Multiple SclerosisNCT03603691

Study to Evaluate the Reliability and Validity of the Modified Manual Muscle Test for Persons With MS

A study testing Modified Manual Muscle Test, BMRC manual muscle test and MicroFET2 handhold dynamometer in people with Multiple Sclerosis, run by Nanco van der Maas at 1 site in Switzerland.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

Many patients with multiple sclerosis (MS) are treated with physiotherapy. Muscle weakness is a common symptom. It can be evaluated with a variety of muscle function tests. In MS patients, testing muscle function can be confounded by many factors, such as spasticity and ataxia, which are not considered by the existing… Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Multiple Sclerosis
What is being tested?
Modified Manual Muscle Test, BMRC manual muscle test and MicroFET2 handhold dynamometer
Could I receive a placebo?
No, everyone receives an active treatment
How many people
28 participants
Where
1 siteSwitzerland
Study dates
1 Aug 2018to 29 Jan 2019
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • diagnosed with MS according to the McDonald criteria
  • EDSS between 0 and 6.5
  • older than 18 years of age.

Who cannot join

  • An acute episode of MS within 3 months prior to the testing,
  • grave cognitive changes
  • distinct fatigue
  • a history of current back, neck or elbow pain

Trial locations and contacts

1 site in 1 country.

Switzerland1 site
  • Felix Platter Hospital

    Basel, 4055

Other trials in Switzerland

About this study

Official title: A Single-centre, Prospective, Cross-sectional Study to Evaluate the Reliability and Validity of the Modified Manual Muscle Test for Persons With Multiple Sclerosis) (MS)

Many patients with multiple sclerosis (MS) are treated with physiotherapy. Muscle weakness is a common symptom. It can be evaluated with a variety of muscle function tests. In MS patients, testing muscle function can be confounded by many factors, such as spasticity and ataxia, which are not considered by the existing tests and may cause biased test results. Steinlin Egli described a Modified Manual Muscle Test (MMMT) that considers spasticity and may provide a more reliable and valid muscle function test for MS patients.

The investigators aim to evaluate the inter- and intra-rater reliability of the Modified Manual Muscle Test in MS and evaluate the validity of the Modified MMT according to the criteria of the 6 level British Medical Research Council (BMRC) manual muscle test and the microFET2 handhold dynamometer.

Read the detailed description

Background In Switzerland, many patients with multiple sclerosis (MS) are treated with physiotherapy. Muscle weakness is a common symptom. Particularly in progressive disease courses, motor dysfunction is a major contributor to the reduced mobility and quality of life. Therefore, improving the motor dysfunction is an important goal for therapy and is evaluated with a variety of muscle function tests. The 6 level British Medical Research Council (BMRC) manual muscle test is the most widely used test in MS and is a part of the Neurostatus-Expanded Disability Status Scale (EDSS) assessment, which has been adopted as the standard in most of the pivotal trials that have led to the approval of the current MS treatments. In MS patients, testing muscle function can be confounded by many factors, such as spasticity and ataxia, which are not considered by the existing tests and may cause biased test results. Steinlin Egli described a Modified Manual Muscle Test (MMMT) that considers spasticity and may provide a more reliable and valid muscle function test for MS patients.

Aim of the study is to evaluate the inter- and intra-rater reliability of the Modified Manual Muscle Test in MS and evaluate the validity of the Modified MMT according to the criteria of the Neurostatus BMRC manual muscle test and the microFET2 handhold dynamometer.

Methods This is a single-centre, prospective cross-sectional study with a test-retest design. The Neurostatus BMRC manual muscle test and the microFET2 hand held dynamometer are used as the criteria. The primary endpoint is the ordinal MMMT level. The intra-class correlation coefficient (ICC) of the ranked MMMT levels will be estimated. The investigator aims to show that the MMMT results are clinically relevant with a high level of ICC.

The secondary endpoints include the Neurostatus BMRC levels, the muscle strength as measured by the microFET2 dynamometer and fatigue using a numeric rating scale (NRS). Subgroup analyses will determine whether the MMMT is less sensitive to the influence of spasticity than the Neurostatus BMRC.

28 patients with MS will be tested in 2 days. In each day, six examiners will test seven patients in the morning and seven patients in the afternoon. Three MS-therapists from the Specialized Group Physiotherapy in MS and three neurologists from the Department of Neurology at the University Hospital Basel will be tester. All examiners will be blinded to the results of the other examiners. All test persons will be blinded to their test results. First, the patient will rate his fatigue using the NRS. Muscle function will be tested with the microFET2 and spasticity will be assessed using the modified Tardieu Scale. Then, the first round of manual testing will be executed. After each test, there will be a break that is long enough for the patients to recuperate. The order of the examiners will be randomized. The retest round will proceed in a different order. Finally, spasticity will be tested again using the modified Tardieu Scale, and the patients will rate their fatigue.

Statistical Analysis The study data will be analysed using the internet-based program R. The descriptive statistics will describe the sample characteristics. The inter-rater reliability will be estimated by the ICC of the MMMT ranks. The ICC will be estimated by a two-way random-effects ANOVA and reported with 95% CI. If the lower limit of the CI is larger than the pre-specified clinically relevant ICC margin, the MMMT will be considered a reliable tool for evaluating muscle strength in MS patients.

The inter-rater reliability the Neurostatus BMRC will be estimated similarly by the ICC. The ICC of both tests will be compared. The test-retest reliability of the MMMT and Neurostatus BMRC will be estimated using two-way random-effects ANOVA and will be reported by ICC estimates. The influence of potential fatigue will be analysed by plotting the test and retest results against the NRS and investigating the associations and interactions using a linear mixed effects model. The pairwise correlations of the MMMT and the Neurostatus BMRC with the microFET2 will be graphically analysed (scatterplots and/or boxplots), and Spearman's ρ will be calculated. The primary and secondary analyses will be repeated for subgroups with high and low spasticity patients to compare the sensitivity of the MMMT and Neurostatus BMRC to spasticity.

Expected results: The investigators expect the MMMT for MS to be a reliable and valid manual muscle function test that can be used in the evaluation of the long-term treatment of MS patients with physical therapy.

Being tested
Modified Manual Muscle Test, BMRC manual muscle test, MicroFET2 handhold dynamometer, Modified Tardieu Scale, Fatigue Scale for Motor and Cognitive Functions
Medical subject headings
Multiple Sclerosis

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Modified Manual Muscle TestDiagnostic test

    Manual Muscle Testing Grading System from 0=No visible or palpable contraction to 5= full range of movement against gravity, maximal resistance

  • BMRC manual muscle testDiagnostic test

    muscle scale grades muscle power on a scale of 0 to 5 in relation to the maximum expected for that muscle

  • MicroFET2 handhold dynamometerDiagnostic test

    MicroFET 2 Handheld Dynamometer is a force evaluation testing device to evaluate the strength of various muscle groups

  • Modified Tardieu ScaleDiagnostic test

    Modified Tardieu Scale is to measure if there is spasticity present in a person's muscle and its response to movement

  • Fatigue Scale for Motor and Cognitive FunctionsDiagnostic test

    Measures cognitive and motor fatigue for people with MS

  • numeric rating scale FatigueDiagnostic test

    visual analog scale that uses a scalar numbering system to objectify a patient's fatigue (0 = no fatigue; 10 = most extreme fatigue)

Study groups in NCT03603691
Study groupRoleWhat they receive
Modified Manual Muscle TestActive comparatorThe MMMT will be tested in a test-retest design
Receives: Diagnostic Test: Modified Manual Muscle Test; Diagnostic Test: BMRC manual muscle test; Diagnostic Test: MicroFET2 handhold dynamometer; Diagnostic Test: Modified Tardieu Scale; Diagnostic Test: Fatigue Scale for Motor and Cognitive Functions; Diagnostic Test: numeric rating scale Fatigue
Neurostatus BMRCActive comparatorThe Neurostatus BMRC measures Strength and will be used in a test-retest design
Receives: Diagnostic Test: Modified Manual Muscle Test; Diagnostic Test: BMRC manual muscle test; Diagnostic Test: MicroFET2 handhold dynamometer; Diagnostic Test: Modified Tardieu Scale; Diagnostic Test: Fatigue Scale for Motor and Cognitive Functions; Diagnostic Test: numeric rating scale Fatigue
MicroFET2Active comparatorThe MicroFET2 is a hand held dynamometer to measure strength
Receives: Diagnostic Test: Modified Manual Muscle Test; Diagnostic Test: BMRC manual muscle test; Diagnostic Test: MicroFET2 handhold dynamometer; Diagnostic Test: Modified Tardieu Scale; Diagnostic Test: Fatigue Scale for Motor and Cognitive Functions; Diagnostic Test: numeric rating scale Fatigue

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Inter-rater reliability Modified Manual Muscle Test

    The primary endpoint is the MMMT level. The ordinal levels (2+, 3-, 3, 4-, 4, 4+ and 5) will be transformed to ranks (1-7), and the intra-class correlation coefficient (ICC) of the ranked MMMT levels will be estimated. The aim is to show that the MMMT results are at a clinically relevant - thus high - level of ICC (that is a pre-specified, clinically relevant level the ICC margin).

    The null hypothesis is that the lower limit of the 95% confidence interval (CI) of the ICC is smaller than or equal to the ICC margin. The alternative hypothesis is that the lower limit of the 95% CI of the ICC is larger than the ICC margin.

    If the lower limit of the CI is larger than the ICC margin, the MMMT will be considered a reliable tool for evaluating muscle strength in MS patients.

    Measured: Baseline

Secondary outcomes (3)
  1. Intra-rater reliability Modified Manual Muscle Test

    The intra-rater reliability will be evaluated by comparing the test results of two tests (test-retest), which will be executed by the same tester. The two tests will be executed on the same day within 4 hours. The intra-class correlation coefficient of the ranked MMMT levels (analogue to the primary endpoint) will be used to describe the intra-rater reliability of the MMMT.

    Measured: Baseline- Baseline plus 4 hours

  2. Validity of the Modified Manual Muscle Test

    The inter-rater reliability of the BMRC will be measured by the ICC. The ICC of the BMRC will be estimated and analysed as previously described for the primary analysis. The ICCs of the MMMT and the BMRC will be compared descriptively.

    The test-retest reliability of the MMMT and the BMRC will be estimated for each rater separately using a two-way random-effects ANOVA, including the patient and "test type" (test/retest) as random effects. The ICC estimates will be reported with the 95% CI and compared descriptively between the raters using the same test and between the tests (MMMT vs BMRC).

    The pairwise correlations of the MMMT and the BMRC with the muscle strength as assessed by the microFET2 will be investigated graphically (scatterplots and/or boxplots). Furthermore, Spearman's rank correlation coefficient will be presented with the corresponding 95% CI and p-values.

    Measured: Baseline- Baseline plus 4 hours

  3. Modified Tardieu Scale level

    The BMRC has 6 levels, which makes the scale robust and reliable but insensitive to change. In addition, the BMRC does not consider spasticity. The MMMT has 12 levels, which makes it more sensitive, but it also might mean that the scale is less reliable. Because the MMMT considers spasticity, we expect that this compensates for the possible loss of reliability. It is assumed that the MMMT is less sensitive to spasticity than the BMRC. Thus, for patients with a different level of spasticity measured by the Modified Tardieu Scale, both the inter-rater and the test-retest reliability of the MMMT are expected to be equal to those of the BMRC.

    The primary and secondary analyses will thus be repeated for the subgroups of high-spasticity and low-spasticity patients separately (assuming that a reasonable number of patients per subgroup and endpoint are available).

    Measured: Baseline- Baseline plus 4 hours

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted27 Jul 2018
  2. Study start1 Aug 2018
  3. Primary completion15 Dec 2018
  4. Study completion29 Jan 2019

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Randomised
Intervention model
Crossover
Model description
Prospective, cross-sectional and test-retest
Masking
Single blind
Who is masked
Outcomes assessor
Enrolment
28 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03603691
Sponsor's protocol ID
2018-01103; me16Kappos3
Other IDs
2018
Status last verified
Jan 2019
First submitted
19 Jun 2018
Last updated
30 Jan 2019
Results posted
No
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Undecided

Questions about this trial

What is this clinical trial studying?

A study testing Modified Manual Muscle Test, BMRC manual muscle test and MicroFET2 handhold dynamometer in people with Multiple Sclerosis, run by Nanco van der Maas at 1 site in Switzerland.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 3 inclusion and 4 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 30 Jan 2019. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Switzerland. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Inter-rater reliability Modified Manual Muscle Test (time frame: Baseline). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?