Pilot Clinical Evaluation of the RHEA Vital Sign Vigilance System in Hospital Patients

A study testing The Rhea Vital Sign Vigilance Device Group and The Earlysense System Device Group in people with Vital Sign Monitoring, run by Darma Inc. at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

This study compares both of the investigational device and reference device to the gold standard Patient Monitor of the accuracy of heart rate and respiratory rate monitoring. The primary variables for the analysis of heart rate (HR) and respiratory rate (RR) are each of the overall mean roots mean square difference… Read more about this study

Who is it for?
Any age
Conditions the study is for
Vital Sign Monitoring
What is being tested?
The Rhea Vital Sign Vigilance Device Group and The Earlysense System Device Group
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
5 participants
Where
1 siteUnited States
Study dates
11 May 2018to 15 May 2018
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages Any age
  • All sexes
  • No healthy volunteers
  • Child, Adult, Older adult

Who may be able to join

  • Male or female, 18 years old and above.
  • Provide written informed consent.
  • Weight ranging from 20 to 150 kg inclusive.
  • Are located in a non-ICU hospital setting.
  • Agree to not eat during the testing period.
  • Agree to keep still.

Who cannot join

  • Are connected to a device which may interfere with the device monitoring in this study.
  • Are receiving any bedside care which may be incompatible with the study procedures.
  • Sleep apnea.
  • Pregnant or breast feeding.
  • A likely need to receive or undergo a procedure during the testing period.
  • Cannot accept a nasal cannula, or have a monitor lead placed on the chest.
  • A significant medical condition in the judgement of the investigator, which may compromise the study testing procedures.
  • Are wearing pacemaker or defibrillator.

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Virginia Commonwealth University School of Medicine

    Richmond, Virginia, 23298-0102

Other trials in United States

About this study

This study compares both of the investigational device and reference device to the gold standard Patient Monitor of the accuracy of heart rate and respiratory rate monitoring. The primary variables for the analysis of heart rate (HR) and respiratory rate (RR) are each of the overall mean roots mean square difference (RMSD). Results of the investigational device and reference device compared with the patient monitor is evaluated to see if there are equivalent. Motion and bed exit accuracy will be evaluated by comparing with the manual observation.

Read the detailed description

Tests (periods 1 and 2), one of the contact-free devices will be used to monitor the HR and RR, while being compared with the Patient Monitor. The RHEA investigational device and reference device will be used to monitor the same subject at different times, as the sensors of the two devices are placed at a similar location of the bed. HR and RR will be recorded for all three devices.

The primary hypotheses for HR, are as follows:

Adults HO: μ = 3.25 vs. HA: μ < 3.25

The primary hypotheses for RR, are as follows:

Adults HO: μ = 2.25 vs. HA: μ < 2.25 where μ represents the population mean HR or RR. If the upper bound of the confidence interval is less than the hypothesized value, the corresponding null hypothesis will be rejected. Comparison of the performance of the two contract-free devices will be performed on the recorded results via statistical analysis. The hypothesis here is that the two devices have equivalent performance. The third test will assess the ability of the RHEA device to accurately detect motion in or exit from the bed, as compared to the manual observation. Motion and no motion accuracy will each be calculated along with their respective two-sided 95% confidence intervals.

Being tested
The Rhea Vital Sign Vigilance Device Group, The Earlysense System Device Group

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • The Rhea Vital Sign Vigilance Device GroupDevice

    The RHEA device and Philips Patient Monitor MX400 are used to monitor the heart rate (HR) and respiratory rate (RR) of participants.

  • The Earlysense System Device GroupDevice

    The reference device Earlysense 2.0 system(vital sign monitoring system) and Philips Patient Monitor MX400 are used to monitor the heart rate (HR) and respiratory rate (RR) of participants.

Study groups in NCT03356184
Study groupRoleWhat they receive
The Rhea Vital Sign Vigilance Device GroupExperimentalThe RHEA device and Philips Patient Monitor Device are used to monitor the heart rate (HR) and respiratory rate (RR) of participants.
Receives: Device: The Rhea Vital Sign Vigilance Device Group
The Earlysense System Device GroupActive comparatorThe reference device -EarlySense System and Philips Patient Monitor Device are used to monitor the heart rate (HR) and respiratory rate (RR) of participants.
Receives: Device: The Earlysense System Device Group

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Root-mean-square Deviation of Heart Rate

    Calculate the root-mean-square deviation (RMSD) of the heart rate using 20 pairs of valid and synchronously recorded heart rate of investigational device and patient monitor.

    Measured: 30 days

  2. Root-mean-square Deviation of Respiratory Rate

    Calculate the root-mean-square deviation (RMSD) of the respiratory rate using 20 pairs of valid and synchronously recorded respiratory rate of investigational device and patient monitor.

    Measured: 30 days

Secondary outcomes (4)
  1. Heart Rate t Test Comparison result between the Investigational Device and the Reference Device

    Record 20 valid heart rate displayed in the reference device and patient monitor at the same time every one minute.

    Calculate the t test results for heart rate for the reference device and the investigational device.

    Measured: 30 days

  2. Respiratory Rate t Test Comparison result between the Investigational Device and the Reference Device

    Record 20 valid respiratory rate displayed in the reference device and patient monitor at the same time every one minute. Calculate the t test results for respiratory rate for the reference device and the investigational device.

    Measured: 30 days

  3. Motion Notification Accuracy(%)

    Record the time for each instructed movements from manual observation and time for movements notification shown in the investigational device. Calculate the times matched between these two to get the accuracy.

    Measured: 30 days

  4. Bed Exit Notification Accuracy(%)

    Record the time for each instructed bed exit from manual observation and time for bed exit notification shown in the investigational device. Calculate the times matched between these two to get the accuracy.

    Measured: 30 days

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted29 Nov 2017
  2. Study start11 May 2018
  3. Primary completion15 May 2018
  4. Study completion15 May 2018

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Screening
Allocation
Non-randomised
Intervention model
Crossover
Model description
The RHEA device and Patient Monitor are used to monitor the heart rate (HR) and respiratory rate (RR) of participants at the same time in Test 1. The reference device and patient monitor are used to monitor the heart rate and respiratory rate of the same participant at the same time in Test 2. The two tests results will be compared and evaluated the accuracy of the two devices. And the RHEA device is used alone to evaluate the accuracy of motion and no motion.
Masking
None (open label)
Enrolment
5 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03356184
Sponsor's protocol ID
RHEA EL30-2017-006
Status last verified
Apr 2019
First submitted
13 Nov 2017
Last updated
16 Apr 2019
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing The Rhea Vital Sign Vigilance Device Group and The Earlysense System Device Group in people with Vital Sign Monitoring, run by Darma Inc. at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 6 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 16 Apr 2019. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Root-mean-square Deviation of Heart Rate (time frame: 30 days). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 16 Apr 2019.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in