Efficacy and Safety of Electric Stimulation-guided Epidural Anesthesia for Cesarean Section

A study testing Epidural electrical stimulation (EES) and Loss of resistance (LOR) in people with Epidural Anesthesia, run by Korea University Guro Hospital at 1 site in South Korea.

Unknown status

Status not confirmed. The sponsor has not verified this record for over two years, so it may be out of date.

Official record
Search all trials

Trial summary

Forty pregnant women (36 to 41 weeks gestation) will randomly allocate to two groups. Groups will be defined based on the method used to identify the epidural space for epidural anesthesia: the loss of resistance group (n=20) and the epidural electric stimulation group (n=20). Read more about this study

Who is it for?
Ages 19 to 64 years, women only
Conditions the study is for
Epidural Anesthesia
What is being tested?
Epidural electrical stimulation (EES) and Loss of resistance (LOR)
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
40 participants
Where
1 siteSouth Korea
Study dates
11 Mar 2015to Mar 2019
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 19 to 64 years
  • Women only
  • No healthy volunteers
  • Adult

Who may be able to join

  • Patients who were at 36 to 41 weeks' gestation and admitted in labor to the university clinic for cesarean section were included. Patients were American Society of Anesthesiologists (ASA) physical status of I or II, and were scheduled to receive epidural anesthesia.

Who cannot join

  • Skin infection at the injection site
  • Difficult catheter placement owing to previous lumbar spinal surgery or deformity
  • Presence of a hemostatic disorder or use of antiplatelet therapy
  • Presence of a cardiac pacemaker.
  • Removal of patient

Trial locations and contacts

1 site in 1 country.

Chung Hun Lee, MD contact

82-2-2626-3240

bodlch@naver.com

Sang Sik Choi, MD, PhD contact

82-2-2626-3238

clonidine@empal.com

South Korea1 site, 1 recruiting
  • Korea University Guro Hospital

    Seoul, 08308

    Seung-hoe Song, MBE, 82-2-2626-1635, ssessong@korea.ac.kr

    Ji-young Lee, 82-2-2626-2279, ljy@kumc.or.kr

    Recruiting

Other trials in South Korea

About this study

Official title: A Comparative Study of Success Rate, Efficacy, Safety Between Electric Stimulation-guided Epidural Catheter Placement and the Loss of Resistance Conventional Method for Cesarean Section

Forty pregnant women (36 to 41 weeks gestation) will randomly allocate to two groups. Groups will be defined based on the method used to identify the epidural space for epidural anesthesia: the loss of resistance group (n=20) and the epidural electric stimulation group (n=20). Pain during the cesarean section will be assessed using a numerical visual analog scale and maternal satisfaction by a post-partum interview. The success rate of epidural anesthesia, maternal satisfaction, and neonatal Apgar scores will be compared between groups.

Read the detailed description

Investigators will place epidural catheter in the epidural space using loss of resistance technique, and will confirm correct placement of the epidural catheter using electric stimulation.

Epidural catheter placement, electric stimulation, and confirmation of response is followed:

Patients will be placed in the left lateral decubitus position. The site will be aseptically prepared and 1% lidocaine will be infiltrated to the skin. An 18-gauge Tuohy needle will be inserted midline of L4/5 interspinous space.

For the LOR group, after identification of the epidural space, the Tuohy needle will be stopped, and a 20-gauge epidural catheter will be advanced through the Touhy needle.

The same process will be followed for the EES group. In addition, the epidural space will be confirmed by epidural electric stimulation using a 20-gauge epidural catheter (RegionalStimTm, Sewoon Medical Co., Ltd, Seoul, Korea, 800 mm) with a conductive guidewire (conductive guidewire, Nitinol, 1100 mm).

After confirming there is no reverse flow of cerebrospinal fluid or blood with aspiration, 3 mL of 1% lidocaine, with 15 mcg of epinephrine (1:200000), will be injected through the epidural catheter as test dose. If there is no response to the test dose, patients will be moved to the operating room. In operating room, 20 mL of 2% lidocaine, 2 mL of bicarbonate (total volume 22 mL) will be administered in divided doses.

Blood pressure (BP), heart rate (HR), oxygen saturation (SpO2), and neurologic assessment findings will be monitored.

Pain relief in cesarean section is assessed in the visual analogue scale (VAS) score. A 10 point VAS, where 0 is no pain and 10 is unbearable pain, is used to assess pain during labor. The scale is assessed after epidural anesthesia. Differences in the VAS we used to assess the efficacy of the epidural anesthesia in decreasing labor pain. Comparison of the change in VAS between groups is used to compare pain control of the two methods. The success of epidural anesthesia is defined by sensory block, without motor block, and a decrease in pain score after adequate dosing of epidural medication. Failure of epidural anesthesia is defined in several ways. Objective outcome include conversion to general anesthesia, conversion to any different form of anesthesia, or pain during surgery.

Patient satisfaction will be evaluated by a postpartum interview. Satisfaction is graded between a score of 1-5, where 1 represent very unsatisfied and 5 represent very satisfied. Patients will indicate a score of 1 to 5.

One- and 5-minute Apgar scores will be compared to assess the effect of epidural electric stimulation on the neonate. Additional time required for epidural electric stimulation will be determined by the difference (in seconds) from LOR to identification of the epidural space through electric stimulation.

Being tested
Epidural electrical stimulation (EES), Loss of resistance (LOR)
Keywords
Electric stimulation, Cesarean section

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Epidural electrical stimulation (EES)Device

    Using loss of resistance technique and electrical stimulation

  • Loss of resistance (LOR)Device

    Using loss of resistance technique only

Study groups in NCT03443466
Study groupRoleWhat they receive
Epidural electrical stimulation (EES)Experimentaln=20
Receives: Device: Epidural electrical stimulation (EES)
Loss of resistance (LOR)Active comparatorn=20
Receives: Device: Loss of resistance (LOR)

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Success rate of epidural anesthesia

    Evaluation parameter : Accuracy comparison between loss of resistance and epidural electrical stimulation

    Measured: Up to 6 months

Secondary outcomes (5)
  1. Maternal satisfaction

    Patient satisfaction will be evaluated by a postpartum interview. Satisfaction was graded by patients from 1 to 5, where 1 represented very unsatisfied and 5 represented very satisfied

    Measured: Up to 6 months

  2. Neonatal Apgar score

    The neonatal apgar scores from 0 to 10 were measured at 1 and 5 minutes to assess the effect of electrical stimulation on the fetus; Scores 7 and above are generally normal, 4 to 6 fairly low, and 3 and below are generally regarded as critically low

    Measured: Up to 6 months

  3. Procedure-related complications

    The number of times dural puncture or intravascular insertion was performed by Tuohy needle of epidural catheter during epidural anesthesia

    Measured: Up to 6 months

  4. Minimum electrical current to elicit a response in the epidural electrical stimulation group

    Minimum current stimulus to confirm that the epidural catheter is located in the epidural space between 0 and 5mA ;mA (frequency, 1 Hz; pulse width, 300 ms)

    Measured: Up to 6 months

  5. Additional time for epidural electrical stimulation

    • Determined by the difference (in seconds) from loss of resistance(LOR) to identification of the epidural space through electrical stimulation
    • Measure : seconds (from observation of LOR to confirmation of the epidural space by electric stimulation)

    Measured: Up to 6 months

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted23 Feb 2018
  2. Study start11 Mar 2015
  3. Primary completionMar 2019 (estimated)
  4. Study completionMar 2019 (estimated)

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Randomised
Intervention model
Parallel groups
Model description
Parallel Assignment
Masking
None (open label)
Enrolment
40 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03443466
Sponsor's protocol ID
2015GR0703 (RegionalStimⓇ)
Status last verified
Feb 2018
First submitted
30 Jan 2018
Last updated
23 Feb 2018
Results posted
No
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
No

Publications

  • Silva M, Halpern SH. Epidural analgesia for labor: Current techniques. Local Reg Anesth. 2010;3:143-53. doi: 10.2147/LRA.S10237. Epub 2010 Dec 8.

    PubMed 23144567
  • Wantman A, Hancox N, Howell PR. Techniques for identifying the epidural space: a survey of practice amongst anaesthetists in the UK. Anaesthesia. 2006 Apr;61(4):370-5. doi: 10.1111/j.1365-2044.2006.04534.x.

    PubMed 16548958
  • Eappen S, Blinn A, Segal S. Incidence of epidural catheter replacement in parturients: a retrospective chart review. Int J Obstet Anesth. 1998 Oct;7(4):220-5. doi: 10.1016/s0959-289x(98)80042-3.

    PubMed 15321183
  • Hermanides J, Hollmann MW, Stevens MF, Lirk P. Failed epidural: causes and management. Br J Anaesth. 2012 Aug;109(2):144-54. doi: 10.1093/bja/aes214. Epub 2012 Jun 26.

    PubMed 22735301
  • Tsui BC, Tarkkila P, Gupta S, Kearney R. Confirmation of caudal needle placement using nerve stimulation. Anesthesiology. 1999 Aug;91(2):374-8. doi: 10.1097/00000542-199908000-00010.

    PubMed 10443599
  • Tsui BC, Gupta S, Finucane B. Confirmation of epidural catheter placement using nerve stimulation. Can J Anaesth. 1998 Jul;45(7):640-4. doi: 10.1007/BF03012093.

    PubMed 9717595
  • Kim YS, Kim HS, Jeong H, Lee CH, Lee MK, Choi SS. Efficacy of electrical stimulation on epidural anesthesia for cesarean section: a randomized controlled trial. BMC Anesthesiol. 2020 Jun 10;20(1):146. doi: 10.1186/s12871-020-01063-1.

    PubMed 32522156

Questions about this trial

What is this clinical trial studying?

A study testing Epidural electrical stimulation (EES) and Loss of resistance (LOR) in people with Epidural Anesthesia, run by Korea University Guro Hospital at 1 site in South Korea.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 1 inclusion and 5 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: status not confirmed, as of the sponsor's last update on 23 Feb 2018. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in South Korea. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Results are not available in the source record.

How can a doctor refer a patient or contact the research team?

Contact Chung Hun Lee, MD (82-2-2626-3240, bodlch@naver.com), or a site near the patient. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Success rate of epidural anesthesia (time frame: Up to 6 months). It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT03443466, which also lists 7 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 23 Feb 2018.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in