Repeatability, Reproducibility and Comparison of Cirrus OCT, RTVue OCT, MS-39 OCT, and Insight 100 VHFDU

A study testing ArcScan Insight 100 very high-frequency digital ultrasound, Carl Zeiss Meditec Cirrus HD OCT 5000 and Optovue RTVue OCT in people with Corneal Thickness Measurement and Epithelial Thickness Measurement, run by London Vision Clinic at 1 site in United Kingdom.

Suspended

Paused. Enrolment is on hold for now and may resume.

Official record
Search all trials

Trial summary

The purpose of this study is to evaluate the repeatability and reproducibility of corneal, epithelial, and LASIK flap thickness using the Carl Zeiss Meditec Cirrus HD OCT 5000, Optovue RTVue OCT, CSO MS-39 OCT, and ArcScan Insight 100 very high-frequency digital ultrasound arc-scanner. Read more about this study

Who is it for?
Ages 21 to 70 yearsHealthy volunteers can take part
What is being tested?
ArcScan Insight 100 very high-frequency digital ultrasound, Carl Zeiss Meditec Cirrus HD OCT 5000 and Optovue RTVue OCT
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
242 participants
Where
1 siteUnited Kingdom
Study dates
1 Sep 2022to 1 Dec 2023
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 21 to 70 years
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Only patients who are medically suitable for corneal laser refractive surgery can be included in the study.
  • As general inclusion criteria the following aspects are defined:
  • Subjects should be 21 years of age or older.
  • Contact lens wearers must stop wearing their contact lenses at least four weeks per decade of wear before baseline measurements in case of hard contact lenses and one week before baseline measurements in case of soft contact lenses.
  • Patient will be able to understand the patient information and willing to sign an informed consent.
  • For the normal eye populations:
Show all 10Show fewer
  • Eyes with no ocular pathology
  • Eyes with no previous ocular or corneal surgery
  • For the post-op eye populations:
  • Eyes between 3 and 9 months after LASIK or SMILE

Who cannot join

  • Previous ocular or corneal surgery (for normal eyes)
  • Patient not being able to tolerate local or topical anesthesia
  • Patient has active conjunctivitis or any active ocular infection or inflammation
  • Patient has a corneal ulcer
  • Patient has an open wound on the corneal epithelium
  • Patient has compromised epithelium (e.g. ≥Grade 3 superficial punctate keratitis)
  • Pregnancy

Trial locations and contacts

1 site in 1 country.

United Kingdom1 site
  • London Vision Clinic

    London, W1G 7LA

Other trials in United Kingdom

About this study

Official title: Repeatability and Reproducibility of Carl Zeiss Meditec Cirrus HD OCT 5000, Optovue RTVue OCT, CSO MS-39 OCT and ArcScan Insight 100 Very High-frequency Digital Ultrasound and Comparison Between Devices

The purpose of this study is to evaluate the repeatability and reproducibility of corneal, epithelial, and LASIK flap thickness using the Carl Zeiss Meditec Cirrus HD OCT 5000, Optovue RTVue OCT, CSO MS-39 OCT, and ArcScan Insight 100 very high-frequency digital ultrasound arc-scanner.

Read the detailed description

Very high-frequency (VHF) digital ultrasound was first used in 1993 to obtain images and measure the thickness of the cornea (the window at the front of the eye) and the layers within the cornea, such as the epithelium (the layer of skin on the surface of the cornea). A commercially available instrument, known as the Artemis, has been used in routine clinical practice since 2001. The ArcScan Insight 100 VHF digital ultrasound scanner is an updated version, which obtained CE mark in 2016.

Optical coherence tomography (OCT) is another method of measuring the thickness of the cornea and the corneal epithelium. OCT has been used for measuring corneal thickness since 1997 and the technology had evolved to also measure epithelial thickness since 2012. There are currently three OCT scanners capable of measuring epithelial thickness; Carl Zeiss Meditec Cirrus HD OCT 5000, Optovue RTVue OCT, CSO MS-39 OCT.

The aim of the study is to establish the repeatability (i.e. the variability in measurements taken by a single examiner during a single visit) and reproducibility (i.e. the variability in measurements taken in the same conditions between two examiners) for each of the four devices for measuring corneal thickness and epithelial thickness.

Being tested
ArcScan Insight 100 very high-frequency digital ultrasound, Carl Zeiss Meditec Cirrus HD OCT 5000, Optovue RTVue OCT, CSO MS-39 OCT
Keywords
VHF digital ultrasound, Optical coherence tomography, Corneal thickness, Epithelial thickness, LASIK, SMILE

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • ArcScan Insight 100 very high-frequency digital ultrasoundDevice

    Also called Artemis very high-frequency digital ultrasound

    The ArcScan Insight 100 VHF digital ultrasound scanner can measure individual layers within the cornea with very high precision, as well as being able to image the interior of the eye including the chamber between the iris and the cornea, the lens and other structures behind the iris. The ArcScan Insight 100 device is a digital ultrasound scanner, meaning that it uses ultrasound waves to measure parameters of the eye. The Insight examination involves the patient sitting in front of the ultrasound unit and looking through a disposable goggle-like eyepiece at a light source. Warm sterile balanced saline solution (like natural tears) are used to fill the space between the light source and the eye, to allow ultrasound transmission. There is no contact between the instrument and the eye other than being immersed in this water bath.

  • Carl Zeiss Meditec Cirrus HD OCT 5000Device

    The Cirrus HD OCT 5000 uses high definition spectral domain optical coherence tomography (OCT) to measure anatomical structures of the eye. OCT is non-contact and uses light, rather than ultrasound, to obtain cross-sectional images. An OCT scan involves the patient sitting in front of the device, placing their chin on a chin-rest and fixating on a light source. Nothing comes into contact with the eye and the patient is only aware of the instrument rotating in front of them.

  • Optovue RTVue OCTDevice

    The Optovue RTVue OCT uses high definition Fourier/spectral domain optical coherence tomography (OCT) to measure anatomical structures of the eye. OCT is non-contact and uses light, rather than ultrasound, to obtain cross-sectional images. An OCT scan involves the patient sitting in front of the device, placing their chin on a chin-rest and fixating on a light source. Nothing comes into contact with the eye and the patient is only aware of the instrument rotating in front of them.

  • CSO MS-39 OCTDevice

    The CSO MS-39 OCT uses high definition Fourier domain optical coherence tomography (OCT) to measure anatomical structures of the eye. OCT is non-contact and uses light, rather than ultrasound, to obtain cross-sectional images. The MS-39 also includes a Placido topographer within the device to simultaneously obtain a measurement of the shape (curvature) of the front surface of the cornea, which is then combined with the OCT thickness measurement to generate the shape of the back surface of the cornea.

    An OCT scan involves the patient sitting in front of the device, placing their chin on a chin-rest and fixating on a light source. Nothing comes into contact with the eye and the patient is only aware of the instrument rotating in front of them.

Study groups in NCT03473847
Study groupRoleWhat they receive
Repeatability and Reproducibility - Normal eyesExperimentalThis arm will include 20 eyes of 20 patients with no previous ocular surgery. The research participant will be scanned a number of times using each of the four devices (the Insight 100 VHF digital ultrasound scanner, the Cirrus HD OCT 5000, the RTVue OCT, and the MS-39 OCT). The scan sequence will be undertaken on a single day as follows: 1. 5 consecutive repeated scans of the cornea will be performed by the first operator using each of the four devices (expected total time approximately 5 minutes for each OCT scan set and 15 minutes for the Insight 100 scan set). 2. There will be a break of about 30 minutes. 3. 5 consecutive repeated scans of the cornea will be performed by the second operator using each of the four devices (expected total time approximately 5 minutes for each OCT scan set and 15 minutes for the Insight 100 scan set).
Receives: Device: ArcScan Insight 100 very high-frequency digital ultrasound; Device: Carl Zeiss Meditec Cirrus HD OCT 5000; Device: Optovue RTVue OCT; Device: CSO MS-39 OCT
Repeatability and Reproducibility - Post-op eyesExperimentalThis arm will include 20 eyes of 20 patients between 3 and 9 months after corneal laser refractive surgery. The research participant will be scanned a number of times using each of the four devices (the Insight 100 VHF digital ultrasound scanner, the Cirrus HD OCT 5000, the RTVue OCT, and the MS-39 OCT). The scan sequence will be undertaken on a single day as follows: 1. 5 consecutive repeated scans of the cornea will be performed by the first operator using each of the four devices (expected total time approximately 5 minutes for each OCT scan set and 15 minutes for the Insight 100 scan set). 2. There will be a break of about 30 minutes. 3. 5 consecutive repeated scans of the cornea will be performed by the second operator using each of the four devices (expected total time approximately 5 minutes for each OCT scan set and 15 minutes for the Insight 100 scan set).
Receives: Device: ArcScan Insight 100 very high-frequency digital ultrasound; Device: Carl Zeiss Meditec Cirrus HD OCT 5000; Device: Optovue RTVue OCT; Device: CSO MS-39 OCT
Comparison between devices - Normal eyesExperimentalThis arm will include 101 eyes of 101 patients with no previous ocular surgery. The research participant will be scanned once using each of the four devices (the Insight 100 VHF digital ultrasound scanner, the Cirrus HD OCT 5000, the RTVue OCT, and the MS-39 OCT). Expected total time approximately 20 minutes to complete all four scans.
Receives: Device: ArcScan Insight 100 very high-frequency digital ultrasound; Device: Carl Zeiss Meditec Cirrus HD OCT 5000; Device: Optovue RTVue OCT; Device: CSO MS-39 OCT
Comparison between devices - Post-op eyesExperimentalThis arm will include 101 eyes of 101 patients between 3 and 9 months after corneal laser refractive surgery. The research participant will be scanned once using each of the four devices (the Insight 100 VHF digital ultrasound scanner, the Cirrus HD OCT 5000, the RTVue OCT, and the MS-39 OCT). Expected total time approximately 20 minutes to complete all four scans.
Receives: Device: ArcScan Insight 100 very high-frequency digital ultrasound; Device: Carl Zeiss Meditec Cirrus HD OCT 5000; Device: Optovue RTVue OCT; Device: CSO MS-39 OCT

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Repeatability and reproducibility of Cirrus HD OCT 5000 corneal scans in normal eyes (standard deviation)

    The within-observer standard deviation (repeatability) and between-observer standard deviation (reproducibility) calculated by ANOVA for 5 repeated measurements by two observers of corneal thickness and epithelial thickness in normal eyes with no previous corneal surgery

    Measured: 10 minutes (5 minutes per observer)

  2. Repeatability and reproducibility of RTVue OCT corneal scans in normal eyes (standard deviation)

    The within-observer standard deviation (repeatability) and between-observer standard deviation (reproducibility) calculated by ANOVA for 5 repeated measurements by two observers of corneal thickness and epithelial thickness in normal eyes with no previous corneal surgery

    Measured: 10 minutes (5 minutes per observer)

  3. Repeatability and reproducibility of MS-39 OCT corneal scans in normal eyes (standard deviation)

    The within-observer standard deviation (repeatability) and between-observer standard deviation (reproducibility) calculated by ANOVA for 5 repeated measurements by two observers of corneal thickness and epithelial thickness in normal eyes with no previous corneal surgery

    Measured: 10 minutes (5 minutes per observer)

  4. Repeatability and reproducibility of Insight 100 VHF digital ultrasound corneal scans in normal eyes (standard deviation)

    The within-observer standard deviation (repeatability) and between-observer standard deviation (reproducibility) calculated by ANOVA for 5 repeated measurements by two observers of corneal thickness and epithelial thickness in normal eyes with no previous corneal surgery

    Measured: 30 minutes (15 minutes per observer)

  5. Repeatability and reproducibility of Cirrus HD OCT 5000 corneal scans in post-op eyes (standard deviation)

    The within-observer standard deviation (repeatability) and between-observer standard deviation (reproducibility) calculated by ANOVA for 5 repeated measurements by two observers of corneal thickness and epithelial thickness in eyes between 3 and 9 months after corneal laser refractive surgery

    Measured: 10 minutes (5 minutes per observer)

  6. Repeatability and reproducibility of RTVue OCT corneal scans in post-op eyes (standard deviation)

    The within-observer standard deviation (repeatability) and between-observer standard deviation (reproducibility) calculated by ANOVA for 5 repeated measurements by two observers of corneal thickness and epithelial thickness in eyes between 3 and 9 months after corneal laser refractive surgery

    Measured: 10 minutes (5 minutes per observer)

  7. Repeatability and reproducibility of MS-39 OCT corneal scans in post-op eyes (standard deviation)

    The within-observer standard deviation (repeatability) and between-observer standard deviation (reproducibility) calculated by ANOVA for 5 repeated measurements by two observers of corneal thickness and epithelial thickness in eyes between 3 and 9 months after corneal laser refractive surgery

    Measured: 10 minutes (5 minutes per observer)

  8. Repeatability and reproducibility of Insight 100 VHF digital ultrasound corneal scans in post-op eyes (standard deviation)

    The within-observer standard deviation (repeatability) and between-observer standard deviation (reproducibility) calculated by ANOVA for 5 repeated measurements by two observers of corneal thickness and epithelial thickness in eyes between 3 and 9 months after corneal laser refractive surgery

    Measured: 30 minutes (15 minutes per observer)

  9. Difference in corneal thickness measurements between Cirrus HD OCT 5000, RTVue OCT, MS-39 OCT, and Insight 100 VHF digital ultrasound in normal eyes

    The difference between each pair of devices for measurements of corneal thickness and epithelial thickness in eyes with no previous ocular surgery.

    Measured: 20 minutes for all four scans

  10. Difference in corneal thickness measurements between Cirrus HD OCT 5000, RTVue OCT, MS-39 OCT, and Insight 100 VHF digital ultrasound in post-op eyes

    The difference between each pair of devices for corneal thickness, epithelial thickness and interface thickness in eyes between 3 and 9 months after corneal laser refractive surgery

    Measured: 20 minutes for all four scans

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted22 Mar 2018
  2. Study start1 Sep 2022 (estimated)
  3. Primary completion1 Dec 2023 (estimated)
  4. Study completion1 Dec 2023 (estimated)

Why it stopped: Devices currently awaiting software updates

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Non-randomised
Intervention model
Parallel groups
Model description
Scans from 4 devices obtained on the same patient population for comparison
Masking
None (open label)
Enrolment
242 participants (estimated)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03473847
Sponsor's protocol ID
LOVC-006
Status last verified
Apr 2022
First submitted
8 Mar 2018
Last updated
26 Apr 2022
Results posted
No
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No: No data will be shared with other researchers

Publications

  • Reinstein DZ, Archer TJ, Gobbe M, Silverman RH, Coleman DJ. Repeatability of layered corneal pachymetry with the artemis very high-frequency digital ultrasound arc-scanner. J Refract Surg. 2010 Sep;26(9):646-59. doi: 10.3928/1081597X-20091105-01. Epub 2009 Nov 16.

    PubMed 19928698
  • Reinstein DZ, Yap TE, Archer TJ, Gobbe M, Silverman RH. Comparison of Corneal Epithelial Thickness Measurement Between Fourier-Domain OCT and Very High-Frequency Digital Ultrasound. J Refract Surg. 2015 Jul;31(7):438-45. doi: 10.3928/1081597X-20150623-01.

    PubMed 26186562
  • Yap TE, Archer TJ, Gobbe M, Reinstein DZ. Comparison of Central Corneal Thickness Between Fourier-Domain OCT, Very High-Frequency Digital Ultrasound, and Scheimpflug Imaging Systems. J Refract Surg. 2016 Feb;32(2):110-6. doi: 10.3928/1081597X-20151223-01.

    PubMed 26856428
  • Kiraly L, Stange J, Kunert KS, Sel S. Repeatability and Agreement of Central Corneal Thickness and Keratometry Measurements between Four Different Devices. J Ophthalmol. 2017;2017:6181405. doi: 10.1155/2017/6181405. Epub 2017 Mar 5.

    PubMed 28357136

Questions about this trial

What is this clinical trial studying?

A study testing ArcScan Insight 100 very high-frequency digital ultrasound, Carl Zeiss Meditec Cirrus HD OCT 5000 and Optovue RTVue OCT in people with Corneal Thickness Measurement and Epithelial Thickness Measurement, run by London Vision Clinic at 1 site in United Kingdom.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 10 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: paused, as of the sponsor's last update on 26 Apr 2022. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United Kingdom. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Results are not available in the source record.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Repeatability and reproducibility of Cirrus HD OCT 5000 corneal scans in normal eyes (standard deviation) (time frame: 10 minutes (5 minutes per observer)). The study lists 10 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT03473847, which also lists 4 linked publications.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.