Clinical Evaluation of a Small Aperture Extended Depth of Focus Intraocular Lens

A study testing IC-8 IOL Group in people with Cataract and Presbyopia, run by AcuFocus, Inc. at 21 sites in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to determine the safety and effectiveness of the IC-8 IOL implanted in one eye and a monofocal or monofocal toric IOL implanted in the fellow eye in accordance with the indication. Read more about this study

Who is it for?
Ages 22 years and olderHealthy volunteers can take part
Conditions the study is for
Cataract, Presbyopia
What is being tested?
IC-8 IOL Group
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
453 participants
Where
21 sitesUnited States
Study dates
4 Dec 2018to 15 Oct 2020
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 22 years and older
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • Minimum 22 years of age;
  • Able to comprehend and have signed a statement of informed consent;
  • Availability, willingness, ability and sufficient cognitive awareness to comply with examination procedures and study visits;
  • Planned crystalline lens removal by phacoemulsification, with or without femtosecond laser-assisted extraction, and posterior chamber IOL implantation in both eyes;
  • Cataractous lens changes as demonstrated by best-corrected visual acuity (BCDVA) of 20/40 or worse either with or without a glare source present;
  • Potential for postoperative BCDVA of 20/25 or better in each eye
  • Clear intraocular media, other than cataract.

Who cannot join

  • Requiring an IC-8 intraocular lens outside the available spherical power range
  • Pharmacologically dilated pupil size less than 6 mm in either eye;
  • Inability to achieve stable keratometric readings for contact lens wearers
  • Irregular astigmatism in either eye;
  • Preoperative corneal astigmatism > 1.50 diopters in either eye
  • Active or recurrent anterior segment pathology
Show all 14Show fewer
  • Presence of ocular abnormalities other than cataract as specified in the protocol
  • Diagnosis of dry eye in which patients are unable to maintain eye comfort or adequate vision even with dry eye medication;
  • Congenital cataracts;
  • Previous corneal or intraocular surgery
  • History of ocular trauma or ocular conditions expected to require retinal laser treatment or other surgical intervention;
  • Systemic conditions as specified in the protocol;
  • Patient is pregnant, plans to become pregnant, or is lactating
  • Concurrent participation or participation in any clinical trial up to 30 days prior to preoperative visit.

Trial locations and contacts

21 sites in 1 country.

21 sites in 1 country

United States21 sites
  • Empire Eye & Laser Center

    Bakersfield, California, 93309

  • Chu Vision Institute

    Bloomington, Minnesota, 55420

  • Cleveland Eye Clinic

    Brecksville, Ohio, 44141

  • Hoopes Vision

    Draper, Utah, 84021

  • Eye Center of Northern Colorado

    Fort Collins, Colorado, 80525

  • Physicians Protocol

    Greensboro, North Carolina, 27408

  • Baylor College of Medicine

    Houston, Texas, 77030

  • Harvard Eye Associates

    Laguna Hills, California, 92653

  • Altos Eye Physicians

    Los Altos, California, 94024

  • Advanced Vision Care

    Los Angeles, California, 90067

  • Virginia Eye Consultants

    Norfolk, Virginia, 23502

  • Utah Eye Centers

    Ogden, Utah, 84403

  • Kugler Vision

    Omaha, Nebraska, 68118

  • Eye Center of North Florida

    Panama City, Florida, 32405

  • Philadelphia Eye Associates

    Philadelphia, Pennsylvania, 19148

  • Alterman, Modi & Wolter

    Poughkeepsie, New York, 12603

  • Parkhurst NuVision

    San Antonio, Texas, 78229

  • Vance Thompson Vision

    Sioux Falls, South Dakota, 57108

  • Pepose Vision Institute

    St Louis, Missouri, 63128

  • Fishkind, Bakewell, Maltzman, Hunter & Associates Eye Care & Surgery Center

    Tucson, Arizona, 85704

  • Bucci Laser Vision

    Wilkes-Barre, Pennsylvania, 18702

Other trials in United States

About this study

The purpose of this study is to determine the safety and effectiveness of the IC-8 IOL implanted in one eye and a monofocal or monofocal toric IOL implanted in the fellow eye in accordance with the indication.

Read the detailed description

Eligible patients will complete up to 12 study visits over a 12-month period. All subjects will complete a preoperative examination of both eyes to assess study eligibility: operative visit (each eye) and up to 9 postoperative visits (Day 1 - each eye, Week 1 - each eye, Month 1 - each eye or combined, Month 3 - both eyes, Month 6 - both eyes, and Month 12 - both eyes). The first eye must meet specific qualification criteria in order to proceed with IOL implantation in the second eye. The second eye should be implanted within 45 days of the first eye.

Conditions
Cataract, Presbyopia
Being tested
IC-8 IOL Group, Control Group
Keywords
Cataract, Intraocular lens
Medical subject headings
Cataract, Presbyopia

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • IC-8 IOL GroupDevice

    Also called Intraocular Lens

    A monofocal IOL (TECNIS ZCB00, ZCT150 or ZCT225 or AcrySof IQ SA60WF, SA6AT3 or SA6AT4) implanted in the first eye for use over the lifetime of the patient. The IC-8 IOL implanted in the second eye for use over the lifetime of the patient.

  • Control GroupDevice

    Also called Intraocular lens

    A monofocal IOL (TECNIS ZCB00, ZCT150 or ZCT225 or AcrySof IQ SA60WF, SA6AT3 or SA6AT4) bilaterally implanted for lifetime use of the patient.

Study groups in NCT03633695
Study groupRoleWhat they receive
IC-8 IOL GroupExperimentalA monofocal IOL (TECNIS ZCB00, ZCT150 or ZCT225 or AcrySof IQ SA60WF, SA6AT3 or SA6AT4) implanted in the first eye. The AcuFocus IC-8 IOL implanted in the second eye.
Receives: Device: IC-8 IOL Group
Control GroupActive comparatorA monofocal IOL (TECNIS ZCB00, ZCT150 or ZCT225 or AcrySof IQ SA60WF, SA6AT3 or SA6AT4) bilaterally implanted.
Receives: Device: Control Group

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Mean Binocular Photopic Uncorrected Intermediate (66 cm) Visual Acuity (UCIVA)

    Visual acuity (VA) was tested binocularly (both eyes) at a distance of 66 cm, without manifest refraction, using an Early Treatment Diabetic Retinopathy Study (ETDRS) chart presenting on a self-calibrating monitor in a computerized testing system. Visual acuity was measured in logarithm of the minimum angle of resolution (logMAR), with a 0.02 logMAR increment corresponding to a single letter on the ETDRS chart.

    Measured: 6 Months (160-210 days post second eye operative visit)

  2. Mean Binocular Photopic Uncorrected Near (40 cm) Visual Acuity (UCNVA)

    Visual acuity (VA) was tested binocularly (both eyes) at a distance of 40 cm, without manifest refraction, using an Early Treatment Diabetic Retinopathy Study (ETDRS) chart presenting on a self-calibrating monitor in a computerized testing system. Visual acuity was measured in logarithm of the minimum angle of resolution (logMAR), with a 0.02 logMAR increment corresponding to a single letter on the ETDRS chart.

    Measured: 6 Months (160-210 days post second eye operative visit)

  3. Mean Binocular Photopic Uncorrected Distance (4 m) Visual Acuity (UCDVA)

    Visual acuity (VA) was tested binocularly (both eyes) at a distance of 4 m, with +0.25 D infinity adjustment lens in front of the eyes, using an Early Treatment Diabetic Retinopathy Study (ETDRS) chart presenting on a self-calibrating monitor in a computerized testing system. Visual acuity was measured in logarithm of the minimum angle of resolution (logMAR), with a 0.02 logMAR increment corresponding to a single letter on the ETDRS chart.

    Measured: 6 Months (160-210 days post second eye operative visit)

  4. Mean Monocular Photopic Distance-Corrected Intermediate (66 cm) Visual Acuity (DCIVA) in IC-8™ IOL Eyes

    Visual acuity (VA) was tested monocularly (each eye) at a distance of 66 cm, with the distance manifest refraction (with infinity adjustment) in place for that eye, using an Early Treatment Diabetic Retinopathy Study (ETDRS) chart presenting on a self-calibrating monitor in a computerized testing system. Visual acuity was measured in logarithm of the minimum angle of resolution (logMAR), with a 0.02 logMAR increment corresponding to a single letter on the ETDRS chart.

    Measured: 6 Months (160-210 days post second eye operative visit)

  5. Mean Monocular Photopic Depth of Focus (DOF) in IC-8™ IOL Eyes

    Depth of focus was performed monocularly in the IC-8™ IOL Group at the far to near range of vision (+2.00 to -5.00 D), using 100% contrast ETDRS charts in a computerized testing system calibrated for a 4 m test distance, with manifest refraction (no infinity adjustment) in place for the eye(s) being tested. The defocus power was progressively introduced in 0.50 D increments from +2.00 D to +0.50 D and from -0.50 D to -5.00 D, then in 0.25 D increments from +0.50 D to -0.50 D, while visual acuity was measured at each successive defocus step. The depth of focus was estimated as the defocus range between zero defocus and the first point on the negative lens induced mean defocus curve that crosses the 0.2 logMAR using a linear interpolation.

    Measured: 3 Months (60-110 days post second eye operative visit)

  6. Mean Monocular Best-corrected Distance (4 m) Visual Acuity (BCDVA)

    Visual acuity (VA) was tested monocularly (each eye) at a distance of 4 m, in photopic conditions, with manifest refraction in place (no infinity adjustment) using an Early Treatment Diabetic Retinopathy Study (ETDRS) chart presenting on a self-calibrating monitor in a computerized testing system. Visual acuity was measured in logarithm of the minimum angle of resolution (logMAR), with a 0.02 logMAR increment corresponding to a single letter on the ETDRS chart.

    Measured: 12 Months (300-420 days post second eye operative visit)

  7. Proportion of IC-8™ IOL Eyes Achieving Best-corrected Distance Visual Acuity (BCDVA) 0.3 logMAR or Better Compared to the Safety and Performance Endpoints (SPE) Rates

    Visual acuity (VA) was tested monocularly (each eye) at a distance of 4 m, in photopic conditions, with manifest refraction in place (no infinity adjustment) using an Early Treatment Diabetic Retinopathy Study (ETDRS) chart presenting on a self-calibrating monitor in a computerized testing system. Visual acuity was measured in logarithm of the minimum angle of resolution (logMAR), with a 0.02 logMAR increment corresponding to a single letter on the ETDRS chart.

    Measured: 12 Months (300-420 days post second eye operative visit)

  8. Proportion of Best-Case IC-8™ IOL Eyes Achieving Best-corrected Distance Visual Acuity (BCDVA) 0.3 logMAR or Better Compared to the Safety and Performance Endpoints (SPE) Rates

    Visual acuity (VA) was tested monocularly (each eye) at a distance of 4 m, in photopic conditions, with manifest refraction in place (no infinity adjustment) using an Early Treatment Diabetic Retinopathy Study (ETDRS) chart presenting on a self-calibrating monitor in a computerized testing system. Visual acuity was measured in logarithm of the minimum angle of resolution (logMAR), with a 0.02 logMAR increment corresponding to a single letter on the ETDRS chart. Monocular BCDVA for the IC-8™ IOL eyes of subjects in the Best-Case population was compared to the safety and performance endpoints (SPE) rate for posterior chamber IOLs.

    Measured: 12 Months (300-420 days post second eye operative visit)

  9. Rates of IC-8™ IOL Eyes With Cumulative Ocular Serious Adverse Events

    The rates of cumulative ocular serious adverse events (SAEs) for IC-8™ IOL eyes were calculated from the time of second eye (IC-8™ IOL eye) implantation through Month 12.

    Measured: Through 12 Months (300-420 days post second eye operative visit)

  10. Rates of IC-8™ IOL Eyes With Cumulative Postoperative Ocular Adverse Events (Secondary Surgical Interventions)

    The rates of cumulative ocular adverse events for IC-8™ IOL eyes were calculated from the time of second eye (IC-8™ IOL eye) implantation through Month 12.

    Measured: Through 12 Months (300-420 days post second eye operative visit)

  11. Rates of IC-8™ IOL Eyes With Persistent Ocular Serious Adverse Events

    The rates of persistent (defined as being unresolved at final scheduled visit) ocular adverse events (AEs) for IC-8™ IOL eyes were calculated from the time of second eye (IC-8™ IOL eye) implantation through Month 12.

    Measured: Through 12 Months (300-420 days post second eye operative visit)

  12. Rates of IC-8™ IOL Eyes With IC-8™ IOL Removals

    The rates of cumulative ocular adverse events for IC-8™ IOL eyes were calculated from the time of second eye (IC-8™ IOL eye) implantation through Month 12.

    Measured: Through 12 Months (300-420 days post second eye operative visit)

  13. Rates of Eyes With Postoperative Ocular Adverse Events (Other Than Secondary Surgical Intervention), Related to Device (Serious and Non-Serious Combined)

    The rates of cumulative ocular adverse events (AEs) for IC-8™ IOL eyes were calculated from the time of second eye (IC-8™ IOL eye) implantation through Month 12.

    Measured: Through 12 Months (300-420 days post second eye operative visit)

Secondary outcomes (9)
  1. Tolerance to Preoperative Corneal Astigmatism in IC-8™ IOL Eyes With BCDVA 20/25 at 3 Months

    Uncorrected-distance visual acuity (UCDVA) in IC-8™ IOL eyes achieving best-corrected distance visual acuity (BCDVA) of 20/25 or better was compared between eyes with preoperative corneal astigmatism < 1.0 D to eyes with 1.0 D to 1.5 D of preoperative corneal astigmatism.

    Measured: 3 Months (60-110 days post second eye operative visit)

  2. Mean Monocular Photopic Contrast Sensitivity Without Glare

    Contrast sensitivity (the ability to distinguish between an object and it's background) was evaluated in a subgroup of subjects and was tested monocularly (each eye) in photopic (well-lit) conditions, without a glare source, using sine waving gratings in a computerized testing system. The photopic spatial frequencies tested were 3, 6, 12, 18 cycles per degree (cpd). When a subject was unable to see the highest contrast grating, the highest contrast score was assigned. Raw scores (0 to 9) from contrast sensitivity testing were transformed to log units. Better contrast sensitivity results are reflected by a higher numeric value.

    Measured: 6 Months (160-210 days post second eye operative visit)

  3. Mean Monocular Photopic Contrast Sensitivity With Glare

    Contrast sensitivity (the ability to distinguish between an object and it's background) was evaluated in a subgroup of subjects, and was tested monocularly (each eye) in photopic (well-lit) conditions, with a glare source, using sine waving gratings in a computerized testing system. The photopic spatial frequencies tested were 3, 6, 12, 18 cycles per degree (cpd). When a subject was unable to see the highest contrast grating, the highest contrast score was assigned. Raw scores (0 to 9) from contrast sensitivity testing were transformed to log units. Better results are reflected by a higher numeric value.

    Measured: 6 Months (160-210 days post second eye operative visit)

  4. Mean Monocular Mesopic Contrast Sensitivity Without Glare

    Contrast sensitivity (the ability to distinguish between an object and it's background) was evaluated in a subgroup of subjects and was tested monocularly (each eye) in mesopic (low-light) conditions, without a glare source, using sine waving gratings in a computerized testing system. The mesopic spatial frequencies tested were 1.5, 3, 6, 12 cycles per degree (cpd). When a subject was unable to see the highest contrast grating, the highest contrast score was assigned. Raw scores (0 to 9) from contrast sensitivity testing were transformed to log units. Better results are reflected by a higher numeric value.

    Measured: 6 Months (160-210 days post second eye operative visit)

  5. Mean Monocular Mesopic Contrast Sensitivity With Glare

    Contrast sensitivity (the ability to distinguish between an object and it's background) was evaluated in a subgroup of subjects and was tested monocularly (each eye) in mesopic (low-light) conditions, with a glare source, using sine waving gratings in a computerized testing system. The mesopic spatial frequencies tested were 1.5, 3, 6, 12 cycles per degree (cpd). When a subject was unable to see the highest contrast grating, the highest contrast score was assigned. Raw scores (0 to 9) from contrast sensitivity testing were transformed to log units. Better results are reflected by a higher numeric value.

    Measured: 6 Months (160-210 days post second eye operative visit)

  6. Mean Binocular Photopic Contrast Sensitivity Without Glare

    Contrast sensitivity (the ability to distinguish between an object and it's background) was evaluated in a subgroup of subjects and was tested binocularly (both eyes) in photopic (well-lit) conditions, without a glare source, using sine waving gratings in a computerized testing system. The photopic spatial frequencies tested were 3, 6, 12, 18 cycles per degree (cpd). When a subject was unable to see the highest contrast grating, the highest contrast score was assigned. Raw scores (0 to 9) from contrast sensitivity testing were transformed to log units. Better results are reflected by a higher numeric value.

    Measured: 6 Months (160-210 days post second eye operative visit)

  7. Mean Binocular Photopic Contrast Sensitivity With Glare

    Contrast sensitivity (the ability to distinguish between an object and it's background) was evaluated in a subgroup of subjects and was tested binocularly (both eyes) in photopic (well-lit) conditions, with a glare source, using sine waving gratings in a computerized testing system. The photopic spatial frequencies tested were 3, 6, 12, 18 cycles per degree (cpd). When a subject was unable to see the highest contrast grating, the highest contrast score was assigned. Raw scores (0 to 9) from contrast sensitivity testing were transformed to log units. Better results are reflected by a higher numeric value.

    Measured: 6 Months (160-210 days post second eye operative visit)

  8. Mean Binocular Mesopic Contrast Sensitivity Without Glare

    Contrast sensitivity (the ability to distinguish between an object and it's background) was evaluated in a subgroup of subjects and was tested binocularly (both eyes) in mesopic (low-light) conditions, without a glare source, using sine waving gratings in a computerized testing system. The mesopic spatial frequencies tested were 1.5, 3, 6, 12 cycles per degree (cpd). When a subject was unable to see the highest contrast, the highest contrast grating, the highest contrast score was assigned. Raw scores (0 to 9) from contrast sensitivity testing were transformed to log units. Better results are reflected by a higher numeric value.

    Measured: 6 Months (160-210 days post second eye operative visit)

  9. Mean Binocular Mesopic Contrast Sensitivity With Glare

    Contrast sensitivity (the ability to distinguish between an object and it's background) was evaluated in a subgroup of subjects and was tested binocularly (both eyes) in mesopic (low-light) conditions, with a glare source, using sine waving gratings in a computerized testing system. The mesopic spatial frequencies tested were 1.5, 3, 6, 12 cycles per degree (cpd). When a subject was unable to see the highest contrast grating, the highest contrast score was assigned. Raw scores (0 to 9) from contrast sensitivity testing were transformed to log units. Better results are reflected by a higher numeric value.

    Measured: 6 Months (160-210 days post second eye operative visit)

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted16 Aug 2018
  2. Study start4 Dec 2018
  3. Primary completion15 Oct 2020
  4. Study completion15 Oct 2020

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Non-randomised
Intervention model
Parallel groups
Masking
Single blind
Who is masked
Outcomes assessor
Enrolment
453 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03633695
Sponsor's protocol ID
SAIL-101-UNI
Status last verified
Sep 2022
First submitted
14 Aug 2018
Last updated
24 Oct 2022
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Publications

  • Vukich J, Thompson V, Yeu E, Wiley WF, Bafna S, Koch DD, Lin L, Michna M. Evaluating the small aperture intraocular lens: depth of focus and the role of refraction and preoperative corneal astigmatism in visual performance. J Cataract Refract Surg. 2024 Nov 1;50(11):1165-1172. doi: 10.1097/j.jcrs.0000000000001524.

    PubMed 39075732

Questions about this trial

What is this clinical trial studying?

A study testing IC-8 IOL Group in people with Cataract and Presbyopia, run by AcuFocus, Inc. at 21 sites in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 7 inclusion and 14 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 24 Oct 2022. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 21 sites in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 24 Oct 2022). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Mean Binocular Photopic Uncorrected Intermediate (66 cm) Visual Acuity (UCIVA) (time frame: 6 Months (160-210 days post second eye operative visit)). The study lists 13 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT03633695, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 24 Oct 2022.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in