AzTrials
AzTrials

AsthmaNCT03894475

Non-inferiority of Portable Versus Desktop Spirometry

A study testing AioCare and Spirometer USB CPFS/D (MGC Diagnostics) in people with Asthma and COPD, run by National Institute for Tuberculosis and Lung Diseases, Poland at 1 site in Poland.

Completed

Finished. The study has ended and is not enrolling. Results may be posted on the official record.

Official record
Search all trials

Trial summary

The investigators compared the ventilation parameters for volume and flow obtained from standard spirometry procedures from patients presently monitored and treated for asthma or chronic obstructive pulmonary disease (COPD) using AioCare (HealthUp Sp. Read more about this study

Who is it for?
Ages 18 years and older
Conditions the study is for
Asthma, COPD
What is being tested?
AioCare and Spirometer USB CPFS/D (MGC Diagnostics)
Type of treatment
Medical device
Could I receive a placebo?
No, everyone receives an active treatment
How many people
62 participants
Where
1 sitePoland
Study dates
18 Dec 2017to 21 Jun 2018
Phase
Interventional
Are there results?
None posted in the registry

Key requirements

Who can take part, as listed by the sponsor. Only the study team can confirm whether you are eligible, so take this list to your doctor.

  • Ages 18 years and older
  • All sexes
  • No healthy volunteers
  • Adult, Older adult

Who may be able to join

  • Age>18.
  • Diagnosed asthma or COPD
  • Signed consent to participation in the research experiment.
  • Ability to comply with the spirometry protocol.

Who cannot join

  • Pregnant females (based on declaration; no pregnancy tests are planned before the test).
  • Recent myocardial infarction (<30 days).
  • Known thoracic, aortic or cerebral aneurysm.
  • Recent stroke, eye surgery, thoracic/abdominal surgery.
  • Haemoptysis.
  • Recent pneumothorax.
Show all 13Show fewer
  • Uncontrolled hypertension.
  • Pulmonary Embolism.
  • Angina.
  • Chest or abdominal pain of any etiology.
  • Oral or facial pain exacerbated by a mouthpiece.
  • Stress incontinence.
  • Dementia or state of confusion. 14, Acute Diarrhea

Trial locations and contacts

1 site in 1 country.

Poland1 site
  • National Institute for Tuberculosis and Lung Diseases

    Warsaw, Masovian Voivodeship, 01-138

Other trials in Poland

About this study

The investigators compared the ventilation parameters for volume and flow obtained from standard spirometry procedures from patients presently monitored and treated for asthma or chronic obstructive pulmonary disease (COPD) using AioCare (HealthUp Sp. z o.o., Serial Number: MS082017005412, software version: MySpiroo app 1.1.14) as the tested device and Spirometer USB CPFS/D (MGC Diagnostics) as the reference, which required calibration prior to each session. Spirometry measurements were performed on sixty-two patients (forty-four females (58±17 years old) and eighteen males (52±19 years old)) at the Institute of Tuberculosis and Lung Disease in Warsaw, Poland. Participants were asked to perform correct spirometry examinations (which means at least three technically correct exhales and meeting repeatability criteria for FEV1 and FVC) on both measuring devices with a five-minute break between devices to prevent respiratory muscle fatigue. The highest value from all acceptable spirometry results was then used for analysis. All spirometry examinations followed ERS/ATS standards.

Conditions
Asthma, COPD
Being tested
AioCare, Spirometer USB CPFS/D (MGC Diagnostics)
Keywords
Asthma, COPD, Mobile Spirometry
Medical subject headings
Asthma, Pulmonary Disease, Chronic Obstructive

What participants receive

What participants receive. In a randomised study each person is assigned to one group by chance.

  • AioCareDevice

    AioCare is an ultraportable, handheld hardware module that contains the MEMS-based flow sensor and electronics, with a dedicated mobile application that works on iOS and Android operating systems. The unit is used with a disposable mouthpiece fitted to the tip of the flow tube, and a nose clip. AioCare is connected to its dedicated mobile application, which contains software that will show flow-volume graphs and results in real time. The device encompasses all of the widely used spirometry parameters, including: peak expiratory flow (PEF), forced expiratory volume in 1 second (FEV1), forced vital capacity (FVC), FEV1/FVC ratio, forced expiratory flow at 25% of expiration (FEF25), forced expiratory flow at 50% of expiration (FEF50), forced expiratory flow at 75% of expiration (FEF75), and forced expiratory volume in 6 second (FEV6).

  • Spirometer USB CPFS/D (MGC Diagnostics)Device

    Spirometer USB CPFS/D (MGC Diagnostics) is a diagnostic desktop spirometer used in hospitals/clinics. CPFS/D USB spirometer is compatible with desktop and laptop computers.

Study groups in NCT03894475
Study groupRoleWhat they receive
Sequence AActive comparatorPatient would first perform an examination with handheld spirometer (AioCare), followed by measurements with the reference spirometer (MGC)
Receives: Device: AioCare; Device: Spirometer USB CPFS/D (MGC Diagnostics)
Sequence BActive comparatorPatient would first perform an examination with the reference spirometer (MGC), followed by measurements with handheld spirometer (AioCare)
Receives: Device: AioCare; Device: Spirometer USB CPFS/D (MGC Diagnostics)

What is being measured

The primary outcome is the main question the study is designed to answer.

  1. Non-inferiority of PEF between AioCare and CPFS/D USB Spirometer

    Comparison of Peak Expiratory Flow obtained during a standard dynamic spirometry measured with AioCare (HealthUp) and reference spirometer (MGC)

    Measured: 6 months

  2. Non-inferiority of FEV1 between AioCare and CPFS/D USB Spirometer

    Comparison of Forced Expiratory Volume in the first second (FEV1) obtained during a standard dynamic spirometry measured with AioCare (HealthUp) and reference spirometer (MGC)

    Measured: 6 months

  3. Non-inferiority of FVC between AioCare and CPFS/D USB Spirometer

    Comparison of Forced Vital Capacity (FVC) obtained during a standard dynamic spirometry measured with AioCare (HealthUp) and reference spirometer (MGC)

    Measured: 6 months

Timeline

From the start of the study to its completion; the mark is today. Estimated dates are the sponsor’s plan.

  1. First posted28 Mar 2019
  2. Study start18 Dec 2017
  3. Primary completion21 Jun 2018
  4. Study completion21 Jun 2018

Resources

More details

For doctors and researchers: how the study is designed, who runs it, and its registry IDs.

Study design
Study type
Interventional
Phase
Phase N/A
Primary purpose
Diagnosis
Allocation
Randomised
Intervention model
Parallel groups
Masking
None (open label)
Enrolment
62 participants (actual)
IDs, dates and publications
ClinicalTrials.gov ID
NCT03894475
Sponsor's protocol ID
KB-6
Status last verified
Mar 2019
First submitted
19 Mar 2019
Last updated
28 Mar 2019
Results posted
No
Data monitoring committee
Yes
FDA-regulated drug
No
FDA-regulated device
No
Individual data sharing
Undecided

Questions about this trial

What is this clinical trial studying?

A study testing AioCare and Spirometer USB CPFS/D (MGC Diagnostics) in people with Asthma and COPD, run by National Institute for Tuberculosis and Lung Diseases, Poland at 1 site in Poland.

The full registered description is in the About this study section and the official record.

Who may be eligible to take part?

The sponsor lists 4 inclusion and 13 exclusion criteria. Eligibility is decided by the study team after screening, so read the full criteria and contact a listed site.

Is this clinical trial currently recruiting?

The record says: finished, as of the sponsor's last update on 28 Mar 2019. Status can change between updates, so confirm with the study contact before making plans.

Where is the study taking place?

The record lists 1 site in Poland. Whether a site is taking part right now should be confirmed with that site. See every site.

What are the possible benefits and risks?

They depend on the study and the treatment, and no study can promise a benefit. The study team explains the known risks during informed consent, before anyone joins; the record itself does not list them.

Are results available for this study?

The study is completed, but results are not available in the source record. That does not mean the treatment did or did not work.

How can a doctor refer a patient or contact the research team?

The record lists no contact. The official record names the sponsor, who can be asked directly. AzTrials does not enrol participants or take referrals.

What does the primary outcome measure?

Non-inferiority of PEF between AioCare and CPFS/D USB Spirometer (time frame: 6 months). The study lists 3 primary outcomes in all. It is what the study was designed to measure, not a finding.

Where can I read the original study information?
Does a listing mean the treatment is approved or proven to work?

No. A listing means the study is registered. It does not show that the treatment is safe, works, or is approved by a regulator.

How often is this information updated?

AzTrials copies changed records from ClinicalTrials.gov every night and reloads the whole registry every week. The sponsor last updated this record on 28 Mar 2019.

Free account

Log in to keep searching

You’ve used your 3 free searches. AzTrials is free: log in or create a free account to keep searching clinical trials and save the ones you want to follow.

Create a free accountLog in