Glaucoma, Open-AngleNCT03804684

visuALL Field Analyzer (vFA) Compared to Standard Humphrey Automated Perimetry

A study testing Standard Automatic Perimetry Humphrey Field Analyzer and visuALL Field Analyzer in people with Glaucoma, Open-Angle, run by Wills Eye at 1 site in United States.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

To determine age-adjusted reference values of the visuALL Field Analyzer (vFA) retinal sensitivity and to assess the repeatability of the measured values and compare them with those measured by a Standard Automatic Perimetry (SAP). Read more about this study

Who is it for?
Ages 21 to 80 yearsHealthy volunteers can take part
Conditions the study is for
Glaucoma, Open-Angle
What is being tested?
Standard Automatic Perimetry Humphrey Field Analyzer and visuALL Field Analyzer
Could I receive a placebo?
No, everyone receives an active treatment
How many people
72 participants
Where
1 siteUnited States
Study dates
8 Jan 2019to 20 Oct 2020
Phase
Interventional
Are there results?
Yes, posted in the registrySummary in the doctors' view

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 21 to 80 years
  • All sexes
  • Healthy volunteers accepted
  • Adult, Older adult

Who may be able to join

  • age 21 to 80 years
  • Healthy subjects with no eye diseases, normal appearing optic nerve and retina, intraocular pressure less than 19 millimeters mercury and normal standard automatic perimetry (SAP).
  • mild glaucoma subjects with less than -6 mean deviation
  • moderate glaucoma subjects with -6 to -12 mean deviation

Who cannot join

  • spherical refraction outside ± 3.0 diopters and cylinder correction outside 2.0 diopters
  • Unreliable SAP (false positives, fixation losses and false negatives greater than 25% and/or observable testing artifacts)
  • Unreliable visuALL Field Analyzer (vFA) (greater than 25% false positive, excessive fixation losses)
  • SAP abnormality with pattern loss consistent with neurologic and/or other ocular diseases than glaucoma
  • Intraocular surgery in study eye (except non-complicated cataract or refractive surgery performed more than 6 months before enrollment and without posterior capsule opacification)
  • History of systemic condition known to affect visual function
  • History of medication known to affect visual function

Trial locations and contacts

1 site in 1 country.

United States1 site
  • Wills Eye Hospital

    Philadelphia, Pennsylvania, 19107

Other trials in United States

About this study

Official title: How Does the visuALL Field Analyzer (vFA) Compare to Standard Automatic Perimetry (SAP) in Patients With Mild and Moderate Chronic Open Angle Glaucoma (COAG) and Subjects Without Eye Problems?

To determine age-adjusted reference values of the visuALL Field Analyzer (vFA) retinal sensitivity and to assess the repeatability of the measured values and compare them with those measured by a Standard Automatic Perimetry (SAP).

Read the detailed description

Standard Automatic Perimetry (SAP) is the gold standard test for the evaluation of central and peripheral retinal sensitivity to light for detection and monitoring optic nerve diseases such as glaucoma. The current device has its own limitation such as positioning for the duration of the test which makes is hard for older patients especially those with back or other musculoskeletal diseases, artifact of the corrective lenses used to accommodate the patients refractive errors, etc.

Several devices have been developed since the advent of the Octopus Perimeter 3-5 and the Humphrey Field Analyzer (HFA), in an effort to improve the test reliability and patient comfort.

The main goal of this study is to evaluate the repeatability of a novel psychophysical platform that takes advantage of a Head Mounted Device (HMD) with eye tracking capabilities. Other objectives of this study include the development of an initial reference database and comparison of the measured parameters with HFA.

Being tested
Standard Automatic Perimetry Humphrey Field Analyzer, visuALL Field Analyzer
Keywords
open angle glaucoma, visual field, Standard Automatic Perimetry
Medical subject headings
Glaucoma, Open-Angle

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • Standard Automatic Perimetry Humphrey Field AnalyzerDiagnostic test

    Also called SAP HFA

    Standard Automatic Perimetry (SAP) Humphrey Field Analyzer (HFA) 24-2, Swedish Interactive Threshold Algorithm (SITA) Standard Strategy measures peripheral and central vision in an ophthalmic setting.

  • visuALL Field AnalyzerDiagnostic test

    Also called vFA

    visuALL Field Analyzer a new portable hardware and software virtual reality system measures peripheral and central vision to improve early detection of glaucoma damage in a non ophthalmic setting.

Study groups in NCT03804684
Study groupRoleWhat they receive
Healthy ControlsExperimentalHealthy subjects between 21 and 80 years of age with no eye diseases (normal appearing optic nerve and retina, intraocular eye pressure less than 19 millimeters of mercury, normal and reliable standard automatic perimetry, and spherical refraction less than 3 diopters and cylinder correction equal to 2 diopters or less. All subjects will perform Standard Automatic Perimetry Humphrey Field Analyzer and visuALL Field Analyzer to measure peripheral and central visual field.
Receives: Diagnostic Test: Standard Automatic Perimetry Humphrey Field Analyzer; Diagnostic Test: visuALL Field Analyzer
Mild and Moderate GlaucomaExperimentalSubjects between 21 and 80 years of age with Mild or Moderate Glaucoma. Eyes will have a reliable standard automatic perimetry with no more than -6 mean deviation (Mild) and between -6 mean and -12 mean deviation (Moderate). Spherical refraction will be less than 3 diopters and cylinder correction equal to 2 diopters or less. All subjects will perform Standard Automatic Perimetry Humphrey Field Analyzer and visuALL Field Analyzer to measure peripheral and central visual field.
Receives: Diagnostic Test: Standard Automatic Perimetry Humphrey Field Analyzer; Diagnostic Test: visuALL Field Analyzer

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Mean Retinal Sensitivity From Two Machines

    Mean (Average) retinal sensitivity to light as measured in decibels in multiple spots across the central and peripheral retina (entire visual field). Higher numbers translate to dimmer light which when detected means better vision.

    Measured: three hours

  2. Global Correlation (Agreement Between Two Machines in the Overall Visual Field)

    Correlation (Agreement) between Humphrey Visual Field and VisuALL retinal sensitivity as measured by mean deviation (light measurement in decibels) in multiple spots across the central and peripheral retina (entire visual field). This is the amount of agreement of the measure threshold, which is the lowest level of light detected in different points across the visual field (overall), between two testing machines.

    Measured: three hours

Secondary outcomes (2)
  1. Inferior Nasal Quadrant Correlation (Agreement Between Two Machines in the Inferior Nasal Visual Field)

    Correlation (Agreement) between Humphrey Visual Field and VisuALL retinal sensitivity as measured by mean deviation (light measurement in decibels) in multiple spots across the inferior nasal visual field. This is the amount of agreement of the measure threshold, which is the lowest level of light detected in different points across the inferior (lower) nasal portion of the visual field, between two testing machines.

    Measured: three hours

  2. Superior Hemifield Correlation (Agreement Between Two Machines in the Superior Hemifield Visual Field

    Correlation (Agreement) between Humphrey Visual Field and VisuALL retinal sensitivity as measured by mean deviation (light measurement in decibels) in multiple spots across the superior (upper) hemifield portion of the visual field. This is the amount of agreement of the measure threshold, which is the lowest level of light detected in different points across the visual field in the superior hemifield, between two testing machines.

    Measured: three hours

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted15 Jan 2019
  2. Study start8 Jan 2019
  3. Primary completion20 Oct 2020
  4. Study completion20 Oct 2020

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Non-randomised
Intervention model
Parallel groups
Masking
None (open label)
Enrolment
72 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03804684
Sponsor's protocol ID
IRB Control #18-768E
Status last verified
Jun 2022
First submitted
11 Jan 2019
Last updated
29 Jun 2022
Data monitoring committee
No
FDA-regulated drug
No
FDA-regulated device
Yes
Individual data sharing
No

Publications

  • Razeghinejad R, Gonzalez-Garcia A, Myers JS, Katz LJ. Preliminary Report on a Novel Virtual Reality Perimeter Compared With Standard Automated Perimetry. J Glaucoma. 2021 Jan 1;30(1):17-23. doi: 10.1097/IJG.0000000000001670.

    PubMed 32941320

Questions about this trial

What is this clinical trial studying?

A study testing Standard Automatic Perimetry Humphrey Field Analyzer and visuALL Field Analyzer in people with Glaucoma, Open-Angle, run by Wills Eye at 1 site in United States.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 7 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 29 Jun 2022. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in United States. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

Yes. Results are posted in the registry (first posted 29 Jun 2022). A summary is in the For doctors view, and the full results are on ClinicalTrials.gov.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Mean Retinal Sensitivity From Two Machines (time frame: three hours). The study lists 2 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?

On the ClinicalTrials.gov record for NCT03804684, which also lists 1 linked publication.

Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 29 Jun 2022; AzTrials last refreshed its copy on 2 Nov 2026.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in