal or surgical procedure in the last 12 hours or planned in the next 12 hours;
- Significant pneumothorax requiring drainage;
- Gas leak around the endotracheal tube >50%; #
- No arterial access;
- The responsible clinician does not agree with recruitment;
- Parents do not consent.
Trial locations and contacts
1 site in 1 country.
United Kingdom1 site
Neonatal Intensive Care Unit, Cambridge University Hospitals NHS Trust
Cambridge, CB20QQ
About this study
Official title: Investigating the Effect of Altering the Slope of the Rise in Pressure During Mechanical Ventilation of Preterm Babies
During neonatal mechanical ventilation inflating pressures, tidal volumes, and inflation and expiration times need to be set and adjusted to optimise oxygenation and carbon dioxide removal. The flow of gas into the ventilator circuit has a big effect on ventilation but is usually set to a constant value (~8 L/min) for all babies regardless of size or severity of illness, based on minimal research. High circuit flow may lead to lung damage and low flow to inadequate ventilation. The investigators recently developed a unique system to capture, record, analyse, and display ventilator data at high resolution over long periods. Using this the investigators will investigate, in within patient cross-over studies, how the level of gas flow affects ventilator parameters and ventilation, in two commonly used ventilation modes. The results will determine the lowest circuit flow that ventilates a baby safely and effectively. It will also provide preliminary data for a randomised trial.
Read the detailed description
- Background of the project
About 1.5% of newborns require mechanical ventilation. Although mechanical ventilation can be life-saving for neonates with respiratory failure, it may cause lung injury due to excess airway pressure (barotrauma), delivery of high tidal volumes (volutrauma) and repetitive closing/re-opening of lung units (atelectotrauma)1. Very preterm neonates are especially vulnerable to ventilator induced lung injury. Ventilator associated lung injury is one of several factors contributing to the burden of chronic lung disease of infancy, also called bronchopulmonary dysplasia (BPD).
Ventilators have several different modes. The most frequently used mode is, time-cycled, pressure limited, where the doctor sets inspired oxygen level, the peak inflation pressure (PIP), rate, and inflation time of the ventilator. In this mode the inflations are usually synchronised with the baby's breathing. This is either Synchronised Intermittent Positive Pressure Ventilation (SIPPV), where the ventilator synchronises inflations with all the baby's breaths, or Synchronised Intermittent Mandatory Ventilation (SIMV), where the ventilator only synchronises with a set number of breaths. These can be delivered with or without targeting the tidal volume delivered to a set value by adjusting the peak inflating pressure. This mode is called volume guarantee (VG). A similar mode of ventilation is flow-cycled or Pressure Support Ventilation (PSV) and can be combined with tidal volume targeting 2. In this mode the inflation is terminated as soon as the flow rate falls to 15% of the maximum flow during inflation.
These modes have continuous gas flow through the ventilation circuit. Inflation starts and pressure rises when the expiratory valve is closed. The rate of circuit flow alters the ventilation waveforms. With time-cycled ventilation the higher the circuit flow the faster the pressure rises, the lung is distended and potentially injured, and the earlier the set PIP is achieved. A high flow, with a relatively long inflation time will result in a "pressure plateau", that is the PIP is sustained after the flow, into the baby, has stopped because the lung volume is maximal at the PIP used (Figure 1). In PSV inflation stops when the flow has decreased to ~15% of the peak inflation flow and so the inflation time is shorter. Increasing the device flow shortens the inflation time and thereby reduces the mean airway pressure.
Despite the effects on ventilation patterns and the speed of lung distension and injury, consideration has rarely been given to the circuit flow. By protocol, it is usually set at 7-10 L/min when the ventilator is turned on and it is not changed.
This is relatively high, and usually produces a "square" pressure waveform with a rapid distention of the lungs and a sustained pressure plateau. (see A in figure 1) With PSV a high flow results in a relatively short inflation time (~0.2 sec; it needs to be at least 0.3 sec for adequate lung aeration). There is no evidence these high flows are best for optimum ventilation and minimum lung damage. In a preterm lamb model there were no adverse effects on gas exchange or cardiovascular parameters until the flow was reduced to 3 l/min 3. In animal studies ventilation with high flow resulted in histologic and molecular changes of lung injury 4. The effect of lowering the ventilator circuit flow rate has never been investigated in clinical studies.
The Dräger Babylog VN500 ventilator has an alternative to setting a gas flow: the user can set the slope time instead, that is time required to reach the set pressure. In Cambridge it is invariably set to 0.08 sec which results in a flow ~7 L/min. As we use an inflation time between 0.33 - 0.45 sec, there is a pressure plateau lasting at least 0.25 sec and sustained inflation with little or no flow.
The investigators are the first to develop a unique system for downloading and analysing data from the Dräger VN500 neonatal ventilator. Using DataGrabber software obtained from Dräger Medical, the investigators can retrieve ventilator parameters at 100Hz frequency over long periods (hours and even days). To analyse and visualise the large datasets the investigators developed a data analysis workflow using the Python programing language and its add-on packages (Figure 2). With this tool the investigators can now study details of each inflation and spontaneous breath. The Investigators have recorded detailed data from 30 babies and shown it is feasible, accurate, and the investigators have the expertise (Belteki et al, submitted).
In this application the investigators propose to investigate the effect of different slope times (and therefore different levels of circuit flow) on ventilation parameters and gas exchange in preterm infants. The investigators hypothesise that a longer slope time (= lower circuit flow) will distend the lungs more gently yet maintain ventilation and gas exchange.
- Intervention:
The study is a within patient crossover design comparing short periods of ventilation with different slope times both in SIPPV-VG and PSV-VG modes with the following interventions:
A ventilator download tool and transcutaneous and expired CO2 monitors is attached to the ventilator and data download commences while the baby is ventilated with the parameters used by the clinical team. An arterial blood gas is performed. The ventilated parameters are changed as shown below. The order of these epochs is randomised. Another arterial blood gas is performed. Ventilator is changed back to the original parameters (or different as appropriate by the blood gas). Ventilator data and CO2 recording will be continued for another 30 minutes.
Interventions
Duration Ventilator More Slope time Inspiratory time[max] 15 min SIPPV-VG 0.08 0.40 15 min PSV-VG 0.08 [0.60] 15 min PSV-VG 0.16 [0.60] 15 min SIPPV-VG 0.16 0.40 15 min SIPPV-VG 0.24 0.40 15 min PSV-VG 0.24 [0.60] 15 min PSV-VG 0.32 [0.60] 15 min SIPPV-VG 0.32 0.40 15 min SIPPV-VG 0.40 0.40 115 min PSV-VG 0.40 [0.60]
Total study duration is 220 minutes. A researcher will be present continuously. FiO2 will be adjusted if needed to maintain saturations between 90-95%. If the FiO2 rises >15% or the end-tidal CO2 rises >1.5kPa over the pre-study level, that intervention will be abandoned.
Comparison:
At each slope time, the following parameters will be determined and compared with those recorded at 0.08 sec slope time: (1) peak inflating pressure, (2) inflation duration, (3) duration of inflation plateau, (4) duration of no gas flow, (5) expired tidal and minute volumes (mandatory/spontaneous, inspiratory/expiratory, (6) ventilator rate, (7) FiO2, (8) transcutaneous and/or end-tidal CO2 and, (9) interaction between ventilator inflations and baby's breaths. Values for SIPPV and PSV will also be compared.
- Being tested
- SIPPV_VG_0_08, SIPPV_VG_0_16, SIPPV_VG_0_24, SIPPV_VG_0_32, SIPPV_VG_0_40
- Keywords
- mechanical ventilation, SIPPV, PSV, Device flow, Circuit flow, Slope time, neonatal intensive care
What participants receive
What participants receive. In a randomised study each person is assigned to one group by chance.
SIPPV_VG_0_08Device
Mechanical ventilation using SIPPV-VG ventilator mode with a slope time of 0.08 seconds, inspiratory time of 0.40 seconds for 15 minutes
SIPPV_VG_0_16Device
Mechanical ventilation using SIPPV-VG ventilator mode with a slope time of 0.16seconds, inspiratory time of 0.40 seconds for 15 minutes
SIPPV_VG_0_24Device
Mechanical ventilation using SIPPV-VG ventilator mode with a slope time of 0.24 seconds, inspiratory time of 0.40 seconds for 15 minutes
SIPPV_VG_0_32Device
Mechanical ventilation using SIPPV-VG ventilator mode with a slope time of 0.32 seconds, inspiratory time of 0.40 seconds for 15 minutes
SIPPV_VG_0_40Device
Mechanical ventilation using SIPPV-VG ventilator mode with a slope time of 0.40 seconds, inspiratory time of 0.40 seconds for 15 minutes
PSV_VG_0_08Device
Mechanical ventilation using PSV-VG ventilator mode with a slope time of 0.08 seconds, maximum inspiratory time of 0.60 seconds for 15 minutes
PSV_VG_0_16Device
Mechanical ventilation using PSV-VG ventilator mode with a slope time of 0.16 seconds, maximum inspiratory time of 0.60 seconds for 15 minutes
PSV_VG_0_24Device
Mechanical ventilation using PSV-VG ventilator mode with a slope time of 0.24 seconds, maximum inspiratory time of 0.60 seconds for 15 minutes
PSV_VG_0_32Device
Mechanical ventilation using PSV-VG ventilator mode with a slope time of 0.32 seconds, maximum inspiratory time of 0.60 seconds for 15 minutes
PSV_VG_0_40Device
Mechanical ventilation using PSV-VG ventilator mode with a slope time of 0.40 seconds, maximum inspiratory time of 0.60 seconds for 15 minutes
| Study group | Role | What they receive |
|---|---|---|
| SIPPV_VG_0_08 | Experimental | SIPPV VG ventilation for 15 minutes slope time = 0.08 seconds inspiratory time = 0.40 seconds Receives: Device: SIPPV_VG_0_08 |
| PSV_VG_0_08 | Experimental | PSV VG ventilation for 15 minutes slope time = 0.08 seconds maximum inspiratory time = 0.60 seconds Receives: Device: PSV_VG_0_08 |
| SIPPV_VG_0_16 | Experimental | SIPPV VG ventilation for 15 minutes slope time = 0.16 seconds inspiratory time = 0.40 seconds Receives: Device: SIPPV_VG_0_16 |
| PSV_VG_0_16 | Experimental | PSV VG ventilation for 15 minutes slope time = 0.16 seconds maximum inspiratory time = 0.60 seconds Receives: Device: PSV_VG_0_16 |
| SIPPV_VG_0_24 | Experimental | SIPPV VG ventilation for 15 minutes slope time = 0.24 seconds inspiratory time = 0.40 seconds Receives: Device: SIPPV_VG_0_24 |
| PSV_VG_0_24 | Experimental | PSV VG ventilation for 15 minutes slope time = 0.24 seconds maximum inspiratory time = 0.60 seconds Receives: Device: PSV_VG_0_24 |
| SIPPV_VG_0_32 | Experimental | SIPPV VG ventilation for 15 minutes slope time = 0.32 seconds inspiratory time = 0.40 seconds Receives: Device: SIPPV_VG_0_32 |
| PSV_VG_0_32 | Experimental | PSV VG ventilation for 15 minutes slope time = 0.32 seconds maximum inspiratory time = 0.60 seconds Receives: Device: PSV_VG_0_32 |
| SIPPV_VG_0_40 | Experimental | SIPPV VG ventilation for 15 minutes slope time = 0.32 seconds inspiratory time = 0.40 seconds Receives: Device: SIPPV_VG_0_40 |
| PSV_VG_0_40 | Experimental | PSV VG ventilation for 15 minutes slope time = 0.40 seconds maximum inspiratory time = 0.60 seconds Receives: Device: PSV_VG_0_40 |
What is being measured
The primary outcome is the main question the study is designed to answer.
End tidal CO2 measurement
Primary outcome will be the difference in end-tidal CO2 concentration during the epochs with slope times of 0.40 and 0.08 sec.
Measured: One year
Timeline
From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.
- First posted11 Oct 2017
- Study start1 Jul 2017
- Primary completion30 Jun 2018
- Study completion30 Jun 2018
Resources
- Official record on ClinicalTrials.govThe registry entry this page is built from, with its full change history.
- Publications (4)Papers the sponsor has linked to this study.
- Other trials for Respiratory Distress Syndrome in Premature InfantStudies of the same condition, open and under way.
- What the trial phases meanHow a treatment moves from first tests in people to approval.
More details
For doctors and researchers: how the study is designed, who runs it, and its registry IDs.
Study design
- Study type
- Interventional
- Phase
- Phase N/A
- Primary purpose
- Treatment
- Allocation
- Randomised
- Intervention model
- Crossover
- Model description
- Ten different interventions (with changing slope time or ventilation mode) will be compared in a crossover design. Each intervention will be given for 15 minutes. For details of the intervention see study design details.
- Masking
- Single blind
- Who is masked
- Participant
- Enrolment
- 12 participants (actual)
Sponsor and investigators
- Lead sponsor
- Cambridge University Hospitals NHS Foundation Trust
- Sponsor type
- University, hospital or other
- Responsible party
- Principal investigator, Gusztav Belteki, Cambridge University Hospitals NHS Foundation Trust
- Investigators
- Gusztav Belteki, M.D., Ph.D., Principal investigator, Cambridge University Hospitals NHS Foundation Trust
IDs, dates and publications
- ClinicalTrials.gov ID
- NCT03306524
- Sponsor's protocol ID
- 16/28
- Acronym
- E-Flow
- Status last verified
- May 2019
- First submitted
- 25 Sep 2017
- Last updated
- 20 May 2019
- Results posted
- No
- Data monitoring committee
- No
- FDA-regulated drug
- No
- FDA-regulated device
- No
- Individual data sharing
- Undecided
Publications
Attar MA, Donn SM. Mechanisms of ventilator-induced lung injury in premature infants. Semin Neonatol. 2002 Oct;7(5):353-60. doi: 10.1053/siny.2002.0129.
PubMed 12464497Nafday SM, Green RS, Lin J, Brion LP, Ochshorn I, Holzman IR. Is there an advantage of using pressure support ventilation with volume guarantee in the initial management of premature infants with respiratory distress syndrome? A pilot study. J Perinatol. 2005 Mar;25(3):193-7. doi: 10.1038/sj.jp.7211233.
PubMed 15674409Bach KP, Kuschel CA, Oliver MH, Bloomfield FH. Ventilator gas flow rates affect inspiratory time and ventilator efficiency index in term lambs. Neonatology. 2009;96(4):259-64. doi: 10.1159/000220765. Epub 2009 May 27.
PubMed 19478530Bach KP, Kuschel CA, Hooper SB, Bertram J, McKnight S, Peachey SE, Zahra VA, Flecknoe SJ, Oliver MH, Wallace MJ, Bloomfield FH. High bias gas flows increase lung injury in the ventilated preterm lamb. PLoS One. 2012;7(10):e47044. doi: 10.1371/journal.pone.0047044. Epub 2012 Oct 8.
PubMed 23056572
Study at a glance
From the registry record, last updated by the sponsor on 20 May 2019. Recruitment status is as the sponsor reported it: confirm with the study team.
- Official title
- Investigating the Effect of Altering the Slope of the Rise in Pressure During Mechanical Ventilation of Preterm Babies
- ClinicalTrials.gov ID
- NCT03306524
- Acronym / sponsor ID
- E-Flow / 16/28
- Keywords
- mechanical ventilation, SIPPV, PSV, Device flow, Circuit flow, Slope time, neonatal intensive care
- Study type
- Interventional
- Recruitment status
- Completed
- Enrolment
- 12 participants (actual)
- Lead sponsor
- Cambridge University Hospitals NHS Foundation TrustUniversity, hospital or other
- Study start
- 1 Jul 2017 (actual)
- Primary completion
- 30 Jun 2018 (actual)
- Study completion
- 30 Jun 2018 (actual)
- First posted
- 11 Oct 2017 (actual)
- Results
- Not posted in the source record
Clinical summary
The research question as the sponsor registered it. This describes what the study plans to test, not what it found.
During neonatal mechanical ventilation inflating pressures, tidal volumes, and inflation and expiration times need to be set and adjusted to optimise oxygenation and carbon dioxide removal. The flow of gas into the ventilator circuit has a big effect on ventilation but is usually set to a constant value (~8 L/min) for all babies regardless of size or severity of illness, based on minimal research. High circuit flow may lead to lung damage and low flow to inadequate ventilation. The investigators recently developed a unique system to capture, record, analyse, and display ventilator data at high resolution over long periods. Using this the investigators will investigate, in within patient cross-over studies, how the level of gas flow affects ventilator parameters and ventilation, in two commonly used ventilation modes. The results will determine the lowest circuit flow that ventilates a baby safely and effectively. It will also provide preliminary data for a randomised trial.
Detailed description
- Background of the project
About 1.5% of newborns require mechanical ventilation. Although mechanical ventilation can be life-saving for neonates with respiratory failure, it may cause lung injury due to excess airway pressure (barotrauma), delivery of high tidal volumes (volutrauma) and repetitive closing/re-opening of lung units (atelectotrauma)1. Very preterm neonates are especially vulnerable to ventilator induced lung injury. Ventilator associated lung injury is one of several factors contributing to the burden of chronic lung disease of infancy, also called bronchopulmonary dysplasia (BPD).
Ventilators have several different modes. The most frequently used mode is, time-cycled, pressure limited, where the doctor sets inspired oxygen level, the peak inflation pressure (PIP), rate, and inflation time of the ventilator. In this mode the inflations are usually synchronised with the baby's breathing. This is either Synchronised Intermittent Positive Pressure Ventilation (SIPPV), where the ventilator synchronises inflations with all the baby's breaths, or Synchronised Intermittent Mandatory Ventilation (SIMV), where the ventilator only synchronises with a set number of breaths. These can be delivered with or without targeting the tidal volume delivered to a set value by adjusting the peak inflating pressure. This mode is called volume guarantee (VG). A similar mode of ventilation is flow-cycled or Pressure Support Ventilation (PSV) and can be combined with tidal volume targeting 2. In this mode the inflation is terminated as soon as the flow rate falls to 15% of the maximum flow during inflation.
These modes have continuous gas flow through the ventilation circuit. Inflation starts and pressure rises when the expiratory valve is closed. The rate of circuit flow alters the ventilation waveforms. With time-cycled ventilation the higher the circuit flow the faster the pressure rises, the lung is distended and potentially injured, and the earlier the set PIP is achieved. A high flow, with a relatively long inflation time will result in a "pressure plateau", that is the PIP is sustained after the flow, into the baby, has stopped because the lung volume is maximal at the PIP used (Figure 1). In PSV inflation stops when the flow has decreased to ~15% of the peak inflation flow and so the inflation time is shorter. Increasing the device flow shortens the inflation time and thereby reduces the mean airway pressure.
Despite the effects on ventilation patterns and the speed of lung distension and injury, consideration has rarely been given to the circuit flow. By protocol, it is usually set at 7-10 L/min when the ventilator is turned on and it is not changed.
This is relatively high, and usually produces a "square" pressure waveform with a rapid distention of the lungs and a sustained pressure plateau. (see A in figure 1) With PSV a high flow results in a relatively short inflation time (~0.2 sec; it needs to be at least 0.3 sec for adequate lung aeration). There is no evidence these high flows are best for optimum ventilation and minimum lung damage. In a preterm lamb model there were no adverse effects on gas exchange or cardiovascular parameters until the flow was reduced to 3 l/min 3. In animal studies ventilation with high flow resulted in histologic and molecular changes of lung injury 4. The effect of lowering the ventilator circuit flow rate has never been investigated in clinical studies.
The Dräger Babylog VN500 ventilator has an alternative to setting a gas flow: the user can set the slope time instead, that is time required to reach the set pressure. In Cambridge it is invariably set to 0.08 sec which results in a flow ~7 L/min. As we use an inflation time between 0.33 - 0.45 sec, there is a pressure plateau lasting at least 0.25 sec and sustained inflation with little or no flow.
The investigators are the first to develop a unique system for downloading and analysing data from the Dräger VN500 neonatal ventilator. Using DataGrabber software obtained from Dräger Medical, the investigators can retrieve ventilator parameters at 100Hz frequency over long periods (hours and even days). To analyse and visualise the large datasets the investigators developed a data analysis workflow using the Python programing language and its add-on packages (Figure 2). With this tool the investigators can now study details of each inflation and spontaneous breath. The Investigators have recorded detailed data from 30 babies and shown it is feasible, accurate, and the investigators have the expertise (Belteki et al, submitted).
In this application the investigators propose to investigate the effect of different slope times (and therefore different levels of circuit flow) on ventilation parameters and gas exchange in preterm infants. The investigators hypothesise that a longer slope time (= lower circuit flow) will distend the lungs more gently yet maintain ventilation and gas exchange.
- Intervention:
The study is a within patient crossover design comparing short periods of ventilation with different slope times both in SIPPV-VG and PSV-VG modes with the following interventions:
A ventilator download tool and transcutaneous and expired CO2 monitors is attached to the ventilator and data download commences while the baby is ventilated with the parameters used by the clinical team. An arterial blood gas is performed. The ventilated parameters are changed as shown below. The order of these epochs is randomised. Another arterial blood gas is performed. Ventilator is changed back to the original parameters (or different as appropriate by the blood gas). Ventilator data and CO2 recording will be continued for another 30 minutes.
Interventions
Duration Ventilator More Slope time Inspiratory time[max] 15 min SIPPV-VG 0.08 0.40 15 min PSV-VG 0.08 [0.60] 15 min PSV-VG 0.16 [0.60] 15 min SIPPV-VG 0.16 0.40 15 min SIPPV-VG 0.24 0.40 15 min PSV-VG 0.24 [0.60] 15 min PSV-VG 0.32 [0.60] 15 min SIPPV-VG 0.32 0.40 15 min SIPPV-VG 0.40 0.40 115 min PSV-VG 0.40 [0.60]
Total study duration is 220 minutes. A researcher will be present continuously. FiO2 will be adjusted if needed to maintain saturations between 90-95%. If the FiO2 rises >15% or the end-tidal CO2 rises >1.5kPa over the pre-study level, that intervention will be abandoned.
Comparison:
At each slope time, the following parameters will be determined and compared with those recorded at 0.08 sec slope time: (1) peak inflating pressure, (2) inflation duration, (3) duration of inflation plateau, (4) duration of no gas flow, (5) expired tidal and minute volumes (mandatory/spontaneous, inspiratory/expiratory, (6) ventilator rate, (7) FiO2, (8) transcutaneous and/or end-tidal CO2 and, (9) interaction between ventilator inflations and baby's breaths. Values for SIPPV and PSV will also be compared.
Study design
- Primary purpose
- Treatment
- Allocation
- Randomised
- Intervention model
- Crossover
- Masking
- Single blind (participant)
- Number of arms
- 10
- Model description
- Ten different interventions (with changing slope time or ventilation mode) will be compared in a crossover design. Each intervention will be given for 15 minutes. For details of the intervention see study design details.
| Arm | Type | Receives |
|---|---|---|
| SIPPV_VG_0_08 | Experimental | Device: SIPPV_VG_0_08SIPPV VG ventilation for 15 minutes slope time = 0.08 seconds inspiratory time = 0.40 seconds |
| PSV_VG_0_08 | Experimental | Device: PSV_VG_0_08PSV VG ventilation for 15 minutes slope time = 0.08 seconds maximum inspiratory time = 0.60 seconds |
| SIPPV_VG_0_16 | Experimental | Device: SIPPV_VG_0_16SIPPV VG ventilation for 15 minutes slope time = 0.16 seconds inspiratory time = 0.40 seconds |
| PSV_VG_0_16 | Experimental | Device: PSV_VG_0_16PSV VG ventilation for 15 minutes slope time = 0.16 seconds maximum inspiratory time = 0.60 seconds |
| SIPPV_VG_0_24 | Experimental | Device: SIPPV_VG_0_24SIPPV VG ventilation for 15 minutes slope time = 0.24 seconds inspiratory time = 0.40 seconds |
| PSV_VG_0_24 | Experimental | Device: PSV_VG_0_24PSV VG ventilation for 15 minutes slope time = 0.24 seconds maximum inspiratory time = 0.60 seconds |
| SIPPV_VG_0_32 | Experimental | Device: SIPPV_VG_0_32SIPPV VG ventilation for 15 minutes slope time = 0.32 seconds inspiratory time = 0.40 seconds |
| PSV_VG_0_32 | Experimental | Device: PSV_VG_0_32PSV VG ventilation for 15 minutes slope time = 0.32 seconds maximum inspiratory time = 0.60 seconds |
| SIPPV_VG_0_40 | Experimental | Device: SIPPV_VG_0_40SIPPV VG ventilation for 15 minutes slope time = 0.32 seconds inspiratory time = 0.40 seconds |
| PSV_VG_0_40 | Experimental | Device: PSV_VG_0_40PSV VG ventilation for 15 minutes slope time = 0.40 seconds maximum inspiratory time = 0.60 seconds |
Who may qualify
- Ages
- 1 day to 2 months
- Sex
- All
- Healthy volunteers
- Not accepted
- Age groups
- Child
Key inclusion criteria (4 listed)
- Birth weight < 2 kg;
- Ventilated with SIPPV-VG modes,
- Informed parental consent,
- Clinician assent.
Key exclusion criteria (8 listed)
- Baby's respiratory condition unstable (Inspired oxygen (FiO2) > 50%, PaCO2 > 8.5kPa or <5kPa in the last 12 hours)
- Extubation planned in the next 12 hours;
- Neonatal or surgical procedure in the last 12 hours or planned in the next 12 hours;
- Significant pneumothorax requiring drainage;
- Gas leak around the endotracheal tube >50%; #
Eligibility is decided by the study team after screening; this overview does not establish that a patient qualifies. Full inclusion and exclusion criteria.
Interventions
| Intervention | Type | Arms |
|---|---|---|
| SIPPV_VG_0_08Mechanical ventilation using SIPPV-VG ventilator mode with a slope time of 0.08 seconds, inspiratory time of 0.40 seconds for 15 minutes | Device | SIPPV_VG_0_08 |
| SIPPV_VG_0_16Mechanical ventilation using SIPPV-VG ventilator mode with a slope time of 0.16seconds, inspiratory time of 0.40 seconds for 15 minutes | Device | SIPPV_VG_0_16 |
| SIPPV_VG_0_24Mechanical ventilation using SIPPV-VG ventilator mode with a slope time of 0.24 seconds, inspiratory time of 0.40 seconds for 15 minutes | Device | SIPPV_VG_0_24 |
| SIPPV_VG_0_32Mechanical ventilation using SIPPV-VG ventilator mode with a slope time of 0.32 seconds, inspiratory time of 0.40 seconds for 15 minutes | Device | SIPPV_VG_0_32 |
| SIPPV_VG_0_40Mechanical ventilation using SIPPV-VG ventilator mode with a slope time of 0.40 seconds, inspiratory time of 0.40 seconds for 15 minutes | Device | SIPPV_VG_0_40 |
| PSV_VG_0_08Mechanical ventilation using PSV-VG ventilator mode with a slope time of 0.08 seconds, maximum inspiratory time of 0.60 seconds for 15 minutes | Device | PSV_VG_0_08 |
| PSV_VG_0_16Mechanical ventilation using PSV-VG ventilator mode with a slope time of 0.16 seconds, maximum inspiratory time of 0.60 seconds for 15 minutes | Device | PSV_VG_0_16 |
| PSV_VG_0_24Mechanical ventilation using PSV-VG ventilator mode with a slope time of 0.24 seconds, maximum inspiratory time of 0.60 seconds for 15 minutes | Device | PSV_VG_0_24 |
| PSV_VG_0_32Mechanical ventilation using PSV-VG ventilator mode with a slope time of 0.32 seconds, maximum inspiratory time of 0.60 seconds for 15 minutes | Device | PSV_VG_0_32 |
| PSV_VG_0_40Mechanical ventilation using PSV-VG ventilator mode with a slope time of 0.40 seconds, maximum inspiratory time of 0.60 seconds for 15 minutes | Device | PSV_VG_0_40 |
Primary outcome
End tidal CO2 measurement
Time frame: One year
Primary outcome will be the difference in end-tidal CO2 concentration during the epochs with slope times of 0.40 and 0.08 sec.
Results and safety
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
Research links
- Official record on ClinicalTrials.govLast updated by the sponsor 20 May 2019.
- Background: PubMed 12464497Attar MA, Donn SM. Mechanisms of ventilator-induced lung injury in premature infants. Semin Neonatol. 2002 Oct;7(5):353-60. doi: 10.1053/siny.2002.0129.
- Background: PubMed 15674409Nafday SM, Green RS, Lin J, Brion LP, Ochshorn I, Holzman IR. Is there an advantage of using pressure support ventilation with volume guarantee in the initial management of premature infants with respiratory distress syndrome? A pilot study. J Perinatol. 2005 Mar;25(3):193-7. doi: 10.1038/sj.jp.7211233.
- Background: PubMed 19478530Bach KP, Kuschel CA, Oliver MH, Bloomfield FH. Ventilator gas flow rates affect inspiratory time and ventilator efficiency index in term lambs. Neonatology. 2009;96(4):259-64. doi: 10.1159/000220765. Epub 2009 May 27.
- Background: PubMed 23056572Bach KP, Kuschel CA, Hooper SB, Bertram J, McKnight S, Peachey SE, Zahra VA, Flecknoe SJ, Oliver MH, Wallace MJ, Bloomfield FH. High bias gas flows increase lung injury in the ventilated preterm lamb. PLoS One. 2012;7(10):e47044. doi: 10.1371/journal.pone.0047044. Epub 2012 Oct 8.
- Individual participant dataSharing planned: Undecided
Related trials for Respiratory Distress Syndrome in Premature Infant
Recruiting
REspiratory MEchanics for Delivering Individualised Exogenous Surfactant
Respiratory Distress Syndrome in Premature Infant
Not yet recruiting
The Hemodynamic Impact of High-frequency Oscillatory Ventilation and Periodic Sighs in Preterm Infants
Preterm, Respiratory Distress Syndrome in Premature Infant
Recruiting
Ventilator Pressure and Optimization of Compliance and Hemodynamics
Bronchopulmonary Dysplasia, Pulmonary Hypertension Due to Lung Diseases and Hypoxia
Active, not recruiting
Measuring Diaphragm Electrical Activity in Neonates Using a Smaller Inter-electrode Distance
Respiratory Distress Syndrome, Newborn, Infant, Premature
Not yet recruitingObservational study
Chest Ultrasound as Diagnostic and Prognostic Tool in Neonatal RDS
Respiratory Distress Syndrome, Newborn, Neonatal Respiratory Distress
Referral and contact
AzTrials does not enrol participants or take referrals. Contact the study team or a site listed below; they decide who can take part. This study is not recruiting at the moment.
Gusztav Belteki, M.D., Ph.D. principal investigator
Cambridge University Hospitals NHS Foundation Trust
1 site in 1 country: see every site and its status.
Questions about this trial
What is this clinical trial studying?
A study testing SIPPV_VG_0_08, SIPPV_VG_0_16 and SIPPV_VG_0_24 in people with Respiratory Distress Syndrome in Premature Infant, run by Cambridge University Hospitals NHS Foundation Trust at 1 site in United Kingdom.
The full registered description is in the About this study section and the official record.
Who may be eligible to take part?
The sponsor lists 4 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.
Is this clinical trial currently recruiting?
The record says: finished, as of the sponsor's last update on 20 May 2019. Status can change between updates, so confirm with the study contact before making plans.
Where is the study taking place?
The record lists 1 site in United Kingdom. Whether a site is taking part right now should be confirmed with that site. See every site.
What are the possible benefits and risks?
They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.
Are results available for this study?
The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.
How can a doctor refer a patient or contact the research team?
The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.
What does the primary outcome measure?
End tidal CO2 measurement (time frame: One year). It is what the study was designed to measure, not a finding.
Where can I read the original study information?
On the ClinicalTrials.gov record for NCT03306524, which also lists 4 linked publications.