A Comparison of Two Devices for Taking Measurements of the Eye to Assist in Lens Selection for Cataract Surgery

A study testing GALILEI G6 Lens Professional and Oculus Pentacam AXL in people with Cataract, run by Ziemer Ophthalmic Systems AG at 1 site in Switzerland.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to compare the Ziemer Ophthalmic Systems AG GALILEI G6 Lens Professional to the Oculus Pentacam® AXL for taking images and measurements of the anterior segment of the eye, including the cornea, pupil, anterior chamber and lens, to assist in determining the power of the intraocular lens for… Read more about this study

Who is it for?
Ages 18 years and olderHealthy volunteers can take part
Conditions the study is for
Cataract
What is being tested?
GALILEI G6 Lens Professional and Oculus Pentacam AXL
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
112 participants
Where
1 siteSwitzerland
Study dates
5 Jan 2018to 1 Aug 2018
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Satisfies one of the subject group categories and the enrollment quota for that category has not been reached:
  • Normal eyes (phakic eyes without cataracts or corneal disease, refraction ≥ -5.5 D or ≤ +5.0 D)
  • Eyes having undergone previous refractive surgery (LASIK or PRK)
  • Eyes with cataracts (LOCS III grading with grades 2-6 in any or all of the posterior subcapsular (P2-P5), nuclear (N2-N5) and cortical (C2-C5), assessed by slit lamp examination; no other known ocular pathology)
  • Eyes with severe myopia (≤ -6 D according to the American Academy of Ophthalmology)
  • Eyes with severe hyperopia (≥ +5.25 D according to the American Academy of Ophthalmology)
Show all 9Show fewer
  • Eyes with previous cross-linking and eyes with advanced keratoconus
  • Manifest refraction sSpherical eEquivalent Refraction (MRSER) between -10dpt and +10dpt as measured by cycloplegic autorefraction assessed less than 12 months prior to enrollment
  • Best corrected visual acuity (BCVA) of LogMAR 0.2 or better

Who cannot join

  • Strabismus, nystagmus, amblyopia, anisometropia (difference in MRSER>3D)
  • Angle closure glaucoma
  • Seizure disorder, brain damage, Down Syndrome, Trisomy 13 or 18, Cerebral Palsy or other serious ophthalmic or central nervous system disorders
  • Severe dry eye
  • Severe corneal scarring
  • Inability to hold gaze
  • Inability to see fixation target

Trial locations and contacts

1 site in 1 country.

Switzerland1 site
  • Ziemer Ophthalmic Systems AG

    Port

Other trials in Switzerland

About this study

Official title: GALILEI G6 Lens Professional vs. a Predicate Device: a Comparison Study

The purpose of this study is to compare the Ziemer Ophthalmic Systems AG GALILEI G6 Lens Professional to the Oculus Pentacam® AXL for taking images and measurements of the anterior segment of the eye, including the cornea, pupil, anterior chamber and lens, to assist in determining the power of the intraocular lens for implantation.

Read the detailed description

A corneal topographer/biometer named GALILEI G6 Lens Professional was developed by Ziemer Ophthalmic Systems AG for the measurement of anterior segment geometry and axial intraocular distances. The measurement principle of the corneal topographer corresponds to that of the commercially available Galilei G4 Dual Scheimpflug Analyzer, whereas the measurement principle of the biometer is based on low coherence interferometry/reflectometry, which has been widely used and applied clinically by a number of predicate devices.

One complete measurement consists of a corneal topography/ tomography scan followed by three consecutive, axial biometry scans of the anterior segment (cornea and crystalline lens) and three consecutive, axial biometry scans of the retina. The measurement process is continuous but divided into three alignment-click-steps.

Such measurements can be applied to given Intra Ocular Lens (IOL) types and IOL equations, to permit the calculation of recommended IOLs to be implanted during cataract surgery in order to achieve the desired vision correction.

The primary objectives of this clinical study are to: 1) evaluate the inter-device repeatability, inter-operator repeatability and reproducibility of the GALILEI G6 Lens Professional in repeated measurements of anterior segment geometry and axial intraocular distances and, 2) demonstrate substantial equivalence through examining agreement of the GALILEI measurements to those of the predicate device, the Pentacam® AXL.

Conditions
Cataract
Being tested
GALILEI G6 Lens Professional, Oculus Pentacam AXL
Medical subject headings
Cataract

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • GALILEI G6 Lens ProfessionalDevice

    Eight measurements of the eye using the GALILEI G6 Lens Professional will be taken using two different devices and 2 different operators

  • Oculus Pentacam AXLDevice

    A single measurement of the eye will be taken using the Oculus Pentacam AXL

Study groups in NCT03396003
Study groupRoleWhat they receive
GALILEI G6 Lens ProfessionalExperimentalThe GALILEI G6 Lens Professional will measure anterior segment geometry and axial intra-ocular distances of the eye.
Receives: Device: GALILEI G6 Lens Professional
Oculus Pentacam AXLActive comparatorThe Oculus Pentacam AXL will measure anterior segment geometry and axial intra-ocular distances of the eye.
Receives: Device: Oculus Pentacam AXL

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Axial Length (AL) Measurement in millimeters (mm)

    Repeatability and reproducability across and between evaluations

    Measured: One day

  2. Central Corneal Thickness (CCT) Measurement in micrometers (µm)

    Repeatability and reproducability across and between evaluations

    Measured: One day

  3. Radius Flat Meridian (R flat) Measurement in millimeters (mm)

    Repeatability and reproducability across and between evaluations

    Measured: One day

  4. Radius Steep Meridian (R steep) Measurement in millimeters (mm)

    Repeatability and reproducability across and between evaluations

    Measured: One day

  5. Mean Radius (Rm) Measurement in millimeters (mm)

    Repeatability and reproducability across and between evaluations

    Measured: One day

  6. Corneal Cylinder Measurement in diopters (D)

    Repeatability and reproducability across and between evaluations

    Measured: One day

  7. Corneal Cylinder Axis (A flat) Measurement in degrees

    Repeatability and reproducability across and between evaluations

    Measured: One day

  8. Anterior Chamber Depth (ACD) Measurement in millimeters (mm)

    Repeatability and reproducability across and between evaluations

    Measured: One day

  9. Horizontal White-to-White distance (WtW) Measurement in millimeters (mm).

    Repeatability and reproducability across and between evaluations

    Measured: One day

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted10 Jan 2018
  2. Study start5 Jan 2018
  3. Primary completion1 Aug 2018
  4. Study completion1 Aug 2018

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Non-randomised
Intervention model
Parallel groups
Masking
None (open label)
Enrolment
112 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03396003
Sponsor's protocol ID
ZIEMER_G6
Status last verified
Sep 2018
First submitted
20 Nov 2017
Last updated
26 Sep 2018
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Questions about this trial

What is this clinical trial studying?

A study testing GALILEI G6 Lens Professional and Oculus Pentacam AXL in people with Cataract, run by Ziemer Ophthalmic Systems AG at 1 site in Switzerland.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 9 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 26 Sep 2018. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Switzerland. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Axial Length (AL) Measurement in millimeters (mm) (time frame: One day). The study lists 9 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 26 Sep 2018.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in