AzTrials
AzTrials

HealthyNCT03783780

Desaturation Validation of INVSENSOR00031

A study testing INVSENSOR00031 in people with Healthy, run by Masimo Corporation at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

This study is designed to compare the accuracy of a noninvasive measurement of oxygen saturation compared to reference values obtained by a laboratory blood gas analyzer. Arterial blood samples will be collected from healthy adult subjects while undergoing a desaturation procedure wherein the concentration of oxygen… Read more about this study

Who is it for?
Ages 18 to 50 yearsHealthy volunteers can take part
Conditions the study is for
Healthy
What is being tested?
INVSENSOR00031
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
21 participants
Where
1 siteUnited States
Study dates
11 Dec 2018to 18 Dec 2018
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 to 50 years
  • All sexes
  • Healthy volunteers accepted
  • Adult

Who may be able to join

  • Subject is 18-50 years of age.
  • Subject weighs a minimum of 110 lbs and no more than 250 lbs unless subject is over 6 feet tall.
  • Hemoglobin value is greater than or equal to 11 g/dL.
  • Baseline heart rate ≥ 45 bpm and ≤ 85 bpm.
  • Carbon monoxide (CO) value ≤ 2.0% fractional carboxyhemoglobin (FCOHb)
  • Subject has a physical status of ASA I or II (American Society of Anesthesiology Class I; Healthy subjects without any systemic disease at all. American Society of Anesthesiology Class II; subjects with mild systemic disease) as it applies to the systemic disease portion of the classification.
  • Systolic Blood Pressure ≤ 140 mmHg and Diastolic Blood Pressure ≤ 90 mmHg.
  • Subject is able to read and communicate in English and understands the study and risks involved.

Who cannot join

  • Subject is pregnant.
  • Subject smokes (smoking includes e-cigarette use).
  • Subject has a BMI > 35 and has been classified as morbidly obese or at an increased risk for participation by a medical professional.
  • Subject has a history of fainting (vasovagal), blacking out or losing consciousness during or after blood draw, or has a fear of blood draws.
  • Subject has open wounds, inflamed tattoos or piercings, or any visible healing wounds that a medical professional renders them at an increased risk for participation.*
  • Subject experiences frequent or severe headaches and/or migraine headaches.
Show all 31Show fewer
  • Subject has known drug or alcohol abuse and/or use of recreational drugs.
  • Subject has experienced a concussion or head injury with loss of consciousness within the last 12 months.
  • Subject has any chronic bleeding disorders (i.e. hemophilia).
  • Subject has any history of a stroke, myocardial infarction, seizures or heart attack.
  • Subject has any cancer or history of cancer (not including skin cancer).*
  • Subject has any cardiac dysrhythmia(s) (i.e. atrial fibrillation) and has not received clearance by their physician to participate.
  • Subject has a known neurological and/or psychiatric disorder (i.e. schizophrenia, bipolar disorder, Multiple Sclerosis, Huntington's Disease) that interferes with the subject's level of consciousness.
  • Subject has Wolff-Parkinson-White Syndrome or Stokes-Adams Syndrome.
  • Subject who has taken anticoagulant medication within the last 30 days (excluding nonsteroidal anti-inflammatory drugs (NSAIDS)).
  • Subject has taken opioid pain medication 24 hours before the study.
  • Subject has any type of infectious disease (i.e. Hepatitis, HIV, Tuberculosis, Flu, Malaria, Measles, etc.)
  • Subject is taking medications known to treat any type of infectious disease.
  • Subject has either signs or history of peripheral ischemia.
  • Subject has had invasive surgery within the past year- including but not limited to major dental surgery*, appendix*, plastic surgery*.
  • Subject has had invasive surgery within the past year- including but not limited to gallbladder, heart, major fracture repairs (involving plates/ screws), jaw surgery, urinary tract surgery, major ears, nose and throat (ENT) surgery, joint replacement or gynecological surgeries, heart surgery or thoracic surgery.
  • Subject has donated blood within the past 4 weeks.
  • Subject has symptoms of congestion, head colds, flu or other illnesses.
  • Subject experiences claustrophobia or has generalized anxiety disorder.
  • Subject has been in severe car accident(s) or a similar type of accident(s) requiring hospitalization within the last 12 months.
  • Subject has chronic unresolved asthma, lung disease or respiratory disease.
  • Subject is allergic to lidocaine, latex, adhesives, or plastic.
  • Subject has heart conditions, insulin-dependent Diabetes or uncontrolled hypertension.
  • Subject has given vaginal delivery, had a pregnancy terminated, a miscarriage with hospitalization, or had a C-section within the last 6 months.
  • Subject intends to participate in any heavy lifting, repetitive movement of their wrist (including riding a motorcycle) or exercise (working out, riding a bike, riding a skate board etc.), or any activity that will put additional stress on the wrist within 24 hours of the study.
  • Subject has any medical condition which in the judgment of the investigator and/or medical staff, renders them ineligible for participation in this study or subject is deemed ineligible by the discretion of the investigator/study.

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Masimo Corporation

    Irvine, California, 92618

Other trials in United States

About this study

This study is designed to compare the accuracy of a noninvasive measurement of oxygen saturation compared to reference values obtained by a laboratory blood gas analyzer. Arterial blood samples will be collected from healthy adult subjects while undergoing a desaturation procedure wherein the concentration of oxygen inhaled is slowly reduced until the subject's arterial oxygen concentration is approximately 70%. During the desaturation, the subject's hand and arm without the arterial line will be subjected to motion at various times during the study. The subject will then be returned to breathing room air.

Conditions
Healthy
Being tested
INVSENSOR00031

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • INVSENSOR00031Device

    Noninvasive pulse oximeter sensor

Study groups in NCT03783780
Study groupRoleWhat they receive
INVSENSOR00031ExperimentalAll subjects who are enrolled into the test group and participate in data collection receive the noninvasive INVSENSOR00031 sensor during motion and non-motion.
Receives: Device: INVSENSOR00031

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Oxygen Saturation (SpO2) Accuracy of Sensor by Arms Calculation During Non-motion

    Performance of the sensors will be determined by comparing the noninvasive oxygen saturation measurement (SpO2) of the pulse oximeter sensors to the arterial oxygen saturation (SaO2) value obtained from a reference blood sample and calculating the Arithmetic root mean square (ARMS) value during non-motion.

    Measured: 1-5 hours

Secondary outcomes (1)
  1. Oxygen Saturation (SpO2) Accuracy of Sensor by Arms Calculation During Motion

    Performance of the sensors will be determined by comparing the noninvasive oxygen saturation measurement (SpO2) of the pulse oximeter sensors to the arterial oxygen saturation (SaO2) value obtained from a reference blood sample and calculating the Arithmetic root mean square (ARMS) value during motion.

    Measured: 1-5 hours

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted21 Dec 2018
  2. Study start11 Dec 2018
  3. Primary completion18 Dec 2018
  4. Study completion18 Dec 2018

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Other
Allocation
Not applicable
Intervention model
Single group
Masking
None (open label)
Enrolment
21 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03783780
Sponsor's protocol ID
TP-19564
Status last verified
Nov 2019
First submitted
19 Dec 2018
Last updated
17 Dec 2019
FDA-regulated drug
No
FDA-regulated device
Yes

Questions about this trial

What is this clinical trial studying?

A study testing INVSENSOR00031 in people with Healthy, run by Masimo Corporation at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 8 inclusion and 31 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 17 Dec 2019. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 2 Jul 2019). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Oxygen Saturation (SpO2) Accuracy of Sensor by Arms Calculation During Non-motion (time frame: 1-5 hours). It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 17 Dec 2019.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in
Desaturation Validation of INVSENSOR00031 | AzTrials