A Pivotal Study to Evaluate Guardian™ Connect CGM System Performance in China

A study testing Guardian™ Sensor (3) in people with Diabetes Mellitus, run by Medtronic MiniMed, Inc. at 3 sites in China.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The purpose of this study is to demonstrate the performance and safety of the Guardian™ Sensor (3) over 170 hours (7 days) when inserted in the abdomen used in subjects age 14 - 75 years with the Guardian™ Connect System. Read more about this study

Who is it for?
Ages 14 to 75 years
Conditions the study is for
Diabetes Mellitus
What is being tested?
Guardian™ Sensor (3)
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
72 participants
Where
3 sitesChina
Study dates
28 Feb 2019to 25 Jul 2019
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 14 to 75 years
  • All sexes
  • No healthy volunteers
  • Child, Adult, Older adult

Who may be able to join

  • Subject is 14 - 75 years of age at time of screening
  • Subject has a clinical diagnosis of type 1 or 2 diabetes as determined via medical record or source documentation by an individual qualified to make a medical diagnosis
  • Subject has adequate venous access as assessed by investigator or appropriate staff
  • Subject is willing to follow the study procedures and willing to come to study visits

Who cannot join

  • Subject will not tolerate tape adhesive in the area of Guardian™ Sensor (3) placement as assessed by qualified individual.
  • Subject has any unresolved adverse skin condition in the area of study device or device placement (e.g., psoriasis, rash, Staphylococcus infection)
  • Subject is actively participating in an investigational study (drug or device) wherein they have received treatment from an investigational study (drug or device) in the last 2 weeks
  • Subject is female and has a positive pregnancy screening test
  • Females of child bearing age and who are sexually active should be excluded if they are not using a form of contraception deemed reliable by investigator
  • Subject is female and plans to become pregnant during the course of the study
  • Subject has a hematocrit (Hct) lower than the normal reference range
  • Subject may not be on the research staff of those performing this study

Trial locations and contacts

3 sites in 1 country.

China3 sites
  • The Seventh Medical Center of PLA General Hospital

    Beijing, 100700

  • Sir Run Run Shaw Hospital

    Hangzhou, 310016

  • Shanghai Sixth People's Hospital

    Shanghai, 200233

Other trials in China

About this study

The purpose of this study is to demonstrate the performance and safety of the Guardian™ Sensor (3) over 170 hours (7 days) when inserted in the abdomen used in subjects age 14 - 75 years with the Guardian™ Connect System.

Read the detailed description

This study is a multi-center, prospective single-arm design without controls. Up to 72 subjects will be enrolled in order to have approximately 60 subjects complete the study. Three investigational centers in China will be used during this study.

During the study, each subject will be randomly assigned to one day of the Yellow Springs Instrument (YSI™*) frequent sample testing (FST) (Day 1, 3-5, or 7).Subjects will wear two Guardian™ Sensor (3)s each connected to a Guardian™ Connect transmitter for approximately 7 days (one which will be paired to the Guardian Connect app and the other will function as a glucose recorder.).

The Guardian Sensor (3)s will be worn in the abdomen area and self-inserted by the subject on same side or opposite sides.

On the evening prior to FST, subjects will be asked to fast for approximately 12 hours and adjust their insulin and medications according to routine care (for example as they would do for fasting lipid panel). Subjects may fast for shorter period of time based on investigator discretion.

The subject should be in fasting status upon arrival hospital to start FST process. The feeding protocol may be modified based on investigator discretion. The duration of FST will be approximately 7 hours.

During the study, subjects will continue with their current diabetes regimen independent of the study devices. Subjects will be instructed by the investigational center that they are not to use the study devices (except for the CONTOUR™* study meter) for the management of their diabetes. The CONTOUR™* study meter may be used for treatment decisions and calibration of Guardian™ Sensor (3).

Being tested
Guardian™ Sensor (3)
Keywords
Diabetes Type I Type II
Medical subject headings
Diabetes Mellitus

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Guardian™ Sensor (3)Device

    Use two Guardian™ Sensor (3)s each connected to a Guardian™ Connect transmitter for approximately 7 days.

Study groups in NCT03710083
Study groupRoleWhat they receive
Study armExperimentalUse two Guardian™ Sensor (3)s each connected to a Guardian™ Connect transmitter for approximately 7 days and undergo one Yellow Springs Instrument (YSI™) frequent sample testing (FST) (Day 1, 3-5, or 7).
Receives: Device: Guardian™ Sensor (3)

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Percentage of Guardian™ Sensor (3) Values That Are Within 20% Agreement of YSI

    Percentage of Guardian™ Sensor (3) Values that are within 20% agreement of Yellow Springs Instrument (YSI) plasma glucose values (±20 mg/dL (1.1 mmol/L) when reference blood glucose (YSI) less than or equal to (≤) 80 mg/dL (4.4 mmol/L) ) during YSI frequent sampling testing (FST) days.

    Measured: 168 Hours

  2. Consensus Error Grid Analysis of Mean Rate of Paired Sensor Values and YSI Values in Zone A+B

    A mean rate in Zone A+B of Consensus Error Grid between Guardian Sensor (3) values and YSI™* plasma glucose values during YSI™* FST days was evaluated. Consensus Error Grid (or Parkes error grid) compared the paired primary sensor and YSI reference glucose values. Zone A is defined in the Parkes error grid as the zone of "clinical accurate measurements with no effect on clinical action", Zone B as "altered clinical action with little or no effect on clinical outcome". Ideal situation is 100% in Zone A + B.

    Measured: 168 Hours

  3. Clarke Error Grid Analysis of Mean Rate of Paired Sensor Values and YSI Values in Zone A+B

    A mean rate in Zone A+B of Clarke Error Grid between Guardian Sensor (3) values and YSI™* plasma glucose values during YSI™* FST days was evaluated. Clarke Error Grid compared the paired primary sensor and YSI reference glucose values. Zone A are "clinically accurate in that they would lead to clinically correct treatment decisions". Zone B "would lead to benign or no treatment". Ideal situation is 100% in Zone A + B.

    Measured: 168 Hours

  4. Mean Absolute Relative Difference (MARD)

    Mean absolute relative difference (MARD) between Guardian Sensor (3) values and YSI™* plasma glucose values during YSI™* FST days was evaluated. MARD = Mean of ((Absolute difference of YSI reference and Sensor glucose values / YSI reference glucose values) * 100). Therefore, the unit of MARD is percentage (%). Note that results from multiple FST days were pooled together for reporting purpose.

    Measured: 168 Hours

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted17 Oct 2018
  2. Study start28 Feb 2019
  3. Primary completion25 Jul 2019
  4. Study completion25 Jul 2019

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Treatment
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
72 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03710083
Sponsor's protocol ID
CIP320
Status last verified
Aug 2020
First submitted
15 Oct 2018
Last updated
9 Sep 2020
FDA-regulated drug
No
FDA-regulated device
Yes

Questions about this trial

What is this clinical trial studying?

A study testing Guardian™ Sensor (3) in people with Diabetes Mellitus, run by Medtronic MiniMed, Inc. at 3 sites in China.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 8 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 9 Sep 2020. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 3 sites in China. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 21 Aug 2020). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Percentage of Guardian™ Sensor (3) Values That Are Within 20% Agreement of YSI (time frame: 168 Hours). The study lists 4 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 9 Sep 2020.

Free account

Keep searching with a free account

You’ve used your 3 free searches for this session. Log in or create a free AzTrials account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in